🇵🇱 ASPEL S.A. PL
电子与电磁兼容建筑
ASPEL S.A. 是由 PCA 认可的实验室,认可编号 AB 1604(Zabierzów,波兰)。认可范围涵盖 36 项检测方法,涉及以下领域:电子与电磁兼容、建筑。检测对象包括:医用电气设备、Electrical, electronic and medical equipment、电气与电子设备、Industrial, scientific and medical equipment。
要点速览
- 认可编号: PCA AB 1604
- 认可范围内方法数: 36
- 检测对象种类: 8
- 城市: Zabierzów, 波兰
- 认可编号
- AB 1604
- 认可机构
- PCA
- 地址
- CENTRALNE LABORATORIUM APARATURY MEDYCZNEJ CELAMED, Os. Henryka Sienkiewicza 33, 32-080 Zabierzów
- 城市
- Zabierzów, 波兰
该实验室检测什么 (8)
- 医用电气设备 — 17 项方法
- Electrical, electronic and medical equipment — 8 项方法
- 电气与电子设备 — 3 项方法
- Industrial, scientific and medical equipment — 2 项方法
- Information technology equipment — 2 项方法
- Multimedia equipment — 2 项方法
- Medical electrical equipment - electrocardiographs (limitation of the dimensions of the tested object to: (540x540x750) mm) — 1 项方法
- Medical electrical equipment - electroencephalographs — 1 项方法
认可范围 — 项检测方法 (36)
| 序号 | 检测对象 | 方法 / 参数 | 技术 | 依据文件 | |
|---|---|---|---|---|---|
| 检测方法 | |||||
| 1 | 医用电气设备 | Power consumption Measuring range: up to 3 kW | PN-EN 60601-1:2011 PN-EN 60601-1/A11:2011 PN-EN 60601-1:2011/A1:2014-02 PN-EN 60601-1:2011/A12:2014-12 PN-EN 60601-1:2011/A2:2022-03 | ||
| 2 | 医用电气设备 | Durability of marking | |||
| 3 | 医用电气设备 | Legibility of marking | |||
| 4 | 医用电气设备 | Residual voltage Measurement range: up to 1 kV | |||
| 5 | 医用电气设备 | Leakage currents: earth leakage current Range: (0 – 8000) μA touch leakage current Range: (0 – 8000) μA patient leakage current Range: (0 – 8000) μA patient auxiliary current Range: (0 – 8000) μA patient leakage current (mains voltage on the applied part) Range: (0 – 8000) μA | |||
| 6 | 医用电气设备 | Electric strength of the insulation Range: (0 – 5) kV AC | |||
| 7 | 医用电气设备 | Resistance to heat Range: (40 – 150) °C | |||
| 8 | 医用电气设备 | Creepage distances and clearances | |||
| 9 | 医用电气设备 | Gaps of trapping zones | |||
| 10 | 医用电气设备 | Instability of the appliance in the transport position | |||
| 11 | 医用电气设备 | Instability of the appliance other than in transport | |||
| 12 | 医用电气设备 | Instability due to horizontal and vertical forces | |||
| 13 | 医用电气设备 | Force required to set the equipment in motion | |||
| 14 | 医用电气设备 | Ability of the device to pass over a threshold | |||
| 15 | 医用电气设备 | Enclosure stiffness (pressure test) Impact resistance Free-fall resistance Ability to: pass over a protruding threshold, descend from a step, withstand the device striking a door frame, | |||
| 16 | 医用电气设备 | Interruptions in the continuity of the power supply | |||
| 17 | 医用电气设备 | Humidity preconditioning Temperature range in climatic mode: (20 – 32) °C Humidity range in climatic mode: 98% ± 3% | |||
| 18 | Medical electrical equipment - electroencephalographs | Defibrillation pulse withstand Multichannel crosstalk Input impedance Range and bandwidth of transmitted frequencies | PN-EN 60601-2-26:2015-09 | ||
| 19 | Medical electrical equipment - electrocardiographs (limitation of the dimensions of the tested object to: (540x540x750) mm) | Defibrillation pulse withstand Baseline recovery time Input impedance and lead network impedance Reproduction of the calibration voltage Gain stability Gain accuracy Common mode rejection Overload withstand Baseline (thermal drift, stability, noise level, recording speed and trace width, channel crosstalk, sensitivity/effect on the baseline) Frequency response Linearity and dynamic range Response to a minimum signal Recording speed Use with a cardiac pacemaker (signal distortion, visibility of pacemaker pulses) | PN-EN 60601-2-25:2004 PN-EN 60601-2-47:2015-09 PN-EN 60601-2-51:2005 PN-EN 60601-2-25:2016-01 | ||
| 20 | Electrical, electronic and medical equipment | Immunity to voltage dips, short interruptions and voltage variations Range: 0%U , 40%U , 70%U , Ext T T T | PN-EN 61000-4-11:2020-11 | ||
| 21 | Electrical, electronic and medical equipment | Immunity to electrostatic discharge contact up to 9 kV air up to 16,5 kV | PN-EN 61000-4-2:2011 | ||
| 22 | Electrical, electronic and medical equipment | Immunity to surges Range: (0,2 – 4,4) kV | PN-EN 61000-4-5:2014-10 PN-EN 61000-4-5:2014- 10/A1:2018-01 z wyłączeniem badań szybkich linii telekomunikacyjnych | ||
| 23 | Electrical, electronic and medical equipment | Immunity to electrical fast transient/burst Range: (0,2 – 4,4) kV | PN-EN 61000-4-4:2013-05 z wyłączeniem punktu 6.4 | ||
| 24 | Electrical, electronic and medical equipment | Immunity to power frequency magnetic field Range: (0,1 – 40) A/m | PN-EN 61000-4-8:2010 | ||
| 25 | Electrical, electronic and medical equipment | Immunity to pulse magnetic field Range: (100 – 2200) A/m | PN-EN 61000-4-9:1998 + A1:2003 | ||
| 26 | Electrical, electronic and medical equipment | Immunity to conducted disturbances induced by radio-frequency fields: Range:150 kHz – 230 MHz | PN-EN IEC 61000-4-6:2024-03 | ||
| 27 | Electrical, electronic and medical equipment | Immunity to radiated radio-frequency electromagnetic field: Range: 80 MHz – 3 GHz, disturbance level up to 10 V/m | PN-EN 61000-4-3:2021-06 | ||
| 28 | 电气与电子设备 | Emission of electromagnetic disturbances | PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12 PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03 | ||
| 29 | 电气与电子设备 | Current harmonics Range: (0 – 16) A, n = 1, 2, ..., 50 | PN-EN IEC 61000-3-2:2019-04 PN-EN IEC 61000-3-2:2019-04/A1:2021-08 PN-EN IEC 61000-3-2:2019-04/A2:2024-12 | ||
| 30 | 电气与电子设备 | Voltage changes, voltage fluctuations and flicker in public low-voltage supply systems | PN-EN 61000-3-3:2013-10 PN-EN 61000-3-3:2013-10/A1:2019-10 PN-EN 61000-3-3:2013-10/A2:2022-04 | ||
| 31 | Industrial, scientific and medical equipment | Continuous radio disturbance voltage on mains supply leads Range: 150 kHz – 30 MHz | PN-EN 55011:2016-05 PN-EN 55011:2016-05/A1:2017-06 PN-EN 55011:2016-05/A2:2021-08 PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12 | ||
| 32 | Industrial, scientific and medical equipment | Electromagnetic field strength Range: 30 MHz – 6 GHz measurement distance 3 m | PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03 PN-EN 55011:2016-05 PN-EN 55011:2016-05/A1:2017-06 PN-EN 55011:2016-05/A2:2021-08 | ||
| 33 | Information technology equipment | Continuous radio disturbance voltage on mains supply leads Range: 150 kHz – 30 MHz | PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12 | ||
| 34 | Information technology equipment | Electromagnetic field strength Range: 30 MHz – 6 GHz measurement distance 3 m | PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03 | ||
| 35 | Multimedia equipment | Continuous radio disturbance voltage on mains supply leads Range: 150 kHz – 30 MHz | PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12 PN-EN 55032:2015-09 PN-EN 55032:2015-09/A1:2021-05 | ||
| 36 | Multimedia equipment | Electromagnetic field strength Range: 30 MHz – 6 GHz measurement distance 3 m | PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03 PN-EN 55032:2015-09 PN-EN 55032:2015-09/A1:2021-05 | ||
引用标准 (20)
PN-EN 55011:2016-05 PN-EN 55011:2016-05/A1:2017-06 PN-EN 55011:2016-05/A2:2021-08 PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12
PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12
PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12 PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03
PN-EN 55016-2-1:2014-09 PN-EN 55016-2-1:2014-09/A1:2017-12 PN-EN 55032:2015-09 PN-EN 55032:2015-09/A1:2021-05
PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03
PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03 PN-EN 55011:2016-05 PN-EN 55011:2016-05/A1:2017-06 PN-EN 55011:2016-05/A2:2021-08
PN-EN 55016-2-3:2017-06 PN-EN 55016-2-3:2017-06/A1:2020-01 PN-EN 55016-2-3:2017-06/A2:2024-03 PN-EN 55032:2015-09 PN-EN 55032:2015-09/A1:2021-05
PN-EN 60601-1:2011 PN-EN 60601-1/A11:2011 PN-EN 60601-1:2011/A1:2014-02 PN-EN 60601-1:2011/A12:2014-12 PN-EN 60601-1:2011/A2:2022-03
PN-EN 60601-2-25:2004 PN-EN 60601-2-47:2015-09 PN-EN 60601-2-51:2005 PN-EN 60601-2-25:2016-01
PN-EN 60601-2-26:2015-09
PN-EN 61000-3-3:2013-10 PN-EN 61000-3-3:2013-10/A1:2019-10 PN-EN 61000-3-3:2013-10/A2:2022-04
PN-EN 61000-4-11:2020-11
PN-EN 61000-4-2:2011
PN-EN 61000-4-3:2021-06
PN-EN 61000-4-4:2013-05 z wyłączeniem punktu 6.4
PN-EN 61000-4-5:2014-10 PN-EN 61000-4-5:2014- 10/A1:2018-01 z wyłączeniem badań szybkich linii telekomunikacyjnych
PN-EN 61000-4-8:2010
PN-EN 61000-4-9:1998 + A1:2003
PN-EN IEC 61000-3-2:2019-04 PN-EN IEC 61000-3-2:2019-04/A1:2021-08 PN-EN IEC 61000-3-2:2019-04/A2:2024-12
PN-EN IEC 61000-4-6:2024-03