🇨🇿 Hexagon Manufacturing Intelligence Czech Republic s.r.o. PL
Metrology & calibration
Hexagon Manufacturing Intelligence Czech Republic s.r.o. is a calibration laboratory accredited by ČIA under number 2397 (Praha 8, Czech Republic). Its accreditation scope covers 23 calibration items in the following areas: Metrology & calibration. Calibrated items include: Laser trackers, Measuring arms, Coordinate measuring machines, Optical measuring machines with a multisensor. Most frequently used calibration principles: Measurement using a calibration ball, Measurement using a step gauge, Comparison with a reference laser tracker.
At a glance
- Accreditation number: ČIA 2397
- Calibration items in scope: 23
- Types of calibrated items: 5
- Calibration principles used: 14
- City: Praha 8, Czech Republic
- Valid until: 28-11-2030
- Accreditation number
- 2397
- Accreditation body
- ČIA
- Address
- Boudníkova 2538/13, Libeň, 180 00 Praha 8
- City
- Praha 8, Czech Republic
- Division
- Laboratorium wzorcujące
- Valid until
- 28-11-2030
What this laboratory calibrates (5)
- Laser trackers — 8 calibration items
- Measuring arms — 6 calibration items
- Coordinate measuring machines — 4 calibration items
- Optical measuring machines with a multisensor — 3 calibration items
- Coordinate measuring machines with a laser sensor — 2 calibration items
Calibration principles (14)
- Measurement using a calibration ball — 8 calibration items
- Measurement using a step gauge — 3 calibration items
- Comparison with a reference laser tracker — 1 calibration item
- Glass ruler measurement — 1 calibration item
- Laser interferometer measurements — 1 calibration item
- Measurement by self-calibration procedure — 1 calibration item
- Measurement using a reference plate by an optical sensor (scanning) — 1 calibration item
- Measurement using a scale bar — 1 calibration item
- Measurement using white plate standard — 1 calibration item
- Measurement with parallel gauge blocks — 1 calibration item
- Measurement with parallel gauge blocks and step gauge — 1 calibration item
- Multi-position measurement with a switching tactile probe — 1 calibration item
- Rubidium frequency standard measurement — 1 calibration item
- Two face measurement by reflector — 1 calibration item
Laboratory sections (3)
- Workplace Prague — 19 calibration items
- Workplace Bratislava — 2 calibration items
- Workplace Budapest — 2 calibration items
Laboratory locations
- Hexagon Manufacturing Intelligence Czech Republic s.r.o. (main site) — Boudníkova 2538/13, Libeň, 180 00 Praha 8
- Pracownia Praga — Boudníkova 2538/13, Libeň, 180 00 Praha 8
- Pracownia Bratysława — Tuhovská, 831 06 Bratislava – Vajnory
- Pracownia Budapeszt — Nagytétényi út 112, 1222 Budapest
Certificate information
- Certificate date
- 23.09.2026
Accreditation scope — calibration items (23)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Workplace Prague | |||||
| 1 | Coordinate measuring machines | Calibration – length Spatial length — 0 m to 6 m · CMC (0,25·L + 0,25) µm CMC for the field of measured quantity: Length |
Measurement with parallel gauge blocks and step gauge | KP1 – SMS1 | |
| 2 | Coordinate measuring machines | Calibration – length Sensing system error — 0 mm to 1 mm · CMC 0,1 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP1 – SMS1 | |
| 3 | Coordinate measuring machines | Calibration – length Spatial length — 0 m to 30 m · CMC (0,5·L + 0,06) µm CMC for the field of measured quantity: Length |
Laser interferometer measurements | KP2 – SMS2 | |
| 4 | Coordinate measuring machines | Calibration – length Sensing system error — 0 mm to 1 mm · CMC 0,1 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP2 – SMS2 | |
| 5 | Coordinate measuring machines with a laser sensor | Calibration – length Dimension sensing — 10 mm to 50 mm · CMC 1,9 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP5 – LASER | |
| 6 | Coordinate measuring machines with a laser sensor | Calibration – length Sensing of shape and scatter of sensing — 0 mm to 1 mm · CMC 2,0 µm CMC for the field of measured quantity: Length |
Measurement using white plate standard | KP5 – LASER | |
| 7 | Optical measuring machines with a multisensor | Calibration – length Spatial length — 0 mm to 600 mm · CMC (1,1·L + 0,3) µm CMC for the field of measured quantity: Length |
Measurement with parallel gauge blocks | KP3 – OPTIKA | |
| 8 | Optical measuring machines with a multisensor | Calibration – length Spatial length — 0 mm to 600 mm · CMC (0,2·L + 0,2) µm CMC for the field of measured quantity: Length |
Glass ruler measurement | KP3 – OPTIKA | |
| 9 | Optical measuring machines with a multisensor | Calibration – length Sensing system error — 0 mm to 1 mm · CMC 0,1 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP3 – OPTIKA | |
| 10 | Measuring arms | Calibration – length Spatial length — 0 m to 4,5 m · CMC (2,1·L + 1,2) µm CMC for the field of measured quantity: Length |
Measurement using a step gauge | KP4 – RAMENA; ČSN EN ISO 10360-12; STN EN ISO 10360-12; MSZ EN ISO 10360-12; EN ISO 10360-12 | |
| 11 | Measuring arms | Calibration – length Form sensing deviation — 0 mm to 1 mm · CMC 1,0 µm Size sensing deviation — -1 mm to 1 mm · CMC 1,0 µm Position sensing deviation — 0 mm to 1 mm · CMC 1,0 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP4 – RAMENA; ČSN EN ISO 10360-12; STN EN ISO 10360-12; MSZ EN ISO 10360-12; EN ISO 10360-12 | |
| 12 | Laser trackers | Calibration – length Distance measurement deviation; length 1,5 m to 53 m — -0,05 mm to 0,05 mm · CMC 5 µm CMC for the field of measured quantity: Length |
Comparison with a reference laser tracker | KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 13 | Laser trackers | Calibration – length Two face measurement deviation; length 1,5 m to 53 m — -0,4 mm to 0,4 mm · CMC 6 µm CMC for the field of measured quantity: Length |
Two face measurement by reflector | KP6 – TRAKČNÍ LASER 1; KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 14 | Laser trackers | Calibration – length Spatial length deviation in probe sensing; length 1,5 m to 25 m — -0,5 mm to 0,5 mm · CMC 12 µm Spatial length deviation in reflector sensing; length 1,5 m to 25 m — -0,5 mm to 0,5 mm · CMC 9 µm CMC for the field of measured quantity: Length |
Measurement using a scale bar | KP6 – TRAKČNÍ LASER 1; KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 15 | Laser trackers | Calibration – length Size sensing deviation; length 2 m to 20 m — -0,2 mm to 0,2 mm · CMC 6 µm Form sensing deviation; length 2 m to 20 m — 0 mm to 0,4 mm · CMC 8 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 16 | Laser trackers | Calibration – length Plane sensing deviation; length 2 m to 20 m — 0 mm to 0,5 mm · CMC 12 µm CMC for the field of measured quantity: Length |
Measurement using a reference plate by an optical sensor (scanning) | KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 17 | Laser trackers | Calibration – length Orientation deviation; length 2 m to 10 m — 0 mm to 0,2 mm · CMC 10 µm CMC for the field of measured quantity: Length |
Multi-position measurement with a switching tactile probe | KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 18 | Laser trackers | Calibration – length Distance offset; length 1,5 m to 15 m — -2,0 mm to 2,0 mm · CMC 7 µm CMC for the field of measured quantity: Length |
Measurement by self-calibration procedure | KP6 – TRAKČNÍ LASER 1; KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| 19 | Laser trackers | Calibration – length Absolute distance meter modulation frequency — 0 kHz to 25 MHz · CMC 0,75 Hz CMC for the field of measured quantity: Length |
Rubidium frequency standard measurement | KP6 – TRAKČNÍ LASER 1; KP7 – TRAKČNÍ LASER 2; ČSN EN ISO 10 360-10; EN ISO 10 360-10 | |
| Workplace Bratislava | |||||
| 20 | Measuring arms | Calibration – length Spatial length — 0 m to 4,5 m · CMC (2,1·L + 1,2) µm CMC for the field of measured quantity: Length |
Measurement using a step gauge | KP4 – RAMENA; ČSN EN ISO 10360-12; STN EN ISO 10360-12; MSZ EN ISO 10360-12; EN ISO 10360-12 | |
| 21 | Measuring arms | Calibration – length Form sensing deviation — 0 mm to 1 mm · CMC 1,0 µm Size sensing deviation — -1 mm to 1 mm · CMC 1,0 µm Position sensing deviation — 0 mm to 1 mm · CMC 1,0 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP4 – RAMENA; ČSN EN ISO 10360-12; STN EN ISO 10360-12; MSZ EN ISO 10360-12; EN ISO 10360-12 | |
| Workplace Budapest | |||||
| 22 | Measuring arms | Calibration – length Spatial length — 0 m to 4,5 m · CMC (2,1·L + 1,2) µm CMC for the field of measured quantity: Length |
Measurement using a step gauge | KP4 – RAMENA; ČSN EN ISO 10360-12; STN EN ISO 10360-12; MSZ EN ISO 10360-12; EN ISO 10360-12 | |
| 23 | Measuring arms | Calibration – length Form sensing deviation — 0 mm to 1 mm · CMC 1,0 µm Size sensing deviation — -1 mm to 1 mm · CMC 1,0 µm Position sensing deviation — 0 mm to 1 mm · CMC 1,0 µm CMC for the field of measured quantity: Length |
Measurement using a calibration ball | KP4 – RAMENA; ČSN EN ISO 10360-12; STN EN ISO 10360-12; MSZ EN ISO 10360-12; EN ISO 10360-12 | |
Reference standards (13)
EN ISO 10 360-10
EN ISO 10360-12
KP1 – SMS1
KP2 – SMS2
KP3 – OPTIKA
KP4 – RAMENA
KP5 – LASER
KP6 – TRAKČNÍ LASER 1
KP7 – TRAKČNÍ LASER 2
MSZ EN ISO 10360-12
STN EN ISO 10360-12
ČSN EN ISO 10 360-10
ČSN EN ISO 10360-12