🇵🇱 SAMSO Grzegorz Nadolny PL

PCA AP 231 • Koszalin, Poland • 33 calibration items

Metrology & calibration

SAMSO Grzegorz Nadolny is a calibration laboratory accredited by PCA under number AP 231 (Koszalin, Poland). Its accreditation scope covers 33 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electrical safety testers, Power network parameter meters, Multimeters, Calibrators. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S), Internal procedure; Indirect method; At permanent location (S), Internal procedure; Method 5.2.1 and 5.2.2. Direct method; At perma….

At a glance

  • Accreditation number: PCA AP 231
  • Calibration items in scope: 33
  • Types of calibrated items: 12
  • Calibration principles used: 3
  • City: Koszalin, Poland
  • Valid until: 04-09-2029
Accreditation number
AP 231
Accreditation body
PCA
Address
ul. Różana 1c/1, 75-220 Koszalin
City
Koszalin, Poland
Division
SAMSO Calibration Laboratory
Valid until
04-09-2029

What this laboratory calibrates (12)

Calibration principles (3)

Certificate information

Issue date
17.07.2026
Issue no.
2
Certificate date
05.09.2025
Accreditation cycle
05.09.2025 — 04.09.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (33)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Calibrators Calibration — DC voltage
0 mV to 100 mV · CMC 7,5 · 10⁻⁶ · U + 0,49 µV 0,1 V to 1 V · CMC 5,2 · 10⁻⁶ · U + 0,85 µV 1 V to 10 V · CMC 5,4 · 10⁻⁶ · U + 7 µV 10 V to 100 V · CMC 7,4 · 10⁻⁶ · U + 93 µV 100 V to 1000 V · CMC 7,4 · 10⁻⁶ · U + 1,4 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.2
2 Electrical safety testers Calibration — DC voltage
0 mV to 100 mV · CMC 7,5 · 10⁻⁶ · U + 0,49 µV 0,1 V to 1 V · CMC 5,2 · 10⁻⁶ · U + 0,85 µV 1 V to 10 V · CMC 5,4 · 10⁻⁶ · U + 7 µV 10 V to 100 V · CMC 7,4 · 10⁻⁶ · U + 93 µV 100 V to 1000 V · CMC 7,4 · 10⁻⁶ · U + 1,4 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
3 Insulation resistance meters
Electrical safety testers
Calibration — DC voltage
0 V to 1000 V · CMC 1,2 · 10⁻² · U + 1,1 V 1000 V to 6000 V · CMC 1,2 · 10⁻² · U + 0,0072 V
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
4 Power network parameter meters Calibration — DC voltage
0 V to 1000 V · CMC 1,2 · 10⁻² · U + 1,1 V 1000 V to 6000 V · CMC 1,2 · 10⁻² · U + 0,0072 V
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
5 Multimeters
Digital voltmeters
Calibration — DC voltage
0 mV to 200 mV · CMC 2,6 · 10⁻⁵ · U + 3,6 µV 0,2 V to 2 V · CMC 1,5 · 10⁻⁵ · U + 3,6 µV 2 V to 20 V · CMC 1,5 · 10⁻⁵ · U + 36 µV 20 V to 200 V · CMC 1,5 · 10⁻⁵ · U + 0,36 mV 200 V to 1000 V · CMC 2,2 · 10⁻⁵ · U + 3,6 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.1 based on EURAMET cg-15 v.3.0
6 Electrical safety testers Calibration — DC voltage
0 mV to 200 mV · CMC 2,6 · 10⁻⁵ · U + 3,6 µV 0,2 V to 2 V · CMC 1,5 · 10⁻⁵ · U + 3,6 µV 2 V to 20 V · CMC 1,5 · 10⁻⁵ · U + 36 µV 20 V to 200 V · CMC 1,5 · 10⁻⁵ · U + 0,36 mV 200 V to 1000 V · CMC 2,2 · 10⁻⁵ · U + 3,6 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
7 Power network parameter meters Calibration — DC voltage
0 mV to 200 mV · CMC 2,6 · 10⁻⁵ · U + 3,6 µV 0,2 V to 2 V · CMC 1,5 · 10⁻⁵ · U + 3,6 µV 2 V to 20 V · CMC 1,5 · 10⁻⁵ · U + 36 µV 20 V to 200 V · CMC 1,5 · 10⁻⁵ · U + 0,36 mV 200 V to 1000 V · CMC 2,2 · 10⁻⁵ · U + 3,6 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
8 Calibrators Calibration — DC current
0 µA to 100 µA · CMC 1 · 10⁻⁵ · I + 0,48 nA 100 µA to 1 mA · CMC 1 · 10⁻⁵ · I + 4,8 nA 1 mA to 10 mA · CMC 1,2 · 10⁻⁵ · I + 49 nA 10 mA to 100 mA · CMC 3,5 · 10⁻⁵ · I + 0,72 µA 100 mA to 1 A · CMC 1,7 · 10⁻⁴ · I + 16 µA 1 A to 10 A · CMC 4,2 · 10⁻⁴ · I + 410 µA 10 A to 30 A · CMC 5,8 · 10⁻⁴ · I + 5,4 mA
DC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.2
9 Electrical safety testers Calibration — DC current
0 µA to 100 µA · CMC 1 · 10⁻⁵ · I + 0,48 nA 100 µA to 1 mA · CMC 1 · 10⁻⁵ · I + 4,8 nA 1 mA to 10 mA · CMC 1,2 · 10⁻⁵ · I + 49 nA 10 mA to 100 mA · CMC 3,5 · 10⁻⁵ · I + 0,72 µA 100 mA to 1 A · CMC 1,7 · 10⁻⁴ · I + 16 µA 1 A to 10 A · CMC 4,2 · 10⁻⁴ · I + 410 µA 10 A to 30 A · CMC 5,8 · 10⁻⁴ · I + 5,4 mA
DC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
10 Multimeters
Digital ammeters
Calibration — DC current
0 µA to 200 µA · CMC 9,4 · 10⁻⁵ · I + 35 nA 0,2 mA to 2 mA · CMC 8,3 · 10⁻⁵ · I + 46 nA 2 mA to 20 mA · CMC 5 · 10⁻⁵ · I + 0,35 µA 20 mA to 200 mA · CMC 7,3 · 10⁻⁵ · I + 3,5 µA 0,2 A to 2 A · CMC 1,1 · 10⁻⁴ · I + 41 µA 2 A to 30 A · CMC 3,8 · 10⁻⁴ · I + 0,41 mA
DC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.1 based on EURAMET cg-15 v.3.0
11 Electrical safety testers Calibration — DC current
0 µA to 200 µA · CMC 9,4 · 10⁻⁵ · I + 35 nA 0,2 mA to 2 mA · CMC 8,3 · 10⁻⁵ · I + 46 nA 2 mA to 20 mA · CMC 5 · 10⁻⁵ · I + 0,35 µA 20 mA to 200 mA · CMC 7,3 · 10⁻⁵ · I + 3,5 µA 0,2 A to 2 A · CMC 1,1 · 10⁻⁴ · I + 41 µA 2 A to 30 A · CMC 3,8 · 10⁻⁴ · I + 0,41 mA
DC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
12 Calibrators Calibration — AC voltage
40 Hz to 200 Hz — 10 mV to 100 mV · CMC 2,5 · 10⁻⁴ · U + 10 µV 40 Hz to 200 Hz — 0,1 V to 1 V · CMC 2,2 · 10⁻⁴ · U + 70 µV 40 Hz to 200 Hz — 1 V to 10 V · CMC 2,2 · 10⁻⁴ · U + 700 µV 40 Hz to 200 Hz — 10 V to 100 V · CMC 2,3 · 10⁻⁴ · U + 10 mV 40 Hz to 200 Hz — 100 V to 1000 V · CMC 2,4 · 10⁻⁴ · U + 100 mV 0,2 kHz to 1kHz — 10 mV to 100 mV · CMC 2,1 · 10⁻⁴ · U + 9,3 µV 0,2 kHz to 1kHz — 0,1 V to 1 V · CMC 1,8 · 10⁻⁴ · U + 70 µV 0,2 kHz to 1kHz — 1 V to 10 V · CMC 1,8 · 10⁻⁴ · U + 700 µV 0,2 kHz to 1kHz — 10 V to 100 V · CMC 2,1 · 10⁻⁴ · U + 8,1 mV 0,2 kHz to 1kHz — 100 V to 1000 V · CMC 2,1 · 10⁻⁴ · U + 81 mV 1 kHz to 2 kHz — 10 mV to 100 mV · CMC 2,1 · 10⁻⁴ · U + 10 µV 1 kHz to 2 kHz — 0,1 V to 1 V · CMC 1,8 · 10⁻⁴ · U + 71 µV 1 kHz to 2 kHz — 1 V to 10 V · CMC 1,8 · 10⁻⁴ · U + 700 µV 1 kHz to 2 kHz — 10 V to 100 V · CMC 2,1 · 10⁻⁴ · U + 8,1 mV 1 kHz to 2 kHz — 100 V to 1000 V · CMC 2,1 · 10⁻⁴ · U + 81mV 2 kHz to 20 kHz — 10 mV to 100 mV · CMC 3 · 10⁻⁴ · U + 12 µV 2 kHz to 20 kHz — 0,1 V to 1 V · CMC 2,9 · 10⁻⁴ · U + 120 µV 2 kHz to 20 kHz — 1 V to 10 V · CMC 2,9 · 10⁻⁴ · U + 1,2 mV 2 kHz to 20 kHz — 10 V to 100 V · CMC 3,5 · 10⁻⁴ · U + 12 mV 2 kHz to 10 kHz — 100 V to 1000 V · CMC 3,5 · 10⁻⁴ · U + 120 mV
AC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.2
13 Electrical safety testers Calibration — AC voltage
50 Hz to 60 Hz — 100 V to 1000 V · CMC 2,4 · 10⁻⁴ · U + 100 mV 50 Hz to 60 Hz — 1000 V to 5000 V · CMC 1,3 · 10⁻² · U + 0,7 V
AC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
14 Multimeters
Digital voltmeters
Calibration — AC voltage
0,045 kHz to 10 kHz — 20 mV to 200 mV · CMC 2,5 · 10⁻⁴ · U + 33 µV 0,045 kHz to 10 kHz — 0,2 V to 2 V · CMC 2,3 · 10⁻⁴ · U + 95 µV 0,045 kHz to 10 kHz — 2 V to 20 V · CMC 2,3 · 10⁻⁴ · U + 0,82 mV 0,045 kHz to 10 kHz — 20 V to 200 V · CMC 5,4 · 10⁻⁴ · U + 8,8 mV 10 kHz to 20 kHz — 20 mV to 200 mV · CMC 5,9 · 10⁻⁴ · U + 42 µV 10 kHz to 20 kHz — 0,2 V to 2 V · CMC 6,4 · 10⁻⁴ · U + 95 µV 10 kHz to 20 kHz — 2 V to 20 V · CMC 6,4 · 10⁻⁴ · U + 0,93 mV 10 kHz to 20 kHz — 20 V to 200 V · CMC 7,4 · 10⁻⁴ · U + 12 mV 0,045 kHz to 1 kHz — 200 V to 1020 V · CMC 5,4 · 10⁻⁴ · U + 28 mV 1 kHz to 5 kHz — 200 V to 1020 V · CMC 7,9 · 10⁻⁴ · U + 28 mV
AC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.1 based on EURAMET cg-15 v.3.0
15 Electrical safety testers Calibration — AC voltage
0,045 kHz to 10 kHz — 20 mV to 200 mV · CMC 2,5 · 10⁻⁴ · U + 33 µV 0,045 kHz to 10 kHz — 0,2 V to 2 V · CMC 2,3 · 10⁻⁴ · U + 95 µV 0,045 kHz to 10 kHz — 2 V to 20 V · CMC 2,3 · 10⁻⁴ · U + 0,82 mV 0,045 kHz to 10 kHz — 20 V to 200 V · CMC 5,4 · 10⁻⁴ · U + 8,8 mV 10 kHz to 20 kHz — 20 mV to 200 mV · CMC 5,9 · 10⁻⁴ · U + 42 µV 10 kHz to 20 kHz — 0,2 V to 2 V · CMC 6,4 · 10⁻⁴ · U + 95 µV 10 kHz to 20 kHz — 2 V to 20 V · CMC 6,4 · 10⁻⁴ · U + 0,93 mV 10 kHz to 20 kHz — 20 V to 200 V · CMC 7,4 · 10⁻⁴ · U + 12 mV 0,045 kHz to 1 kHz — 200 V to 1020 V · CMC 5,4 · 10⁻⁴ · U + 28 mV 1 kHz to 5 kHz — 200 V to 1020 V · CMC 7,9 · 10⁻⁴ · U + 28 mV
AC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
16 Power network parameter meters Calibration — AC voltage
0,045 kHz to 10 kHz — 20 mV to 200 mV · CMC 2,5 · 10⁻⁴ · U + 33 µV 0,045 kHz to 10 kHz — 0,2 V to 2 V · CMC 2,3 · 10⁻⁴ · U + 95 µV 0,045 kHz to 10 kHz — 2 V to 20 V · CMC 2,3 · 10⁻⁴ · U + 0,82 mV 0,045 kHz to 10 kHz — 20 V to 200 V · CMC 5,4 · 10⁻⁴ · U + 8,8 mV 10 kHz to 20 kHz — 20 mV to 200 mV · CMC 5,9 · 10⁻⁴ · U + 42 µV 10 kHz to 20 kHz — 0,2 V to 2 V · CMC 6,4 · 10⁻⁴ · U + 95 µV 10 kHz to 20 kHz — 2 V to 20 V · CMC 6,4 · 10⁻⁴ · U + 0,93 mV 10 kHz to 20 kHz — 20 V to 200 V · CMC 7,4 · 10⁻⁴ · U + 12 mV 0,045 kHz to 1 kHz — 200 V to 1020 V · CMC 5,4 · 10⁻⁴ · U + 28 mV 1 kHz to 5 kHz — 200 V to 1020 V · CMC 7,9 · 10⁻⁴ · U + 28 mV
AC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
17 Calibrators Calibration — AC current
0,04 kHz to 1 kHz — 10 µA to 100 µA · CMC 3,7 · 10⁻⁴ · I + 15 nA 0,04 kHz to 1 kHz — 0,1 mA to 1 mA · CMC 3,7 · 10⁻⁴ · I + 150 nA 0,04 kHz to 1 kHz — 1 mA to 10 mA · CMC 3,7 · 10⁻⁴ · I + 1,5 µA 0,04 kHz to 1 kHz — 10 mA to 100 mA · CMC 3,8 · 10⁻⁴ · I + 14 µA 0,04 kHz to 1 kHz — 0,1 A to 1 A · CMC 4,9 · 10⁻⁴ · I + 180 µA 0,04 kHz to 1 kHz — 1 A to 10 A · CMC 8,3 · 10⁻⁴ · I + 3,5 mA 0,04 kHz to 1 kHz — 10 A to 30 A · CMC 8,3 · 10⁻⁴ · I + 10 mA
AC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.2
18 Electrical safety testers Calibration — AC current
0,04 kHz to 1 kHz — 10 µA to 100 µA · CMC 3,7 · 10⁻⁴ · I + 15 nA 0,04 kHz to 1 kHz — 0,1 mA to 1 mA · CMC 3,7 · 10⁻⁴ · I + 150 nA 0,04 kHz to 1 kHz — 1 mA to 10 mA · CMC 3,7 · 10⁻⁴ · I + 1,5 µA 0,04 kHz to 1 kHz — 10 mA to 100 mA · CMC 3,8 · 10⁻⁴ · I + 14 µA 0,04 kHz to 1 kHz — 0,1 A to 1 A · CMC 4,9 · 10⁻⁴ · I + 180 µA 0,04 kHz to 1 kHz — 1 A to 10 A · CMC 8,3 · 10⁻⁴ · I + 3,5 mA 0,04 kHz to 1 kHz — 10 A to 30 A · CMC 8,3 · 10⁻⁴ · I + 10 mA
AC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
19 Multimeters
Digital ammeters
Calibration — AC current
0,045 kHz to 1 kHz — 20 µA to 200 µA · CMC 6,3 · 10⁻⁴ · I + 0,17 µA 0,045 kHz to 1 kHz — 0,2 mA to 2 mA · CMC 5,8 · 10⁻⁴ · I + 0,24 µA 0,045 kHz to 1 kHz — 2 mA to 20 mA · CMC 4,3 · 10⁻⁴ · I + 2,4 µA 0,045 kHz to 1 kHz — 20 mA to 200 mA · CMC 4,3 · 10⁻⁴ · I + 24 µA 0,045 kHz to 1 kHz — 0,2 A to 2 A · CMC 4,9 · 10⁻⁴ · I + 0,24 mA 45 Hz to 100 Hz — 2 A to 30 A · CMC 4,9 · 10⁻⁴ · I + 2,4 mA
AC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.1 based on EURAMET cg-15 v.3.0
20 Electrical safety testers Calibration — AC current
0,045 kHz to 1 kHz — 20 µA to 200 µA · CMC 6,3 · 10⁻⁴ · I + 0,17 µA 0,045 kHz to 1 kHz — 0,2 mA to 2 mA · CMC 5,8 · 10⁻⁴ · I + 0,24 µA 0,045 kHz to 1 kHz — 2 mA to 20 mA · CMC 4,3 · 10⁻⁴ · I + 2,4 µA 0,045 kHz to 1 kHz — 20 mA to 200 mA · CMC 4,3 · 10⁻⁴ · I + 24 µA 0,045 kHz to 1 kHz — 0,2 A to 2 A · CMC 4,9 · 10⁻⁴ · I + 0,24 mA 45 Hz to 100 Hz — 2 A to 30 A · CMC 4,9 · 10⁻⁴ · I + 2,4 mA
AC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
21 Power network parameter meters (measurement of RCD trip test current) Calibration — AC current
50 Hz to 60 Hz — 2 mA to 10 mA · CMC 1,4 · 10⁻² · I + 0,093 mA 50 Hz to 60 Hz — 10 mA to 30 mA · CMC 1,4 · 10⁻² · I + 0,09 mA 50 Hz to 60 Hz — 30 mA to 100 mA · CMC 1,4 · 10⁻² · I + 0, 09 mA 50 Hz to 60 Hz — 100 mA to 300 mA · CMC 1,4 · 10⁻² · I + 0, 09 mA 50 Hz to 60 Hz — 300 mA to 1 A · CMC 1,4 · 10⁻² · I + 0, 09 mA 50 Hz to 60 Hz — 1 A to 3 A · CMC 1,4 · 10⁻² · I + 0, 09 mA
AC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
22 Resistance calibrators Calibration — DC resistance
0 Ω to 1 Ω · CMC 1,7 · 10⁻⁵ · R + 6,9 µΩ 1 Ω to 10 Ω · CMC 1,3 · 10⁻⁵ · R + 35 µΩ 10 Ω to 100 Ω · CMC 1,1 · 10⁻⁵ · R + 120 µΩ 0,1 kΩ to 1 kΩ · CMC 9,4 · 10⁻⁶ · R + 930 µΩ 1 kΩ to 10 kΩ · CMC 1,1 · 10⁻⁵ · R + 9,3 mΩ 10 kΩ to 100 kΩ · CMC 1,2 · 10⁻⁵ · R + 93 mΩ 0,1 MΩ to 1 MΩ · CMC 1,4 · 10⁻⁵ · R + 2,3 Ω 1 MΩ to 10 MΩ · CMC 2,2 · 10⁻⁵ · R + 92 Ω 10 MΩ to 100 MΩ · CMC 5,8 · 10⁻⁴ · R + 230 Ω 0,1 GΩ to 1 GΩ · CMC 5,8 · 10⁻³ · R + 9,2 kΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.2
23 Multimeters
Digital ohmmeters
Calibration — DC resistance
0,1 Ω · CMC 5,8 mΩ 1 Ω · CMC 5,9 mΩ 10 Ω · CMC 6,8 mΩ 100 Ω · CMC 11 mΩ 1 kΩ · CMC 87 mΩ 10 kΩ · CMC 0,87 Ω 100 kΩ · CMC 8,1 Ω 1 MΩ · CMC 0,15 kΩ 10 MΩ · CMC 3,9 kΩ 100 MΩ · CMC 0,47 MΩ 1 GΩ · CMC 10 MΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.1 based on EURAMET cg-15 v.3.0
24 Electrical safety testers
Insulation resistance meters
Calibration — DC resistance
0 Ω to 9,999 Ω · CMC 2,9 · 10⁻³ · R + 12 mΩ 10 Ω to 99,99 Ω · CMC 2,9 · 10⁻³ · R + 14 mΩ 100 Ω to 999,9 Ω · CMC 2,9 · 10⁻³ · R + 140 mΩ 1 kΩ to 4,999 kΩ · CMC 2,9 · 10⁻³ · R + 1,4 Ω 5 kΩ to 50 kΩ · CMC 2,9 · 10⁻³ · R + 14 Ω
DC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
25 Power network parameter meters Calibration — DC resistance
0 Ω to 9,999 Ω · CMC 2,9 · 10⁻³ · R + 12 mΩ 10 Ω to 99,99 Ω · CMC 2,9 · 10⁻³ · R + 14 mΩ 100 Ω to 999,9 Ω · CMC 2,9 · 10⁻³ · R + 140 mΩ 1 kΩ to 4,999 kΩ · CMC 2,9 · 10⁻³ · R + 1,4 Ω 5 kΩ to 50 kΩ · CMC 2,9 · 10⁻³ · R + 14 Ω 0,05 Ω · CMC 5 mΩ 0,1 Ω · CMC 5,4 mΩ 0,2 Ω · CMC 5,8 mΩ 0,3 Ω · CMC 6,4 mΩ 0,4 Ω · CMC 7,1 mΩ 0,5 Ω · CMC 7,9 mΩ 1Ω · CMC 11 mΩ 5Ω · CMC 34 mΩ 9Ω · CMC 57 mΩ 90 Ω · CMC 520 mΩ 900 Ω · CMC 5,2 Ω 0,01 MΩ to 1 MΩ · CMC 1,2 · 10⁻³ · R + 8,2 kΩ 1 MΩ to 10 MΩ · CMC 1,3 · 10⁻³ · R + 0,058 kΩ 10 MΩ to 100 MΩ · CMC 1,2 · 10⁻² · R + 0,58 kΩ 100 MΩ to 1 GΩ · CMC 1,2 · 10⁻² · R + 5,8 kΩ 1 GΩ to 10 GΩ · CMC 1,2 · 10⁻² · R + 0,058 MΩ 10 GΩ to 100 GΩ · CMC 1,2 · 10⁻² · R + 0,58 MΩ 100 GΩ to 1,1 TΩ · CMC 3,6 · 10⁻² · R + 5,8 MΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
26 Electrical safety testers
Earth continuity testers
Calibration — AC resistance
50 Hz to 150 Hz — 0,05 Ω · CMC 5 mΩ 50 Hz to 150 Hz — 0,1 Ω · CMC 5,4 mΩ 50 Hz to 150 Hz — 0,2 Ω · CMC 5,8 mΩ 50 Hz to 150 Hz — 0,3 Ω · CMC 6,4 mΩ 50 Hz to 150 Hz — 0,4 Ω · CMC 7,1 mΩ 50 Hz to 150 Hz — 0,5 Ω · CMC 7,9 mΩ 50 Hz to 150 Hz — 1Ω · CMC 11 mΩ 50 Hz to 150 Hz — 5Ω · CMC 34 mΩ 50 Hz to 150 Hz — 9Ω · CMC 57 mΩ 50 Hz to 150 Hz — 90 Ω · CMC 520 mΩ 50 Hz to 150 Hz — 900 Ω · CMC 5,2 Ω 50 Hz to 150 Hz — 0 Ω to 9,999 Ω · CMC 2,9 · 10⁻³ · R + 12 mΩ 50 Hz to 150 Hz — 10 Ω to 99,99 Ω · CMC 2,9 · 10⁻³ · R + 14 mΩ 50 Hz to 150 Hz — 100 Ω to 999,9 Ω · CMC 2,9 · 10⁻³ · R + 140 mΩ 50 Hz to 150 Hz — 1 kΩ to 4,999 kΩ · CMC 2,9 · 10⁻³ · R + 1,4 Ω 50 Hz to 150 Hz — 5 kΩ to 50 kΩ · CMC 2,9 · 10⁻³ · R + 14 Ω
AC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.3
27 Power network parameter meters Calibration — AC resistance
50 Hz to 150 Hz — 0 Ω to 9,999 Ω · CMC 2,9 · 10⁻³ · R + 12 mΩ 50 Hz to 150 Hz — 10 Ω to 99,99 Ω · CMC 2,9 · 10⁻³ · R + 14 mΩ 50 Hz to 150 Hz — 100 Ω to 999,9 Ω · CMC 2,9 · 10⁻³ · R + 140 mΩ 50 Hz to 150 Hz — 1 kΩ to 4,999 kΩ · CMC 2,9 · 10⁻³ · R + 1,4 Ω 50 Hz to 150 Hz — 5 kΩ to 50 kΩ · CMC 2,9 · 10⁻³ · R + 14 Ω 50 Hz to 150 Hz — 0,1 Ω · CMC 5,8 mΩ 50 Hz to 150 Hz — 1Ω · CMC 5,9 mΩ 50 Hz to 150 Hz — 10 Ω · CMC 6,8 mΩ 50 Hz to 150 Hz — 100Ω · CMC 11 mΩ 50 Hz to 150 Hz — 1 kΩ · CMC 87 mΩ 50 Hz to 150 Hz — 10 kΩ · CMC 0,87 Ω
AC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
28 Power network parameter meters Calibration — Impedance
0,05 Ω · CMC 5,9 · 10⁻³ · R + 4,8 mΩ 0,1 Ω · CMC 5,9 · 10⁻³ · R + 4,8 mΩ 0,22 Ω · CMC 5,8 · 10⁻³ · R + 4,8 mΩ 0,33 Ω · CMC 5,9 · 10⁻³ · R + 4,8 mΩ 0,5 Ω · CMC 5,8 · 10⁻³ · R + 4,8 mΩ 1Ω · CMC 5,8 · 10⁻³ · R + 4,9 mΩ 10 Ω · CMC 5,8 · 10⁻³ · R + 7,6 mΩ 100 Ω · CMC 5,8 · 10⁻³ · R + 19 mΩ 1000 Ω · CMC 5,8 · 10⁻³ · R + 36 mΩ
Impedance
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
29 Electrical safety testers Calibration — AC power
100 VA to 2300 VA · CMC 3,9 · 10⁻³ · P + 0,58 VA
AC power
Internal procedure; Indirect method; At permanent location (S) Internal procedure PW.3
30 Power network parameter meters Calibration — Time (time interval)
20 ms to 5 s · CMC 5,6 · 10⁻⁴ · t + 1,1 ms
Time (time interval)
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.5
31 Calibrators Calibration — Frequency
1 Hz to 100 Hz · CMC 2,6 · 10⁻⁶ · f + 62 µHz 100 Hz to 1000 Hz · CMC 2,6 · 10⁻⁶ · f + 0,62 mHz 1 kHz to 10 kHz · CMC 2,6 · 10⁻⁶ · f + 6,2 mHz 0,01 MHz to 1 MHz · CMC 2,6 · 10⁻⁶ · f + 6,2 Hz
Frequency
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.2
32 Multimeters Calibration — Frequency
1 Hz to 1 kHz · CMC 1,2 · 10⁻⁶ · f + 0,0000058 Hz 1 kHz to 10 kHz · CMC 1,2 · 10⁻⁶ · f + 0,00058 Hz 10 kHz to 10 MHz · CMC 1,2 · 10⁻⁶ · f + 0,058 Hz
Frequency
Internal procedure; Direct method; At permanent location (S) Internal procedure PW.1 based on EURAMET cg-15 v.3.0
33 Electronic pressure gauges Calibration — Pressure
0 to 200 kPa · CMC 1,3 · 10⁻⁴ · p + 0,033 kPa 0 to 2000 kPa · CMC 1,3 · 10⁻⁴ · p + 0,66 kPa
Pressure
Internal procedure; Method 5.2.1 and 5.2.2. Direct method; At permanent location (S) Internal procedure PW.4 based on EURAMET cg-17 v.4.1

Reference standards (7)

EURAMET cg-15 v.3.0
EURAMET cg-17 v.4.1
PW.1
PW.2
PW.3
PW.4
PW.5

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20231