🇵🇱 Politechnika Rzeszowska im. Ignacego Łukasiewicza PL
Metrology & calibration
Politechnika Rzeszowska im. Ignacego Łukasiewicza is a calibration laboratory accredited by PCA under number AP 226 (Rzeszów, Poland). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Data acquisition cards, Multimeters, Calibrators, Digital ammeters. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S).
At a glance
- Accreditation number: PCA AP 226
- Calibration items in scope: 5
- Types of calibrated items: 8
- Calibration principles used: 1
- City: Rzeszów, Poland
- Valid until: 12-03-2029
- Accreditation number
- AP 226
- Accreditation body
- PCA
- Address
- ul. Wincentego Pola 2, 35-021 Rzeszów
- City
- Rzeszów, Poland
- Division
- LABBiKAL Testing and Calibration Laboratory
- Valid until
- 12-03-2029
What this laboratory calibrates (8)
- Data acquisition cards — 5 calibration items
- Multimeters — 5 calibration items
- Calibrators — 4 calibration items
- Digital ammeters — 2 calibration items
- Digital voltmeters — 2 calibration items
- Power supplies — 2 calibration items
- Digital ohmmeters — 1 calibration item
- Resistance calibrators — 1 calibration item
Calibration principles (1)
- Internal procedure; Direct method; At permanent location (S) — 5 calibration items
Certificate information
- Issue date
- 27.01.2026
- Issue no.
- 2
- Certificate date
- 13.03.2025
- Accreditation cycle
- 13.03.2025 — 12.03.2029
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (5)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Calibrators Data acquisition cards Digital voltmeters Multimeters Power supplies |
Calibration — DC voltage 0 mV to 100 mV · CMC 15 · 10⁻⁶ · U + 0,2 µV 0,1 V to 1 V · CMC 5,8 · 10⁻⁶ · U + 0,3 µV 1 V to 10 V · CMC 5,8 · 10⁻⁶ · U + 0,5 µV 10 V to 100 V · CMC 8,5 · 10⁻⁶ · U + 30 µV 100 V to 1000 V · CMC 8,9 · 10⁻⁶ · U + 0,5 mV DC voltage |
Internal procedure; Direct method; At permanent location (S) | PW-01-1, PW-01-2, PW-01-3, PW-01-4 based on EURAMET cg-15 v.3.0 | |
| 2 | Calibrators Data acquisition cards Digital ammeters Multimeters Power supplies |
Calibration — DC current 0 µA to 10 µA · CMC 40 · 10⁻⁶ · I + 0,4 nA 10 µA to 100 µA · CMC 15 · 10⁻⁶ · I + 0,4 nA 0,1 mA to 1 mA · CMC 14 · 10⁻⁶ · I + 4 nA 1 mA to 10 mA · CMC 20 · 10⁻⁶ · I + 40 nA 10 mA to 100 mA · CMC 86 · 10⁻⁶ · I + 1 µA 0,1 A to 1 A · CMC 0,2 · 10⁻³ · I + 0,1 mA 1 A to 10 A · CMC 0,35 · 10⁻³ · I + 0,4 mA 10 A to 20 A · CMC 0,83 · 10⁻³ · I + 4,4 mA DC current |
Internal procedure; Direct method; At permanent location (S) | PW-01-1, PW-01-2, PW-01-3, PW-01-4 based on EURAMET cg-15 v.3.0 | |
| 3 | Calibrators Data acquisition cards Digital voltmeters Multimeters |
Calibration — AC voltage 10 Hz to 2 kHz — 1 mV to 10 mV · CMC 0,33 · 10⁻³ · U + 1,1 µV 2 kHz to 10 kHz — 1 mV to 10 mV · CMC 0,48 · 10⁻³ · U + 1,1 µV 10 kHz to 30 kHz — 1 mV to 10 mV · CMC 0,49 · 10⁻³ · U + 1,1 µV 30 kHz to 100 kHz — 1 mV to 10 mV · CMC 3,5 · 10⁻³ · U + 1,1 µV 10 Hz to 2 kHz — 10 mV to 100 mV · CMC 0,13 · 10⁻³ · U + 0,5 µV 2 kHz to 10 kHz — 10 mV to 100 mV · CMC 0,2 · 10⁻³ · U + 0,5 µV 10 kHz to 30 kHz — 10 mV to 100 mV · CMC 0,38 · 10⁻³ · U + 1 µV 30 kHz to 100 kHz — 10 mV to 100 mV · CMC 0,92 · 10⁻³ · U + 5 µV 10 Hz to 2 kHz — 0,1 V to 1 V · CMC 0,12 · 10⁻³ · U + 5 µV 2 kHz to 10 kHz — 0,1 V to 1 V · CMC 0,19 · 10⁻³ · U + 5 µV 10 kHz to 30 kHz — 0,1 V to 1 V · CMC 0,38 · 10⁻³ · U + 10 µV 30 kHz to 100 kHz — 0,1 V to 1 V · CMC 0,92 · 10⁻³ · U + 50 µV 10 Hz to 2 kHz — 1 V to 10 V · CMC 0,12 · 10⁻³ · U + 0,05 mV 2 kHz to 10 kHz — 1 V to 10 V · CMC 0,19 · 10⁻³ · U + 0,05 mV 10 kHz to 30 kHz — 1 V to 10 V · CMC 0,38 · 10⁻³ · U + 0,1 mV 30 kHz to 100 kHz — 1 V to 10 V · CMC 0,92 · 10⁻³ · U + 0,5 mV 10 Hz to 2 kHz — 10 V to 33 V · CMC 0,13 · 10⁻³ · U + 0,5 mV 2 kHz to 10 kHz — 10 V to 33 V · CMC 0,16 · 10⁻³ · U + 0,5 mV 10 kHz to 30 kHz — 10 V to 33 V · CMC 0,38 · 10⁻³ · U + 1 mV 30 kHz to 100 kHz — 10 V to 33 V · CMC 0,98 · 10⁻³ · U + 5 mV 45 Hz to 2 kHz — 33 V to 100 V · CMC 0,13 · 10⁻³ · U + 0,5 mV 2 kHz to 10 kHz — 33 V to 100 V · CMC 0,16 · 10⁻³ · U + 0,5 mV 10 kHz to 30 kHz — 33 V to 100 V · CMC 0,38 · 10⁻³ · U + 1 mV 30 kHz to 100 kHz — 33 V to 100 V · CMC 0,98 · 10⁻³ · U + 5 mV 45 Hz to 2 kHz — 100 V to 330 V · CMC 0,16 · 10⁻³ · U + 25 mV 2 kHz to 10 kHz — 100 V to 330 V · CMC 0,16 · 10⁻³ · U + 25 mV 10 kHz to 30 kHz — 100 V to 330 V · CMC 0,38 · 10⁻³ · U + 25 mV 30 kHz to 100 kHz — 100 V to 330 V · CMC 0,98 · 10⁻³ · U + 0,1 V 45 Hz to 2 kHz — 330 V to 1000 V · CMC 0,16 · 10⁻³ · U + 25 mV 2 kHz to 10 kHz — 330 V to 1000 V · CMC 0,16 · 10⁻³ · U + 25 mV AC voltage |
Internal procedure; Direct method; At permanent location (S) | PW-01-1, PW-01-2, PW-01-4 based on EURAMET cg-15 v.3.0 | |
| 4 | Calibrators Data acquisition cards Digital ammeters Multimeters |
Calibration — AC current 10 Hz to 2 kHz — 30 µA to 100 µA · CMC 0,39 · 10⁻³ · I + 5 nA 2 kHz to 10 kHz — 30 µA to 100 µA · CMC 0,75 · 10⁻³ · I + 5 nA 10 Hz to 2 kHz — 0,1 mA to 1 mA · CMC 0,39 · 10⁻³ · I + 50 nA 2 kHz to 10 kHz — 0,1 mA to 1 mA · CMC 0,75 · 10⁻³ · I + 50 nA 10 Hz to 2 kHz — 1 mA to 10 mA · CMC 0,39 · 10⁻³ · I + 0,5 µA 2 kHz to 10 kHz — 1 mA to 10 mA · CMC 0,75 · 10⁻³ · I + 0,5 µA 10 Hz to 2 kHz — 10 mA to 100 mA · CMC 0,38 · 10⁻³ · I + 5 µA 2 kHz to 10 kHz — 10 mA to 100 mA · CMC 0,74 · 10⁻³ · I + 5 µA 10 Hz to 2 kHz — 0,1 A to 1 A · CMC 0,4 · 10⁻³ · I + 0,1 mA 2 kHz to 10 kHz — 0,1 A to 1 A · CMC 0,77 · 10⁻³ · I + 0,1 mA 10 Hz to 2 kHz — 1 A to 3 A · CMC 1,55 · 10⁻³ · I + 0,5 mA 2 kHz to 10 kHz — 1 A to 3 A · CMC 1,41 · 10⁻³ · I + 0,5 mA 45 Hz to 2 kHz — 3 A to 10 A · CMC 1,55 · 10⁻³ · I + 0,5 mA 2 kHz to 5 kHz — 3 A to 10 A · CMC 1,41 · 10⁻³ · I + 0,5 mA 45 Hz to 2 kHz — 10 A to 20 A · CMC 1,26 · 10⁻³ · I + 12 mA 2 kHz to 5 kHz — 10 A to 20 A · CMC 1,5 · 10⁻³ · I + 12 mA AC current |
Internal procedure; Direct method; At permanent location (S) | PW-01-1, PW-01-2, PW-01-4 based on EURAMET cg-15 v.3.0 | |
| 5 | Resistance calibrators Data acquisition cards Digital ohmmeters Multimeters |
Calibration — DC resistance 0 Ω to 100 Ω · CMC 14 · 10⁻⁶ · R + 50 µΩ 0,1 kΩ to 1 kΩ · CMC 14 · 10⁻⁶ · R + 0,5 mΩ 1 kΩ to 10 kΩ · CMC 14 · 10⁻⁶ · R + 5 mΩ 10 kΩ to 100 kΩ · CMC 14 · 10⁻⁶ · R + 50 mΩ 0,1 MΩ to 1 MΩ · CMC 16 · 10⁻⁶ · R + 1 Ω 1 MΩ to 10 MΩ · CMC 28 · 10⁻⁶ · R + 0,1 kΩ 10 MΩ to 100 MΩ · CMC 0,18 · 10⁻³ · R + 10 kΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | PW-01-1, PW-01-2, PW-01-4 based on EURAMET cg-15 v.3.0 | |
Reference standards (5)
EURAMET cg-15 v.3.0
PW-01-1
PW-01-2
PW-01-3
PW-01-4