🇵🇱 Politechnika Rzeszowska im. Ignacego Łukasiewicza PL

PCA AP 226 • Rzeszów, Poland • 5 calibration items

Metrology & calibration

Politechnika Rzeszowska im. Ignacego Łukasiewicza is a calibration laboratory accredited by PCA under number AP 226 (Rzeszów, Poland). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Data acquisition cards, Multimeters, Calibrators, Digital ammeters. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S).

At a glance

  • Accreditation number: PCA AP 226
  • Calibration items in scope: 5
  • Types of calibrated items: 8
  • Calibration principles used: 1
  • City: Rzeszów, Poland
  • Valid until: 12-03-2029
Accreditation number
AP 226
Accreditation body
PCA
Address
ul. Wincentego Pola 2, 35-021 Rzeszów
City
Rzeszów, Poland
Division
LABBiKAL Testing and Calibration Laboratory
Valid until
12-03-2029

What this laboratory calibrates (8)

Calibration principles (1)

Certificate information

Issue date
27.01.2026
Issue no.
2
Certificate date
13.03.2025
Accreditation cycle
13.03.2025 — 12.03.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (5)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Calibrators
Data acquisition cards
Digital voltmeters
Multimeters
Power supplies
Calibration — DC voltage
0 mV to 100 mV · CMC 15 · 10⁻⁶ · U + 0,2 µV 0,1 V to 1 V · CMC 5,8 · 10⁻⁶ · U + 0,3 µV 1 V to 10 V · CMC 5,8 · 10⁻⁶ · U + 0,5 µV 10 V to 100 V · CMC 8,5 · 10⁻⁶ · U + 30 µV 100 V to 1000 V · CMC 8,9 · 10⁻⁶ · U + 0,5 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) PW-01-1, PW-01-2, PW-01-3, PW-01-4 based on EURAMET cg-15 v.3.0
2 Calibrators
Data acquisition cards
Digital ammeters
Multimeters
Power supplies
Calibration — DC current
0 µA to 10 µA · CMC 40 · 10⁻⁶ · I + 0,4 nA 10 µA to 100 µA · CMC 15 · 10⁻⁶ · I + 0,4 nA 0,1 mA to 1 mA · CMC 14 · 10⁻⁶ · I + 4 nA 1 mA to 10 mA · CMC 20 · 10⁻⁶ · I + 40 nA 10 mA to 100 mA · CMC 86 · 10⁻⁶ · I + 1 µA 0,1 A to 1 A · CMC 0,2 · 10⁻³ · I + 0,1 mA 1 A to 10 A · CMC 0,35 · 10⁻³ · I + 0,4 mA 10 A to 20 A · CMC 0,83 · 10⁻³ · I + 4,4 mA
DC current
Internal procedure; Direct method; At permanent location (S) PW-01-1, PW-01-2, PW-01-3, PW-01-4 based on EURAMET cg-15 v.3.0
3 Calibrators
Data acquisition cards
Digital voltmeters
Multimeters
Calibration — AC voltage
10 Hz to 2 kHz — 1 mV to 10 mV · CMC 0,33 · 10⁻³ · U + 1,1 µV 2 kHz to 10 kHz — 1 mV to 10 mV · CMC 0,48 · 10⁻³ · U + 1,1 µV 10 kHz to 30 kHz — 1 mV to 10 mV · CMC 0,49 · 10⁻³ · U + 1,1 µV 30 kHz to 100 kHz — 1 mV to 10 mV · CMC 3,5 · 10⁻³ · U + 1,1 µV 10 Hz to 2 kHz — 10 mV to 100 mV · CMC 0,13 · 10⁻³ · U + 0,5 µV 2 kHz to 10 kHz — 10 mV to 100 mV · CMC 0,2 · 10⁻³ · U + 0,5 µV 10 kHz to 30 kHz — 10 mV to 100 mV · CMC 0,38 · 10⁻³ · U + 1 µV 30 kHz to 100 kHz — 10 mV to 100 mV · CMC 0,92 · 10⁻³ · U + 5 µV 10 Hz to 2 kHz — 0,1 V to 1 V · CMC 0,12 · 10⁻³ · U + 5 µV 2 kHz to 10 kHz — 0,1 V to 1 V · CMC 0,19 · 10⁻³ · U + 5 µV 10 kHz to 30 kHz — 0,1 V to 1 V · CMC 0,38 · 10⁻³ · U + 10 µV 30 kHz to 100 kHz — 0,1 V to 1 V · CMC 0,92 · 10⁻³ · U + 50 µV 10 Hz to 2 kHz — 1 V to 10 V · CMC 0,12 · 10⁻³ · U + 0,05 mV 2 kHz to 10 kHz — 1 V to 10 V · CMC 0,19 · 10⁻³ · U + 0,05 mV 10 kHz to 30 kHz — 1 V to 10 V · CMC 0,38 · 10⁻³ · U + 0,1 mV 30 kHz to 100 kHz — 1 V to 10 V · CMC 0,92 · 10⁻³ · U + 0,5 mV 10 Hz to 2 kHz — 10 V to 33 V · CMC 0,13 · 10⁻³ · U + 0,5 mV 2 kHz to 10 kHz — 10 V to 33 V · CMC 0,16 · 10⁻³ · U + 0,5 mV 10 kHz to 30 kHz — 10 V to 33 V · CMC 0,38 · 10⁻³ · U + 1 mV 30 kHz to 100 kHz — 10 V to 33 V · CMC 0,98 · 10⁻³ · U + 5 mV 45 Hz to 2 kHz — 33 V to 100 V · CMC 0,13 · 10⁻³ · U + 0,5 mV 2 kHz to 10 kHz — 33 V to 100 V · CMC 0,16 · 10⁻³ · U + 0,5 mV 10 kHz to 30 kHz — 33 V to 100 V · CMC 0,38 · 10⁻³ · U + 1 mV 30 kHz to 100 kHz — 33 V to 100 V · CMC 0,98 · 10⁻³ · U + 5 mV 45 Hz to 2 kHz — 100 V to 330 V · CMC 0,16 · 10⁻³ · U + 25 mV 2 kHz to 10 kHz — 100 V to 330 V · CMC 0,16 · 10⁻³ · U + 25 mV 10 kHz to 30 kHz — 100 V to 330 V · CMC 0,38 · 10⁻³ · U + 25 mV 30 kHz to 100 kHz — 100 V to 330 V · CMC 0,98 · 10⁻³ · U + 0,1 V 45 Hz to 2 kHz — 330 V to 1000 V · CMC 0,16 · 10⁻³ · U + 25 mV 2 kHz to 10 kHz — 330 V to 1000 V · CMC 0,16 · 10⁻³ · U + 25 mV
AC voltage
Internal procedure; Direct method; At permanent location (S) PW-01-1, PW-01-2, PW-01-4 based on EURAMET cg-15 v.3.0
4 Calibrators
Data acquisition cards
Digital ammeters
Multimeters
Calibration — AC current
10 Hz to 2 kHz — 30 µA to 100 µA · CMC 0,39 · 10⁻³ · I + 5 nA 2 kHz to 10 kHz — 30 µA to 100 µA · CMC 0,75 · 10⁻³ · I + 5 nA 10 Hz to 2 kHz — 0,1 mA to 1 mA · CMC 0,39 · 10⁻³ · I + 50 nA 2 kHz to 10 kHz — 0,1 mA to 1 mA · CMC 0,75 · 10⁻³ · I + 50 nA 10 Hz to 2 kHz — 1 mA to 10 mA · CMC 0,39 · 10⁻³ · I + 0,5 µA 2 kHz to 10 kHz — 1 mA to 10 mA · CMC 0,75 · 10⁻³ · I + 0,5 µA 10 Hz to 2 kHz — 10 mA to 100 mA · CMC 0,38 · 10⁻³ · I + 5 µA 2 kHz to 10 kHz — 10 mA to 100 mA · CMC 0,74 · 10⁻³ · I + 5 µA 10 Hz to 2 kHz — 0,1 A to 1 A · CMC 0,4 · 10⁻³ · I + 0,1 mA 2 kHz to 10 kHz — 0,1 A to 1 A · CMC 0,77 · 10⁻³ · I + 0,1 mA 10 Hz to 2 kHz — 1 A to 3 A · CMC 1,55 · 10⁻³ · I + 0,5 mA 2 kHz to 10 kHz — 1 A to 3 A · CMC 1,41 · 10⁻³ · I + 0,5 mA 45 Hz to 2 kHz — 3 A to 10 A · CMC 1,55 · 10⁻³ · I + 0,5 mA 2 kHz to 5 kHz — 3 A to 10 A · CMC 1,41 · 10⁻³ · I + 0,5 mA 45 Hz to 2 kHz — 10 A to 20 A · CMC 1,26 · 10⁻³ · I + 12 mA 2 kHz to 5 kHz — 10 A to 20 A · CMC 1,5 · 10⁻³ · I + 12 mA
AC current
Internal procedure; Direct method; At permanent location (S) PW-01-1, PW-01-2, PW-01-4 based on EURAMET cg-15 v.3.0
5 Resistance calibrators
Data acquisition cards
Digital ohmmeters
Multimeters
Calibration — DC resistance
0 Ω to 100 Ω · CMC 14 · 10⁻⁶ · R + 50 µΩ 0,1 kΩ to 1 kΩ · CMC 14 · 10⁻⁶ · R + 0,5 mΩ 1 kΩ to 10 kΩ · CMC 14 · 10⁻⁶ · R + 5 mΩ 10 kΩ to 100 kΩ · CMC 14 · 10⁻⁶ · R + 50 mΩ 0,1 MΩ to 1 MΩ · CMC 16 · 10⁻⁶ · R + 1 Ω 1 MΩ to 10 MΩ · CMC 28 · 10⁻⁶ · R + 0,1 kΩ 10 MΩ to 100 MΩ · CMC 0,18 · 10⁻³ · R + 10 kΩ
DC resistance
Internal procedure; Direct method; At permanent location (S) PW-01-1, PW-01-2, PW-01-4 based on EURAMET cg-15 v.3.0

Reference standards (5)

EURAMET cg-15 v.3.0
PW-01-1
PW-01-2
PW-01-3
PW-01-4

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20226