🇵🇱 Przedsiębiorstwo Usługowo-Handlowe - EnergoSystem - inż. Zbigniew Szczyrbak PL
Metrology & calibration
Przedsiębiorstwo Usługowo-Handlowe - EnergoSystem - inż. Zbigniew Szczyrbak is a calibration laboratory accredited by PCA under number AP 223 (Skawina, Poland). Its accreditation scope covers 8 calibration items in the following areas: Metrology & calibration. Calibrated items include: Temperature indicators (meters), including temperature controllers, Temperature simulators, Base metal thermocouples type J, K, N, Electrical thermometers (including electronic). Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S), Internal procedure; Electric indirect method; At permanent location…, Internal procedure; Electric indirect method; At permanent location….
At a glance
- Accreditation number: PCA AP 223
- Calibration items in scope: 8
- Types of calibrated items: 9
- Calibration principles used: 4
- City: Skawina, Poland
- Valid until: 21-01-2029
- Accreditation number
- AP 223
- Accreditation body
- PCA
- Address
- ul. Piłsudskiego 3A, 32-050 Skawina
- City
- Skawina, Poland
- Division
- Laboratorium Pomiarowe EnergoSystem
- Valid until
- 21-01-2029
What this laboratory calibrates (9)
- Temperature indicators (meters), including temperature controllers — 2 calibration items
- Temperature simulators — 2 calibration items
- Base metal thermocouples type J, K, N — 1 calibration item
- Electrical thermometers (including electronic) — 1 calibration item
- Furnaces — 1 calibration item
- Noble metal thermocouples type S, R — 1 calibration item
- Temperature indicators (meters) compatible with resistance sensors — 1 calibration item
- Temperature indicators (meters) compatible with thermoelectric sensors — 1 calibration item
- Thermostatic chambers — 1 calibration item
Calibration principles (4)
- Internal procedure; Direct method; At permanent location (S) — 3 calibration items
- Internal procedure; Electric indirect method; At permanent location (S) — 2 calibration items
- Internal procedure; Electric indirect method; At permanent location (S) and outside of it (P) — 2 calibration items
- Internal procedure; Direct method; Outside of permanent location (P) — 1 calibration item
Certificate information
- Issue date
- 22.01.2025
- Issue no.
- 1
- Certificate date
- 22.01.2025
- Accreditation cycle
- 22.01.2025 — 21.01.2029
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (8)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Temperature indicators (meters), including temperature controllers Temperature indicators (meters) compatible with thermoelectric sensors |
Calibration — Electric quantity simulation -270 °C to 1820 °C · CMC 1,0 °C Electric quantity simulation |
Internal procedure; Electric indirect method; At permanent location (S) and outside of it (P) | IL-7-04 based on EURAMET cg-11 v 2.0 | |
| 2 | Temperature simulators | Calibration — Electric quantity simulation -270 °C to 1820 °C · CMC 1,0 °C Electric quantity simulation |
Internal procedure; Electric indirect method; At permanent location (S) | IL-7-04 based on EURAMET cg-11 v 2.0 | |
| 3 | Temperature indicators (meters), including temperature controllers Temperature indicators (meters) compatible with resistance sensors |
Calibration — Electric quantity simulation -200 °C to 850 °C · CMC 1,0 °C Electric quantity simulation |
Internal procedure; Electric indirect method; At permanent location (S) and outside of it (P) | IL-7-04 based on EURAMET cg-11 v 2.0 | |
| 4 | Temperature simulators | Calibration — Electric quantity simulation -200 °C to 850 °C · CMC 1,0 °C Electric quantity simulation |
Internal procedure; Electric indirect method; At permanent location (S) | IL-7-04 based on EURAMET cg-11 v 2.0 | |
| 5 | Noble metal thermocouples type S, R | Calibration — Temperature (electrical thermometry) 25 °C to 1100 °C · CMC 2,0 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | IL-7-01 based on EURAMET cg-8 v 3.1 | |
| 6 | Base metal thermocouples type J, K, N | Calibration — Temperature (electrical thermometry) 25 °C to 1100 °C · CMC 2,0 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | IL-7-01 based on EURAMET cg-8 v 3.1 | |
| 7 | Electrical thermometers (including electronic) | Calibration — Temperature (electrical thermometry) 0 °C · CMC 1,5 °C 25 °C to 200 °C · CMC 1,5 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | IL-7-02 based on EURAMET cg-8 v 3.1 DKD-R 5-1 (11/2023) | |
| 8 | Thermostatic chambers Furnaces |
Calibration — Temperature (electrical thermometry) 30 °C to 330 °C · CMC 2,4 °C 330 °C to 500 °C · CMC 3,3 °C 500 °C to 850 °C · CMC 4,6 °C 850 °C to 1000 °C · CMC 5,2 °C 1000 °C to 1200 °C · CMC 6,0 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; Outside of permanent location (P) | IL-7-09 based on EURAMET cg-20 v 5.0 | |
Reference standards (8)
DKD-R 5-1 (11/2023)
EURAMET cg-11 v 2.0
EURAMET cg-20 v 5.0
EURAMET cg-8 v 3.1
IL-7-01
IL-7-02
IL-7-04
IL-7-09