🇵🇱 SFERATIC Sp. z o.o. PL
Metrology & calibration
SFERATIC Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 180 (Gdynia, Poland). Its accreditation scope covers 56 calibration items in the following areas: Metrology & calibration. Calibrated items include: Analogue dial gauges with a scale interval of:, Digital dial gauges with a resolution of:, Ultrasonic thickness gauges, 90° squares. Most frequently used calibration principles: Internal procedure; Direct method using gauge blocks; At permanent…, Internal procedure; Direct method using a horizontal length measuri…, Internal procedure; At permanent location (S).
At a glance
- Accreditation number: PCA AP 180
- Calibration items in scope: 56
- Types of calibrated items: 57
- Calibration principles used: 20
- City: Gdynia, Poland
- Valid until: 22-01-2030
- Accreditation number
- AP 180
- Accreditation body
- PCA
- Address
- ul. Kadłubowców 4, 81-336 Gdynia
- City
- Gdynia, Poland
- Division
- Calibration Laboratory
- Valid until
- 22-01-2030
What this laboratory calibrates (57)
- Analogue dial gauges with a scale interval of: — 2 calibration items
- Digital dial gauges with a resolution of: — 2 calibration items
- Ultrasonic thickness gauges — 2 calibration items
- 90° squares — 1 calibration item
- Bourdon tube pressure gauges — 1 calibration item
- Brinell hardness tester — 1 calibration item
- Builder's spirit levels — 1 calibration item
- Coating thickness gauges — 1 calibration item
- Cylindrical thread plug gauges — 1 calibration item
- Cylindrical thread ring gauges — 1 calibration item
- Dial bore gauges — 1 calibration item
- Dial calipers for external measurements — 1 calibration item
- Dial calipers for internal measurements — 1 calibration item
- Dial depth gauges — 1 calibration item
- Dial thickness gauges — 1 calibration item
- and 42 more
Calibration principles (20)
- Internal procedure; Direct method using gauge blocks; At permanent location (S) — 15 calibration items
- Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) — 10 calibration items
- Internal procedure; At permanent location (S) — 4 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 3 calibration items
- Internal procedure; Direct method using a micrometer head; At permanent location (S) — 3 calibration items
- Internal procedure; Direct method using angle gauge blocks; At permanent location (S) — 3 calibration items
- At permanent location (S) — 2 calibration items
- group 2 tests; At permanent location (S) — 2 calibration items
- Internal procedure; Comparison method using a rigid rule; At permanent location (S) — 2 calibration items
- Internal procedure; Direct method using an optical dividing head; At permanent location (S) — 2 calibration items
- Comparison method using a two-probe comparator; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in the temperature range 10 °C to 35 °C; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method using a horizontal length measuring machine and a setting ring; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method using a micrometer; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method using a step wedge; At permanent location (S) — 1 calibration item
- and 5 more
Certificate information
- Issue date
- 12.10.2026
- Issue no.
- 20
- Certificate date
- 13.09.2022
- Accreditation cycle
- 22.12.2025 — 22.01.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (56)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Lever-type analogue dial gauges with a scale interval of: | Calibration — Length 0,01 mm; 0,001 mm; 0,002 mm — 0 mm to 3 mm · CMC 3,1 µm Length |
Internal procedure; Direct method using a micrometer head; At permanent location (S) | Internal procedure PP TENSLAB/CZ/5/24 | |
| 2 | Analogue dial gauges with a scale interval of: | Calibration — Length 0,1 mm — 0 mm to 50 mm · CMC 7 µm 0,01 mm — 0 mm to 50 mm · CMC 3 µm 0,001 mm — 0 mm to 50 mm · CMC 3 µm Length |
Internal procedure; Direct method using a micrometer head; At permanent location (S) | Internal procedure PP TENSLAB/CZ/5/24 | |
| 3 | Analogue dial gauges with a scale interval of: | Calibration — Length 0,1 mm — 0 mm to 100 mm · CMC 17 µm 0,01 mm — 0 mm to 100 mm · CMC 3 µm 0,001 mm — 0 mm to 100 mm · CMC 3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PCZ/1/25 | |
| 4 | Digital dial gauges with a resolution of: | Calibration — Length 0,01 mm — 0 mm to 50 mm · CMC 7 µm 0,001 mm — 0 mm to 50 mm · CMC 4 µm Length |
Internal procedure; Direct method using a micrometer head; At permanent location (S) | Internal procedure PP TENSLAB/CZ/5/24 | |
| 5 | Digital dial gauges with a resolution of: | Calibration — Length 0,01 mm — 0 mm to 100 mm · CMC 17 µm 0,001 mm — 0 mm to 100 mm · CMC 3 µm 0,0005 mm — 0 mm to 100 mm · CMC 3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PCZ/1/25 | |
| 6 | Reference foils | Calibration — Length 0 mm to 1,5 mm · CMC 4,3 µm 1,5 mm to 5 mm · CMC 8,0 µm 5 mm to 10 mm · CMC 16 µm 10 mm to 15 mm · CMC 24 µm 15 mm to 20 mm · CMC 32 µm 20 mm to 25 mm · CMC 40 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/WZ.GR/2/25 | |
| 7 | Dial depth gauges | Calibration — Length 0 mm to 100 mm · CMC 3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PCZ/1/25 | |
| 8 | Micrometer depth gauges | Calibration — Length 0 mm to 200 mm · CMC 1 + 10 · L µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/GŁ/3/18 | |
| 9 | Vernier depth gauges | Calibration — Length 0 mm to 150 mm · CMC 20 µm 0 mm to 300 mm · CMC 27 µm 0 mm to 400 mm · CMC 28 µm 0 mm to 500 mm · CMC 30 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/GŁ/3/18 | |
| 10 | Dial thickness gauges | Calibration — Length 0 mm to 100 mm · CMC 3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PCZ/1/25 | |
| 11 | Ultrasonic thickness gauges | Calibration — Length 3 mm to 48 mm · CMC 0,03 mm Length |
Internal procedure; Direct method using a step wedge; At permanent location (S) | Internal procedure PP TENSLAB/UMG/2/19 | |
| 12 | Gauge block comparators | Calibration — Length 0 mm to 100 mm · CMC 30 nm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) | Internal procedure PP TENSLAB/KOMP/1/22 based on EURAMET cg-2 Version 2.1 (02/2023) | |
| 13 | Dial calipers for internal measurements | Calibration — Length 0 mm to 100 mm · CMC 4 µm 100 mm to 150 mm · CMC 5 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PCZ/1/25 | |
| 14 | Dial calipers for external measurements | Calibration — Length 0 mm to 50 mm · CMC 3 µm 50 mm do 150 mm · CMC 4 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PCZ/1/25 | |
| 15 | Internal micrometers | Calibration — Length 5 mm to 50 mm · CMC 3,0 µm 50 mm to 150 mm · CMC 3,2 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/PMwew/1/19 | |
| 16 | Coating thickness gauges | Calibration — Length 0 µm to 300 µm · CMC 2,1 µm 300 µm to 1000 µm · CMC 4,1 µm Length |
Internal procedure; Direct method using reference foils; At permanent location (S) | Internal procedure PP TENSLAB/MGR/2/24 | |
| 17 | External micrometers | Calibration — Length 0 mm to 1000 mm · CMC 1 + 10 · L µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/MIKR.ZEW/5/18 | |
| 18 | Gauge blocks of class (0, 1, 2) | Calibration — Length 0,5 mm to 100 mm · CMC 0,08 + 0,4 · L µm Length |
Comparison method using a two-probe comparator; At permanent location (S) | PN-EN ISO 3650:2000 | |
| 19 | Rigid rules Semi-rigid rules |
Calibration — Length 0 m to 3 m · CMC 0,148 + 0,0022 · L mm Length |
Internal procedure; Comparison method using a rigid rule; At permanent location (S) | Internal procedure PP TENSLAB/PRZ/2/20 | |
| 20 | Measuring tapes | Calibration — Length 0 m to 30 m · CMC 0,148 + 0,022 · L mm Length |
Internal procedure; Comparison method using a rigid rule; At permanent location (S) | Internal procedure PP TENSLAB/PRZ/2/20 | |
| 21 | Special vernier instruments Weld gauges |
Calibration — Length - butt weld height — 0 mm to 15 mm · CMC 0,13 mm - fillet weld height — 0 mm to 20 mm · CMC 0,13 mm - undercut depth — up to 7 mm · CMC 0,07 mm Length |
Internal procedure; Direct method using gauge blocks and measuring rollers; At permanent location (S) | Internal procedure PP TENSLAB/SP/2/18 | |
| 22 | Setting rings | Calibration — Length 2 mm do 8 mm · CMC 2,0 µm 8 mm do 30 mm · CMC 2,1 µm 30 mm do 60 mm · CMC 2,2 µm 60 mm do 100 mm · CMC 2,4 µm 100 mm do 150 mm · CMC 2,7 µm 150 mm to 175 mm · CMC 4,0 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine and a setting ring; At permanent location (S) | Internal procedure PP TENSLAB/PWz/4/25 based on EURAMET cg-6 Version 2.0 (03/2011) | |
| 23 | Cylindrical thread plug gauges | Calibration — Length 0 mm to 40 mm · CMC 4 µm 40 mm to 60 mm · CMC 4,1 µm 60 mm to 100 mm · CMC 4,4 µm 100 mm to 200 mm · CMC 4,8 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/SGT/4/25 | |
| 24 | Cylindrical thread ring gauges | Calibration — Length 2,5 mm to 4 mm · CMC 3,1 µm 4 mm to 80 mm · CMC 3,6 µm 80 mm to 100 mm · CMC 3,7 µm 100 mm to 200 mm · CMC 4,1 µm 200 mm to 225 mm · CMC 4,5 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/SGP/5/25 based on EURAMET cg-10 Version 2.1 (12/2012) | |
| 25 | Taper thread plug gauges | Calibration — Length 3 mm to 100 mm · CMC 4,0 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/SGS/1/20 | |
| 26 | Taper thread ring gauges | Calibration — Length 3 mm to 100 mm · CMC 4,0 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/SGS/1/20 | |
| 27 | Plain ring gauges | Calibration — Length 8 mm do 30 mm · CMC 2,0 µm 30 mm do 60 mm · CMC 2,1 µm 60 mm do 100 mm · CMC 2,6 µm 100 mm do 150 mm · CMC 2,8 µm 150 mm to 175 mm · CMC 3,0 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/WZ.WEW/2/25 | |
| 28 | Plug gauges | Calibration — Length 1 mm to 50 mm · CMC 1,1 µm 50 mm to 100 mm · CMC 1,9 µm 100 mm to 200 mm · CMC 2,5 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/SO/3/25 | |
| 29 | Vernier calipers | Calibration — Length 0 mm do 150 mm · CMC 20 µm 0 mm do 300 mm · CMC 27 µm 0 mm do 400 mm · CMC 28 µm 0 mm do 500 mm · CMC 30 µm 0 mm to 600 mm · CMC 33 µm 0 mm to 1000 mm · CMC 50 µm 0 mm do 1500 mm · CMC 70 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/SUW/4/18 | |
| 30 | Two-point inside micrometers | Calibration — Length 0 mm to 100 mm · CMC 3,6 µm 0 mm to 175 mm · CMC 5,3 µm 0 mm to 275 mm · CMC 6,7 µm 0 mm to 375 mm · CMC 8,5 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/SM/1/18 | |
| 31 | Three-point inside micrometers | Calibration — Length 2 mm to 100 mm · CMC 3,7 µm 100 mm to 200 mm · CMC 4,7 µm Length |
Internal procedure; Direct method using setting rings; At permanent location (S) | Internal procedure PP TENSLAB/PMwew/1/19 | |
| 32 | Dial bore gauges | Calibration — Length 0 mm to 100 mm · CMC 3,5 µm 100 mm to 300 mm · CMC 6,5 µm 300 mm to 500 mm · CMC 10,0 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/Ścz/1/22 | |
| 33 | Feeler gauges | Calibration — Length - leaf type — 0,02 mm to 2 mm · CMC 2,4 µm Length |
Internal procedure; Direct method using a micrometer; At permanent location (S) | Internal procedure PP TENSLAB/SzL/1/23 | |
| 34 | Measuring rollers | Calibration — Length 0 mm to 30 mm · CMC 0,9 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/WP/3/25 | |
| 35 | Setting standards for internal dimensions | Calibration — Length 8 mm do 30 mm · CMC 2,0 µm 30 mm do 60 mm · CMC 2,1 µm 60 mm do 100 mm · CMC 2,6 µm 100 mm do 150 mm · CMC 2,8 µm 150 mm to 175 mm · CMC 3,0 µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/WZ.WEW/2/25 | |
| 36 | Vernier height gauges | Calibration — Length 0 mm do 300 mm · CMC 27 µm 0 mm do 400 mm · CMC 29 µm 0 mm do 500 mm · CMC 30 µm 0 mm to 600 mm · CMC 34 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/WYS.SUW/1/18 | |
| 37 | Setting standards for external dimensions | Calibration — Length 0 mm to 500 mm · CMC 0,5 + 10 · L µm Length |
Internal procedure; Direct method using a horizontal length measuring machine; At permanent location (S) | Internal procedure PP TENSLAB/WZ.ZEW/4/25 | |
| 38 | Builder's spirit levels | Calibration — Angle 0 m to 2,5 m · CMC 0,1° Angle |
Internal procedure; Direct method using an optical dividing head; At permanent location (S) | Internal procedure PP TENSLAB/Poz/1/25 | |
| 39 | Digital levels | Calibration — Angle 4 x 90° · CMC 0,1° Angle |
Internal procedure; Direct method using an optical dividing head; At permanent location (S) | Internal procedure PP TENSLAB/Poz/1/25 | |
| 40 | 90° squares | Calibration — Angle up to 400 mm longer arm length · CMC 7 µm Angle |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/KK/1/18 | |
| 41 | Protractors | Calibration — Angle - marking-out — 4 x 90° · CMC 5’ - universal; - digital — 0° to 360° · CMC 5’ Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/KĄT/1/18 | |
| 42 | Squares | Calibration — Angle 0° to 210° · CMC 10’ Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/KĄT/1/18 | |
| 43 | Weld gauges (bevel angle) | Calibration — Angle 80° to 180° · CMC 0,6° Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure PP TENSLAB/SP/2/18 | |
| 44 | Stylus profilometers | Calibration — Length (surface geometry) Ra: — 0 µm to 0,4 µm · CMC 0,10 µm Ra: — 0 µm to 2,93 µm · CMC 0,15 µm Ra: — 0 µm to 6,5 µm · CMC 0,26 µm Rz: — 0 µm to 1,5 µm · CMC 0,15 µm Rz: — 0 µm to 11,3 µm · CMC 0,63 µm Rz: — 0 µm to 19,81 µm · CMC 1,12 µm Length (surface geometry) |
Internal procedure; Direct method using roughness standards; At permanent location (S) | Internal procedure PP TENSLAB/MCH/2/20 | |
| 45 | Torque wrenches | Calibration — Torque 0,4 Nm to 20 Nm · CMC 1,7 % 20 Nm to 3000 Nm · CMC 0,9 % Torque |
At permanent location (S) | PN EN ISO 6789-1:2017-03 PN EN ISO 6789-2:2017-03 | |
| 46 | Torque screwdrivers | Calibration — Torque 0,4 Nm to 20 Nm · CMC 1,7 % Torque |
At permanent location (S) | PN EN ISO 6789-1:2017-03 PN EN ISO 6789-2:2017-03 | |
| 47 | Brinell hardness tester | Calibration — Hardness Hardness — HBW 2,5/187,5 · CMC 2,0 % Hardness — HBW 10/3000 · CMC 2,0 % Force — 1839 N to 29420 N · CMC 0,22 % Length — 0 mm to 1 mm · CMC 0,6 µm Length — 1 mm to 8 mm · CMC 1,5 µm Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PP TENSLAB/TB/1/24 based on PN-EN ISO 6506-2:2019 | |
| 48 | Rockwell hardness testers | Calibration — Hardness Hardness — HRBW · CMC 1 HRBW Hardness — HRC · CMC 0,8 HRC Force — 98,07 N to 1471 N · CMC 0,22 % Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PP TENSLAB/TR/2/24 based on PN-EN ISO 6508- 2:2024 | |
| 49 | Vickers hardness testers | Calibration — Hardness Hardness — 200 HV30 to 800 HV30 · CMC 2 % Hardness — 200 HV10 to 800 HV10 · CMC 2 % Hardness — 200 HV5 to 800 HV5 · CMC 2 % Force — 49,03 N to 294,2 N · CMC 0,22 % Length — 0 mm to 1 mm · CMC 0,6 µm Hardness |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PP TENSLAB/TV/1/22 based on PN-EN ISO 6507-2:2018 | |
| 50 | Thermohygrometers Hygrometers |
Calibration — Relative humidity 28 %rh to 78 %rh · CMC 2,1 %rh Relative humidity |
Internal procedure; Calibration in the temperature range 10 °C to 35 °C; At permanent location (S) | Internal procedure PP TENSLAB/TH/3/22 | |
| 51 | Bourdon tube pressure gauges Electronic pressure gauges |
Calibration — Pressure -0,1 MPa do 0,25 MPa · CMC 1,2 · 10⁻⁴ MPa 0,25 MPa do 1 MPa · CMC 4,1 · 10⁻⁴ MPa 1 MPa do 4 MPa · CMC 1,2 · 10⁻³ MPa 4 MPa do 10 MPa · CMC 2,9 · 10⁻³ MPa 10 MPa do 35 MPa · CMC 1,0 ∙ 10⁻² MPa 35 MPa do 70 MPa · CMC 2,3 ∙ 10⁻² MPa 70 MPa do 100 MPa · CMC 5,8 · 10⁻² MPa 100 MPa to 160 MPa · CMC 2,0 ∙ 10⁻¹ MPa 160 MPa do 200 MPa · CMC 2,4 ∙ 10⁻¹ MPa 200 MPa do 250 MPa · CMC 3,0 ∙ 10⁻¹ MPa Pressure |
Internal procedure; At permanent location (S) | Internal procedure PP TENSLAB/CE/1/18 PN EN 837-1:2000 | |
| 52 | Electrical thermometers (including electronic) for air temperature measurement (calibration in a climatic chamber) | Calibration — Temperature (electrical thermometry) 10 °C to 35 °C · CMC 0,2 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | Internal procedure PP TENSLAB/TH/3/22 | |
| 53 | Electrical thermometers (including electronic): | Calibration — Temperature (electrical thermometry) - calibrated in a liquid bath — -25 °C to 0 °C · CMC 0,59 °C - calibrated in a liquid bath — 0 °C to 100 °C · CMC 0,10 °C - calibrated in a temperature calibrator — 50 °C to 150 °C · CMC 0,24 °C - calibrated in a temperature calibrator — 150 °C to 300 °C · CMC 0,28 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | Internal procedure PP TENSLAB/TE/2/20 | |
| 54 | Pyrometers (including radiation pyrometers, thermal imaging cameras) | Calibration — Temperature (radiation thermometry) 0 °C do 100 °C · CMC 1,4 °C 100 °C to 200 °C · CMC 1,6 °C 200 °C do 350 °C · CMC 2,0 °C 350 °C do 500 °C · CMC 2,4 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) | Internal procedure PP TENSLAB/BPT/1/20 | |
| 55 | Ultrasonic flaw detectors | Calibration — RF electrical quantities; Time (time interval); Frequency - measurement of transmitter pulse voltage — 30 V to 500 V · CMC 4 % - measurement of transmitter pulse time interval — 2 ns to 10 ns · CMC 2,4 % + 0,43 ns - measurement of transmitter pulse time interval — 10 ns to 10 ms · CMC 4,81 % + 0,13 ns - measurement of amplifier frequency response — 0,01 MHz to 25 MHz · CMC 0,0001 MHz - measurement of equivalent input noise level — 2 · 10⁻¹¹ V/√Hz to 1 · 10⁻⁴ V/√Hz · CMC 0,2 · 10⁻⁸ V/√Hz - measurement of attenuator accuracy — 0 dB to 100 dB · CMC 0,3 dB - measurement of time base linearity — 0,01 % WE to 100 % WE1) · CMC 0,15 % WE1) RF electrical quantities; Time (time interval); Frequency |
group 2 tests; At permanent location (S) | PN-EN ISO 22232-1:2021-02 | |
| 56 | Ultrasonic thickness gauges | Calibration — RF electrical quantities; Time (time interval); Frequency - measurement of transmitter pulse voltage — 30 V to 500 V · CMC 4 % - measurement of transmitter pulse time interval — 2 ns to 10 ms · CMC 2,4 % + 0,43 ns - measurement of transmitter pulse time interval — 10 ns to 10 ms · CMC 4,81 % + 0,13 ns RF electrical quantities; Time (time interval); Frequency |
group 2 tests; At permanent location (S) | PN-EN 15317:2014-02 excluding cl. 9.4, 9.6, 9.7, 9.9, 9.12, 9.13, 9.16 | |
Reference standards (47)
EURAMET cg-10 Version 2.1 (12/2012)
EURAMET cg-2 Version 2.1 (02/2023)
EURAMET cg-6 Version 2.0 (03/2011)
PN EN 837-1:2000
PN EN ISO 6789-1:2017-03
PN EN ISO 6789-2:2017-03
PN-EN 15317:2014-02
PN-EN ISO 22232-1:2021-02
PN-EN ISO 3650:2000
PN-EN ISO 6506-2:2019
PN-EN ISO 6507-2:2018
PN-EN ISO 6508- 2:2024
PP TENSLAB/BPT/1/20
PP TENSLAB/CE/1/18
PP TENSLAB/CZ/5/24
PP TENSLAB/GŁ/3/18
PP TENSLAB/KK/1/18
PP TENSLAB/KOMP/1/22
PP TENSLAB/KĄT/1/18
PP TENSLAB/MCH/2/20
PP TENSLAB/MGR/2/24
PP TENSLAB/MIKR.ZEW/5/18
PP TENSLAB/PCZ/1/25
PP TENSLAB/PMwew/1/19
PP TENSLAB/PRZ/2/20
PP TENSLAB/PWz/4/25
PP TENSLAB/Poz/1/25
PP TENSLAB/SGP/5/25
PP TENSLAB/SGS/1/20
PP TENSLAB/SGT/4/25
PP TENSLAB/SM/1/18
PP TENSLAB/SO/3/25
PP TENSLAB/SP/2/18
PP TENSLAB/SUW/4/18
PP TENSLAB/SzL/1/23
PP TENSLAB/TB/1/24
PP TENSLAB/TE/2/20
PP TENSLAB/TH/3/22
PP TENSLAB/TR/2/24
PP TENSLAB/TV/1/22
PP TENSLAB/UMG/2/19
PP TENSLAB/WP/3/25
PP TENSLAB/WYS.SUW/1/18
PP TENSLAB/WZ.GR/2/25
PP TENSLAB/WZ.WEW/2/25
PP TENSLAB/WZ.ZEW/4/25
PP TENSLAB/Ścz/1/22