🇵🇱 Krajowe Usługi Metrologiczne M. Koziorowski, P. Koziorowski s.c. PL
Metrology & calibration
Krajowe Usługi Metrologiczne M. Koziorowski, P. Koziorowski s.c. is a calibration laboratory accredited by PCA under number AP 162 (Łódź, Poland). Its accreditation scope covers 111 calibration items in the following areas: Metrology & calibration. Calibrated items include: Digital dial gauges, Analogue dial gauges, Feeler gauges, Analogue lever dial gauges. Most frequently used calibration principles: Internal procedure; Direct method using gauge blocks; At permanent…, Internal procedure; Direct method using a MAHR OPTIMAR 100 instrume…, Internal procedure; Direct method using a Mitutoyo micrometer head;….
At a glance
- Accreditation number: PCA AP 162
- Calibration items in scope: 111
- Types of calibrated items: 95
- Calibration principles used: 42
- City: Łódź, Poland
- Valid until: 28-04-2028
- Accreditation number
- AP 162
- Accreditation body
- PCA
- Address
- ul. Demokratyczna 117, 93-348 Łódź
- City
- Łódź, Poland
- Division
- Calibration Laboratory
- Valid until
- 28-04-2028
What this laboratory calibrates (95)
- Digital dial gauges — 9 calibration items
- Analogue dial gauges — 5 calibration items
- Feeler gauges — 5 calibration items
- Analogue lever dial gauges — 4 calibration items
- Gauge blocks (classes 0, 1, 2) — 4 calibration items
- Measuring rollers — 4 calibration items
- Two-point dial bore gauges — 4 calibration items
- – resolution 0,0005 mm — 3 calibration items
- – resolution 0,001 mm — 3 calibration items
- – resolution 0,01 mm — 3 calibration items
- – scale interval 0,01 mm — 3 calibration items
- Analogue inch dial gauges — 3 calibration items
- Special vernier instruments — 3 calibration items
- – resolution 0,005 mm — 2 calibration items
- – scale interval (0,002; 0,001) mm — 2 calibration items
- and 80 more
Calibration principles (42)
- Internal procedure; Direct method using gauge blocks; At permanent location (S) — 24 calibration items
- Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) — 16 calibration items
- Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) — 6 calibration items
- Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) — 6 calibration items
- Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) — 5 calibration items
- Internal procedure; Direct method using an Abbe P01 vertical length measuring machine; At permanent location (S) — 4 calibration items
- Internal procedure; Direct method using angle gauge blocks; At permanent location (S) — 4 calibration items
- Internal procedure; Direct method using an ISD-V300A microscope; At permanent location (S) — 3 calibration items
- Internal procedure; Direct method using gauge blocks and measuring rollers; At permanent location (S) — 3 calibration items
- Internal procedure; Indirect method using a Mitutoyo measuring microscope; At permanent location (S) — 3 calibration items
- Internal procedure; Comparative method using a MAHR two-probe comparator; At permanent location (S) — 2 calibration items
- Internal procedure; Direct method using a CARL ZEISS JENA optical dividing head; At permanent location (S) — 2 calibration items
- Internal procedure; Direct method using a Mitutoyo Type 521-105 instrument; At permanent location (S) — 2 calibration items
- Internal procedure; Direct method using a ULM 520 S-E horizontal length measuring machine; At permanent location (S) — 2 calibration items
- Internal procedure; Indirect method using Talyvel 6 Taylor-Hobson differential levels; At permanent location (S) and outside of it (P) — 2 calibration items
- and 27 more
Certificate information
- Issue date
- 10.03.2026
- Issue no.
- 13
- Certificate date
- 23.04.2020
- Accreditation cycle
- 18.03.2024 — 28.04.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (111)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Analogue dial gauges – scale interval 0,01 mm |
Calibration — Length 0 mm do 10 mm · CMC 2,5 µm 0 mm do 30 mm · CMC 2,9 µm 0 mm to 50 mm · CMC 3,3 µm Length |
Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) | Internal procedure IW/KUM/05/01 | |
| 2 | Analogue lever dial gauges – scale interval 0,01 mm |
Calibration — Length 0 mm to 3 mm · CMC 2,5 µm Length |
Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) | Internal procedure IW/KUM/05/01 | |
| 3 | Analogue dial gauges – scale interval (0,002; 0,001) mm |
Calibration — Length 0 mm to 5 mm · CMC 1,8 µm Length |
Internal procedure; Direct method using a Mitutoyo Type 521-105 instrument; At permanent location (S) | Internal procedure IW/KUM/06/01 | |
| 4 | Analogue lever dial gauges – scale interval (0,002; 0,001) mm |
Calibration — Length 0 mm to 0,6 mm · CMC 1,8 µm Length |
Internal procedure; Direct method using a Mitutoyo Type 521-105 instrument; At permanent location (S) | Internal procedure IW/KUM/06/01 | |
| 5 | Analogue dial gauges – scale interval 0,1 mm |
Calibration — Length 0 mm to 50 mm · CMC 6 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 6 | Analogue dial gauges – scale interval (0,02; 0,01) mm |
Calibration — Length 0 mm do 10 mm · CMC 1,0 µm 0 mm do 30 mm · CMC 1,5 µm 0 mm to 50 mm · CMC 2,1 µm 0 mm do 80 mm · CMC 3,1 µm 0 mm do 100 mm · CMC 3,9 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 7 | Analogue dial gauges – scale interval (0,005; 0,002; 0,001) mm |
Calibration — Length 0 mm to 5 mm · CMC 0,5 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 8 | Analogue lever dial gauges – scale interval (0,02; 0,01) mm |
Calibration — Length 0 mm to 3 mm · CMC 0,9 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 9 | Analogue lever dial gauges – scale interval (0,005; 0,002; 0,001) mm |
Calibration — Length 0 mm to 0,6 mm · CMC 0,5 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 10 | Analogue inch dial gauges | Calibration — Length – scale int. 0,00127 mm / 0,00005” — 0 mm to 6,35 mm / 0” to 0,250” · CMC 0,4 µm / 0,000014” – scale int. 0,00254 mm / 0,00010” — 0 mm to 6,35 mm / 0” to 0,250” · CMC 0,4 µm / 0,000015” – scale int. 0,00635 mm / 0,00025” — 0 mm to 6,35 mm / 0” to 0,250” · CMC 0,5 µm / 0,000020” – scale int. 0,0127 mm / 0,0005” — 0 mm to 6,35 mm / 0” to 0,250” · CMC 0,8 µm / 0,000031” – scale int. 0,0254 mm / 0,001” — 0 mm to 6,35 mm / 0” to 0,250” · CMC 1,5 µm / 0,000060” Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 11 | Analogue inch dial gauges | Calibration — Length – scale int. 0,00254 mm / 0,0001” — 0 mm to 25,4 mm / 0” to 1,000” · CMC 0,8 µm / 0,000030” – scale int. 0,0254 mm / 0,001” — 0 mm to 25,4 mm / 0” to 1,000” · CMC 1,7 µm / 0,000065” – scale int. 0,254 mm / 0,01” — 0 mm to 25,4 mm / 0” to 1,000” · CMC 14,7 µm / 0,00058” Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 12 | Analogue inch dial gauges – scale int. 0,0254 mm / 0,001” – scale int. 0,127 mm / 0,005” |
Calibration — Length 0 mm to 50,8 mm / 0” to 2,000” · CMC 2,0 µm / 0,00008” 0 mm to 76,2 mm / 0” to 3,000” · CMC 2,5 µm / 0,00010” 0 mm to 101,6 mm / 0” to 4,000” · CMC 3,0 µm / 0,00012” Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 13 | Analogue inch lever dial gauges – scale int. 0,00254 mm/ 0,0001” – scale int. 0,00127 mm / 0,00005” – scale int. 0,0127 mm / 0,0005” |
Calibration — Length -0,030” to 0,030” · CMC 0,3 µm / 0,000013” -0,030” to 0,030” · CMC 0,8 µm / 0,000030” Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 14 | Digital dial gauges – resolution 0,0005 mm – resolution 0,0002 mm |
Calibration — Length 0 mm to 3 mm · CMC 0,5 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 15 | Digital dial gauges – resolution 0,0005 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 0,8 µm 0 mm do 25,4 mm · CMC 1,1 µm 0 mm to 50,8 mm · CMC 2,0 µm 0 mm do 100 mm · CMC 3,8 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 16 | Digital dial gauges – resolution 0,001 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 2,8 µm 0 mm do 25,4 mm · CMC 2,9 µm 0 mm to 50,8 mm · CMC 3,4 µm Length |
Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) | Internal procedure IW/KUM/05/01 | |
| 17 | Digital dial gauges – resolution 0,002 mm – resolution 0,001 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 1,0 µm 0 mm do 25,4 mm · CMC 1,3 µm 0 mm to 50,8 mm · CMC 2,1 µm 0 mm do 100 mm · CMC 3,9 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 18 | Digital dial gauges – resolution 0,005 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 3,1 µm 0 mm do 25,4 mm · CMC 3,2 µm 0 mm to 50,8 mm · CMC 3,6 µm Length |
Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) | Internal procedure IW/KUM/05/01 | |
| 19 | Digital dial gauges – resolution 0,005 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 1,8 µm 0 mm do 25,4 mm · CMC 2,0 µm 0 mm to 50,8 mm · CMC 2,5 µm 0 mm do 100 mm · CMC 4,1 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 20 | Digital dial gauges – resolution 0,01 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 3,6 µm 0 mm do 25,4 mm · CMC 3,7 µm 0 mm to 50,8 mm · CMC 4,1 µm Length |
Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) | Internal procedure IW/KUM/05/01 | |
| 21 | Digital dial gauges – resolution 0,01 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 2,5 µm 0 mm do 25,4 mm · CMC 2,7 µm 0 mm to 50,8 mm · CMC 3,1 µm 0 mm do 100 mm · CMC 4,5 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 22 | Digital dial gauges – resolution 0,02 mm |
Calibration — Length 0 mm do 12,7 mm · CMC 3,3 µm 0 mm do 25,4 mm · CMC 3,4 µm 0 mm to 50,8 mm · CMC 3,8 µm 0 mm do 100 mm · CMC 5,0 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 23 | Digital lever dial gauges | Calibration — Length – resolution 0,01 mm — 0 mm to 1 mm · CMC 3,3 µm – resolution 0,001 mm — 0 mm to 1 mm · CMC 0,9 µm Length |
Internal procedure; Direct method using a MAHR OPTIMAR 100 instrument; At permanent location (S) | Internal procedure IW/KUM/18/01 | |
| 24 | Digital lever dial gauges | Calibration — Length – resolution 0,01 mm — 0 mm to 1 mm · CMC 3,4 µm – resolution 0,005 mm — 0 mm to 1 mm · CMC 2,9 µm – resolution 0,001 mm — 0 mm to 1 mm · CMC 2,5 µm Length |
Internal procedure; Direct method using a Mitutoyo micrometer head; At permanent location (S) | Internal procedure IW/KUM/05/01 | |
| 25 | Optical probes | Calibration — Length – MOP 02/20 — -20 µm to 20 µm · CMC 0,12 µm – MOP 1/100 — -100 µm to 100 µm · CMC 0,13 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/12/01 | |
| 26 | Abbe vertical length measuring machines | Calibration — Length – length measuring machine measurement error — 0 mm to 100 mm · CMC Q[0,11; 3,4∙L] µm – reading microscope measurement error — 0 mm to 100 mm · CMC 0,12 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/12/02 | |
| 27 | Abbe horizontal length measuring machines | Calibration — Length – reading microscope measurements — 0 mm to 3 mm · CMC 0,1 µm – external measurements — 0 mm to 100 mm · CMC 0,5 µm – internal measurements — 0 mm to 100 mm · CMC 1,0 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/12/03 | |
| 28 | Horizontal length measuring machines (1-D machines) | Calibration — Length – measurements with the quill — 0 mm to 100 mm · CMC Q[0,10; 1,27∙L] µm – measurements with the length measuring machine — 100 mm to 500 mm · CMC Q[0,40; 2,50∙L] µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/12/05 | |
| 29 | Reference foils | Calibration — Length 0 µm to 8000 µm · CMC Q[1,4; 1,2∙L] µm Length |
Internal procedure; Direct method using a thickness gauge; At permanent location (S) | Internal procedure IW/KUM/07/06 | |
| 30 | Dial depth gauges | Calibration — Length – scale int. (0,001; 0,002; 0,005) mm — 0 mm to 5 mm · CMC 0,3 µm – scale int. 0,01 mm — 0 mm to 10 mm · CMC 0,7 µm – scale int. 0,01 mm — 0 mm to 30 mm · CMC 0,8 µm – scale int. 0,01 mm — 0 mm to 50 mm · CMC 1,0 µm – scale int. 0,1 mm — 0 mm to 50 mm · CMC 6 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/02 | |
| 31 | Dial depth gauges | Calibration — Length – resolution (0,001; 0,005; 0,01)mm — 0 mm to 10 mm · CMC 0,7 µm – resolution (0,001; 0,005; 0,01)mm — 0 mm to 30 mm · CMC 0,8 µm – resolution (0,001; 0,005; 0,01)mm — 0 mm to 50 mm · CMC 1,0 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/02 | |
| 32 | Micrometer depth gauges | Calibration — Length 0 mm to 50 mm · CMC 0,8 µm 50 mm do 100 mm · CMC 1,3 µm 100 mm do 150 mm · CMC 1,9 µm 150 mm do 200 mm · CMC 2,6 µm 200 mm do 250 mm · CMC 3,3 µm 250 mm do 300 mm · CMC 3,9 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/02 | |
| 33 | Vernier depth gauges | Calibration — Length 0 mm do 150 mm · CMC 8 µm 0 mm do 300 mm · CMC 11 µm 0 mm do 500 mm · CMC 16 µm 0 mm to 600 mm · CMC 18 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/01/01 | |
| 34 | Ultrasonic thickness gauges | Calibration — Length - resolution 0,001 mm — 0,5 mm to 50 mm · CMC 4 µm - resolution 0,001 mm — 50 mm to 100 mm · CMC 5 µm - resolution 0,001 mm — 100 mm to 125 mm · CMC 7 µm - resolution 0,01 mm — 0,5 mm to 50 mm · CMC 7 µm - resolution 0,01 mm — 50 mm to 100 mm · CMC 9 µm - resolution 0,01 mm — 100 mm to 125 mm · CMC 10 µm - resolution 0,1 mm — 0,5 mm to 125 mm · CMC 64 µm Length |
Internal procedure; Direct method using step standards for ultrasonic thickness gauges; At permanent location (S) | Internal procedure IW/KUM/24/02 | |
| 35 | Micrometer heads | Calibration — Length 0 mm do 15 mm · CMC 0,7 µm 0 mm to 25 mm · CMC 0,8 µm 0 mm to 50 mm · CMC 1,0 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/02 | |
| 36 | Dial thickness gauges | Calibration — Length – scale int. (0,001; 0,002; 0,005) mm — 0 mm to 5 mm · CMC 0,3 µm – scale int. 0,01 mm — 0 mm to 10 mm · CMC 0,7 µm – scale int. 0,01 mm — 0 mm to 30 mm · CMC 0,8 µm – scale int. 0,01 mm — 0 mm to 50 mm · CMC 1,0 µm – scale int. 0,1 mm — 0 mm to 50 mm · CMC 6 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/02 | |
| 37 | Dial thickness gauges | Calibration — Length – resolution (0,001; 0,005; 0,01)mm — 0 mm to 10 mm · CMC 0,7 µm – resolution (0,001; 0,005; 0,01)mm — 0 mm to 30 mm · CMC 0,8 µm – resolution (0,001; 0,005; 0,01)mm — 0 mm to 50 mm · CMC 1,0 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/02 | |
| 38 | External measuring calipers Internal measuring calipers |
Calibration — Length External measuring calipers — 0 mm to 10 mm · CMC 0,5 µm External measuring calipers — 0 mm to 20 mm · CMC 0,7 µm External measuring calipers — 0 mm to 40 mm · CMC 1,2 µm External measuring calipers — 0 mm to 60 mm · CMC 1,7 µm External measuring calipers — 0 mm to 100 mm · CMC 2,8 µm Internal measuring calipers — 2,5 mm to 12,5 mm · CMC 0,6 µm Internal measuring calipers — 10 mm to 22 mm · CMC 0,8 µm Internal measuring calipers — 10 mm to 35 mm · CMC 1,1 µm Internal measuring calipers — 30 mm to 55 mm · CMC 1,6 µm Internal measuring calipers — 40 mm to 90 mm · CMC 2,5 µm Internal measuring calipers — 70 mm to 120 mm · CMC 3,3 µm Internal measuring calipers — 90 mm to 140 mm · CMC 3,9 µm Internal measuring calipers — 130 mm to 180 mm · CMC 4,9 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/03 | |
| 39 | Coating thickness meters | Calibration — Length 0 µm to 200 µm · CMC 1,5 µm 200 µm to 500 µm · CMC 1,7 µm 500 µm to 1000 µm · CMC 2,0 µm 1000 µm to 1500 µm · CMC 2,4 µm 1500 µm to 2000 µm · CMC 2,9 µm Length |
Internal procedure; Direct method using reference foils; At permanent location (S) | Internal procedure IW/KUM/24/01 | |
| 40 | Laser scan micrometers | Calibration — Length 0,1 mm to 25 mm · CMC 0,32 µm Length |
Internal procedure; Direct method using measuring rollers; At permanent location (S) | Internal procedure IW/KUM/25/01 | |
| 41 | Internal micrometers | Calibration — Length 5 mm do 30 mm · CMC 1,0 µm 30 mm do 55 mm · CMC 1,2 µm 50 mm to 75 mm · CMC 1,5 µm 75 mm to 100 mm · CMC 1,8 µm 100 mm to 125 mm · CMC 2,0 µm 125 mm to 150 mm · CMC 2,5 µm 150 mm to 175 mm · CMC 2,8 µm 175 mm to 200 mm · CMC 3,2 µm 200 mm do 225 mm · CMC 3,5 µm 225 mm do 250 mm · CMC 3,8 µm 250 mm do 275 mm · CMC 4,2 µm 275 mm do 300 mm · CMC 4,5 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/01 | |
| 42 | Micrometers with built-in dial indicator | Calibration — Length – indicator unit — -140 µm to 140 µm · CMC 0,3 µm – micrometer unit — 0 mm to 25 mm · CMC 0,8 µm – micrometer unit — 25 mm to 50 mm · CMC 1,0 µm – micrometer unit — 50 mm to 75 mm · CMC 1,3 µm – micrometer unit — 75 mm to 100 mm · CMC 1,6 µm – micrometer unit — 100 mm to 125 mm · CMC 1,9 µm – micrometer unit — 125 mm to 150 mm · CMC 2,3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/03 | |
| 43 | External micrometers | Calibration — Length 0 mm to 25 mm · CMC 0,9 µm 25 mm to 50 mm · CMC 1,1 µm 50 mm to 75 mm · CMC 1,3 µm 75 mm to 100 mm · CMC 1,7 µm 100 mm to 125 mm · CMC 2,0 µm 125 mm to 150 mm · CMC 2,4 µm 150 mm to 175 mm · CMC 2,7 µm 175 mm to 200 mm · CMC 3,1 µm 200 mm do 225 mm · CMC 3,4 µm 225 mm do 250 mm · CMC 3,7 µm 250 mm do 275 mm · CMC 4,0 µm 275 mm do 300 mm · CMC 4,4 µm 300 mm do 400 mm · CMC 4,7 µm 400 mm do 500 mm · CMC 6,1 µm 500 mm do 600 mm · CMC 7,5 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/01 | |
| 44 | Digital external micrometers – resolution 0,0001 mm |
Calibration — Length 0 mm to 25 mm · CMC 0,6 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/01 | |
| 45 | Setting rings: | Calibration — Length – with the head — 3 mm to 50 mm · CMC 0,7 µm – with the head — 50 mm to 90 mm · CMC 0,8 µm – with the bows — 12 mm to 50 mm · CMC 0,6 µm – with the bows — 50 mm to 100 mm · CMC 0,7 µm – with the bows — 100 mm to 150 mm · CMC 1,0 µm – with the bows — 150 mm to 200 mm · CMC 1,4 µm – with the bows — 200 mm to 250 mm · CMC 1,8 µm – with the bows — 250 mm to 300 mm · CMC 2,3 µm Length |
Internal procedure; Direct method using a MAHR ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/01 | |
| 46 | Plane-parallel optical flats | Calibration — Length – length deviation — 12 mm · CMC 0,5 µm – length deviation — 15 mm · CMC 0,5 µm – length deviation — 27 mm · CMC 0,7 µm – length deviation — 40 mm · CMC 0,9 µm – length deviation — 52 mm · CMC 1,1 µm – length deviation — 65 mm · CMC 1,3 µm – length deviation — 77 mm · CMC 1,5 µm Length |
Internal procedure; Comparative method using gauge blocks and a MOP 02/20 optical probe; At permanent location (S) | Internal procedure IW/KUM/08/01 | |
| 47 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 0,5 mm to 100 mm · CMC Q[0,06; 1,05∙L] µm Length |
Internal procedure; Comparative method using a TESA two-probe comparator; At permanent location (S) | Internal procedure IW/KUM/16/01 based on PN-EN ISO 3650:2000 | |
| 48 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 0,5 mm to 100 mm · CMC Q[0,06; 1,05∙L] µm Length |
Internal procedure; Comparative method using a MAHR two-probe comparator; At permanent location (S) | Internal procedure IW/KUM/16/02 based on PN-EN ISO 3650:2000 | |
| 49 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 125 mm to 175 mm · CMC Q[0,07; 1,00∙L] µm Length |
Internal procedure; Comparative method using a MAHR two-probe comparator; At permanent location (S) | Internal procedure IW/KUM/16/02 based on PN-EN ISO 3650:2000 | |
| 50 | Gauge blocks (classes 0, 1, 2) | Calibration — Length 150 mm to 500 mm · CMC Q[0,14; 1,03∙L] µm Length |
Internal procedure; Comparative method using a MAHR ULM 520 S-E horizontal length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/07 | |
| 51 | Measuring tapes | Calibration — Length 0 m to 5 m · CMC Q[0,11; 0,028∙L] mm 5 m do 10 m · CMC Q[0,16; 0,028∙(L-5)] mm 10 m do 15 m · CMC Q[0,20; 0,028∙(L-10)] mm 15 m do 20 m · CMC Q[0,23; 0,028∙(L-15)] mm 20 m do 25 m · CMC Q[0,26; 0,028∙(L-20)] mm Length |
Internal procedure; Comparative method using a POLMIAR rigid rule; At permanent location (S) | Internal procedure IW/KUM/13/01 | |
| 52 | Rigid rules Semi-rigid rules |
Calibration — Length 0 m do 3 m · CMC Q[0,11; 0,028∙L] mm 0 m to 5 m · CMC Q[0,11; 0,028∙L] mm Length |
Internal procedure; Comparative method using a POLMIAR rigid or semi-rigid rule; At permanent location (S) | Internal procedure IW/KUM/13/02 | |
| 53 | Rigid rules Semi-rigid rules |
Calibration — Length 0 mm to 300 mm · CMC Q[19; 0,04∙L] µm Length |
Internal procedure; Direct method using an ISD-V300A microscope; At permanent location (S) | Internal procedure IW/KUM/23/01 | |
| 54 | Special vernier instruments Digital vernier weld gauges |
Calibration — Length – welds on a plane — 0 mm to 30 mm · CMC 6 µm – corner welds — 0 mm to 30 mm · CMC 10 µm Length |
Internal procedure; Direct method using gauge blocks and measuring rollers; At permanent location (S) | Internal procedure IW/KUM/01/02 | |
| 55 | Special vernier instruments Analogue vernier weld gauges |
Calibration — Length – welds on a plane — 0 mm to 30 mm · CMC 0,06 mm – corner welds — 0 mm to 30 mm · CMC 0,06 mm Length |
Internal procedure; Direct method using gauge blocks and measuring rollers; At permanent location (S) | Internal procedure IW/KUM/01/02 | |
| 56 | Weld gauges | Calibration — Length – butt weld height — 0 mm to 15 mm · CMC 0,06 mm – fillet weld height — 0 mm to 20 mm · CMC 0,06 mm – fillet weld throat thickness — 0 mm to 15 mm · CMC 0,06 mm – butt weld width — 0 mm to 60 mm · CMC 0,21 mm – undercut depth — 0 mm to 6 mm · CMC 0,065 mm Length |
Internal procedure; Direct method using gauge blocks and measuring rollers; At permanent location (S) | Internal procedure IW/KUM/01/02 | |
| 57 | Special vernier instruments | Calibration — Length Vernier chamfer gauges — 0 mm to 10 mm · CMC 20 µm Taper vernier gap gauges — 0,2 mm to 40 mm · CMC 13 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/01/04 | |
| 58 | Cylindrical thread ring gauges | Calibration — Length pitch diameter D₂ — 2,6 mm to 90 mm · CMC 3,1 µm pitch diameter D₂ — 90 mm to 125 mm · CMC 3,2 µm Length |
Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/02 | |
| 59 | Cylindrical thread plug gauges | Calibration — Length pitch diameter d₂ — 1 mm to 100 mm · CMC 3,0 µm pitch diameter d₂ — 100 mm to 200 mm · CMC 3,1 µm Length |
Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/03 | |
| 60 | Taper thread ring gauges | Calibration — Length pitch diameter — 4 mm to 100 mm · CMC 5,0 µm Length |
Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/05 | |
| 61 | Taper thread plug gauges | Calibration — Length pitch diameter — 4 mm to 100 mm · CMC 5,0 µm Length |
Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/06 | |
| 62 | Plain ring gauges: | Calibration — Length – with the head — 3 mm to 50 mm · CMC 0,7 µm – with the head — 50 mm to 90 mm · CMC 0,8 µm – with the bows — 12 mm to 50 mm · CMC 0,6 µm – with the bows — 50 mm to 100 mm · CMC 0,7 µm – with the bows — 100 mm to 150 mm · CMC 1,0 µm – with the bows — 150 mm to 200 mm · CMC 1,4 µm – with the bows — 200 mm to 250 mm · CMC 1,8 µm – with the bows — 250 mm to 300 mm · CMC 2,3 µm Length |
Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/01 | |
| 63 | Taper hole gauges | Calibration — Length 0,1 mm to 5 mm · CMC 9 µm 1 mm to 60 mm · CMC 17 µm Length |
Internal procedure; Direct method using an ISD-V300A microscope; At permanent location (S) | Internal procedure IW/KUM/23/02 | |
| 64 | Plug gauges | Calibration — Length 1 mm to 50 mm · CMC 0,5 µm 50 mm to 100 mm · CMC 0,9 µm Length |
Internal procedure; Direct method using an Abbe P01 vertical length measuring machine; At permanent location (S) | Internal procedure IW/KUM/07/02 | |
| 65 | Plug gauges | Calibration — Length 1 mm to 50 mm · CMC 0,4 µm 50 mm to 100 mm · CMC 0,8 µm Length |
Internal procedure; Direct method using a ULM 520 S-E horizontal length measuring machine; At permanent location (S) | Internal procedure IW/KUM/07/02 | |
| 66 | Vernier calipers | Calibration — Length 0 mm do 150 mm · CMC 8 µm 0 mm do 300 mm · CMC 11 µm 0 mm do 500 mm · CMC 16 µm 0 mm to 600 mm · CMC 18 µm 0 mm do 800 mm · CMC 23 µm 0 mm to 1000 mm · CMC 29 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/01/01 | |
| 67 | Feeler gauges | Calibration — Length – leaf — 0,02 mm to 3,0 mm · CMC 0,3 µm Length |
Internal procedure; Direct method using an Abbe P01 vertical length measuring machine; At permanent location (S) | Internal procedure IW/KUM/07/01 | |
| 68 | Feeler gauges | Calibration — Length – leaf — 0,02 mm to 3,0 mm · CMC 0,8 µm Length |
Internal procedure; Direct method using a digital micrometer with 0,0001 mm resolution; At permanent location (S) | Internal procedure IW/KUM/07/01 | |
| 69 | Feeler gauges | Calibration — Length – wide wedge — 0,5 mm to 27 mm · CMC 12 µm Length |
Internal procedure; Direct method using an Abbe P01 vertical length measuring machine; At permanent location (S) | Internal procedure IW/KUM/07/01 | |
| 70 | Feeler gauges | Calibration — Length – wide wedge — 0,5 mm to 27 mm · CMC 14 µm Length |
Internal procedure; Direct method using a digital dial gauge with 0,0005 mm resolution; At permanent location (S) | Internal procedure IW/KUM/07/01 | |
| 71 | Feeler gauges | Calibration — Length – wedge – flat wedge — 0,5 mm to 45 mm · CMC 8 µm Length |
Internal procedure; Direct method using a digital measuring microscope; At permanent location (S) | Internal procedure IW/KUM/07/01 | |
| 72 | Two-point dial bore gauges – scale interval 0,01 mm – resolution 0,001 mm |
Calibration — Length 3,5 mm to 35 mm · CMC 2,1 µm 35 mm to 60 mm · CMC 2,1 µm 50 mm to 150 mm · CMC 2,2 µm 150 mm to 250 mm · CMC 3,4 µm 250 mm to 400 mm · CMC 5,1 µm 400 mm to 600 mm · CMC 7,8 µm 600 mm to 800 mm · CMC 11,5 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/01 | |
| 73 | Two-point dial bore gauges – scale interval 0,001 mm – scale interval 0,0005 mm |
Calibration — Length 3,5 mm to 35 mm · CMC 0,8 µm 35 mm to 60 mm · CMC 0,9 µm 50 mm to 150 mm · CMC 1,1 µm 150 mm to 250 mm · CMC 2,9 µm 250 mm to 400 mm · CMC 4,7 µm 400 mm to 600 mm · CMC 7,5 µm 600 mm to 800 mm · CMC 11,3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/01 | |
| 74 | Two-point dial bore gauges – resolution 0,01 mm |
Calibration — Length 3,5 mm to 35 mm · CMC 8,7 µm 35 mm to 60 mm · CMC 8,7 µm 50 mm to 150 mm · CMC 8,7 µm 150 mm to 250 mm · CMC 9,0 µm 250 mm to 400 mm · CMC 9,8 µm 400 mm to 600 mm · CMC 11,4 µm 600 mm to 800 mm · CMC 14,3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/01 | |
| 75 | Two-point dial bore gauges – resolution 0,0005 mm |
Calibration — Length 3,5 mm to 18 mm · CMC 1,2 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/14/01 | |
| 76 | Two-point micrometer bore gauges | Calibration — Length 30 mm · CMC 0,8 µm 35 mm · CMC 0,9 µm 50 mm · CMC 1,0 µm 63 mm · CMC 1,2 µm 75 mm · CMC 1,3 µm 88 mm · CMC 1,5 µm 100 mm · CMC 1,6 µm 125 mm · CMC 1,9 µm 150 mm · CMC 2,3 µm 175 mm · CMC 2,6 µm 200 mm · CMC 2,9 µm 225 mm · CMC 3,3 µm 250 mm · CMC 3,6 µm 275 mm · CMC 3,9 µm 300 mm · CMC 4,3 µm 325 mm · CMC 4,6 µm 350 mm · CMC 4,9 µm 375 mm · CMC 5,3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/04 | |
| 77 | Three-point micrometer bore gauges Three-point dial bore gauges |
Calibration — Length 3 mm do 100 mm · CMC 1,9 µm 100 mm do 175 mm · CMC 2,9 µm 175 mm to 200 mm · CMC 3,0 µm 200 mm do 250 mm · CMC 4,3 µm 250 mm do 300 mm · CMC 4,8 µm Length |
Internal procedure; Direct method using setting rings; At permanent location (S) | Internal procedure IW/KUM/02/05 | |
| 78 | Transameters | Calibration — Length – transameter indicator range — -140 µm to 140 µm · CMC 0,3 µm – transameter range — 0 mm to 150 mm · CMC 0,3 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/02/03 | |
| 79 | Measuring rollers | Calibration — Length – for threads — 0,17 mm to 6,35 mm · CMC 0,40 µm – for holes — 0,05 mm to 10,0 mm · CMC 0,36 µm – for holes — 10,0 mm to 20,0 mm · CMC 0,37 µm – for gears — 1,7 mm to 17 mm · CMC 0,37 µm Length |
Internal procedure; Comparative method using gauge blocks and a CARL ZEISS JENA MOP 02/20 probe; At permanent location (S) | Internal procedure IW/KUM/09/01 | |
| 80 | Measuring rollers | Calibration — Length 0,05 mm to 25 mm · CMC 0,27 µm Length |
Internal procedure; Direct method using an Abbe P01 vertical length measuring machine; At permanent location (S) | Internal procedure IW/KUM/09/02 | |
| 81 | Measuring rollers | Calibration — Length 0,05 mm to 25 mm · CMC 0,24 µm Length |
Internal procedure; Direct method using a ULM 520 S-E horizontal length measuring machine; At permanent location (S) | Internal procedure IW/KUM/09/02 | |
| 82 | Measuring rollers | Calibration — Length 0,1 mm to 25 mm · CMC 0,51 µm Length |
Internal procedure; Direct method using a laser scan micrometer; At permanent location (S) | Internal procedure IW/KUM/09/02 | |
| 83 | Line standards | Calibration — Length 0 mm to 300 mm · CMC Q[4,5; 0,0066∙L] µm Length |
Internal procedure; Direct method using an ISD-V300A microscope; At permanent location (S) | Internal procedure IW/KUM/23/01 | |
| 84 | Line standards of measuring magnifiers | Calibration — Length 0 mm do 30 mm · CMC 3,2 µm 0 mm to 50 mm · CMC 4,0 µm Length |
Internal procedure; Indirect method using a Mitutoyo measuring microscope; At permanent location (S) | Internal procedure IW/KUM/10/04 | |
| 85 | Circular arc standards (radius gauges) | Calibration — Length 1 mm to 7 mm · CMC 4 µm 7,5 mm to 15 mm · CMC 15 µm 15,5 mm to 25 mm · CMC 21 µm Length |
Internal procedure; Indirect method using a Mitutoyo measuring microscope; At permanent location (S) | Internal procedure IW/KUM/10/01 | |
| 86 | Circular arc standards (radius gauges) | Calibration — Length 0,1 mm do 7 mm · CMC 3 µm 7 mm do 15 mm · CMC 4 µm 15 mm do 25 mm · CMC 13 µm 25 mm to 50 mm · CMC 20 µm 50 mm do 100 mm · CMC 33 µm Length |
Internal procedure; Direct method using an ISD-V300A measuring microscope; At permanent location (S) | Internal procedure IW/KUM/23/03 | |
| 87 | Setting standards for external dimensions | Calibration — Length 25 mm to 50 mm · CMC 0,5 µm 50 mm do 100 mm · CMC 0,6 µm 100 mm do 150 mm · CMC 0,8 µm Length |
Internal procedure; Comparative method using a transameter; At permanent location (S) | Internal instruction IW/KUM/02/06 | |
| 88 | Setting standards for external dimensions | Calibration — Length 25 mm to 100 mm · CMC 0,4 µm 100 mm to 200 mm · CMC 1,0 µm 200 mm to 300 mm · CMC 1,4 µm 300 mm to 400 mm · CMC 1,8 µm 400 mm to 500 mm · CMC 2,2 µm Length |
Internal procedure; Direct method using a ULM 520 S-E length measuring machine; At permanent location (S) | Internal procedure IW/KUM/22/04 | |
| 89 | Step standards for ultrasonic thickness gauges | Calibration — Length – steel — 0,5 mm to 50 mm · CMC 3 µm – steel — 50 mm to 100 mm · CMC 5 µm – steel — 100 mm to 125 mm · CMC 6 µm – aluminium — 0,5 mm to 50 mm · CMC 4 µm – aluminium — 50 mm to 100 mm · CMC 6 µm – aluminium — 100 mm to 125 mm · CMC 7 µm Length |
Internal procedure; Direct method using digital external micrometers; At permanent location (S) | Internal procedure IW/KUM/26/01 | |
| 90 | Thread profile standards | Calibration — Length – metric — 0,25 mm to 6 mm · CMC 4 µm – inch — 4 to 62 threads per inch · CMC 4 µm Length |
Internal procedure; Indirect method using a Mitutoyo measuring microscope; At permanent location (S) | Internal procedure IW/KUM/10/02 | |
| 91 | Digital height gauges | Calibration — Length – resolution 0,01 mm — 0 mm to 1000 mm · CMC Q[5,86; 3,24∙L] µm – resolution 0,005 mm — 0 mm to 1000 mm · CMC Q[3,03; 3,24∙L] µm – resolution 0,001 mm — 0 mm to 1000 mm · CMC Q[1,11; 3,24∙L] µm – resolution 0,0005 mm — 0 mm to 1000 mm · CMC Q[1,00; 3,24∙L] µm – resolution 0,0001 mm — 0 mm to 1000 mm · CMC Q[0,95; 3,24∙L] µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/12/04 | |
| 92 | Height gauges for measuring vehicle wheel height | Calibration — Length 0 mm to 1000 mm · CMC 0,5 mm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/01/03 | |
| 93 | Vernier height gauges | Calibration — Length 0 mm do 300 mm · CMC 11 µm 0 mm to 600 mm · CMC 18 µm 0 mm do 750 mm · CMC 22 µm 0 mm to 1000 mm · CMC 29 µm Length |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/01/01 | |
| 94 | Analogue universal protractors | Calibration — Angle 4 x 90° · CMC 3,0' Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure IW/KUM/03/01 | |
| 95 | Digital universal protractors | Calibration — Angle 0° to 360° · CMC 1,5' Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure IW/KUM/03/01 | |
| 96 | Digital protractors | Calibration — Angle 0° to 360° · CMC 0,04° Angle |
Internal procedure; Direct method using an angle prism; At permanent location (S) | Internal procedure IW/KUM/03/02 | |
| 97 | Marking-out protractors | Calibration — Angle 0° to 180° · CMC 0,12° Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure IW/KUM/03/01 | |
| 98 | 90° two-leg squares | Calibration — Angle longer leg length; – flatness — 40 mm to 500 mm · CMC 2 µm longer leg length; – straightness — 40 mm to 500 mm · CMC 1 µm longer leg length; – parallelism — 40 mm to 500 mm · CMC 1,4 µm longer leg length; – perpendicularity — 40 mm to 500 mm · CMC 1,1 µm Angle |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/04/01 | |
| 99 | 90° cylindrical squares | Calibration — Angle 0 mm to 400 mm · CMC 1,0 µm Angle |
Internal procedure; Indirect method using an inductive probe; At permanent location (S) | Internal procedure IW/KUM/15/01 | |
| 100 | Optical angle levels | Calibration — Angle – zero setting error — -120° to 120° · CMC 0,2 dz. el. – scale interval value error — -120° to 120° · CMC 4″ – measurement error over the full range — -120° to 120° · CMC 12″ Angle |
Internal procedure; Direct method using an optical dividing head; At permanent location (S) | Internal procedure IW/KUM/21/01 | |
| 101 | Angle gauge blocks with matt surfaces | Calibration — Angle 0° to 30° · CMC 31″ Angle |
Internal procedure; Indirect method using a microscope; At permanent location (S) | Internal procedure IW/KUM/10/03 | |
| 102 | Builder's levels | Calibration — Angle – vial zero indication setting error — 0 m to 2 m · CMC 0,2 mm Angle |
Internal procedure; Direct method using a CARL ZEISS JENA measuring bed; At permanent location (S) | Internal procedure IW/KUM/11/02 | |
| 103 | Digital levels | Calibration — Angle 4 x 90° · CMC 0,04° Angle |
Internal procedure; Direct method using a CARL ZEISS JENA optical dividing head; At permanent location (S) | Internal procedure IW/KUM/11/02 | |
| 104 | Coincidence levels | Calibration — Angle – measurement error over the full range — 0 mm/m to 20 mm/m · CMC 0,014 mm/m – zero indication setting error — 0 mm/m to 20 mm/m · CMC 0,002 mm/m Angle |
Internal procedure; Direct method using a CARL ZEISS JENA optical dividing head; At permanent location (S) | Internal procedure IW/KUM/21/02 | |
| 105 | Straightedge levels | Calibration — Angle – zero indication setting error — 0 mm/m to 1 mm/m · CMC 0,2 dz. el. – scale interval value error — 0 mm/m to 1 mm/m · CMC 0,019 mm/m Angle |
Internal procedure; Direct method using a coincidence level; At permanent location (S) | Internal procedure IW/KUM/11/01 | |
| 106 | Weld gauges | Calibration — Angle – bevel angle measurements — 80° to 160° · CMC 0,29° Angle |
Internal procedure; Direct method using angle gauge blocks; At permanent location (S) | Internal procedure IW/KUM/01/02 | |
| 107 | Knife-edge straightedges | Calibration — Surface geometry 0 mm to 100 mm · CMC 0,7 µm 100 mm to 400 mm · CMC 2,3 µm 400 mm to 700 mm · CMC 3,9 µm Surface geometry |
Internal procedure; Direct method using gauge blocks; At permanent location (S) | Internal procedure IW/KUM/17/03 | |
| 108 | Optical flats | Calibration — Surface geometry diameter — 0 mm to 100 mm · CMC 0,04 µm Surface geometry |
Internal procedure; Indirect method using optical flats; At permanent location (S) | Internal procedure IW/KUM/08/01 | |
| 109 | Plane-parallel optical flats | Calibration — Surface geometry flat length; – surface flatness deviation — 12,00 mm to 80,00 mm · CMC 0,06 µm flat length; – surface parallelism deviation — 12,00 mm to 80,00 mm · CMC 0,16 µm Surface geometry |
Internal procedure; Indirect method using optical flats and a MOP 02/20 probe; At permanent location (S) | Internal procedure IW/KUM/08/01 | |
| 110 | Surface plates | Calibration — Surface geometry longer side length — 250 mm · CMC 1,1 µm longer side length — 400 mm · CMC 1,3 µm longer side length — 630 mm · CMC 1,8 µm longer side length — 800 mm · CMC 2,0 µm longer side length — 1000 mm · CMC 2,2 µm longer side length — 1200 mm · CMC 2,5 µm longer side length — 1600 mm · CMC 2,9 µm longer side length — 2000 mm · CMC 3,2 µm longer side length — 2500 mm · CMC 3,6 µm Surface geometry |
Internal procedure; Indirect method using Talyvel 6 Taylor-Hobson differential levels; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/20/01 based on PN-ISO 8512-1:1998 PN-ISO 8512-2:1999 | |
| 111 | Surface straightedges | Calibration — Surface geometry straightedge length — 250 mm · CMC 1,0 µm straightedge length — 500 mm · CMC 1,5 µm straightedge length — 1000 mm · CMC 2,2 µm straightedge length — 1500 mm · CMC 2,8 µm straightedge length — 2000 mm · CMC 3,2 µm straightedge length — 2500 mm · CMC 3,6 µm straightedge length — 3000 mm · CMC 4,0 µm straightedge length — 4000 mm · CMC 4,6 µm straightedge length — 5000 mm · CMC 5,2 µm Surface geometry |
Internal procedure; Indirect method using Talyvel 6 Taylor-Hobson differential levels; At permanent location (S) and outside of it (P) | Internal procedure IW/KUM/20/01 | |
Reference standards (62)
IW/KUM/01/01
IW/KUM/01/02
IW/KUM/01/03
IW/KUM/01/04
IW/KUM/02/01
IW/KUM/02/02
IW/KUM/02/03
IW/KUM/02/04
IW/KUM/02/05
IW/KUM/02/06
IW/KUM/03/01
IW/KUM/03/02
IW/KUM/04/01
IW/KUM/05/01
IW/KUM/06/01
IW/KUM/07/01
IW/KUM/07/02
IW/KUM/07/06
IW/KUM/08/01
IW/KUM/09/01
IW/KUM/09/02
IW/KUM/10/01
IW/KUM/10/02
IW/KUM/10/03
IW/KUM/10/04
IW/KUM/11/01
IW/KUM/11/02
IW/KUM/12/01
IW/KUM/12/02
IW/KUM/12/03
IW/KUM/12/04
IW/KUM/12/05
IW/KUM/13/01
IW/KUM/13/02
IW/KUM/14/01
IW/KUM/14/02
IW/KUM/14/03
IW/KUM/15/01
IW/KUM/16/01
IW/KUM/16/02
IW/KUM/17/03
IW/KUM/18/01
IW/KUM/20/01
IW/KUM/21/01
IW/KUM/21/02
IW/KUM/22/01
IW/KUM/22/02
IW/KUM/22/03
IW/KUM/22/04
IW/KUM/22/05
IW/KUM/22/06
IW/KUM/22/07
IW/KUM/23/01
IW/KUM/23/02
IW/KUM/23/03
IW/KUM/24/01
IW/KUM/24/02
IW/KUM/25/01
IW/KUM/26/01
PN-EN ISO 3650:2000
PN-ISO 8512-1:1998
PN-ISO 8512-2:1999