🇵🇱 Instytut Automatyki Systemów Energetycznych Sp. z o.o. PL

PCA AP 145 • Wrocław, Poland • 3 calibration items

Metrology & calibration

Instytut Automatyki Systemów Energetycznych Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 145 (Wrocław, Poland). Its accreditation scope covers 3 calibration items in the following areas: Metrology & calibration. Calibrated items include: Combination wave (surge) generators, Electrical fast transient/burst (EFT/B) generators, Electrostatic discharge (ESD) generators. Most frequently used calibration principles: Internal procedure; At permanent location (S).

At a glance

  • Accreditation number: PCA AP 145
  • Calibration items in scope: 3
  • Types of calibrated items: 3
  • Calibration principles used: 1
  • City: Wrocław, Poland
  • Valid until: 18-11-2028
Accreditation number
AP 145
Accreditation body
PCA
Address
ul. Wystawowa 1, 51-618 Wrocław
City
Wrocław, Poland
Division
Laboratorium Badawcze i Wzorcujące
Valid until
18-11-2028

What this laboratory calibrates (3)

Calibration principles (1)

Certificate information

Issue date
16.10.2025
Issue no.
8
Certificate date
27.10.2020
Accreditation cycle
28.10.2024 — 18.11.2028
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (3)

The measurement uncertainty value for CMC expressed as a percentage is a relative measurement uncertainty and refers to a percentage contribution in the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Electrostatic discharge (ESD) generators Calibration — High-frequency electrical quantities
output voltage — 2 kV to 30 kV · CMC 3,0 % peak value of discharge current — 5 A to 35 A · CMC 6,0 % discharge current rise time — 0,5 ns to 1,5 ns · CMC 8,0 % discharge current at 30 ns — 2 A to 21 A · CMC 8,0 % discharge current at 60 ns — 1 A to 11 A · CMC 8,0 %
High-frequency electrical quantities
Internal procedure; At permanent location (S) LB-IT07 w oparciu o PN-EN 61000-4-2:2011
2 Electrical fast transient/burst (EFT/B) generators Calibration — High-frequency electrical quantities
peak voltage — 0,2 kV to 4,8 kV · CMC 6,0 % pulse rise time — 3 ns to 7 ns · CMC 5,5 % pulse duration — 30 ns to 200 ns · CMC 4,0 % burst duration — 0,5 ms to 20 ms · CMC 4,0 % burst period — 200 ms to 400 ms · CMC 4,0 % pulse repetition frequency within a burst — 2 Hz to 120 kHz · CMC 4,0 %
High-frequency electrical quantities
Internal procedure; At permanent location (S) LB-IT05 w oparciu o PN-EN 61000-4-4:2010 +A1:2010; LB-IT05A w oparciu o PN-EN 61000-4-4:2013-05
3 Combination wave (surge) generators Calibration — High-frequency electrical quantities
open-circuit peak voltage — 0,2 kV to 7 kV · CMC 5 % open-circuit voltage front time — 0,5 μs to 20 μs · CMC 7 % open-circuit voltage duration — 10 μs to 1000 μs · CMC 6 % short-circuit peak current — 5 A to 3800 A · CMC 5 % short-circuit current front time — 1,5 μs to 15 μs · CMC 5 % short-circuit current duration — 15 μs to 400 μs · CMC 6 % effective output impedance — 1,5 Ω to 45 Ω · CMC 6 %
High-frequency electrical quantities
Internal procedure; At permanent location (S) LB-IT06 w oparciu o PN-EN 61000-4-5:2010; LB-IT06A w oparciu o PN-EN 61000-4-5:2014-10

Reference standards (10)

LB-IT05
LB-IT05A
LB-IT06
LB-IT06A
LB-IT07
PN-EN 61000-4-2:2011
PN-EN 61000-4-4:2010 +A1:2010
PN-EN 61000-4-4:2013-05
PN-EN 61000-4-5:2010
PN-EN 61000-4-5:2014-10

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20145