🇵🇱 LABOSERWIS Sp. z o.o. PL
Metrology & calibration
LABOSERWIS Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 137 (Katowice, Poland). Its accreditation scope covers 11 calibration items in the following areas: Metrology & calibration. Calibrated items include: Aspirators, Electronic manometers, Electronic thermometers, - absolute pressure (medium - gas). Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Calibration in a thermostatic chamber; At perma…, Internal procedure; Calibration in real conditions; At permanent lo….
At a glance
- Accreditation number: PCA AP 137
- Calibration items in scope: 11
- Types of calibrated items: 10
- Calibration principles used: 5
- City: Katowice, Poland
- Valid until: 16-11-2030
- Accreditation number
- AP 137
- Accreditation body
- PCA
- Address
- ul. Chorzowska 73 A, 40-101 Katowice
- City
- Katowice, Poland
- Division
- Calibration Laboratory
- Valid until
- 16-11-2030
What this laboratory calibrates (10)
- Aspirators — 2 calibration items
- Electronic manometers — 2 calibration items
- Electronic thermometers — 2 calibration items
- - absolute pressure (medium - gas) — 1 calibration item
- - relative pressure (medium - gas) — 1 calibration item
- Bubble flow meters — 1 calibration item
- Gas dividers — 1 calibration item
- Multi-gas analyzers — 1 calibration item
- Total organic carbon (TOC) in the form of propane in syntetic air — 1 calibration item
- Total organic carbon analyzers — 1 calibration item
Calibration principles (5)
- Internal procedure; At permanent location (S) — 7 calibration items
- Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in real conditions; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in the KT-1 calibrator; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration using C3H8 in synthetic air; At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 30.10.2026
- Issue no.
- 20
- Certificate date
- 27.09.2019
- Accreditation cycle
- 30.10.2026 — 16.11.2030
- Approved by
- Tadeusz Matras
Accreditation scope — calibration items (11)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Total organic carbon analyzers | Calibration — Mole fraction (gas analysis) 0 µmol/mol to 80 µmol/mol · CMC 1,5 % 80 µmol/mol to 822 µmol/mol · CMC 1,2 % Mole fraction (gas analysis) |
Internal procedure; Calibration using C3H8 in synthetic air; At permanent location (S) | PW/J/A-01 | |
| 2 | Multi-gas analyzers | Calibration — Mole fraction (gas analysis) CO — 0 µmol/mol to 5000 µmol/mol · CMC 2,0 % CO₂ — 0 µmol/mol to 50·10⁴ µmol/mol · CMC 2,0 % NO — 0 µmol/mol to 2500 µmol/mol · CMC 2,5 % SO₂ — 0 µmol/mol to 3000 µmol/mol · CMC 3,0 % O₂ — 0 µmol/mol to 25·10⁴ µmol/mol · CMC 2,0 % NO₂1) — 0 µmol/mol to 200 µmol/mol · CMC 3,0 % Mole fraction (gas analysis) |
Internal procedure; At permanent location (S) | PW/J/A-02 | |
| 3 | Total organic carbon (TOC) in the form of propane in syntetic air | Calibration — Reference materials - measurement 50 µmol/mol to 55 µmol/mol · CMC 0,9 % 2) 71 µmol/mol to 79 µmol/mol · CMC 1,1 % 2) 236 µmol/mol to 262 µmol/mol · CMC 0,7 % 2) 473 µmol/mol to 523 µmol/mol · CMC 0,6 % 2) Reference materials - measurement |
Internal procedure; At permanent location (S) | PW/J/G-01 | |
| 4 | Bubble flow meters | Calibration — Volume flow (flow - gases) 1,2 dm³/h to 300 dm³/h · CMC 1,0 % Volume flow (flow - gases) |
Internal procedure; At permanent location (S) | PW/J/P-02 | |
| 5 | Aspirators | Calibration — Volume flow (flow - gases) 10 dm³/h to 50 dm³/h · CMC 1,5 % 50 dm³/h to 300 dm³/h · CMC 1,0 % Volume flow (flow - gases) |
Internal procedure; At permanent location (S) | PW/J/P-01 | |
| 6 | Aspirators | Calibration — Volume flow (flow - gases) 300 dm³/h to 750 dm³/h · CMC 1,0 % Volume flow (flow - gases) |
Internal procedure; At permanent location (S) | PW/J/P-04 | |
| 7 | Gas dividers | Calibration — Volume flow (flow - gases) 5 dm³/h to 10 dm³/h · CMC 3,0 % 10 dm³/h to 100 dm³/h · CMC 1,0 % Volume flow (flow - gases) |
Internal procedure; Calibration in real conditions; At permanent location (S) | PW/J/P-03 | |
| 8 | Electronic manometers - absolute pressure (medium - gas) |
Calibration — Pressure 700 hPa to 1100 hPa · CMC 0,55 hPa Pressure |
Internal procedure; At permanent location (S) | PW/J/C-01 based on EURAMET cg-17 v.4.1 clause 5.2.2 | |
| 9 | Electronic manometers - relative pressure (medium - gas) |
Calibration — Pressure 0 Pa to 250 Pa · CMC 0,50 Pa 250 Pa to 2500 Pa · CMC 2,0 Pa 2,5 kPa to 10 kPa · CMC 0,007 kPa 10 kPa to 100 kPa · CMC 0,07 kPa 100 kPa to 700 kPa · CMC 0,7 kPa Pressure |
Internal procedure; At permanent location (S) | PW/J/C-02 based on EURAMET cg-17 v.4.1 clause 5.2.2 | |
| 10 | Electronic thermometers | Calibration — Temperature (electric thermometry) 5 °C to 45 °C · CMC 0,5 °C Temperature (electric thermometry) |
Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) | PW/J/T-01 | |
| 11 | Electronic thermometers | Calibration — Temperature (electric thermometry) 0 °C to 50 °C · CMC 1,0 °C Temperature (electric thermometry) |
Internal procedure; Calibration in the KT-1 calibrator; At permanent location (S) | PW/J/T-02 | |
Reference standards (12)
EURAMET cg-17 v.4.1 pkt 5.2.2
PW/J/A-01
PW/J/A-02
PW/J/C-01
PW/J/C-02
PW/J/G-01
PW/J/P-01
PW/J/P-02
PW/J/P-03
PW/J/P-04
PW/J/T-01
PW/J/T-02