🇵🇱 Urząd Dozoru Technicznego PL
Metrology & calibration
Urząd Dozoru Technicznego is a calibration laboratory accredited by PCA under number AP 136 (Poznań, Poland). Its accreditation scope covers 28 calibration items in the following areas: Metrology & calibration. Calibrated items include: Digital tachometers, Technological devices, Universal Testing Machines, Digital displacement sensors / transducers. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S), At permanent location (S) and outside of it (P), Internal procedure; Direct method; At permanent location (S) and ou….
At a glance
- Accreditation number: PCA AP 136
- Calibration items in scope: 28
- Types of calibrated items: 30
- Calibration principles used: 9
- City: Poznań, Poland
- Valid until: 23-06-2030
- Accreditation number
- AP 136
- Accreditation body
- PCA
- Address
- ul. Ostrowska 388, 61-312 Poznań
- City
- Poznań, Poland
- Division
- Department of Metrology and Management WBP
- Valid until
- 23-06-2030
What this laboratory calibrates (30)
- Digital tachometers — 3 calibration items
- Technological devices — 3 calibration items
- Universal Testing Machines — 3 calibration items
- Digital displacement sensors / transducers — 2 calibration items
- Electric thermometers — 2 calibration items
- - Crosshead /Actuator Speed — 1 calibration item
- - Crosshead Displacement or Actuator Stroke — 1 calibration item
- - for compressive forces — 1 calibration item
- - for tensile forces — 1 calibration item
- Caliper depth gauges — 1 calibration item
- Calipers — 1 calibration item
- Circometer (diameter tapes) — 1 calibration item
- Coating thickness gauges — 1 calibration item
- Digital manometers — 1 calibration item
- Extensometers — 1 calibration item
- and 15 more
Calibration principles (9)
- Internal procedure; Direct method; At permanent location (S) — 16 calibration items
- At permanent location (S) and outside of it (P) — 4 calibration items
- Internal procedure; Direct method; At permanent location (S) and outside of it (P) — 2 calibration items
- At permanent location (S) — 1 calibration item
- Internal procedure; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method Contact method; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method Optical method; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method using a gauge blocks; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method using an interferometer; At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 16.06.2026
- Issue no.
- 18
- Accreditation cycle
- 16.06.2026 — 23.06.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (28)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Digital tachometers | Calibration — Speed 10,00 rpm to 99,99 rpm · CMC 0,1 rpm 100,0 rpm to 999,9 rpm · CMC 0,2 rpm 1000 rpm to 33000 rpm · CMC 2 rpm 33001 rpm to 50000 rpm · CMC 3 rpm Speed |
Internal procedure; Direct method Optical method; At permanent location (S) | LW-1/IN/203 | |
| 2 | Digital tachometers | Calibration — Speed 10,00 rpm to 99,99 rpm · CMC 0,1 rpm 100,0 rpm to 999,9 rpm · CMC 0,2 rpm 100 0 rpm to 20000 rpm · CMC 2 rpm Speed |
Internal procedure; Direct method Contact method; At permanent location (S) | LW-1/IN/203 | |
| 3 | Digital tachometers | Calibration — Speed 0,89 mpm to 5,00 mpm · CMC 0,02 mpm 5,01 mpm to 300,0 mpm · CMC 0,3 % Speed |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/203 | |
| 4 | Digital displacement sensors / transducers | Calibration — Length 0 mm to 150 mm · CMC 0,06 mm Length |
Internal procedure; Direct method using a gauge blocks; At permanent location (S) | LW-1/IN/202 | |
| 5 | Digital displacement sensors / transducers | Calibration — Length 0 mm to 3,5 mm · CMC 0,005 mm 3,5 mm to 500 mm · CMC 0,15 % Length |
Internal procedure; Direct method using an interferometer; At permanent location (S) | LW-1/IN/202 | |
| 6 | Laser distance meters | Calibration — Length 0 m to 5 m · CMC 1,0 mm 5,001 m do 30 m · CMC 2,5 mm 30,0001 m do 50 m · CMC 2,8 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/205 | |
| 7 | Caliper depth gauges | Calibration — Length 0 mm to 150 mm · CMC 0,06 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/202 | |
| 8 | Ultrasonic thickness gauges | Calibration — Length 0 mm to 2 mm · CMC 0,01 mm 2 mm to 6 mm · CMC 0,02 mm 6 mm to 100 mm · CMC 0,3 % Length |
At permanent location (S) | PN-EN ISO 16831:2025-11 Group 2 tests cl. 9.3, 9.13 | |
| 9 | Internal / External measuring gauge | Calibration — Length 0 mm to 50 mm · CMC 0,005 mm 50 mm do 100 mm · CMC 0,008 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/202 | |
| 10 | Coating thickness gauges | Calibration — Length 0 mm to 0,5 mm · CMC 0,003 mm 0,5 mm to 3 mm · CMC 0,7 % Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/210 | |
| 11 | Tape measures | Calibration — Length 0 m to 60 m · CMC 0,2 mm + 0,2 · (n – 1) mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/205 | |
| 12 | Semi-rigid rules Circometer (diameter tapes) |
Calibration — Length 0 m to 5 m · CMC 0,2 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/205 | |
| 13 | Weld gauges | Calibration — Length 0 mm to 35 mm · CMC 0,06 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/202 | |
| 14 | Calipers | Calibration — Length 0 mm to 500 mm · CMC 0,06 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/202 | |
| 15 | Feeler gauges | Calibration — Length 0,05 mm to 2 mm · CMC 0,007 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/209 | |
| 16 | Step wedges (calibration blocks) for ultrasonic thickness gauges | Calibration — Length 0,5 mm to 25 mm · CMC 0,007 mm Length |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/209 | |
| 17 | Extensometers | Calibration — Force 0 mm to 0,35 mm · CMC 0,5 µm 0,35 to 2500 mm · CMC 0,15 % Force |
At permanent location (S) and outside of it (P) | PN-EN ISO 9513:2013-06 ASTM E83-25 | |
| 18 | Universal Testing Machines Technological devices - for tensile forces - for compressive forces |
Calibration — Force 0,5 N to 250 kN · CMC 0,13 % Force |
At permanent location (S) and outside of it (P) | PN-EN ISO 7500-1:2018-05 ASTM E4-24 | |
| 19 | Universal Testing Machines Technological devices - Crosshead Displacement or Actuator Stroke |
Calibration — Force 0 mm to 0,35 mm · CMC 0,5 µm 0,35 mm to 5000 mm · CMC 0,15 % Force |
At permanent location (S) and outside of it (P) | PN-EN ISO 9513:2013-06 Annex H ASTM E2309/E2309M-2 | |
| 20 | Universal Testing Machines Technological devices - Crosshead /Actuator Speed |
Calibration — Force 0 mm/min to 0,5 mm/min · CMC 1 µm/min 0,5 mm/min to 3000 mm/min · CMC 0,2 % Force |
At permanent location (S) and outside of it (P) | ASTM E2658-15 | |
| 21 | Force gauges | Calibration — Force 0,5 N to 250 kN · CMC 0,13 % Force |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | LW-1/IN/212 based on PN-EN ISO 7500-1:2018-05 | |
| 22 | Portable hardness testers | Calibration — Hardness HV5 · CMC 3 % HV10 · CMC 3 % Hardness |
Internal procedure; At permanent location (S) | LW-1/IN/213 based on DIN 50159-2:2022-06 PN-EN ISO 6507-2:2018-05 | |
| 23 | UV-A irradiance meters for non-destructive testing (NDT) | Calibration — Radiometric quantities 0,1 mW/cm² to 5 mW/cm² · CMC 12 % Radiometric quantities |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/207 based on CIE 220:2016 | |
| 24 | Lux meters (illuminance meters) for non-destructive testing (NDT) |
Calibration — Illuminance 20 lx to 5000 lx · CMC 3,5 % Illuminance |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/207 | |
| 25 | Mechanical manometers Digital manometers |
Calibration — Pressure gauge pressure (gas medium) — -0,095 MPa to 0 MPa · CMC 0,00005 MPa gauge pressure (gas medium) — 0 MPa to 0,1 MPa · CMC 0,00003 MPa gauge pressure (gas medium) — 0,1 MPa to 3,5 MPa · CMC 4,1 · 10⁻⁴ · p gauge pressure (liquid medium) — 0,1 MPa to 1,4 MPa · CMC 6,2 · 10⁻⁴ MPa gauge pressure (liquid medium) — 1,4 MPa to 12 MPa · CMC 4,1 · 10⁻⁴ · p gauge pressure (liquid medium) — 12 MPa to 40 MPa · CMC 2,1 · 10⁻⁴ · p gauge pressure (liquid medium) — 40 MPa to 100 MPa · CMC 3,5 · 10⁻⁴ · p Pressure |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/201 based on EURAMET Calibration Guide No. 17 Version 4.1 (09/2022) | |
| 26 | Electric thermometers | Calibration — Temperature (electrical thermometry) -20 °C to 110 °C · CMC 0,18 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) and outside of it (P) | LW-1/IN/206 based on DKD-R 5-1 (11/2023) | |
| 27 | Electric thermometers | Calibration — Temperature (electrical thermometry) 110 °C to 200 °C · CMC 0,40 °C 200 °C to 550 °C · CMC 0,65 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/206 based on DKD-R 5-1 (11/2023) | |
| 28 | Pyrometers | Calibration — Temperature (radiation thermometry) -23 °C do 100 °C · CMC 0,9 °C 100 °C to 200 °C · CMC 1,5 °C 200 °C do 300 °C · CMC 1,8 °C 300 °C do 400 °C · CMC 2,1 °C 400 °C do 500 °C · CMC 2,3 °C 500 °C do 550 °C · CMC 2,6 °C Temperature (radiation thermometry) |
Internal procedure; Direct method; At permanent location (S) | LW-1/IN/206 based on MSL Technical Guide 22 (06.2009) | |
Reference standards (24)
ASTM E2309/E2309M-2
ASTM E2658-15
ASTM E4-24
ASTM E83-25
CIE 220:2016
DIN 50159-2:2022-06
DKD-R 5-1 (11/2023)
EURAMET Calibration Guide No. 17 Version 4.1 (09/2022)
LW-1/IN/201
LW-1/IN/202
LW-1/IN/203
LW-1/IN/205
LW-1/IN/206
LW-1/IN/207
LW-1/IN/209
LW-1/IN/210
LW-1/IN/212
LW-1/IN/213
MSL Technical Guide 22 (06.2009)
PN-EN ISO 16831:2025-11
PN-EN ISO 6507-2:2018-05
PN-EN ISO 7500-1:2018-05
PN-EN ISO 9513:2013-06
PN-EN ISO 9513:2013-06 Annex H