🇵🇱 Instytut Techniki Budowlanej PL

PCA AP 113 • Warszawa, Poland • 20 calibration items

Metrology & calibration

Instytut Techniki Budowlanej is a calibration laboratory accredited by PCA under number AP 113 (Warszawa, Poland). Its accreditation scope covers 20 calibration items in the following areas: Metrology & calibration. Calibrated items include: Bar sieves, Electrical thermometers (including electronic), Electrical thermometers (with temperature recording), Laboratory test sieves. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; Direct method; At permanent location (S) and ou….

At a glance

  • Accreditation number: PCA AP 113
  • Calibration items in scope: 20
  • Types of calibrated items: 22
  • Calibration principles used: 10
  • City: Warszawa, Poland
  • Valid until: 20-01-2027
Accreditation number
AP 113
Accreditation body
PCA
Address
ul. Filtrowa 1, 00-611 Warszawa
City
Warszawa, Poland
Division
Calibration Laboratory
Valid until
20-01-2027

What this laboratory calibrates (22)

Calibration principles (10)

Certificate information

Issue date
19.12.2025
Issue no.
23
Certificate date
23.11.2018
Accreditation cycle
21.12.2023 — 20.01.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (20)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Digital displacement sensors Calibration — Length
0 mm to 100 mm · CMC 34,0 µm
Length
Internal procedure; Outside of permanent location (P) PW-06
2 Analogue dial gauges with scale interval Calibration — Length
scale interval 0,01 mm — 0 mm to 10 mm · CMC 2,6 µm scale interval 0,01 mm — 0 mm to 30 mm · CMC 2,9 µm scale interval 0,01 mm — 0 mm to 50 mm · CMC 3,4 µm scale interval 0,01 mm — 0 mm to 100 mm · CMC 4,9 µm scale interval 0,001 mm — 0 mm to 5 mm · CMC 2,0 µm scale interval 0,002 mm — 0 mm to 12 mm · CMC 2,2 µm
Length
Internal procedure; At permanent location (S) PW-20 PW-29
3 Digital dial gauges with resolution Calibration — Length
resolution 0,01 mm — 0 mm to 13 mm · CMC 3,0 µm resolution 0,01 mm — 0 mm to 25 mm · CMC 3,1 µm resolution 0,01 mm — 0 mm to 50 mm · CMC 3,4 µm resolution 0,01 mm — 0 mm to 100 mm · CMC 4,2 µm resolution 0,001 mm — 0 mm to 13 mm · CMC 2,0 µm resolution 0,001 mm — 0 mm to 25 mm · CMC 2,1 µm resolution 0,001 mm — 0 mm to 50 mm · CMC 2,4 µm resolution 0,001 mm — 0 mm to 100 mm · CMC 3,5 µm resolution 0,0005 mm — 0 mm to 30 mm · CMC 1,7 µm resolution 0,0005 mm — 0 mm to 60 mm · CMC 2,2 µm resolution 0,0005 mm — 0 mm to 100 mm · CMC 3,5 µm
Length
Internal procedure; At permanent location (S) PW-20 PW-29
4 Callipers Calibration — Length
0 mm to 500 mm · CMC 30 µm
Length
Internal procedure; At permanent location (S) PW-01
5 External micrometers Calibration — Length
0 mm to 25 mm · CMC 1,8 µm 25 mm to 50 mm · CMC 2,2 µm
Length
Internal procedure; At permanent location (S) PW-05
6 Laboratory test sieves Calibration — Length
0,02 mm to 10 mm · CMC 3,0 μm
Length
Internal procedure; Measurement of sieve apertures with a microscope; At permanent location (S) PW-18
7 Laboratory test sieves Calibration — Length
10 mm to 50 mm · CMC 50 μm 50 mm to 200 mm · CMC 110 μm
Length
Internal procedure; Measurement of apertures with a calliper; At permanent location (S) PW-18
8 Bar sieves Calibration — Length
2,5 mm to 10 mm · CMC 3,0 μm
Length
Internal procedure; Measurement of sieve slots with a microscope; At permanent location (S) PW-18
9 Bar sieves Calibration — Length
10 mm to 50 mm · CMC 50 μm
Length
Internal procedure; Measurement of sieve slots with a calliper; At permanent location (S) PW-18
10 Geometric dimension testers Calibration — Length
0 mm do 250 mm · CMC 47 µm 250 mm do 500 mm · CMC 66 µm 0 mm do 300 mm · CMC 32 µm 0 mm to 25 mm · CMC 3,7 µm 0 mm do 30 mm · CMC 5,6 µm 0 mm do 60 mm · CMC 5,6 µm 0 mm do 100 mm · CMC 7,7 µm 0 mm do 10 mm · CMC 3,0 µm
Length
Internal procedure; At permanent location (S) PW-23
11 Electronic non-automatic weighing instruments Calibration — Mass (weighing instruments)
up to 610 g · CMC 8,2 · 10⁻⁵ % powyżej 610 g do 6,2 kg · CMC 3,3 · 10⁻⁴ % powyżej 6,2 kg do 31 kg · CMC 6,5 · 10⁻⁴ % powyżej 31 kg do 160 kg · CMC 2,7 · 10⁻³ %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PW-12 based on EURAMET cg-18 v. 4.0
12 Mass standards M1
Weights of accuracy class M1
Calibration — Mass (weights and mass standards)
20 kg · CMC 200 mg 10 kg · CMC 140 mg 5 kg · CMC 50 mg 2 kg · CMC 20 mg 1 kg · CMC 16 mg 500 g · CMC 5 mg 200 g · CMC 2 mg 100 g · CMC 1 mg 50 g · CMC 0,6 mg 20 g · CMC 0,5 mg 10 g · CMC 0,4 mg 5 g · CMC 0,4 mg 2 g · CMC 0,3 mg 1 g · CMC 0,2 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW-24 based on OIML R111-1:2004 Annex C
13 Mass standards M2
Weights of accuracy class M2
Calibration — Mass (weights and mass standards)
20 kg · CMC 1000 mg 10 kg · CMC 500 mg 5 kg · CMC 250 mg 2 kg · CMC 100 mg 1 kg · CMC 50 mg 500 g · CMC 25 mg 200 g · CMC 10 mg 100 g · CMC 5,0 mg 50 g · CMC 3,0 mg 20 g · CMC 2,5 mg 10 g · CMC 2,0 mg 5 g · CMC 1,6mg 2 g · CMC 1,2 mg 1 g · CMC 1,0 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW-24 based on OIML R111-1:2004 Annex C
14 Mass standards and weights 25 kg Calibration — Mass (weights and mass standards)
25 kg · CMC 250 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) PW-24 based on OIML R111-1:2004 Annex C
15 Load weights Calibration — Mass (weights and mass standards)
from 1 g to 600 g · CMC 0,0010 g
Mass (weights and mass standards)
Internal procedure; Direct method; At permanent location (S) PW-42
16 Load weights Calibration — Mass (weights and mass standards)
above 600 g to 6000 g · CMC 0,049 g above 6000 g to 31000 g · CMC 0,35 g
Mass (weights and mass standards)
Internal procedure; Direct method; At permanent location (S) and outside of it (P) PW-42
17 Electronic pressure gauges
Pressure transmitters
- gauge pressure (gas)
Calibration — Pressure
-9000 Pa to -1000 Pa · CMC 7 Pa -1000 Pa to 1000 Pa · CMC 1,4 Pa 1000 Pa to 9000 Pa · CMC 7 Pa
Pressure
Internal procedure; At permanent location (S) and outside of it (P) PW-39
18 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-30 °C to 140 °C · CMC 0,08 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PW-08
19 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-30 °C to 130 °C · CMC 0,10 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) and outside of it (P) PW-21
20 Thermostatic chambers
Liquid thermostats
Calibration — Temperature (electrical thermometry)
-30 °C to 140 °C · CMC 0,60 °C 1)
Temperature (electrical thermometry)
Internal procedure; Measurement of the distribution in the measurement space; At permanent location (S) and outside of it (P) PW-38

Reference standards (16)

EURAMET cg-18 v. 4.0
OIML R111-1:2004 Załącznik C
PW-01
PW-05
PW-06
PW-08
PW-12
PW-18
PW-20
PW-21
PW-23
PW-24
PW-29
PW-38
PW-39
PW-42

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20113