🇵🇱 Limatherm Sensor Sp. z o.o. PL

PCA AP 108 • Limanowa, Poland • 14 calibration items

Metrology & calibration

Limatherm Sensor Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 108 (Limanowa, Poland). Its accreditation scope covers 14 calibration items in the following areas: Metrology & calibration. Calibrated items include: Bimetallic dial thermometers, Electrical thermometers (including electronic), Electrical thermometers (with temperature recording), Expansion dial thermometers. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Outside of permanent location (P), Internal procedure; At permanent location (S) and outside of it (P).

At a glance

  • Accreditation number: PCA AP 108
  • Calibration items in scope: 14
  • Types of calibrated items: 13
  • Calibration principles used: 3
  • City: Limanowa, Poland
  • Valid until: 23-09-2027
Accreditation number
AP 108
Accreditation body
PCA
Address
ul. Skrudlak 1, 34-600 Limanowa
City
Limanowa, Poland
Division
Measurement Laboratory
Valid until
23-09-2027

What this laboratory calibrates (13)

Calibration principles (3)

Certificate information

Issue date
16.06.2025
Issue no.
16
Certificate date
09.09.2019
Accreditation cycle
08.09.2023 — 23.09.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (14)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Temperature transmitters Calibration — Electric quantity simulation
type: R, S, B — 0 °C to 1760 °C · CMC 0,55 °C type: J, K, N, T, E — -200 °C to 1370 °C · CMC 0,25 °C type: RTD — -200 °C to 0 °C · CMC 0,03 °C type: RTD — 0 °C to 850 °C · CMC 0,07 °C
Electric quantity simulation
Internal procedure; At permanent location (S) and outside of it (P) PPL-09, PPLW-01 based on EURAMET cg -11 v.2.0
2 Temperature simulators Calibration — Electric quantity simulation
type: R, S, B — -30 °C to 0 °C · CMC 0,6 °C type: R, S, B — 0 °C to 1820 °C · CMC 0,5 °C type: J, K, N, T, E — -270 °C to 1370 °C · CMC 0,3 °C type: RTD — -200 °C to 200 °C · CMC 0,03 °C type: RTD — 200 °C to 850 °C · CMC 0,05 °C
Electric quantity simulation
Internal procedure; At permanent location (S) and outside of it (P) PPL-09, PPLW-01 based on EURAMET cg -11 v.2.0
3 Temperature indicators (meters) (including temperature controllers)
Temperature recorders
Calibration — Electric quantity simulation
type: R, S, B — -30 °C to 0 °C · CMC 0,7 °C type: R, S, B — 0 °C to 1820 °C · CMC 0,4 °C type: J, K, N, T, E — -270 °C to 1370 °C · CMC 0,35 °C type: RTD — -200 °C to 200 °C · CMC 0,05 °C type: RTD — 200 °C to 850 °C · CMC 0,07 °C
Electric quantity simulation
Internal procedure; At permanent location (S) and outside of it (P) PPL-09, PPLW-01 based on EURAMET cg -11 v.2.0
4 Noble and base metal thermocouples Calibration — Temperature (electrical thermometry)
-30 °C do 250 °C · CMC 0,4 °C 250 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,0 °C 1200 °C to 1300 °C · CMC 1,4 °C 1300 °C do 1500 °C · CMC 1,9 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PPL-01
5 Noble and base metal thermocouples Calibration — Temperature (electrical thermometry)
-30 °C to 250 °C · CMC 0,6 °C 250 °C to 500 °C · CMC 0,7 °C 500 °C to 1000 °C · CMC 0,9 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PPL-01, PPLW-01
6 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-50 °C to -30 °C · CMC 0,08 °C -30 °C to 100 °C · CMC 0,02 °C 100 °C to 250 °C · CMC 0,04 °C 250 °C to 500 °C · CMC 0,18 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PPL-03
7 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-50 °C to 150 °C · CMC 0,10 °C 150 °C to 250 °C · CMC 0,15 °C 250 °C to 500 °C · CMC 0,22 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PPL-03
8 Temperature transmitters (containing temperature sensors) Calibration — Temperature (electrical thermometry)
-50 °C do 250 °C · CMC 0,10 °C 250 °C do 500 °C · CMC 0,20 °C 500 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,0 °C 1200 °C to 1300 °C · CMC 1,4 °C 1300 °C do 1500 °C · CMC 1,9 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PPL-05
9 Temperature transmitters (containing temperature sensors) Calibration — Temperature (electrical thermometry)
-50 °C to 150 °C · CMC 0,15 °C 150 °C to 250 °C · CMC 0,20 °C 250 °C to 500 °C · CMC 0,25 °C 500 °C to 1000 °C · CMC 0,9 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PPL-05, PPLW-01
10 Temperature calibrators Calibration — Temperature (electrical thermometry)
-50 °C to 150 °C · CMC 0,5 °C 150 °C to 400 °C · CMC 1,2 °C 400 °C to 1000 °C · CMC 2,5 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PPL-07
11 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-50 °C do -30 °C · CMC 0,08 °C -30 °C do 100 °C · CMC 0,02 °C 100 °C do 250 °C · CMC 0,04 °C 250 °C do 500 °C · CMC 0,18 °C 500 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,0 °C 1200 °C to 1300 °C · CMC 1,4 °C 1300 °C do 1500 °C · CMC 1,9 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PPL-04
12 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-50 °C to 150 °C · CMC 0,10 °C 150 °C to 250 °C · CMC 0,15 °C 250 °C to 500 °C · CMC 0,22 °C 500 °C to 1000 °C · CMC 0,9 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PPL-04, PPLW-01
13 Bimetallic dial thermometers
Pressure-spring dial thermometers
Expansion dial thermometers
Calibration — Temperature (non-electrical thermometry)
-50 °C to 50 °C · CMC 0,4 °C 50 °C to 400 °C · CMC 1,0 °C 400 °C to 600 °C · CMC 2,0 °C
Temperature (non-electrical thermometry)
Internal procedure; At permanent location (S) PPL-06
14 Bimetallic dial thermometers
Pressure-spring dial thermometers
Expansion dial thermometers
Calibration — Temperature (non-electrical thermometry)
-50 °C to 50 °C · CMC 0,5 °C 50 °C to 250 °C · CMC 1,0 °C 250 °C to 600 °C · CMC 2,0 °C
Temperature (non-electrical thermometry)
Internal procedure; Outside of permanent location (P) PPL-06, PPLW-01

Reference standards (9)

EURAMET cg -11 v.2.0
PPL-01
PPL-03
PPL-04
PPL-05
PPL-06
PPL-07
PPL-09
PPLW-01

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20108