🇵🇱 ERG95 Spółka z ograniczoną odpowiedzialnością PL

PCA AP 092 • Wrocław, Poland • 74 calibration items

Metrology & calibration

ERG95 Spółka z ograniczoną odpowiedzialnością is a calibration laboratory accredited by PCA under number AP 092 (Wrocław, Poland). Its accreditation scope covers 74 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Power network parameter meters, Calibrators, Clamp meters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Outside of permanent location (P), Internal procedure; Direct method; At permanent location (S).

At a glance

  • Accreditation number: PCA AP 092
  • Calibration items in scope: 74
  • Types of calibrated items: 54
  • Calibration principles used: 13
  • City: Wrocław, Poland
  • Valid until: 18-01-2030
Accreditation number
AP 092
Accreditation body
PCA
Address
ul. Gen. Juliana Filipowicza 7, 52-208 Wrocław
City
Wrocław, Poland
Division
Measurement Laboratory
Valid until
18-01-2030

What this laboratory calibrates (54)

Calibration principles (13)

Certificate information

Issue date
16.01.2026
Issue no.
25
Certificate date
20.12.2024
Accreditation cycle
16.01.2026 — 18.01.2030
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (74)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Transmitters Calibration — DC voltage
0,01 mV to 100 mV · CMC 2 µV 0,1 V to 1 V · CMC 5,0·10⁻⁶ · U + 9 µV 1 V to 10 V · CMC 5,0·10⁻⁶ · U + 0,09 mV 10 V to 100 V · CMC 4,0·10⁻⁶ · U + 0,9 mV
DC voltage
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 401
2 Transmitters Calibration — DC voltage
0,01 mV to 100 mV · CMC 2 µV 0,1 V to 1 V · CMC 14 µV 1 V to 10 V · CMC 0,14 mV 10 V to 100 V · CMC 1,4 mV
DC voltage
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 401
3 Multimeters
Analogue voltmeters
Digital voltmeters
Power network parameter meters
Clamp meters
Data acquisition cards
Breakdown testers
Electronic loads
Calibration — DC voltage
0,0001 mV to 10 mV · CMC 0,9 µV 10 mV to 100 mV · CMC 1,0·10⁻⁵ · U + 0,9 µV 0,1 V to 1 V · CMC 9,0·10⁻⁶ · U + 2 µV 1 V to 10 V · CMC 9,0·10⁻⁶ · U + 7 µV 10 V to 100 V · CMC 7,0·10⁻⁶ · U + 0,095 mV 100 V to 1050 V · CMC 5,0·10⁻⁶ · U + 1 mV
DC voltage
Internal procedure; At permanent location (S) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
4 Multimeters
Analogue voltmeters
Digital voltmeters
Power network parameter meters
Clamp meters
Data acquisition cards
Breakdown testers
Electronic loads
Calibration — DC voltage
0,001 mV to 100 mV · CMC 1,9 µV 0,1 V to 1 V · CMC 11 µV 1 V to 10 V · CMC 0,11 mV 10 V to 100 V · CMC 0,92 mV 100 V to 1050 V · CMC 12 mV
DC voltage
Internal procedure; Outside of permanent location (P) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
5 Calibrators
Power supplies
Calibration — DC voltage
0,0001 mV to 10 mV · CMC 0,9 µV 10 mV to 100 mV · CMC 1,0·10⁻⁵ · U + 0,9 µV 0,1 V to 1 V · CMC 9,0·10⁻⁶ · U + 2 µV 1 V to 10 V · CMC 9,0·10⁻⁶ · U + 7 µV 10 V to 100 V · CMC 7,0·10⁻⁶ · U + 0,095 mV 100 V to 1050 V · CMC 5,0·10⁻⁶ · U + 1 mV
DC voltage
Internal procedure; At permanent location (S) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
6 Calibrators
Power supplies
Calibration — DC voltage
0,001 mV to 100 mV · CMC 1,9 µV 0,1 V to 1 V · CMC 11 µV 1 V to 10 V · CMC 0,11 mV 10 V to 100 V · CMC 0,92 mV 100 V to 1050 V · CMC 12 mV
DC voltage
Internal procedure; Outside of permanent location (P) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
7 Transmitters Calibration — DC current
0,1 µA to 100 µA · CMC 1,2·10⁻⁴ · I + 4 nA 0,1 mA to 1 mA · CMC 6,0·10⁻⁵ · I + 11 nA 1 mA to 10 mA · CMC 6,0·10⁻⁵ · I + 0,15 µA 10 mA to 100 mA · CMC 6,0·10⁻⁵ · I + 1,2 µA 0,1 A to 1 A · CMC 1,6·10⁻⁴ · I + 1 µA 1 A to 3 A · CMC 1,0·10⁻⁴ · I + 0,07 mA 3 A to 10 A · CMC 5,0·10⁻⁴ · I + 0,04 mA
DC current
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 401
8 Transmitters Calibration — DC current
0,001 mA to 20 mA · CMC 3,5·10⁻⁴ · I + 1,2 µA 20 mA to 200 mA · CMC 3,5·10⁻⁴ · I + 12 µA 0,2 A to 2 A · CMC 3,5·10⁻⁴ · I + 0,2 mA 2 A to 10 A · CMC 3,5·10⁻⁴ · I + 1,2 mA
DC current
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 401
9 Multimeters
Analogue ammeters
Digital ammeters
Power network parameter meters
Data acquisition cards
Electronic loads
Calibration — DC current
0,1 µA to 100 µA · CMC 1,2·10⁻⁴ · I + 4 nA 0,1 mA to 1 mA · CMC 6,0·10⁻⁵ · I + 11 nA 1 mA to 10 mA · CMC 6,0·10⁻⁵ · I + 0,15 µA 10 mA to 100 mA · CMC 6,0·10⁻⁵ · I + 1,2 µA 0,1 A to 1 A · CMC 1,6·10⁻⁴ · I + 1 µA 1 A to 3 A · CMC 1,0·10⁻⁴ · I + 0,07 mA 3 A to 10 A · CMC 5,0·10⁻⁴ · I + 0,04 mA
DC current
Internal procedure; At permanent location (S) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
10 Multimeters
Analogue ammeters
Digital ammeters
Power network parameter meters
Data acquisition cards
Electronic loads
Calibration — DC current
0,001 mA to 20 mA · CMC 3,5·10⁻⁴ · I + 1,2 µA 20 mA to 200 mA · CMC 3,5·10⁻⁴ · I + 12 µA 0,2 A to 2 A · CMC 3,5·10⁻⁴ · I + 0,2 mA 2 A to 10 A · CMC 3,5·10⁻⁴ · I + 1,2 mA
DC current
Internal procedure; Outside of permanent location (P) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
11 Clamp meters Calibration — DC current
0,1 A to 1 A · CMC 0,14 mA 1 A to 10 A · CMC 1,4 mA 10 A to 100 A · CMC 2,4·10⁻³ · I + 0,4 A 100 A to 1000 A · CMC 7,0·10⁻⁴ · I + 1,2 A
DC current
Internal procedure; At permanent location (S) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
12 Clamp meters Calibration — DC current
2 A to 20 A · CMC 2 mA 20 A to 1000 A · CMC 7,0·10⁻⁴ · I + 1,2 A
DC current
Internal procedure; Outside of permanent location (P) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
13 Calibrators
Power supplies
Calibration — DC current
0,1 µA to 100 µA · CMC 1,2·10⁻⁴ · I + 4 nA 0,1 mA to 1 mA · CMC 6,0·10⁻⁵ · I + 11 nA 1 mA to 10 mA · CMC 6,0·10⁻⁵ · I + 0,15 µA 10 mA to 100 mA · CMC 6,0·10⁻⁵ · I + 1,2 µA 0,1 A to 1 A · CMC 1,6·10⁻⁴ · I + 1 µA 1 A to 3 A · CMC 1,0·10⁻⁴ · I + 0,07 mA 3 A to 10 A · CMC 5,0·10⁻⁴ · I + 0,04 mA 10 A to 200 A · CMC 4,0·10⁻⁴ · I + 5 mA
DC current
Internal procedure; At permanent location (S) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
14 Calibrators
Power supplies
Calibration — DC current
0,001 mA to 20 mA · CMC 3,5·10⁻⁴ · I + 1,2 µA 20 mA to 200 mA · CMC 3,5·10⁻⁴ · I + 12 µA 0,2 A to 2 A · CMC 3,5·10⁻⁴ · I + 0,2 mA 2 A to 10 A · CMC 3,5·10⁻⁴ · I + 1,2 mA 10 A to 200 A · CMC 4,0·10⁻⁴ · I + 5 mA
DC current
Internal procedure; Outside of permanent location (P) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
15 Multimeters
Clamp meters
Analogue voltmeters
Digital voltmeters
Leakage current meters
Power network parameter meters
Breakdown voltage meters
Residual current device testers
Breakdown testers
Data acquisition cards
Calibration — AC voltage
40 Hz to 5000 Hz — 0,1 mV to 10 mV · CMC 9 µV 40 Hz to 5000 Hz — 10 mV to 100 mV · CMC 1,1·10⁻⁴ · U + 8 µV 40 Hz to 5000 Hz — 0,1 V to 1 V · CMC 6,0·10⁻⁵ · U + 0,03 mV 40 Hz to 5000 Hz — 1 V to 10 V · CMC 1,1·10⁻⁴ · U + 0,07 mV 40 Hz to 5000 Hz — 10 V to 100 V · CMC 1,2·10⁻⁴ · U + 0,7 mV 40 Hz to 5000 Hz — 100 V to 1050 V · CMC 9,0·10⁻⁵ · U + 9 mV 5 kHz to 20 kHz — 0,1 mV to 100 mV · CMC 3,0·10⁻⁴ · U + 7 µV 5 kHz to 20 kHz — 0,1 V to 1 V · CMC 6,0·10⁻⁵ · U + 0,04 mV 5 kHz to 20 kHz — 1 V to 10 V · CMC 1,1·10⁻⁴ · U + 0,07 mV 5 kHz to 20 kHz — 10 V to 110 V · CMC 1,2·10⁻⁴ · U + 0,7 mV 20 kHz to 50 kHz — 0,1 mV to 100 mV · CMC 7,0·10⁻⁴ · U + 8 µV 20 kHz to 50 kHz — 0,1 V to 1 V · CMC 1,0·10⁻⁴ · U + 0,07 mV 20 kHz to 50 kHz — 1 V to 10 V · CMC 0,044 %
AC voltage
Internal procedure; At permanent location (S) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
16 Multimeters
Clamp meters
Analogue voltmeters
Digital voltmeters
Leakage current meters
Power network parameter meters
Breakdown voltage meters
Residual current device testers
Breakdown testers
Data acquisition cards
Calibration — AC voltage
40 Hz to 5000 Hz — 0,1 mV to 200 mV · CMC 7,0·10⁻³ · U + 0,05 mV 40 Hz to 5000 Hz — 0,2 V to 2 V · CMC 1,2·10⁻³ · U + 0,5 mV 40 Hz to 5000 Hz — 2 V to 20 V · CMC 1,2·10⁻³ · U + 5 mV 40 Hz to 5000 Hz — 20 V to 200 V · CMC 1,2·10⁻³ · U + 0,05 V 40 Hz to 2000 Hz — 200 V to 1050 V · CMC 1,2·10⁻³ · U + 0,24 V
AC voltage
Internal procedure; Outside of permanent location (P) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
17 Calibrators
Generators
Power supplies
Calibration — AC voltage
40 Hz to 5000 Hz — 0,1 mV to 10 mV · CMC 9 µV 40 Hz to 5000 Hz — 10 mV to 100 mV · CMC 1,1·10⁻⁴ · U + 8 µV 40 Hz to 5000 Hz — 0,1 V to 1 V · CMC 6,0·10⁻⁵ · U + 0,03 mV 40 Hz to 5000 Hz — 1 V to 10 V · CMC 1,1·10⁻⁴ · U + 0,07 mV 40 Hz to 5000 Hz — 10 V to 100 V · CMC 1,2·10⁻⁴ · U + 0,7 mV 40 Hz to 5000 Hz — 100 V to 1050 V · CMC 9,0·10⁻⁵ · U + 9 mV 5 kHz to 20 kHz — 0,1 mV to 100 mV · CMC 3,0·10⁻⁴ · U + 7 µV 5 kHz to 20 kHz — 0,1 V to 1 V · CMC 6,0·10⁻⁵ · U + 0,04 mV 5 kHz to 20 kHz — 1 V to 10 V · CMC 1,1·10⁻⁴ · U + 0,07 mV 5 kHz to 20 kHz — 10 V to 110 V · CMC 1,2·10⁻⁴ · U + 0,7 mV 20 kHz to 50 kHz — 0,1 mV to 100 mV · CMC 7,0·10⁻⁴ · U + 8 µV 20 kHz to 50 kHz — 0,1 V to 1 V · CMC 1,0·10⁻⁴ · U + 0,07 mV 20 kHz to 50 kHz — 1 V to 10 V · CMC 0,044 %
AC voltage
Internal procedure; At permanent location (S) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
18 Calibrators
Generators
Power supplies
Calibration — AC voltage
40 Hz to 5000 Hz — 0,1 mV to 200 mV · CMC 1,2·10⁻³ · U + 0,05 mV 40 Hz to 5000 Hz — 0,2 V to 2 V · CMC 1,2·10⁻³ · U + 0,5 mV 40 Hz to 5000 Hz — 2 V to 20 V · CMC 1,2·10⁻³ · U + 5 mV 40 Hz to 5000 Hz — 20 V to 200 V · CMC 1,2·10⁻³ · U + 0,05 V 40 Hz to 2000 Hz — 200 V to 1050 V · CMC 1,2·10⁻³ · U + 0,24 V
AC voltage
Internal procedure; Outside of permanent location (P) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
19 Multimeters
Analogue ammeters
Digital ammeters
Leakage current meters
Leakage current simulators
Power network parameter meters
Residual current device testers
Data acquisition cards
Calibration — AC current
40 Hz to 5000 Hz — 10 µA to 100 µA · CMC 1,7·10⁻⁴ · I + 19 nA 40 Hz to 5000 Hz — 0,1 mA to 1 mA · CMC 2,7·10⁻⁴ · I + 0,02 µA 40 Hz to 5000 Hz — 1 mA to 10 mA · CMC 2,7·10⁻⁴ · I + 0,14 µA 40 Hz to 5000 Hz — 10 mA to 100 mA · CMC 2,7·10⁻⁴ · I + 1,4 µA 40 Hz to 5000 Hz — 0,1 A to 1 A · CMC 0,075 % 40 Hz to 5000 Hz — 1 A to 10 A · CMC 0,12 %
AC current
Internal procedure; At permanent location (S) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
20 Multimeters
Analogue ammeters
Digital ammeters
Leakage current meters
Leakage current simulators
Power network parameter meters
Residual current device testers
Data acquisition cards
Calibration — AC current
40 Hz to 5000 Hz — 0,1 mA to 20 mA · CMC 2 µA 40 Hz to 5000 Hz — 20 mA to 200 mA · CMC 0,02 mA 40 Hz to 5000 Hz — 0,2 A to 2 A · CMC 0,15 mA 40 Hz to 5000 Hz — 2 A to 10 A · CMC 1 mA
AC current
Internal procedure; Outside of permanent location (P) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
21 Clamp meters Calibration — AC current
45 Hz to 400 Hz — 0,1 A to 1 A · CMC 0,14 mA 45 Hz to 400 Hz — 1 A to 10 A · CMC 1,4 mA 45 Hz to 400 Hz — 10 A to 100 A · CMC 2,4·10⁻³ · I + 0,4 A 45 Hz to 400 Hz — 100 A to 1000 A · CMC 7,0·10⁻⁴ · I + 1,2 A
AC current
Internal procedure; At permanent location (S) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
22 Clamp meters Calibration — AC current
45 Hz to 400 Hz — 2 A to 20 A · CMC 2 mA 45 Hz to 400 Hz — 20 A to 1000 A · CMC 7,0·10⁻⁴ · I + 1,2 A
AC current
Internal procedure; Outside of permanent location (P) Internal procedure PT 402 based on EURAMET cg-15 v. 3.0
23 Calibrators
Generators
Power supplies
Calibration — AC current
40 Hz to 5000 Hz — 10 µA to 100 µA · CMC 1,7·10⁻⁴ · I + 19 nA 40 Hz to 5000 Hz — 0,1 mA to 1 mA · CMC 2,7·10⁻⁴ · I + 0,02 µA 40 Hz to 5000 Hz — 1 mA to 10 mA · CMC 2,7·10⁻⁴ · I + 0,14 µA 40 Hz to 5000 Hz — 10 mA to 100 mA · CMC 2,7·10⁻⁴ · I + 1,4 µA 40 Hz to 5000 Hz — 0,1 A to 1 A · CMC 0,075 % 40 Hz to 5000 Hz — 1 A to 10 A · CMC 0,12 %
AC current
Internal procedure; At permanent location (S) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
24 Calibrators
Generators
Power supplies
Calibration — AC current
40 Hz to 5000 Hz — 0,1 mA to 20 mA · CMC 2 µA 40 Hz to 5000 Hz — 20 mA to 200 mA · CMC 0,02 mA 40 Hz to 5000 Hz — 0,2 A to 2 A · CMC 0,15 mA 40 Hz to 5000 Hz — 2 A to 10 A · CMC 1 mA
AC current
Internal procedure; Outside of permanent location (P) Internal procedure PT 403 based on EURAMET cg-15 v. 3.0
25 Analogue ohmmeters
Digital ohmmeters
Multimeters
Transmitters
Calibration — DC resistance
0,001 Ω to 1 Ω · CMC 2,0·10⁻⁵ · R + 0,6 µΩ 1 Ω to 10 Ω · CMC 2,0·10⁻⁵ · R + 1 µΩ 10 Ω to 100 Ω · CMC 2,0·10⁻⁵ · R + 0,01 mΩ 0,1 kΩ to 1 kΩ · CMC 0,002 % 1 kΩ to 10 kΩ · CMC 0,003 % 10 kΩ to 100 kΩ · CMC 0,003 % 0,1 MΩ to 1 MΩ · CMC 0,004 % 1 MΩ to 10 MΩ · CMC 0,008 % 10 MΩ to 100 MΩ · CMC 0,07 % 0,1 GΩ to 100 GΩ · CMC 0,6 %
DC resistance
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 401
26 Analogue ohmmeters
Digital ohmmeters
Multimeters
Transmitters
Calibration — DC resistance
0,001 Ω to 100 Ω · CMC 1,2·10⁻⁴ · R + 0,005 Ω 0,1 kΩ to 1 kΩ · CMC 0,023 % 1 kΩ to 10 kΩ · CMC 0,023 % 10 kΩ to 100 kΩ · CMC 0,023 % 0,1 MΩ to 1 MΩ · CMC 0,023 % 1 MΩ to 10 MΩ · CMC 0,06 % 10 MΩ to 100 MΩ · CMC 0,4 % 0,1 GΩ to 100 GΩ · CMC 1,8 %
DC resistance
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 401
27 Resistance calibrators
Power network parameter meters
Insulation resistance meters
Calibration — DC resistance
0,001 Ω to 1 Ω · CMC 2,0·10⁻⁵ · R + 0,6 µΩ 1 Ω to 10 Ω · CMC 2,0·10⁻⁵ · R + 1 µΩ 10 Ω to 100 Ω · CMC 2,0·10⁻⁵ · R + 0,01 mΩ 0,1 kΩ to 1 kΩ · CMC 0,002 % 1 kΩ to 10 kΩ · CMC 0,003 % 10 kΩ to 100 kΩ · CMC 0,003 % 0,1 MΩ to 1 MΩ · CMC 0,004 % 1 MΩ to 10 MΩ · CMC 0,008 % 10 MΩ to 100 MΩ · CMC 0,07 % 0,1 GΩ to 100 GΩ · CMC 0,6 %
DC resistance
Internal procedure; At permanent location (S) Internal procedure PT 402, PT 403 based on EURAMET cg-15 v. 3.0
28 Resistance calibrators
Power network parameter meters
Insulation resistance meters
Calibration — DC resistance
0,001 Ω to 100 Ω · CMC 1,2·10⁻⁴ · R + 0,005 Ω 0,1 kΩ to 1 kΩ · CMC 0,023 % 1 kΩ to 10 kΩ · CMC 0,023 % 10 kΩ to 100 kΩ · CMC 0,023 % 0,1 MΩ to 1 MΩ · CMC 0,023 % 1 MΩ to 10 MΩ · CMC 0,06 % 10 MΩ to 100 MΩ · CMC 0,4 % 0,1 GΩ to 100 GΩ · CMC 1,8 %
DC resistance
Internal procedure; Outside of permanent location (P) Internal procedure PT 402, PT 403 based on EURAMET cg-15 v. 3.0
29 Fixed resistors
Variable resistors
Resistance standards
Shunts
Calibration — DC resistance
0,001 Ω to 0,1 Ω · CMC 5,0·10⁻³ · R + 0,1 µΩ 0,1 Ω to 10 Ω · CMC 6,0·10⁻⁵ · R + 0,02 mΩ 10 Ω to 100 Ω · CMC 6,0·10⁻⁵ · R + 0,3 mΩ 0,1 kΩ to 1 kΩ · CMC 6,0·10⁻⁵ · R 1 kΩ to 10 kΩ · CMC 6,0·10⁻⁵ · R 10 kΩ to 100 kΩ · CMC 6,0·10⁻⁵ · R 0,1 MΩ to 1 MΩ · CMC 7,0·10⁻⁵ · R 1 MΩ to 10 MΩ · CMC 1,0·10⁻⁴ · R 10 MΩ to 100 MΩ · CMC 6,0·10⁻⁴ · R 100 MΩ to 1000 MΩ · CMC 2,0·10⁻³ · R 1 GΩ to 2 GΩ · CMC 2,0·10⁻³ · R 2 GΩ to 20 GΩ · CMC 5,0·10⁻³ · R 20 GΩ to 100 GΩ · CMC 5,0·10⁻³ · R
DC resistance
Internal procedure; Direct and indirect method; At permanent location (S) Internal procedure PT 404
30 Fixed resistors
Variable resistors
Resistance standards
Shunts
Calibration — DC resistance
0,001 Ω to 0,1 Ω · CMC 5,0·10⁻³ · R + 0,1 µΩ 0,1 Ω to 10 Ω · CMC 6,0·10⁻⁵ · R + 0,02 mΩ 10 Ω to 100 Ω · CMC 6,0·10⁻⁵ · R + 0,3 mΩ 0,1 kΩ to 1 kΩ · CMC 6,0·10⁻⁵ · R 1 kΩ to 10 kΩ · CMC 6,0·10⁻⁵ · R 10 kΩ to 100 kΩ · CMC 6,0·10⁻⁵ · R 0,1 MΩ to 1 MΩ · CMC 7,0·10⁻⁵ · R 1 MΩ to 10 MΩ · CMC 1,0·10⁻⁴ · R 10 MΩ to 100 MΩ · CMC 6,0·10⁻⁴ · R 100 MΩ to 1000 MΩ · CMC 2,0·10⁻³ · R
DC resistance
Internal procedure; Direct and indirect method; Outside of permanent location (P) Internal procedure PT 404
31 Temperature transmitters Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,07 oC -200 °C do 1820 °C · CMC 0,24 oC
Electrical simulation of quantities
Internal procedure; Calibration by the electrical method; At permanent location (S) Internal procedure PT 401
32 Temperature transmitters Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,36 °C -200 °C do 1820 °C · CMC 0,98 °C
Electrical simulation of quantities
Internal procedure; Calibration by the electrical method; Outside of permanent location (P) Internal procedure PT 401
33 Temperature indicators (meters) (including temperature controllers)
Temperature recorders
Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,07 °C -200 °C do 1820 °C · CMC 0,24 °C
Electrical simulation of quantities
Internal procedure; Calibration by the electrical method; At permanent location (S) Internal procedure PT 402
34 Temperature indicators (meters) (including temperature controllers)
Temperature recorders
Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,36 °C -200 °C do 1820 °C · CMC 0,98 °C
Electrical simulation of quantities
Internal procedure; Calibration by the electrical method; Outside of permanent location (P) Internal procedure PT 402
35 Temperature simulators Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,07 oC -200 °C do 1820 °C · CMC 0,24 oC
Electrical simulation of quantities
Internal procedure; Calibration by the electrical method; At permanent location (S) Internal procedure PT 403
36 Temperature simulators Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,36 oC -200 °C do do 1820 °C · CMC 0,98 oC
Electrical simulation of quantities
Internal procedure; Calibration by the electrical method; Outside of permanent location (P) Internal procedure PT 403
37 Transmitters Calibration — Relative humidity
zakres temperatur 5 °C to 60 °C — 10 %rh to 90 %rh · CMC 2,0 %rh
Relative humidity
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 501
38 Transmitters Calibration — Relative humidity
zakres temperatur 5 °C to 60 °C — 10 %rh to 90 %rh · CMC 3,0 %rh
Relative humidity
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 501
39 Hygrometers
Thermohygrometers
Thermohygrobarometers
Calibration — Relative humidity
zakres temperatur 5 °C to 60 °C — 10 %rh to 90 %rh · CMC 1,7 %rh
Relative humidity
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 502
40 Hygrometers
Thermohygrometers
Thermohygrobarometers
Calibration — Relative humidity
zakres temperatur 5 °C to 60 °C — 10 %rh to 90 %rh · CMC 3,0 %rh
Relative humidity
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 502
41 Climatic chambers Calibration — Relative humidity
zakres temperatur 5 °C to 60 °C — 10 % to 90 %rh · CMC 2,5 %rh
Relative humidity
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 503 based on EURAMET cg-20 v. 5.0
42 Climatic chambers Calibration — Relative humidity
zakres temperatur 5 °C to 60 °C — 10 %rh to 90 %rh · CMC 3,0 %rh
Relative humidity
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 503 based on EURAMET cg-20 v. 5.0
43 Bourdon tube pressure gauges
Electronic pressure gauges
Pressure transmitters
Calibration — Pressure
- gauge pressure, negative and positive pressure (gas) — -0,1 MPa to -0,03 MPa · CMC 3,0·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — -0,03 MPa to -0,0024 MPa · CMC 1,1·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — -0,0024 MPa to 0,0024 MPa · CMC 2,0·10⁻⁶ MPa - gauge pressure, negative and positive pressure (gas) — 0,0024 MPa to 0,01 MPa · CMC 6,2·10⁻⁶ MPa - gauge pressure, negative and positive pressure (gas) — 0,01 MPa to 0,03 MPa · CMC 1,1·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — 0,03 MPa to 0,2 MPa · CMC 4,4·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — 0,2 MPa to 1,8 MPa · CMC 6·10⁻⁶ P + 0,0002 MPa
Pressure
Internal procedure; At permanent location (S) Internal procedure PT 301, PT 302 based on EURAMET cg-17 v. 4.0
44 Bourdon tube pressure gauges
Electronic pressure gauges
Pressure transmitters
Calibration — Pressure
- gauge pressure, negative and positive pressure (gas) — -0,1 MPa to -0,005 MPa · CMC 1,2·10⁻⁴ MPa - gauge pressure, negative and positive pressure (gas) — -0,005 MPa to 0 MPa · CMC 5,5·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — 0 MPa to 0,002 MPa · CMC 0,6·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — 0,002 MPa to 0,005 MPa · CMC 5,5·10⁻⁵ MPa - gauge pressure, negative and positive pressure (gas) — 0,005 MPa to 0,3 MPa · CMC 1,2·10⁻⁴ MPa - gauge pressure, negative and positive pressure (gas) — 0,3 MPa to 1,0 MPa · CMC 1,2·10⁻³ MPa
Pressure
Internal procedure; Outside of permanent location (P) Internal procedure PT 301, PT 302 based on EURAMET cg-17 v. 4.0
45 Bourdon tube pressure gauges
Electronic pressure gauges
Pressure transmitters
Calibration — Pressure
- gauge pressure, negative and positive pressure (liquid) — 0 MPa to 0,1 MPa · CMC 3,4·10⁻⁵ MPa - gauge pressure, negative and positive pressure (liquid) — 0,1 MPa to 0,6 MPa · CMC 6,3·10⁻⁵ MPa - gauge pressure, negative and positive pressure (liquid) — 0,6 MPa to 2,0 MPa · CMC 6·10⁻⁶ · P + 0,0002 MPa - gauge pressure, negative and positive pressure (liquid) — 2,0 MPa to 70,0 MPa · CMC 3·10⁻⁶ · P + 0,002 MPa
Pressure
Internal procedure; At permanent location (S) Internal procedure PT 301, PT 302 based on EURAMET cg-17 v. 4.0
46 Bourdon tube pressure gauges
Electronic pressure gauges
Pressure transmitters
Calibration — Pressure
- gauge pressure, negative and positive pressure (liquid) — 0 MPa to 1,0 MPa · CMC 2,3·10⁻⁴ MPa - gauge pressure, negative and positive pressure (liquid) — 1,0 MPa to 3,0 MPa · CMC 4,7·10⁻⁴ MPa - gauge pressure, negative and positive pressure (liquid) — 3,0 MPa to 10,0 MPa · CMC 2,3·10⁻³ MPa - gauge pressure, negative and positive pressure (liquid) — 10,0 MPa to 30,0 MPa · CMC 4,7·10⁻³ MPa - gauge pressure, negative and positive pressure (liquid) — 30,0 MPa to 70,0 MPa · CMC 1,0·10⁻² MPa
Pressure
Internal procedure; Outside of permanent location (P) Internal procedure PT 301, PT 302 based on EURAMET cg-17 v. 4.0
47 Bourdon tube pressure gauges
Electronic pressure gauges (including barometers)
Pressure transmitters
Barometers
Calibration — Pressure
- absolute pressure — 0,0015 MPa abs to 0,12 MPa abs · CMC 1,5·10⁻⁴ MPa - absolute pressure — 0,12 MPa abs to 0,3 MPa abs · CMC 1,8·10⁻⁴ MPa - absolute pressure — 0,3 MPa abs to 0,6 MPa abs · CMC 2,0·10⁻⁴ MPa - absolute pressure — 0,6 MPa abs to 1,0 MPa abs · CMC 2,3·10⁻⁴ MPa - absolute pressure — 1,0 MPa abs to 3,0 MPa abs · CMC 4,7·10⁻⁴ MPa
Pressure
Internal procedure; At permanent location (S) Internal procedure PT 301, PT 302 based on EURAMET cg-17 v. 4.0
48 Bourdon tube pressure gauges
Electronic pressure gauges (including barometers)
Pressure transmitters
Barometers
Calibration — Pressure
- absolute pressure — 0,0015 MPa abs to 0,3 MPa abs · CMC 2,4·10⁻⁴ MPa - absolute pressure — 0,3 MPa abs to 1,0 MPa abs · CMC 6,0·10⁻⁴ MPa - absolute pressure — 1,0 MPa abs to 3,0 MPa abs · CMC 1,8·10⁻³ MPa
Pressure
Internal procedure; Outside of permanent location (P) Internal procedure PT 301, PT 302 based on EURAMET cg-17 v. 4.0
49 Barometers Calibration — Pressure
- absolute pressure — 300 hPa abs to 1200 hPa abs · CMC 1,5 hPa
Pressure
Internal procedure; At permanent location (S) Internal procedure PT 302 based on EURAMET cg-17 v. 4.0
50 Barometers Calibration — Pressure
- absolute pressure — 300 hPa abs to 1200 hPa abs · CMC 2,4 hPa
Pressure
Internal procedure; Outside of permanent location (P) Internal procedure PT 302 based on EURAMET cg-17 v. 4.0
51 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,03 °C 100 °C to 200 °C · CMC 0,05 °C 200 °C do 650 °C · CMC 0,15 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 201
52 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-35 °C to 150 °C · CMC 0,05 °C 150 °C to 650 °C · CMC 0,15 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 201
53 Noble metal thermocouples
Base metal thermocouples
Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,3 °C 100 °C to 200 °C · CMC 0,4 °C 200 °C do 650 °C · CMC 0,5 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 201 based on EURAMET cg-8 v. 3.1
54 Noble metal thermocouples
Base metal thermocouples
Calibration — Temperature (electrical thermometry)
-35 °C do 150 °C · CMC 0,5 °C 150 °C do 650 °C · CMC 0,7 °C 650 °C do 1085 °C · CMC 1,0 °C 1085 °C to 1200 °C · CMC 1,5 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 201 based on EURAMET cg-8 v. 3.1
55 Temperature transmitters (with temperature sensor) Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,03 °C 100 °C to 200 °C · CMC 0,05 °C 200 °C do 650 °C · CMC 0,15 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 202
56 Temperature transmitters (with temperature sensor) Calibration — Temperature (electrical thermometry)
-35 °C do 150 °C · CMC 0,1 °C 150 °C do 650 °C · CMC 0,2 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 202
57 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,03 °C 100 °C to 200 °C · CMC 0,05 °C 200 °C do 650 °C · CMC 0,15 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 203
58 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-35 °C do 150 °C · CMC 0,1 °C 150 °C do 650 °C · CMC 0,2 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 203
59 Liquid baths Calibration — Temperature (electrical thermometry)
-100 °C to 100 °C · CMC 0,02 °C 100 °C to 250 °C · CMC 0,03 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 206
60 Liquid baths Calibration — Temperature (electrical thermometry)
-100 °C to 250 °C · CMC 0,03 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 206
61 Thermostatic chambers Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,05 °C 100 °C to 200 °C · CMC 0,1 °C 200 °C to 400 °C · CMC 0,2 °C
Temperature (electrical thermometry)
Internal procedure; Calibration and measurement. Indirect method; At permanent location (S) Internal procedure PT 206 based on EURAMET cg-20 v. 5.0
62 Thermostatic chambers Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,05 °C 100 °C to 200 °C · CMC 0,1 °C 200 °C to 400 °C · CMC 0,2 °C
Temperature (electrical thermometry)
Internal procedure; Calibration and measurement. Indirect method; Outside of permanent location (P) Internal procedure PT 206 based on EURAMET cg-20 v. 5.0
63 Steam sterilisation chambers (e.g. autoclaves) Calibration — Temperature (electrical thermometry)
0 °C to 200 °C · CMC 0,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 206
64 Steam sterilisation chambers (e.g. autoclaves) Calibration — Temperature (electrical thermometry)
0 °C to 200 °C · CMC 0,2 °C
Temperature (electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 206
65 Climatic chambers Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,05 °C 100 °C to 200 °C · CMC 0,1 °C 200 °C to 400 °C · CMC 0,2 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 206 based on EURAMET cg-20 v. 5.0
66 Climatic chambers Calibration — Temperature (electrical thermometry)
-100 °C do 100 °C · CMC 0,05 °C 100 °C to 200 °C · CMC 0,1 °C 200 °C to 400 °C · CMC 0,2 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 206 based on EURAMET cg-20 v. 5.0
67 Temperature calibrators Calibration — Temperature (electrical thermometry)
-100 °C do 150 °C · CMC 0,02 °C 150 °C do 420 °C · CMC 0,03 °C 420 °C do 650 °C · CMC 0,07 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 206 based on EURAMET cg-13 v. 4.0
68 Temperature calibrators Calibration — Temperature (electrical thermometry)
-100 °C do 420 °C · CMC 0,1 °C 420 °C do 650 °C · CMC 0,2 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 206 based on EURAMET cg-13 v. 4.0
69 Furnaces Calibration — Temperature (electrical thermometry)
0 °C do 420 °C · CMC 0,03 °C 420 °C do 650 °C · CMC 0,07 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) Internal procedure PT 206
70 Furnaces Calibration — Temperature (electrical thermometry)
0 °C do 650 °C · CMC 0,2 °C 650 °C do 1085 °C · CMC 0,7 °C 1085 °C to 1200 °C · CMC 1,2 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; Outside of permanent location (P) Internal procedure PT 206
71 Dial thermometers (pressure-spring and bimetallic) Calibration — Temperature (non-electrical thermometry)
0 °C to 120 °C · CMC 0,1 °C
Temperature (non-electrical thermometry)
Internal procedure; Indirect method; At permanent location (S) Internal procedure PT 204
72 Dial thermometers (pressure-spring and bimetallic) Calibration — Temperature (non-electrical thermometry)
0 °C to 120 °C · CMC 0,2 °C
Temperature (non-electrical thermometry)
Internal procedure; Indirect method; Outside of permanent location (P) Internal procedure PT 204
73 Pyrometers (including radiation, photoelectric and multiband pyrometers, non-contact temperature measuring systems) Calibration — Temperature (radiation thermometry)
30 °C to 350 °C · CMC 2,9 °C
Temperature (radiation thermometry)
Internal procedure; At permanent location (S) Internal procedure PT 205
74 Thermal imaging cameras
Line scanners
Calibration — Temperature (radiation thermometry)
30 °C to 350 °C · CMC 2,9 °C
Temperature (radiation thermometry)
Internal procedure; Calibration of the temperature measurement error only; At permanent location (S) Internal procedure PT 205

Reference standards (20)

EURAMET cg-13 v. 4.0
EURAMET cg-15 v. 3.0
EURAMET cg-17 v. 4.0
EURAMET cg-20 v. 5.0
EURAMET cg-8 v. 3.1
PT 201
PT 202
PT 203
PT 204
PT 205
PT 206
PT 301
PT 302
PT 401
PT 402
PT 403
PT 404
PT 501
PT 502
PT 503