🇵🇱 PLUM Sp. z o.o. PL
Metrology & calibration
PLUM Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 074 (Kleosin, Poland). Its accreditation scope covers 32 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Calibrators, Electrical thermometers, Breakdown testers. Most frequently used calibration principles: Internal procedure; Direct and indirect method; At permanent locati…, Internal procedure; Comparison method; At permanent location (S), Internal procedure; At permanent location (S).
At a glance
- Accreditation number: PCA AP 074
- Calibration items in scope: 32
- Types of calibrated items: 44
- Calibration principles used: 10
- City: Kleosin, Poland
- Valid until: 23-01-2029
- Accreditation number
- AP 074
- Accreditation body
- PCA
- Address
- Ignatki, ul. Wspólna 19, 16-001 Kleosin
- City
- Kleosin, Poland
- Division
- Measurement Laboratory
- Valid until
- 23-01-2029
What this laboratory calibrates (44)
- Multimeters — 7 calibration items
- Calibrators — 6 calibration items
- Electrical thermometers — 6 calibration items
- Breakdown testers — 4 calibration items
- Breakdown voltage meters — 4 calibration items
- Clamp meters — 4 calibration items
- Reference sources — 4 calibration items
- (including electronic) — 3 calibration items
- (with temperature recording) — 3 calibration items
- Electrical safety testers — 3 calibration items
- Oscilloscopes — 3 calibration items
- Resistance thermometer sensors — 3 calibration items
- (containing temperature sensors) — 2 calibration items
- Bourdon tube pressure gauges — 2 calibration items
- Digital ammeters — 2 calibration items
- and 29 more
Calibration principles (10)
- Internal procedure; Direct and indirect method; At permanent location (S) — 14 calibration items
- Internal procedure; Comparison method; At permanent location (S) — 5 calibration items
- Internal procedure; At permanent location (S) — 3 calibration items
- Internal procedure; Calibration at fixed points; At permanent location (S) — 2 calibration items
- Internal procedure; Direct method; At permanent location (S) — 2 calibration items
- Internal procedure; Indirect electrical method; At permanent location (S) — 2 calibration items
- Internal procedure; Calibration in a climatic chamber Comparison method; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a climatic chamber; At permanent location (S) — 1 calibration item
- Internal procedure; Indirect electrical method Using the characteristics of temperature transmitters with an analogue voltage or current electrical signal; At permanent location (S) — 1 calibration item
- Internal procedure; Indirect method; At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 12.11.2025
- Issue no.
- 26
- Certificate date
- 14.11.2019
- Accreditation cycle
- 03.01.2025 — 23.01.2029
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (32)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Calibrators Power supplies Breakdown testers Reference sources Breakdown voltage meters |
Calibration — DC voltage 0 mV to 100 mV · CMC 5 · 10⁻⁶ · U + 0,0001 mV 0,1 V to 199,99 V · CMC 0,0004 % 1 V · CMC 0,0002 % 10 V · CMC 0,00015 % 199,99 V to 1000 V · CMC 0,0007 % 1000 V to 4000 V · CMC 4,5 · 10⁻³ · U + 10 V DC voltage |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 2 | Multimeters Digital voltmeters Oscilloscopes Electronic loads |
Calibration — DC voltage 0 mV · CMC 0,01 µV 0,01 mV to 100 mV · CMC 5 · 10⁻⁶ · U + 0,0001 mV 0,1 V to 19,99 V · CMC 0,0004 % 1 V · CMC 0,0003 % 10 V · CMC 0,0002 % 19,99 V to 199,99 V · CMC 0,0005 % 199,99 V to 1100 V · CMC 0,0006 % DC voltage |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 3 | Calibrators Power supplies Electronic loads Breakdown testers Breakdown voltage meters |
Calibration — DC current 0 µA to 219,99 µA · CMC 1,2 · 10⁻⁵ · I + 0,5 nA 0,21999 mA to 220 mA · CMC 0,0015 % 0,22 A to 2,2 A · CMC 0,0055 % 2,2 A to 20 A · CMC 0,059 % DC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 4 | Multimeters Digital ammeters Gas volume converters Clamp meters |
Calibration — DC current 0 µA · CMC 0,01 nA 0,01 µA to 10 µA · CMC 1 · 10⁻⁵ · I + 0,43 nA 10 µA to 219,99 µA · CMC 1 · 10⁻⁵ · I + 1,2 nA 0,21999 mA to 219,99 mA · CMC 0,0015 % 0,21999 A to 2,2 A · CMC 0,0057 % 2,2 A to 20 A · CMC 4,5 · 10⁻⁴ · I + 0,0003 A DC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 5 | Clamp meters | Calibration — DC current 20 A to 1000 A · CMC 5 · 10⁻³ · I + 0,2 A DC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 6 | Multimeters Digital voltmeters Oscilloscopes Power quality meters |
Calibration — AC voltage 10 Hz to 20 Hz — 1 mV to 2,2 mV · CMC 0,23 % 10 Hz to 20 Hz — 2,2 mV to 22 mV · CMC 0,049 % 10 Hz to 20 Hz — 22 mV to 220 mV · CMC 0,037 % 10 Hz to 20 Hz — 0,22 V to 220 V · CMC 0,029 % 20 Hz to 20 kHz — 1 mV to 2,2 mV · CMC 0,21 % 20 Hz to 20 kHz — 2,2 mV to 22 mV · CMC 0,060 % 20 Hz to 20 kHz — 22 mV to 220 mV · CMC 8 · 10⁻⁵ · U + 0,007 mV 20 Hz to 40 Hz — 0,22 V to 220 V · CMC 0,011 % 40 Hz to 20 kHz — 0,22 V to 219,99 V · CMC 0,007 % 15 Hz to 50 Hz — 219,99 V to 250 V · CMC 0,043 % 50 Hz to 1 kHz — 219,99 V to 1100 V · CMC 7 · 10⁻⁵ · U + 0,004 V 1 kHz to 10 kHz — 219,99 V to 1000 V · CMC 2,5 · 10⁻⁴ · U + 0,035 mV 20 kHz to 50 kHz — 1 mV to 2,2 mV · CMC 0,22 % 20 kHz to 50 kHz — 2,2 mV to 220 mV · CMC 2,5 · 10⁻⁴ · U + 0,004 mV 20 kHz to 50 kHz — 0,22 V to 219,99 V · CMC 1,2 · 10⁻⁴ · U + 0,00005 V 50 kHz to 100 kHz — 1 mV to 2,2 mV · CMC 0,31 % 50 kHz to 100 kHz — 2,2 mV to 220 mV · CMC 7 · 10⁻⁴ · U + 0,004 mV 50 kHz to 100 kHz — 0,22 V to 219,99 V · CMC 2,5 · 10⁻⁴ · U + 0,00005 V 100 kHz to 300 kHz — 1 mV to 2,2 mV · CMC 0,65 % 100 kHz to 300 kHz — 2,2 mV to 220 mV · CMC 0,16 % 100 kHz to 300 kHz — 0,22 V to 2,2 V · CMC 0,051 % 100 kHz to 300 kHz — 2,2 V to 22 V · CMC 0,034 % 300 kHz to 1 MHz — 1 mV to 2,2 mV · CMC 1,6 % 300 kHz to 1 MHz — 2,2 mV to 220 mV · CMC 0,38 % 300 kHz to 1 MHz — 0,22 V to 22 V · CMC 3 · 10⁻³ · U + 0,003 V AC voltage |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 7 | Calibrators Reference sources Breakdown testers Breakdown voltage meters Generators |
Calibration — AC voltage 10 Hz to 40 Hz — 1 mV to 199,99 mV · CMC 2,2 · 10⁻⁴ · U + 0,0057 mV 10 Hz to 40 Hz — 0,2 V to 199,99 V · CMC 0,013 % 10 Hz to 40 Hz — 200 V to 1000 V · CMC 0,021 % 40 Hz to 10 kHz — 1 mV to 199,99 mV · CMC 1,4 · 10⁻⁴ · U + 0,0060 mV 40 Hz to 10 kHz — 0,2 V to 199,99 V · CMC 0,012 % 40 Hz to 10 kHz — 200 V to 1000 V · CMC 1,5 · 10⁻⁴ · U + 0,01 V 10 kHz to 30 kHz — 1 mV to 199,99 mV · CMC 3,5 · 10⁻⁴ · U + 0,0089 mV 10 kHz to 30 kHz — 0,2 V to 199,99 V · CMC 0,025 % 10 kHz to 30 kHz — 200 V to 1000 V · CMC 0,042 % 30 kHz to 100 kHz — 1 mV to 199,99 mV · CMC 8,0 · 10⁻⁴ · U + 0,021 mV 30 kHz to 100 kHz — 0,2 V to 19,99 V · CMC 0,072 % 30 kHz to 100 kHz — 20 V to 199,99 V · CMC 0,085 % 100 kHz to 300 kHz — 0,2 V to 19,99 V · CMC 3,1 % 300 kHz to 500 kHz — 2 V to 19,99 V · CMC 1 · 10⁻² · U + 0,2 V 45 Hz to 65 Hz — 1000 V to 5200 V · CMC 4,5 · 10⁻³ · U + 10 V AC voltage |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 8 | Multimeters Digital ammeters Clamp meters Power quality meters |
Calibration — AC current 10 Hz to 20 Hz — 10 µA to 220 µA · CMC 0,042 % 10 Hz to 20 Hz — 0,22 mA to 220 mA · CMC 0,030 % 20 Hz to 1 kHz — 10 µA to 220 µA · CMC 1,3 · 10⁻⁴ · I + 0,009 µA 20 Hz to 1 kHz — 0,22 mA to 220 mA · CMC 0,020 % 20 Hz to 1 kHz — 0,22 A to 2,2 A · CMC 0,031 % 1 kHz to 5 kHz — 10 µA to 220 µA · CMC 2,9 · 10⁻⁴ · I + 0,013 µA 1 kHz to 5 kHz — 0,22 mA to 220 mA · CMC 0,027 % 1 kHz to 5 kHz — 0,22 A to 2,2 A · CMC 0,056 % 5 kHz to 10 kHz — 10 µA to 22 mA · CMC 0,18 % 5 kHz to 10 kHz — 22 mA to 220 mA · CMC 0,12 % 5 kHz to 10 kHz — 0,22 A to 2,2 A · CMC 0,75 % 45 Hz to 2 kHz — 2,2 A to 20 A · CMC 1 · 10⁻³ · I + 0,002 A 2 kHz to 5 kHz — 2,2 A to 20 A · CMC 0,72% AC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 9 | Clamp meters | Calibration — AC current 50 Hz — 20 A to 1000 A · CMC 2,5 · 10⁻³ · I + 0,1 A AC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 10 | Calibrators Reference sources Electrical safety testers Breakdown testers Breakdown voltage meters |
Calibration — AC current 40 Hz to 2 kHz — 10 µA to 1,99 mA · CMC 4,1· 10⁻⁴ · I + 0,015 µA 40 Hz to 2 kHz — 2 mA to 199,99 mA · CMC 4,1· 10⁻⁴ · I + 0,0008 mA 40 Hz to 2 kHz — 0,2 A to 1,99 A · CMC 7,6· 10⁻⁴ · I + 0,00013 A 40 Hz to 2 kHz — 2 A to 19,99 A · CMC 0,11 % 2 kHz to 10 kHz — 10 µA to 19,99 mA · CMC 5,1· 10⁻⁴ · I + 0,012 µA 2 kHz to 10 kHz — 20 mA to 199,99 mA · CMC 3,8· 10⁻⁴ · I + 0,015 mA 2 kHz to 10 kHz — 0,2 A to 1,99 A · CMC 8,9· 10⁻⁴ · I + 0,00011 A 2 kHz to 10 kHz — 2 A to 19,99 A · CMC 0,36 % AC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 11 | Calibrators Reference sources Electrical safety testers |
Calibration — AC current 50 Hz to 60 Hz — 20 A to 25 A · CMC 0,1 A 50 Hz to 60 Hz — 25 A to 30 A · CMC 0,2 A AC current |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 12 | Fixed resistors Variable resistors |
Calibration — DC resistance 0,001 Ω to 0,2 Ω · CMC 0,005 mΩ 0,2 Ω to 5 Ω · CMC 0,001 % 5 Ω to 19,99 Ω · CMC 0,0005 % 20 Ω to 199,99 Ω · CMC 0,0003 % 0,2 kΩ to 1,999 kΩ · CMC 0,0005 % 2 kΩ to 19,99 kΩ · CMC 0,0003 % 20 kΩ to 199,99 kΩ · CMC 0,0005 % 0,2 MΩ to 1,999 MΩ · CMC 0,0012 % 2 MΩ to 19,99 MΩ · CMC 0,003 % 20 MΩ to 199,99 MΩ · CMC 0,01 % 0,2 GΩ to 1,999 GΩ · CMC 0,04 % DC resistance |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/7 based on EURAMET cg-15 v.3.0 | |
| 13 | Resistance calibrators | Calibration — DC resistance 0,001 Ω to 0,2 Ω · CMC 0,005 mΩ 0,2 Ω to 5 Ω · CMC 0,001 % 5 Ω to 19,99 Ω · CMC 0,0005 % 20 Ω to 199,99 Ω · CMC 0,0003 % 0,2 kΩ to 1,999 kΩ · CMC 0,0005 % 2 kΩ to 19,99 kΩ · CMC 0,0003 % 20 kΩ to 199,99 kΩ · CMC 0,0005 % 0,2 MΩ to 1,999 MΩ · CMC 0,0012 % 2 MΩ to 19,99 MΩ · CMC 0,003 % 20 MΩ to 199,99 MΩ · CMC 0,01 % 0,2 GΩ to 1,999 GΩ · CMC 0,04 % DC resistance |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 14 | Multimeters Digital ohmmeters |
Calibration — DC resistance 0 Ω · CMC 0,00005 mΩ 0,001 Ω to 0,2 Ω · CMC 0,008 mΩ 0,001 Ω · CMC 0,12 % 0,01 Ω · CMC 0,012 % 0,1 Ω · CMC 0,0025 % 0,2 Ω to 2 Ω · CMC 0,004 % 1 Ω · CMC 0,00025 % 1,9 Ω · CMC 0,0031 % 2 Ω to 10 Ω · CMC 0,002 % 10 Ω · CMC 0,0008 % 10 Ω to 199,99 Ω · CMC 0,0015 % 19 Ω · CMC 0,00065 % 25 Ω · CMC 0,00028 % 100 Ω · CMC 0,00025 % 190 Ω · CMC 0,00038 % 0,2 kΩ to 199,99 kΩ · CMC 0,0015 % 1 kΩ · CMC 0,00055 % 1,9 kΩ · CMC 0,00058 % 10 kΩ · CMC 0,00025 % 19 kΩ · CMC 0,00038 % 100 kΩ · CMC 0,00055 % 190 kΩ · CMC 0,001 % 0,2 MΩ to 1,999 MΩ · CMC 0,002 % 1 MΩ · CMC 0,0013% 1,9 MΩ · CMC 0,0013 % 2 MΩ to 19,99 MΩ · CMC 0,0035 % 10 MΩ · CMC 0,0028 % 19 MΩ · CMC 0,0034 % 20 MΩ to 199,99 MΩ · CMC 0,015 % 100 MΩ · CMC 0,012 % 0,2 GΩ to 1 GΩ · CMC 0,07 % 1 GΩ · CMC 0,015 % 1 GΩ to 20 GΩ · CMC 1,5· 10⁻³ · R + 0,01 GΩ DC resistance |
Internal procedure; Direct and indirect method; At permanent location (S) | ILAJ 5.4/6 based on EURAMET cg-15 v.3.0 | |
| 15 | Digital ohmmeters Electrical safety testers |
Calibration — AC resistance current up to 30 A, 50 Hz to 60 Hz — 10 mΩ to 102 mΩ · CMC 1 mΩ current up to 30 A, 50 Hz to 60 Hz — 102 mΩ to 210 mΩ · CMC 2 mΩ AC resistance |
Internal procedure; Indirect method; At permanent location (S) | ILAJ 5.4/6 | |
| 16 | Multimeters Capacitance meters |
Calibration — Capacitance 0,4 nF to 11 nF · CMC 2 · 10⁻³ · C + 0,02 nF 11 nF to 330 nF · CMC 3 · 10⁻³ · C + 0,1 nF 0,33 µF to 11 µF · CMC 3 · 10⁻³ · C + 1 nF 11 µF to 100 µF · CMC 5 · 10⁻³ · C + 10 nF Capacitance |
Internal procedure; Direct method; At permanent location (S) | ILAJ 5.4/6 | |
| 17 | Temperature simulators | Calibration — Electrical simulation of quantities -200 °C to 200 °C · CMC 0,005 °C 200 °C to 500 °C · CMC 0,010 °C 500 °C to 850 °C · CMC 0,015 °C Electrical simulation of quantities |
Internal procedure; Indirect electrical method; At permanent location (S) | ILAJ 5.4/7 based on EURAMET cg-11 v.2.0 PN-EN 60751:2022 | |
| 18 | Temperature indicators (meters) (for use with resistance thermometer sensors) |
Calibration — Electrical simulation of quantities -200 °C to 200 °C · CMC 0,005 °C 200 °C to 500 °C · CMC 0,010 °C 500 °C to 850 °C · CMC 0,015 °C Electrical simulation of quantities |
Internal procedure; Indirect electrical method; At permanent location (S) | ILAJ 5.4/8 based on EURAMET cg-11 v.2.0 PN-EN 60751:2022 | |
| 19 | Temperature indicators (meters) including temperature controllers |
Calibration — Electrical simulation of quantities -200 °C to 1820 °C · CMC (0,04 · Z + 0,5) · 10⁻³ °C Electrical simulation of quantities |
Internal procedure; Indirect electrical method Using the characteristics of temperature transmitters with an analogue voltage or current electrical signal; At permanent location (S) | ILAJ 5.4/8 based on EURAMET cg-11 v.2.0 PN-EN 60751:2022 | |
| 20 | Multimeters Calibrators Digital frequency meters Oscilloscopes Frequency generators |
Calibration — Frequency 1 Hz to 1000 Hz · CMC 3 · 10⁻⁶ · f + 0,00001 Hz 1 kHz do 100 MHz · CMC 0,0003 % Frequency |
Internal procedure; Direct method; At permanent location (S) | ILAJ 5.4/6 | |
| 21 | Hygrometers Thermohygrometers Transmitters |
Calibration — Relative humidity at t = 22 °C to 60 °C and 10 %rh — 10 %rh to 95 %rh · CMC 0,9 %rh at t = 22 °C to 60 °C and 95 %rh — 10 %rh to 95 %rh · CMC 1,2 %rh at t = 10 °C and 22 %rh — 22 %rh to 95 %rh · CMC 1,2 %rh at t = 10 °C and 95 %rh — 22 %rh to 95 %rh · CMC 1,2 %rh Relative humidity |
Internal procedure; Calibration in a climatic chamber; At permanent location (S) | ILAJ 5.4/11 | |
| 22 | Electronic pressure gauges (calibrators) Gas volume converters with pressure measurement function Pressure transmitters Bourdon tube pressure gauges - gauge pressure (gas, liquid) |
Calibration — Pressure -100 kPa to -5 kPa · CMC -4,4·10⁻⁵ · p + 0,0007 kPa -5 kPa to 10 kPa · CMC 0,001 kPa 10 kPa to 170 kPa · CMC 4,1 · 10⁻⁵ · p + 0,0006 kPa 170 kPa to 700 kPa · CMC 3,0 · 10⁻⁵ · p + 0,006 kPa 700 kPa to 1000 kPa · CMC 1,7 · 10⁻⁴ · p - 0,09 kPa 1000 kPa to 7000 kPa · CMC 2,9 · 10⁻⁵ · p + 0,05 kPa 7000 kPa to 10000 kPa · CMC 0,0002 · p 10000 kPa to 14000 kPa · CMC 3,5 kPa 14000 kPa to 35000 kPa · CMC 9,0 kPa 35000 kPa to 70000 kPa · CMC 18 kPa Pressure |
Internal procedure; At permanent location (S) | ILAJ 5.4/4 ILAJ 5.4/5 ILAJ 5.4/14 | |
| 23 | Electronic pressure gauges (calibrators) Gas volume converters with pressure measurement function Pressure transmitters Bourdon tube pressure gauges Barometers - absolute pressure (gas, liquid) |
Calibration — Pressure 1,4 kPa to 80 kPa · CMC 0,004 kPa 80 kPa to 170 kPa · CMC 2,9 · 10⁻⁵ · p + 0,0017 kPa 170 kPa to 700 kPa · CMC 3,5 · 10⁻⁵ · p + 0,0025 kPa 700 kPa to 1000 kPa · CMC 1,4 · 10⁻⁴ · p - 0,07 kPa 1000 kPa to 7000 kPa · CMC 3 · 10⁻⁵ · p + 0,04 kPa 7000 kPa to 10000 kPa · CMC 0,0002 · p 10000 kPa to 14000 kPa · CMC 3,5 kPa 14000 kPa to 35000 kPa · CMC 9,0 kPa 35000 kPa to 70000 kPa · CMC 18 kPa Pressure |
Internal procedure; At permanent location (S) | ILAJ 5.4/4 ILAJ 5.4/5 ILAJ 5.4/14 | |
| 24 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -40 °C to 230 °C · CMC 0,008 °C Temperature (electrical thermometry) |
Internal procedure; Comparison method; At permanent location (S) | ILAJ 5.4/1 | |
| 25 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -40 °C to 230 °C · CMC 0,010 °C Temperature (electrical thermometry) |
Internal procedure; Comparison method; At permanent location (S) | ILAJ 5.4/1 | |
| 26 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -38,8344 °C · CMC 0,0030 °C 0,01 °C · CMC 0,0025 °C 29,7646 °C · CMC 0,0030 °C 156,5985 °C · CMC 0,0040 °C 231,9280 °C · CMC 0,0050 °C Temperature (electrical thermometry) |
Internal procedure; Calibration at fixed points; At permanent location (S) | ILAJ 5.4/1 | |
| 27 | Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (electrical thermometry) -40 °C to 230 °C · CMC 0,010 °C Temperature (electrical thermometry) |
Internal procedure; Comparison method; At permanent location (S) | ILAJ 5.4/2 | |
| 28 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) |
Calibration — Temperature (electrical thermometry) -40 °C to 230 °C · CMC 0,010 °C Temperature (electrical thermometry) |
Internal procedure; Comparison method; At permanent location (S) | ILAJ 5.4/3 | |
| 29 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) |
Calibration — Temperature (electrical thermometry) -38,8344 °C · CMC 0,0040 °C 0,01 °C · CMC 0,0035 °C 29,7646 °C · CMC 0,0040 °C 156,5985 °C · CMC 0,0050 °C 231,9280 °C · CMC 0,0060 °C Temperature (electrical thermometry) |
Internal procedure; Calibration at fixed points; At permanent location (S) | ILAJ 5.4/3 | |
| 30 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (electrical thermometry) -30 °C to 80 °C · CMC 0,1 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a climatic chamber Comparison method; At permanent location (S) | ILAJ 5.4/11 | |
| 31 | Temperature calibrators | Calibration — Temperature (electrical thermometry) -40 °C to 160 °C · CMC 0,14 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | ILAJ 5.4/16 based on EURAMET cg-13 v.4.0 | |
| 32 | Photoelectric and radiation pyrometers | Calibration — Temperature (radiation thermometry) -15 °C do 100 °C · CMC 0,8 °C 100 °C to 200 °C · CMC 1,2 °C 200 °C do 300 °C · CMC 1,5 °C 300 °C do 400 °C · CMC 2,0 °C 400 °C do 500 °C · CMC 2,4 °C Temperature (radiation thermometry) |
Internal procedure; Comparison method; At permanent location (S) | ILAJ 5.4/15 | |
Reference standards (16)
EURAMET cg-11 v.2.0
EURAMET cg-13 v.4.0
EURAMET cg-15 v.3.0
ILAJ 5.4/1
ILAJ 5.4/11
ILAJ 5.4/14
ILAJ 5.4/15
ILAJ 5.4/16
ILAJ 5.4/2
ILAJ 5.4/3
ILAJ 5.4/4
ILAJ 5.4/5
ILAJ 5.4/6
ILAJ 5.4/7
ILAJ 5.4/8
PN-EN 60751:2022