🇵🇱 LAB-EL Elektronika Laboratoryjna Sp. z o.o. PL
Metrology & calibration
LAB-EL Elektronika Laboratoryjna Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 067 (Reguły, Poland). Its accreditation scope covers 15 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electrical thermometers (including electronic), Electrical thermometers (with temperature recording), Temperature transmitters (containing temperature sensors), Climatic chambers. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; Calibration in a climatic chamber; At permanent….
At a glance
- Accreditation number: PCA AP 067
- Calibration items in scope: 15
- Types of calibrated items: 21
- Calibration principles used: 7
- City: Reguły, Poland
- Valid until: 03-06-2028
- Accreditation number
- AP 067
- Accreditation body
- PCA
- Address
- ul. Herbaciana 9, 05-816 Reguły
- City
- Reguły, Poland
- Division
- LAB-EL Humidity, Temperature and Pressure Laboratory
- Valid until
- 03-06-2028
What this laboratory calibrates (21)
- Electrical thermometers (including electronic) — 4 calibration items
- Electrical thermometers (with temperature recording) — 4 calibration items
- Temperature transmitters (containing temperature sensors) — 4 calibration items
- Climatic chambers — 2 calibration items
- Electronic pressure gauges — 2 calibration items
- Hygrometers — 2 calibration items
- Pressure transmitters — 2 calibration items
- Resistance thermometer sensors — 2 calibration items
- Spring pressure gauges — 2 calibration items
- - absolute pressure (gas medium) — 1 calibration item
- - gauge pressure (gas medium) — 1 calibration item
- Anemometers — 1 calibration item
- Barometers — 1 calibration item
- Liquid thermostats — 1 calibration item
- Liquid-in-glass thermometers — 1 calibration item
- and 6 more
Calibration principles (7)
- Internal procedure; At permanent location (S) — 6 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 3 calibration items
- Internal procedure; Calibration in a climatic chamber; At permanent location (S) — 2 calibration items
- Internal procedure; Calibration at the fixed temperature of liquid nitrogen; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a climatic chamber in the temperature range -5 °C to 60 °C; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a liquid thermostat; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a temperature calibrator; At permanent location (S) and outside of it (P) — 1 calibration item
Certificate information
- Issue date
- 01.09.2025
- Issue no.
- 22
- Certificate date
- 29.05.2020
- Accreditation cycle
- 15.05.2024 — 03.06.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (15)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Anemometers | Calibration — Air velocity (flow - gases) 1,0 m/s to 1,5m/s · CMC 0,1 m/s 1,5 m/s to 15,0 m/s · CMC 0,2 m/s 15,0 m/s to 40,0 m/s · CMC 0,2 + 0,025 · (V – 15) m/s; V – measured value (m/s) Air velocity (flow - gases) |
Internal procedure; At permanent location (S) | SOP-29 SOP-10.IR.15 | |
| 2 | Thermal anemometers | Calibration — Air velocity (flow - gases) 0,3 m/s to 5,0 m/s · CMC 0,1 m/s 5,0 m/s to 15,0 m/s · CMC 0,0469 + 0,0102 · V m/s; V – measured value (m/s) 15,0 m/s to 40,0 m/s · CMC 0,2 + 0,025 · (V – 15) m/s; V – measured value (m/s) Air velocity (flow - gases) |
Internal procedure; At permanent location (S) | SOP-29 SOP-10.IR.15 | |
| 3 | Hygrometers | Calibration — Dew-point temperature -10 °C to 40 °C · CMC 0,2 °C Dew-point temperature |
Internal procedure; Calibration in a climatic chamber; At permanent location (S) | SOP-22 SOP-10.IR.09 | |
| 4 | Hygrometers Thermohygrometers Psychrometers Transmitters |
Calibration — Relative humidity at temperature -5 °C — 68 %rh · CMC 1,1 %rh at temperature -5 °C — 98 %rh · CMC 1,2 %rh at temperature 0 °C — 47 %rh · CMC 0,6 %rh at temperature 0 °C — 98 %rh · CMC 1,1 %rh at temperature 21 °C — 12 %rh · CMC 0,5 %rh at temperature 21 °C — 98 %rh · CMC 0,8 %rh at temperature 40 °C — 10 %rh · CMC 0,4 %rh at temperature 40 °C — 98 %rh · CMC 0,8 %rh at temperature 60 °C — 10 %rh · CMC 0,5 %rh at temperature 60 °C — 38 %rh · CMC 0,8 %rh Relative humidity |
Internal procedure; Calibration in a climatic chamber in the temperature range -5 °C to 60 °C; At permanent location (S) | SOP-14 SOP-10.IR.01 | |
| 5 | Climatic chambers | Calibration — Relative humidity temperature range 5 °C to 60 °C — 10 %rh to 98 %rh · CMC 2,7 %rh Relative humidity |
Internal procedure; At permanent location (S) and outside of it (P) | SOP-25 SOP-10.IR.11 | |
| 6 | Barometers Electronic pressure gauges Spring pressure gauges Pressure transmitters - absolute pressure (gas medium) |
Calibration — Pressure 700 hPa to 1100 hPa · CMC 0,2 hPa Pressure |
Internal procedure; At permanent location (S) | SOP-18 SOP-10.IR.06 | |
| 7 | Electronic pressure gauges Spring pressure gauges Pressure transmitters - gauge pressure (gas medium) |
Calibration — Pressure -1470 Pa to 1470 Pa · CMC 2,0 Pa -200 Pa to 200 Pa · CMC 0,5 Pa Pressure |
Internal procedure; At permanent location (S) | SOP-18 SOP-10.IR.06 | |
| 8 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (electrical thermometry) -10 °C to 60 °C · CMC 0,12 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a climatic chamber; At permanent location (S) | SOP-14 SOP-10.IR.02 SOP-10.IR.03 | |
| 9 | Resistance thermometer sensors Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (electrical thermometry) -80 °C to 250 °C · CMC 0,02 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a liquid thermostat; At permanent location (S) | SOP-13 SOP-15 SOP-10.IR.02 SOP-10.IR.03 SOP-10.IR.04 SOP-10.IR.05 | |
| 10 | Resistance thermometer sensors Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (non-electrical thermometry) -90 °C to 250 °C · CMC 0,14 °C 250 °C to 420 °C · CMC 0,22 °C 420 °C to 650 °C · CMC 0,37 °C Temperature (non-electrical thermometry) |
Internal procedure; Calibration in a temperature calibrator; At permanent location (S) and outside of it (P) | SOP-13 SOP-15 SOP-21 SOP-10.IR.02 SOP-10.IR.03 SOP-10.IR.04 SOP-10.IR.05 | |
| 11 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (non-electrical thermometry) -195,8 °C · CMC 0,5 °C Temperature (non-electrical thermometry) |
Internal procedure; Calibration at the fixed temperature of liquid nitrogen; At permanent location (S) | SOP-13 SOP-15 SOP-21 SOP-27 SOP-10.IR.13 | |
| 12 | Climatic chambers Thermostatic chambers1) |
Calibration — Temperature (non-electrical thermometry) -90 °C to 250 °C · CMC 0,2 °C Temperature (non-electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | SOP-21 SOP-25 SOP-10.IR.11 | |
| 13 | Liquid thermostats | Calibration — Temperature (non-electrical thermometry) -30 °C to 80 °C · CMC 0,2 °C Temperature (non-electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | SOP-21 SOP-25 SOP-10.IR.11 | |
| 14 | Liquid-in-glass thermometers | Calibration — Temperature (non-electrical thermometry) -50 °C to 150 °C · CMC 0,1 °C Temperature (non-electrical thermometry) |
Internal procedure; At permanent location (S) | SOP-30 SOP-10.IR.16 | |
| 15 | Pyrometers (including radiation photoelectric and multiband pyrometers, thermal imaging cameras) | Calibration — Temperature (radiation thermometry) -15 °C do 0 °C · CMC 1,6 °C 0 °C do 50 °C · CMC 1,1 °C 50 °C do 100 °C · CMC 1,4 °C 100 °C to 200 °C · CMC 1,8 °C 200 °C do 350 °C · CMC 2,3 °C 350 °C do 500 °C · CMC 3,0 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) | SOP-23 SOP-10.IR.10 | |
Reference standards (23)
SOP-10.IR.01
SOP-10.IR.02
SOP-10.IR.03
SOP-10.IR.04
SOP-10.IR.05
SOP-10.IR.06
SOP-10.IR.09
SOP-10.IR.10
SOP-10.IR.11
SOP-10.IR.13
SOP-10.IR.15
SOP-10.IR.16
SOP-13
SOP-14
SOP-15
SOP-18
SOP-21
SOP-22
SOP-23
SOP-25
SOP-27
SOP-29
SOP-30