🇵🇱 Centralny Instytut Ochrony Pracy - Państwowy Instytut Badawczy PL
Metrology & calibration
Centralny Instytut Ochrony Pracy - Państwowy Instytut Badawczy is a calibration laboratory accredited by PCA under number AP 061 (Warszawa, Poland). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Meters for measuring electromagnetic fields, Anemometers, Aspirators, Electric field strength meters. Most frequently used calibration principles: Internal procedure; At permanent location (S).
At a glance
- Accreditation number: PCA AP 061
- Calibration items in scope: 5
- Types of calibrated items: 9
- Calibration principles used: 1
- City: Warszawa, Poland
- Valid until: 27-01-2028
- Accreditation number
- AP 061
- Accreditation body
- PCA
- Address
- ul. Czerniakowska 16, 00-701 Warszawa
- City
- Warszawa, Poland
- Division
- Zespół Laboratoriów Wzorcujących
- Valid until
- 27-01-2028
What this laboratory calibrates (9)
- Meters for measuring electromagnetic fields — 2 calibration items
- Anemometers — 1 calibration item
- Aspirators — 1 calibration item
- Electric field strength meters — 1 calibration item
- Flowmeters of energy expenditure meters — 1 calibration item
- Magnetic field strength meters — 1 calibration item
- Magnetic induction meters — 1 calibration item
- Meters for measuring magnetostatic fields — 1 calibration item
- Thermal anemometers — 1 calibration item
Calibration principles (1)
- Internal procedure; At permanent location (S) — 5 calibration items
Certificate information
- Issue date
- 26.11.2025
- Issue no.
- 22
- Certificate date
- 23.07.2020
- Accreditation cycle
- 04.01.2024 — 27.01.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (5)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Meters for measuring magnetostatic fields Meters for measuring electromagnetic fields Magnetic field strength meters Magnetic induction meters |
Calibration — Magnetic quantities for magnetostatic field — 0,0001 mT to 100 mT · CMC 2,8 % for magnetostatic field — 100 mT to 500 mT · CMC 0,1 % for alternating magnetic field, 20 Hz to 1000 Hz — 0,1 A/m to 2000 A/m · CMC 3,4 % for alternating magnetic field, 20 Hz to 1000 Hz — 0,1 A/m to 30000 A/m · CMC 4,0% for alternating magnetic field, 1 kHz to 50 kHz — 0,1 A/m to 150 A/m · CMC 3,0% for alternating magnetic field, 50 kHz to 100 kHz — 0,1 A/m to 150 A/m · CMC 5,0% Magnetic quantities |
Internal procedure; At permanent location (S) | PWS/NA4-7 w oparciu o IEEE Std 1309:2013; PWS/NA4-30 | |
| 2 | Meters for measuring electromagnetic fields Electric field strength meters |
Calibration — Electromagnetic quantities 15 Hz to 1000 Hz — 100 V/m to 25000 V/m · CMC 2,2 % 50 Hz to 100 kHz — 1 V/m to 17000 V/m · CMC 5,9 % 0,1 MHz to 0,5 MHz — 0,5 V/m to 450 V/m · CMC 19,2 % 0,5 MHz to 130 MHz — 0,5 V/m to 450 V/m · CMC 11,8 % 130 MHz to 230 MHz — 0,5 V/m to 450 V/m · CMC 15,6 % 230 MHz to 410 MHz — 0,5 V/m to 100 V/m · CMC 12,0 % 410 MHz 890 MHz — 0,5 V/m to 60 V/m · CMC 36,3 % 900 MHz — 0,5 V/m to 100 V/m · CMC 15,8 % 910 MHz to 2500 MHz — 0,5 V/m to 60 V/m · CMC 44,6 % 2500 MHz to 5000 MHz — 0,5 V/m to 45 V/m · CMC 68,3 % Electromagnetic quantities |
Internal procedure; At permanent location (S) | PWS/NA4-1 w oparciu o IEEE Std 1309:2013; PWS/NA4-13 w oparciu o IEEE Std 1309:2013 | |
| 3 | Flowmeters of energy expenditure meters | Calibration — Volume flow rate (flow – gases) 10 dm³/min to 65 dm³/min · CMC 2,3 % Volume flow rate (flow – gases) |
Internal procedure; At permanent location (S) | PWS/LM-2 | |
| 4 | Aspirators | Calibration — Volume flow rate (flow – gases) 1 dm³/min to 5 dm³/min · CMC 0,1 dm³/min Volume flow rate (flow – gases) |
Internal procedure; At permanent location (S) | PWS/LM-7 | |
| 5 | Anemometers Thermal anemometers |
Calibration — Air velocity (flow – gases) 0,15 m/s to 1,25 m/s · CMC 0,06 m/s 1,25 m/s to 7,5 m/s · CMC 0,21 m/s 7,5 m/s to 40 m/s · CMC 0,70 m/s Air velocity (flow – gases) |
Internal procedure; At permanent location (S) | PWS/NC5-1 | |
Reference standards (8)
IEEE Std 1309:2013
PWS/LM-2
PWS/LM-7
PWS/NA4-1
PWS/NA4-13
PWS/NA4-30
PWS/NA4-7