🇵🇱 PGE Górnictwo i Energetyka Konwencjonalna S.A. Oddział Elektrownia Bełchatów PL
Metrology & calibration
PGE Górnictwo i Energetyka Konwencjonalna S.A. Oddział Elektrownia Bełchatów is a calibration laboratory accredited by PCA under number AP 059 (Bełchatów, Poland). Its accreditation scope covers 18 calibration items in the following areas: Metrology & calibration. Calibrated items include: Calibrators, Multimeters, Electronic pressure gauges, Pressure transducers. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S), Internal procedure; Direct and indirect method; At permanent locati…, Internal procedure; At permanent location (S).
At a glance
- Accreditation number: PCA AP 059
- Calibration items in scope: 18
- Types of calibrated items: 17
- Calibration principles used: 3
- City: Bełchatów, Poland
- Valid until: 03-02-2028
- Accreditation number
- AP 059
- Accreditation body
- PCA
- Address
- Wola Grzymalina 3, 97-406 Bełchatów
- City
- Bełchatów, Poland
- Division
- Measurement Laboratory
- Valid until
- 03-02-2028
What this laboratory calibrates (17)
- Calibrators — 5 calibration items
- Multimeters — 5 calibration items
- Electronic pressure gauges — 3 calibration items
- Pressure transducers — 3 calibration items
- Spring pressure gauges — 3 calibration items
- Analogue current meters — 2 calibration items
- Analogue voltage meters — 2 calibration items
- Digital current meters — 2 calibration items
- Digital voltage meters — 2 calibration items
- Analogue resistance meters — 1 calibration item
- Dial thermometers (bimetallic, expansion, pressure-spring) — 1 calibration item
- Digital resistance meters — 1 calibration item
- Electrical thermometers (including electronic) — 1 calibration item
- Resistance thermometer sensors — 1 calibration item
- Temperature transmitters (containing temperature sensors) — 1 calibration item
- and 2 more
Calibration principles (3)
- Internal procedure; Direct method; At permanent location (S) — 9 calibration items
- Internal procedure; Direct and indirect method; At permanent location (S) — 5 calibration items
- Internal procedure; At permanent location (S) — 4 calibration items
Certificate information
- Issue date
- 21.11.2025
- Issue no.
- 20
- Certificate date
- 30.06.2022
- Accreditation cycle
- 05.01.2024 — 03.02.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (18)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Digital voltage meters Multimeters Calibrators |
Calibration — DC voltage 1 mV to 100 mV · CMC 0,001% + 0,6 µV 0,1 V to 1 V · CMC 0,001 % + 0,3 µV 1 V to 10 V · CMC 0,0014 % + 0,6 µV 10 V to 100 V · CMC 0,0014 % + 0,03 mV 100 V to 1000 V · CMC 0,0026 % + 1 mV DC voltage |
Internal procedure; Direct and indirect method; At permanent location (S) | PP7 based on EURAMET cg-15 v. 3.0 | |
| 2 | Analogue voltage meters | Calibration — DC voltage 1 mV to 100 mV · CMC 0,001% + 0,6 µV 0,1 V to 1 V · CMC 0,001 % + 0,3 µV 1 V to 10 V · CMC 0,0014 % + 0,6 µV 10 V to 100 V · CMC 0,0014 % + 0,03 mV 100 V to 1000 V · CMC 0,0026 % + 1 mV DC voltage |
Internal procedure; Direct method; At permanent location (S) | PP18 based on PN-EN IEC 60051-9:2021-09 | |
| 3 | Digital current meters Multimeters Calibrators |
Calibration — DC current 10 µA to 100 µA · CMC 0,005 % + 0,9 nA 100 µA to 1 mA · CMC 0,004 % + 6 nA 1 mA to 10 mA · CMC 0,004 % + 60 nA 10 mA to 100 mA · CMC 0,006 % + 0,6 µA 100 mA to 1 A · CMC 0,021 % + 12 µA 1 to 2 A · CMC 0,035 % + 0,12 mA 2 to 10 A · CMC 0,03 % + 0,7 mA 10 to 20 A · CMC 0,013 % + 5 mA DC current |
Internal procedure; Direct and indirect method; At permanent location (S) | PP7 based on EURAMET cg-15 v. 3.0 | |
| 4 | Analogue current meters | Calibration — DC current 10 µA to 100 µA · CMC 0,005 % + 0,9 nA 100 µA to 1 mA · CMC 0,004 % + 6 nA 1 mA to 10 mA · CMC 0,004 % + 60 nA 10 mA to 100 mA · CMC 0,006 % + 0,6 µA 100 mA to 1 A · CMC 0,021 % + 12 µA 1 to 2 A · CMC 0,035 % + 0,12 mA 2 to 10 A · CMC 0,03 % + 0,7 mA 10 to 20 A · CMC 0,013 % + 5 mA DC current |
Internal procedure; Direct method; At permanent location (S) | PP18 based on PN-EN IEC 60051-9:2021-09 | |
| 5 | Digital voltage meters Multimeters Calibrators |
Calibration — AC voltage 0,040 kHz to 1 kHz — 1 mV to 10 mV · CMC 0,08 % + 1,5 µV 0,040 kHz to 1 kHz — 10 mV to 100 mV · CMC 0,03 % + 2 µV 0,040 kHz to 1 kHz — 100 mV to 1 V · CMC 0,015 % + 25 µV 0,040 kHz to 1 kHz — 1 V to 10 V · CMC 0,014 % + 0,25 mV 0,040 kHz to 1 kHz — 10 V to 100 V · CMC 0,035 % + 2,3 mV 0,040 kHz to 1 kHz — 100 V to 700 V · CMC 0,055 % + 23 mV 0,040 kHz to 1 kHz — 700 V to 1000 V · CMC 0,15 % 1 kHz to 5 kHz — 1 mV to 10 mV · CMC 0,09 % + 1,5 µV 1 kHz to 5 kHz — 10 V to 100 mV · CMC 0,04 % + 2 µV 1 kHz to 5 kHz — 100 mV to 1 V · CMC 0,025 % + 25 µV 1 kHz to 5 kHz — 1 V to 10 V · CMC 0,025 % + 0,25 mV 1 kHz to 5 kHz — 10 µA to 100 V · CMC 0,035 % + 2,3 mV 1 kHz to 5 kHz — 100 V to 700 V · CMC 0,08 % + 23 mV 1 kHz to 5 kHz — 700 V to 1000 V · CMC 0,15 % 5 kHz to 20 kHz — 1 mV to 10 mV · CMC 0,09 % + 1,5 µV 5 kHz to 20 kHz — 10 mV to 100 mV · CMC 0,04 % + 2 µV 5 kHz to 20 kHz — 100 mV to 1 V · CMC 0,025 % + 25 µV 5 kHz to 20 kHz — 1 V to 10 V · CMC 0,025 % + 0,25 mV 5 kHz to 20 kHz — 10 V to 100 V · CMC 0,035 % + 2,3 mV 20 kHz to 50 kHz — 100 mV to 1 V · CMC 0,045 % + 25 µV 20 kHz to 50 kHz — 1 V to 10 V · CMC 0,045 % + 0,22 mV 20 kHz to 50 kHz — 10 V to 100 V · CMC 0,052 % + 2,3 mV 50 kHz to 100 kHz — 100 mV to 1 V · CMC 0,12 % + 15 µV 50 kHz to 100 kHz — 1 V to 10 V · CMC 0,11 % + 0,2 mV 50 kHz to 100 kHz — 10 V to 100 V · CMC 0,17 % + 2,3 mV AC voltage |
Internal procedure; Direct and indirect method; At permanent location (S) | PP7 based on EURAMET cg-15 v. 3.0 | |
| 6 | Analogue voltage meters | Calibration — AC voltage 0,040 kHz to 1 kHz — 1 mV to 10 mV · CMC 0,08 % + 1,5 µV 0,040 kHz to 1 kHz — 10 mV to 100 mV · CMC 0,03 % + 2 µV 0,040 kHz to 1 kHz — 100 mV to 1 V · CMC 0,015 % + 25 µV 0,040 kHz to 1 kHz — 1 V to 10 V · CMC 0,014 % + 0,25 mV 0,040 kHz to 1 kHz — 10 V to 100 V · CMC 0,035 % + 2,3 mV 0,040 kHz to 1 kHz — 100 V to 700 V · CMC 0,055 % + 23 mV 0,040 kHz to 1 kHz — 700 V to 1000 V · CMC 0,15 % 1 kHz to 5 kHz — 1 mV to 10 mV · CMC 0,09 % + 1,5 µV 1 kHz to 5 kHz — 10 V to 100 mV · CMC 0,04 % + 2 µV 1 kHz to 5 kHz — 100 mV to 1 V · CMC 0,025 % + 25 µV 1 kHz to 5 kHz — 1 V to 10 V · CMC 0,025 % + 0,25 mV 1 kHz to 5 kHz — 10 µA to 100 V · CMC 0,035 % + 2,3 mV 1 kHz to 5 kHz — 100 V to 700 V · CMC 0,08 % + 23 mV 1 kHz to 5 kHz — 700 V to 1000 V · CMC 0,15 % 5 kHz to 20 kHz — 1 mV to 10 mV · CMC 0,09 % + 1,5 µV 5 kHz to 20 kHz — 10 mV to 100 mV · CMC 0,04 % + 2 µV 5 kHz to 20 kHz — 100 mV to 1 V · CMC 0,025 % + 25 µV 5 kHz to 20 kHz — 1 V to 10 V · CMC 0,025 % + 0,25 mV 5 kHz to 20 kHz — 10 V to 100 V · CMC 0,035 % + 2,3 mV 20 kHz to 50 kHz — 100 mV to 1 V · CMC 0,045 % + 25 µV 20 kHz to 50 kHz — 1 V to 10 V · CMC 0,045 % + 0,22 mV 20 kHz to 50 kHz — 10 V to 100 V · CMC 0,052 % + 2,3 mV 50 kHz to 100 kHz — 100 mV to 1 V · CMC 0,12 % + 15 µV 50 kHz to 100 kHz — 1 V to 10 V · CMC 0,11 % + 0,2 mV 50 kHz to 100 kHz — 10 V to 100 V · CMC 0,17 % + 2,3 mV AC voltage |
Internal procedure; Direct method; At permanent location (S) | PP18 based on PN-EN IEC 60051-9:2021-09 | |
| 7 | Digital current meters Multimeters Calibrators |
Calibration — AC current 0,040 kHz to 1 kHz — 20 µA to 200 µA · CMC 0,1 % + 0,06 µA 0,040 kHz to 1 kHz — 0,2 mA to 2 mA · CMC 0,1 % + 0,24 µA 0,040 kHz to 1 kHz — 2 mA to 20 mA · CMC 0,1 % + 2,3 µA 0,040 kHz to 1 kHz — 20 mA to 200 mA · CMC 0,1 % + 22 µA 0,040 kHz to 1 kHz — 0,2 A to 2 A · CMC 0,12 % + 0,4 mA 0,040 kHz to 1 kHz — 2 A to 20 A · CMC 0,14 % + 9 mA 1 kHz to 5 kHz — 20 µA to 200 µA · CMC 0,1 % + 0,06 µA 1 kHz to 5 kHz — 0,2 mA to 2 mA · CMC 0,1 % + 0,24 µA 1 kHz to 5 kHz — 2 mA to 20 mA · CMC 0,1 % + 2,3 µA 1 kHz to 5 kHz — 20 mA to 200 mA · CMC 0,1 % + 22 µA 1 kHz to 5 kHz — 0,2 A to 2 A · CMC 0,12 % + 0,4 mA AC current |
Internal procedure; Direct and indirect method; At permanent location (S) | PP7 based on EURAMET cg-15 v. 3.0 | |
| 8 | Analogue current meters | Calibration — AC current 0,040 kHz to 1 kHz — 20 µA to 200 µA · CMC 0,1 % + 0,06 µA 0,040 kHz to 1 kHz — 0,2 mA to 2 mA · CMC 0,1 % + 0,24 µA 0,040 kHz to 1 kHz — 2 mA to 20 mA · CMC 0,1 % + 2,3 µA 0,040 kHz to 1 kHz — 20 mA to 200 mA · CMC 0,1 % + 22 µA 0,040 kHz to 1 kHz — 0,2 A to 2 A · CMC 0,12 % + 0,4 mA 0,040 kHz to 1 kHz — 2 A to 20 A · CMC 0,14 % + 9 mA 1 kHz to 5 kHz — 20 µA to 200 µA · CMC 0,1 % + 0,06 µA 1 kHz to 5 kHz — 0,2 mA to 2 mA · CMC 0,1 % + 0,24 µA 1 kHz to 5 kHz — 2 mA to 20 mA · CMC 0,1 % + 2,3 µA 1 kHz to 5 kHz — 20 mA to 200 mA · CMC 0,1 % + 22 µA 1 kHz to 5 kHz — 0,2 A to 2 A · CMC 0,12 % + 0,4 mA AC current |
Internal procedure; Direct method; At permanent location (S) | PP18 based on PN-EN IEC 60051-9:2021-09 | |
| 9 | Digital resistance meters Multimeters Calibrators |
Calibration — DC resistance 1 Ω to 10 Ω · CMC 0,002 % + 60 µΩ 10 Ω to 100 Ω · CMC 0,0016 % + 0,6 mΩ 100 Ω to 1 kΩ · CMC 0,0015 % + 0,6 mΩ 1 kΩ to 10 kΩ · CMC 0,0015 % + 6 mΩ 10 kΩ to 100 kΩ · CMC 0,0014 % + 60 mΩ 100 kΩ to 1 MΩ · CMC 0,0021 % + 2,3 Ω 1 MΩ to 10 MΩ · CMC 0,006 % + 120 Ω 10 MΩ to 100 MΩ · CMC 0,062 % + 1100 Ω DC resistance |
Internal procedure; Direct and indirect method; At permanent location (S) | PP7 based on EURAMET cg-15 v. 3.0 | |
| 10 | Analogue resistance meters | Calibration — DC resistance 1 Ω to 10 Ω · CMC 0,002 % + 60 µΩ 10 Ω to 100 Ω · CMC 0,0016 % + 0,6 mΩ 100 Ω to 1 kΩ · CMC 0,0015 % + 0,6 mΩ 1 kΩ to 10 kΩ · CMC 0,0015 % + 6 mΩ 10 kΩ to 100 kΩ · CMC 0,0014 % + 60 mΩ 100 kΩ to 1 MΩ · CMC 0,0021 % + 2,3 Ω 1 MΩ to 10 MΩ · CMC 0,006 % + 120 Ω 10 MΩ to 100 MΩ · CMC 0,062 % + 1100 Ω DC resistance |
Internal procedure; Direct method; At permanent location (S) | PP18 based on PN-EN IEC 60051-9:2021-09 | |
| 11 | Electronic pressure gauges Pressure transducers Spring pressure gauges |
Calibration — Pressure - gauge pressure (gas medium) — -100 kPa to -10 kPa · CMC 2 · 10⁻⁴ · IpI - gauge pressure (gas medium) — -10 kPa to 0 kPa · CMC 0,0015 kPa - gauge pressure (gas medium) — 0 kPa to 1,6 kPa · CMC 0,0003 kPa - gauge pressure (gas medium) — 1,6 kPa to 6000 kPa · CMC 1,6 · 10⁻⁴ · p Pressure |
Internal procedure; Direct method; At permanent location (S) | PP6 based on EURAMET cg-17 v. 4.1 | |
| 12 | Electronic pressure gauges Pressure transducers Spring pressure gauges |
Calibration — Pressure - gauge pressure (liquid medium) — 0 MPa to 1,5 MPa · CMC 2,3 · 10⁻⁴ MPa - gauge pressure (liquid medium) — 1,5 MPa to 70 MPa · CMC 1,6 · 10⁻⁴ · p - gauge pressure (liquid medium) — 70 MPa to 250 MPa · CMC 3 · 10⁻⁴ · p Pressure |
Internal procedure; Direct method; At permanent location (S) | PP6 based on EURAMET cg-17 v. 4.1 | |
| 13 | Electronic pressure gauges Pressure transducers Spring pressure gauges |
Calibration — Pressure - absolute pressure (gas medium) — 5 kPa to 115 kPa · CMC 0,015 kPa - absolute pressure (gas medium) — 115 kPa to 3000 kPa · CMC 1,8 · 10⁻⁴ · p Pressure |
Internal procedure; Direct method; At permanent location (S) | PP6 based on EURAMET cg-17 v. 4.1 | |
| 14 | Type S noble metal thermocouples | Calibration — Temperature (electrical thermometry) 200 °C to 1100 °C · CMC 0,7 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PP2, PP12 based on EURAMET cg-8 v. 3.1 | |
| 15 | Type K base metal thermocouples | Calibration — Temperature (electrical thermometry) 200 °C to 1100 °C · CMC 1,1 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PP2, PP12 based on EURAMET cg-8 v. 3.1 | |
| 16 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -40 °C do -20 °C · CMC 0,05 °C -20 °C to 150 °C · CMC 0,02 °C 150 °C do 280 °C · CMC 0,03 °C 280 °C do 400 °C · CMC 0,10 °C 400 °C do 500 °C · CMC 0,22 °C 500 °C do 600 °C · CMC 0,34 °C 600 °C do 661 °C · CMC 0,41 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PP3 based on DKD-R 5.1 (11/2023) | |
| 17 | Electrical thermometers (including electronic) Temperature transmitters (containing temperature sensors) |
Calibration — Temperature (electrical thermometry) -40 °C do -20 °C · CMC 0,05 °C -20 °C to 150 °C · CMC 0,02 °C 150 °C do 280 °C · CMC 0,03 °C 280 °C do 400 °C · CMC 0,10 °C 400 °C do 500 °C · CMC 0,22 °C 500 °C do 600 °C · CMC 0,34 °C 600 °C do 661 °C · CMC 0,41 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PP15 based on DKD-R 5.1 (11/2023) EURAMET cg-8 v. 3.1 | |
| 18 | Dial thermometers (bimetallic, expansion, pressure-spring) | Calibration — Temperature (non-electrical thermometry) -40 °C to 120 °C · CMC 0,2 °C 120 °C do 280 °C · CMC 0,6 °C Temperature (non-electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PP11 | |
Reference standards (13)
DKD-R 5.1 (11/2023)
EURAMET cg-15 v. 3.0
EURAMET cg-17 v. 4.1
EURAMET cg-8 v. 3.1
PN-EN IEC 60051-9:2021-09
PP11
PP12
PP15
PP18
PP2
PP3
PP6
PP7