🇵🇱 Introl Sp. z o.o. PL

PCA AP 053 • Katowice, Poland • 15 calibration items

Metrology & calibration

Introl Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 053 (Katowice, Poland). Its accreditation scope covers 15 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electronic pressure gauges, Pressure transmitters, Spring pressure gauges, Hygrometers. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; Outside of permanent location (P).

At a glance

  • Accreditation number: PCA AP 053
  • Calibration items in scope: 15
  • Types of calibrated items: 22
  • Calibration principles used: 6
  • City: Katowice, Poland
  • Valid until: 20-05-2027
Accreditation number
AP 053
Accreditation body
PCA
Address
ul. T. Kościuszki 112, 40-519 Katowice
City
Katowice, Poland
Division
Measurement Laboratory
Valid until
20-05-2027

What this laboratory calibrates (22)

Calibration principles (6)

Certificate information

Issue date
14.04.2026
Issue no.
28
Certificate date
19.02.2020
Accreditation cycle
27.04.2023 — 20.05.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (15)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Hygrometers
Thermohygrometers
Transmitters
Calibration — Relative humidity
temperature — 10 °C to 20 °C · CMC 0,4 °C relative humidity (block 10 °C to 20 °C) — 65 %rh to 95 %rh · CMC 2,0 %rh relative humidity (block 10 °C to 20 °C) — 65 %rh to 95 %rh · CMC 2,9 %rh temperature — 20 °C to 40 °C · CMC 0,4 °C relative humidity (block 20 °C to 40 °C) — 20 %rh to 95 %rh · CMC 1,5 %rh relative humidity (block 20 °C to 40 °C) — 20 %rh to 95 %rh · CMC 2,9 %rh
Relative humidity
Internal procedure; Calibration in a climatic chamber; At permanent location (S) PIL - P.09
2 Hygrometers
Thermohygrometers
Transmitters
Calibration — Relative humidity
temperature — 23 °C · CMC 0,3 °C relative humidity (block 23 °C) — 10 %rh to 80 %rh · CMC 1,5 %rh relative humidity (block 23 °C) — 80 %rh to 90 %rh · CMC 2,0 %rh
Relative humidity
Internal procedure; Calibration in a two-pressure generator; At permanent location (S) PIL - P.09
3 Hygrometers
Thermohygrometers
Transmitters
Calibration — Relative humidity
temperature — 20 °C to 30 °C · CMC 0,3 °C relative humidity (block 20 °C to 30 °C) — 10 %rh to 80 %rh · CMC 1,5 %rh relative humidity (block 20 °C to 30 °C) — 80 %rh to 90 %rh · CMC 2,0 %rh
Relative humidity
Internal procedure; Calibration in a two-pressure generator; Outside of permanent location (P) PIL - P.13
4 Electronic pressure gauges
Spring pressure gauges
Pressure transmitters
- gauge pressure (gas, liquid)
Calibration — Pressure
-0,1 MPa do -0,01 MPa · CMC 0,01 % · p + 0,000001 MPa; p - wielkość mierzona (MPa) -0,01 MPa do 0 MPa · CMC 0,0000011 MPa 0 MPa to 0,0025 MPa · CMC 0,0000006 MPa 0,0025 MPa do 0,017 MPa · CMC 0,0000011 MPa 0,017 MPa do 0,17 MPa · CMC 0,006 % · p + 0,000001 MPa; p - wielkość mierzona (MPa) 0,17 MPa do 0,7 MPa · CMC 0,000043 MPa 0,7 MPa do 7 MPa · CMC 0,006 % · p + 0,00001 MPa; p - wielkość mierzona (MPa)
Pressure
Internal procedure; At permanent location (S) PIL - P.01 PIL - P.02 PIL - P.03 based on EURAMET cg-17 v. 4.1
5 Electronic pressure gauges
Spring pressure gauges
Pressure transmitters
- gauge pressure (liquid)
Calibration — Pressure
7 MPa to 16 MPa · CMC 0,01 % · p + 0,0001 MPa; p - measured value (MPa) 16 MPa to 70 MPa · CMC 0,01 % · p + 0,001 MPa; p - measured value (MPa)
Pressure
Internal procedure; At permanent location (S) PIL - P.01 PIL - P.02 PIL - P.03 based on EURAMET cg-17 v. 4.1
6 Electronic pressure gauges
Spring pressure gauges
Pressure transmitters
- absolute pressure (gas, liquid)
Calibration — Pressure
0,005 MPa to 0,27 MPa · CMC 0,00002 MPa 0,27 MPa to 0,8 MPa · CMC 0,00006 MPa 0,8 MPa to 7,1 MPa · CMC 0,008 % · p + 0,00001 MPa; p - measured value (MPa) 7 MPa to 70 MPa · CMC 0,01 % · p + 0,001 MPa; p - measured value (MPa)
Pressure
Internal procedure; At permanent location (S) PIL - P.11 based on EURAMET cg-17 v. 4.1
7 Electronic pressure gauges (barometers)
- absolute pressure (gas)
Calibration — Pressure
800 hPa to 1200 hPa · CMC 0,2 hPa
Pressure
Internal procedure; At permanent location (S) PIL - P.11 based on EURAMET cg-17 v. 4.1
8 Electronic pressure gauges
Spring pressure gauges
Pressure transmitters
- gauge pressure
Calibration — Pressure
-0,1 MPa to -0,04 MPa · CMC 0,04 % · p + 0,00004 MPa; p - measured value (MPa) -0,04 MPa to 0,04 MPa · CMC 0,04 % · p + 0,00002 MPa; p - measured value (MPa) 0,04 MPa to 0,2 MPa · CMC 0,04 % · p + 0,00005MPa; p - measured value (MPa) 0,2 MPa to 2 MPa · CMC 0,045% · p + 0,0003 MPa; p - measured value (MPa) 2 MPa to 3 MPa · CMC 0,12 % · p; p - measured value (MPa) 3 MPa to 10 MPa · CMC 0,0047 MPa 10 MPa to 30 MPa · CMC 0,12 % · p; p - measured value (MPa)
Pressure
Internal procedure; Outside of permanent location (P) PIL - P.12 based on EURAMET cg-17 v. 4.1
9 Electronic pressure gauges (barometers)
- absolute pressure (gas)
Calibration — Pressure
800 hPa to 1200 hPa · CMC 0,6 hPa
Pressure
Internal procedure; Outside of permanent location (P) PIL - P.12 based on EURAMET cg-17 v. 4.1
10 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-45 °C to -30 °C · CMC 0,20 °C -30 °C to 110 °C · CMC 0,05 °C 110 °C to 200 °C · CMC 0,06 °C 200 °C to 1200 °C · CMC 0,6 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PIL - P.08
11 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-30 °C to 110 °C · CMC 0,05 °C 110 °C to 200 °C · CMC 0,06 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PIL - P.05
12 Temperature transmitters Calibration — Temperature (electrical thermometry)
-30 °C to 110 °C · CMC 0,05 °C 110 °C to 200 °C · CMC 0,06 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PIL - P.08
13 Noble metal thermocouples (type S)
Base metal thermocouples (type K, J)
Calibration — Temperature (electrical thermometry)
200 °C do 1085 °C · CMC 0,6 °C 1085 °C to 1200 °C · CMC 1,5 °C
Temperature (electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PIL - P.06 based on EURAMET cg-8 v. 3.1
14 Liquid-in-glass thermometers Calibration — Temperature (non-electrical thermometry)
-30 °C to 110 °C · CMC 0,05 °C
Temperature (non-electrical thermometry)
Internal procedure; Direct method; At permanent location (S) PIL - P.07
15 Radiation pyrometers
Thermal imaging cameras
Line scanners
Calibration — Temperature (radiation thermometry)
-10 °C to 80 °C · CMC 1,5 °C 150 °C to 1085 °C · CMC 2,0 °C 1085 °C to 1500 °C · CMC 4,7 °C
Temperature (radiation thermometry)
Internal procedure; At permanent location (S) PIL - P.04

Reference standards (14)

EURAMET cg-17 v. 4.1
EURAMET cg-8 v. 3.1
PIL - P.01
PIL - P.02
PIL - P.03
PIL - P.04
PIL - P.05
PIL - P.06
PIL - P.07
PIL - P.08
PIL - P.09
PIL - P.11
PIL - P.12
PIL - P.13

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20053