🇵🇱 Introl Sp. z o.o. PL
Metrology & calibration
Introl Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 053 (Katowice, Poland). Its accreditation scope covers 15 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electronic pressure gauges, Pressure transmitters, Spring pressure gauges, Hygrometers. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; Outside of permanent location (P).
At a glance
- Accreditation number: PCA AP 053
- Calibration items in scope: 15
- Types of calibrated items: 22
- Calibration principles used: 6
- City: Katowice, Poland
- Valid until: 20-05-2027
- Accreditation number
- AP 053
- Accreditation body
- PCA
- Address
- ul. T. Kościuszki 112, 40-519 Katowice
- City
- Katowice, Poland
- Division
- Measurement Laboratory
- Valid until
- 20-05-2027
What this laboratory calibrates (22)
- Electronic pressure gauges — 4 calibration items
- Pressure transmitters — 4 calibration items
- Spring pressure gauges — 4 calibration items
- Hygrometers — 3 calibration items
- Thermohygrometers — 3 calibration items
- Transmitters — 3 calibration items
- - absolute pressure (gas) — 2 calibration items
- Electronic pressure gauges (barometers) — 2 calibration items
- - absolute pressure (gas, liquid) — 1 calibration item
- - gauge pressure — 1 calibration item
- - gauge pressure (gas, liquid) — 1 calibration item
- - gauge pressure (liquid) — 1 calibration item
- Base metal thermocouples (type K, J) — 1 calibration item
- Electrical thermometers (including electronic) — 1 calibration item
- Electrical thermometers (with temperature recording) — 1 calibration item
- and 7 more
Calibration principles (6)
- Internal procedure; At permanent location (S) — 5 calibration items
- Internal procedure; Direct method; At permanent location (S) — 5 calibration items
- Internal procedure; Outside of permanent location (P) — 2 calibration items
- Internal procedure; Calibration in a climatic chamber; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a two-pressure generator; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a two-pressure generator; Outside of permanent location (P) — 1 calibration item
Certificate information
- Issue date
- 14.04.2026
- Issue no.
- 28
- Certificate date
- 19.02.2020
- Accreditation cycle
- 27.04.2023 — 20.05.2027
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (15)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Hygrometers Thermohygrometers Transmitters |
Calibration — Relative humidity temperature — 10 °C to 20 °C · CMC 0,4 °C relative humidity (block 10 °C to 20 °C) — 65 %rh to 95 %rh · CMC 2,0 %rh relative humidity (block 10 °C to 20 °C) — 65 %rh to 95 %rh · CMC 2,9 %rh temperature — 20 °C to 40 °C · CMC 0,4 °C relative humidity (block 20 °C to 40 °C) — 20 %rh to 95 %rh · CMC 1,5 %rh relative humidity (block 20 °C to 40 °C) — 20 %rh to 95 %rh · CMC 2,9 %rh Relative humidity |
Internal procedure; Calibration in a climatic chamber; At permanent location (S) | PIL - P.09 | |
| 2 | Hygrometers Thermohygrometers Transmitters |
Calibration — Relative humidity temperature — 23 °C · CMC 0,3 °C relative humidity (block 23 °C) — 10 %rh to 80 %rh · CMC 1,5 %rh relative humidity (block 23 °C) — 80 %rh to 90 %rh · CMC 2,0 %rh Relative humidity |
Internal procedure; Calibration in a two-pressure generator; At permanent location (S) | PIL - P.09 | |
| 3 | Hygrometers Thermohygrometers Transmitters |
Calibration — Relative humidity temperature — 20 °C to 30 °C · CMC 0,3 °C relative humidity (block 20 °C to 30 °C) — 10 %rh to 80 %rh · CMC 1,5 %rh relative humidity (block 20 °C to 30 °C) — 80 %rh to 90 %rh · CMC 2,0 %rh Relative humidity |
Internal procedure; Calibration in a two-pressure generator; Outside of permanent location (P) | PIL - P.13 | |
| 4 | Electronic pressure gauges Spring pressure gauges Pressure transmitters - gauge pressure (gas, liquid) |
Calibration — Pressure -0,1 MPa do -0,01 MPa · CMC 0,01 % · p + 0,000001 MPa; p - wielkość mierzona (MPa) -0,01 MPa do 0 MPa · CMC 0,0000011 MPa 0 MPa to 0,0025 MPa · CMC 0,0000006 MPa 0,0025 MPa do 0,017 MPa · CMC 0,0000011 MPa 0,017 MPa do 0,17 MPa · CMC 0,006 % · p + 0,000001 MPa; p - wielkość mierzona (MPa) 0,17 MPa do 0,7 MPa · CMC 0,000043 MPa 0,7 MPa do 7 MPa · CMC 0,006 % · p + 0,00001 MPa; p - wielkość mierzona (MPa) Pressure |
Internal procedure; At permanent location (S) | PIL - P.01 PIL - P.02 PIL - P.03 based on EURAMET cg-17 v. 4.1 | |
| 5 | Electronic pressure gauges Spring pressure gauges Pressure transmitters - gauge pressure (liquid) |
Calibration — Pressure 7 MPa to 16 MPa · CMC 0,01 % · p + 0,0001 MPa; p - measured value (MPa) 16 MPa to 70 MPa · CMC 0,01 % · p + 0,001 MPa; p - measured value (MPa) Pressure |
Internal procedure; At permanent location (S) | PIL - P.01 PIL - P.02 PIL - P.03 based on EURAMET cg-17 v. 4.1 | |
| 6 | Electronic pressure gauges Spring pressure gauges Pressure transmitters - absolute pressure (gas, liquid) |
Calibration — Pressure 0,005 MPa to 0,27 MPa · CMC 0,00002 MPa 0,27 MPa to 0,8 MPa · CMC 0,00006 MPa 0,8 MPa to 7,1 MPa · CMC 0,008 % · p + 0,00001 MPa; p - measured value (MPa) 7 MPa to 70 MPa · CMC 0,01 % · p + 0,001 MPa; p - measured value (MPa) Pressure |
Internal procedure; At permanent location (S) | PIL - P.11 based on EURAMET cg-17 v. 4.1 | |
| 7 | Electronic pressure gauges (barometers) - absolute pressure (gas) |
Calibration — Pressure 800 hPa to 1200 hPa · CMC 0,2 hPa Pressure |
Internal procedure; At permanent location (S) | PIL - P.11 based on EURAMET cg-17 v. 4.1 | |
| 8 | Electronic pressure gauges Spring pressure gauges Pressure transmitters - gauge pressure |
Calibration — Pressure -0,1 MPa to -0,04 MPa · CMC 0,04 % · p + 0,00004 MPa; p - measured value (MPa) -0,04 MPa to 0,04 MPa · CMC 0,04 % · p + 0,00002 MPa; p - measured value (MPa) 0,04 MPa to 0,2 MPa · CMC 0,04 % · p + 0,00005MPa; p - measured value (MPa) 0,2 MPa to 2 MPa · CMC 0,045% · p + 0,0003 MPa; p - measured value (MPa) 2 MPa to 3 MPa · CMC 0,12 % · p; p - measured value (MPa) 3 MPa to 10 MPa · CMC 0,0047 MPa 10 MPa to 30 MPa · CMC 0,12 % · p; p - measured value (MPa) Pressure |
Internal procedure; Outside of permanent location (P) | PIL - P.12 based on EURAMET cg-17 v. 4.1 | |
| 9 | Electronic pressure gauges (barometers) - absolute pressure (gas) |
Calibration — Pressure 800 hPa to 1200 hPa · CMC 0,6 hPa Pressure |
Internal procedure; Outside of permanent location (P) | PIL - P.12 based on EURAMET cg-17 v. 4.1 | |
| 10 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) |
Calibration — Temperature (electrical thermometry) -45 °C to -30 °C · CMC 0,20 °C -30 °C to 110 °C · CMC 0,05 °C 110 °C to 200 °C · CMC 0,06 °C 200 °C to 1200 °C · CMC 0,6 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PIL - P.08 | |
| 11 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -30 °C to 110 °C · CMC 0,05 °C 110 °C to 200 °C · CMC 0,06 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PIL - P.05 | |
| 12 | Temperature transmitters | Calibration — Temperature (electrical thermometry) -30 °C to 110 °C · CMC 0,05 °C 110 °C to 200 °C · CMC 0,06 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PIL - P.08 | |
| 13 | Noble metal thermocouples (type S) Base metal thermocouples (type K, J) |
Calibration — Temperature (electrical thermometry) 200 °C do 1085 °C · CMC 0,6 °C 1085 °C to 1200 °C · CMC 1,5 °C Temperature (electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PIL - P.06 based on EURAMET cg-8 v. 3.1 | |
| 14 | Liquid-in-glass thermometers | Calibration — Temperature (non-electrical thermometry) -30 °C to 110 °C · CMC 0,05 °C Temperature (non-electrical thermometry) |
Internal procedure; Direct method; At permanent location (S) | PIL - P.07 | |
| 15 | Radiation pyrometers Thermal imaging cameras Line scanners |
Calibration — Temperature (radiation thermometry) -10 °C to 80 °C · CMC 1,5 °C 150 °C to 1085 °C · CMC 2,0 °C 1085 °C to 1500 °C · CMC 4,7 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) | PIL - P.04 | |
Reference standards (14)
EURAMET cg-17 v. 4.1
EURAMET cg-8 v. 3.1
PIL - P.01
PIL - P.02
PIL - P.03
PIL - P.04
PIL - P.05
PIL - P.06
PIL - P.07
PIL - P.08
PIL - P.09
PIL - P.11
PIL - P.12
PIL - P.13