🇵🇱 Sieć Badawcza Łukasiewicz - Instytut Sztucznej Inteligencji i Cyberbezpieczeństwa PL
Metrology & calibration
Sieć Badawcza Łukasiewicz - Instytut Sztucznej Inteligencji i Cyberbezpieczeństwa is a calibration laboratory accredited by PCA under number AP 051 (Katowice, Poland). Its accreditation scope covers 14 calibration items in the following areas: Metrology & calibration. Calibrated items include: Measurement cards, Multimeters, Analogue current meters, Analogue voltage meters. Most frequently used calibration principles: Internal procedure; At permanent location (S).
At a glance
- Accreditation number: PCA AP 051
- Calibration items in scope: 14
- Types of calibrated items: 17
- Calibration principles used: 1
- City: Katowice, Poland
- Valid until: 13-02-2027
- Accreditation number
- AP 051
- Accreditation body
- PCA
- Address
- ul. Leopolda 31, 40-189 Katowice
- City
- Katowice, Poland
- Division
- Testing and Certification Centre, Calibration Laboratory
- Valid until
- 13-02-2027
What this laboratory calibrates (17)
- Measurement cards — 5 calibration items
- Multimeters — 5 calibration items
- Analogue current meters — 2 calibration items
- Analogue voltage meters — 2 calibration items
- Digital current meters — 2 calibration items
- Digital voltage meters — 2 calibration items
- Artificial networks — 1 calibration item
- Attenuators, matching networks, adapters, RF cables — 1 calibration item
- CDN networks — 1 calibration item
- Combination wave (SURGE) generators — 1 calibration item
- Digital insulation resistance meters — 1 calibration item
- Digital resistance meters — 1 calibration item
- Electrical fast transient/burst (EFT/B) generators — 1 calibration item
- Electrostatic discharge (ESD) generator — 1 calibration item
- Fixed resistors — 1 calibration item
- and 2 more
Calibration principles (1)
- Internal procedure; At permanent location (S) — 14 calibration items
Certificate information
- Issue date
- 05.08.2025
- Issue no.
- 23
- Certificate date
- 05.08.2025
- Accreditation cycle
- 23.01.2023 — 13.02.2027
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (14)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Analogue voltage meters Digital voltage meters Multimeters Measurement cards |
Calibration — DC voltage 2 mV to 20 mV · CMC 0,07 % 20 mV to 200 mV · CMC 0,009 % 0,2 V to 1000 V · CMC 0,005 % DC voltage |
Internal procedure; At permanent location (S) | L-PW-E01 L-PW-E02 based on EURAMET cg-15 v 3.0 | |
| 2 | Analogue current meters Digital current meters Multimeters Measurement cards |
Calibration — DC current 0,02 mA to 0,2 mA · CMC 0,12 % 0,2 mA to 200 mA · CMC 0,02 % 0,2 A to 2 A · CMC 0,03 % 2 A to 20 A · CMC 0,12 % DC current |
Internal procedure; At permanent location (S) | L-PW-E01 L-PW-E02 based on EURAMET cg-15 v 3.0 | |
| 3 | Analogue voltage meters Digital voltage meters Multimeters Measurement cards |
Calibration — AC voltage 50 Hz to 1 kHz — 2 mV to 20 mV · CMC 0,32 % 50 Hz to 1 kHz — 0,02 V to 200 V · CMC 0,045 % 50 Hz to 1 kHz — 200 V to 1000 V · CMC 0,06 % 1 kHz to 5 kHz — 2 mV to 20 mV · CMC 0,54 % 1 kHz to 5 kHz — 0,02 V to 200 V · CMC 0,06 % 1 kHz to 5 kHz — 200 V to 1000 V · CMC 0,08 % 5 kHz to 100 kHz — 2 mV to 20 mV · CMC 1,5 % 5 kHz to 100 kHz — 0,02 V to 200 V · CMC 0,05 % AC voltage |
Internal procedure; At permanent location (S) | L-PW-E01 L-PW-E02 based on EURAMET cg-15 v 3.0 | |
| 4 | Analogue current meters Digital current meters Multimeters Measurement cards |
Calibration — AC current 50 Hz to 1 kHz — 0,02 mA to 0,2 mA · CMC 3,0 % 50 Hz to 1 kHz — 0,2 mA to 2 mA · CMC 0,2 % 50 Hz to 1 kHz — 2 mA to 20 mA · CMC 0,08 % 50 Hz to 1 kHz — 0,02 A to 20 A · CMC 0,14 % 1 kHz to 5 kHz — 0,2 mA to 2 mA · CMC 1,3 % 1 kHz to 5 kHz — 2 mA to 20 mA · CMC 0,27 % 1 kHz to 5 kHz — 0,02 A to 2 A · CMC 0,14 % AC current |
Internal procedure; At permanent location (S) | L-PW-E01 L-PW-E02 based on EURAMET cg-15 v 3.0 | |
| 5 | Multimeters Digital resistance meters Measurement cards |
Calibration — DC resistance 0,01 Ω to 0,1 Ω · CMC 0,0058 Ω 0,1 Ω to 5 Ω · CMC 0,1 % 5 Ω MΩ to 1 MΩ · CMC 0,012 % 1 MΩ to 19 MΩ · CMC 0,035 % 19 MΩ to 1 GΩ · CMC 1,8 % DC resistance |
Internal procedure; At permanent location (S) | L-PW-E02 based on EURAMET cg-15 v 3.0 | |
| 6 | Digital insulation resistance meters | Calibration — DC resistance 0,1 MΩ to 1 GΩ · CMC 1,8 % 1 GΩ to 1 TΩ · CMC 2,1 % DC resistance |
Internal procedure; At permanent location (S) | L-PW-E03 | |
| 7 | Fixed resistors Variable resistors |
Calibration — DC resistance 1 Ω to 10 Ω · CMC 0,5 % 10 Ω to 1 kΩ · CMC 0,02 % 1 kΩ to 1 MΩ · CMC 0,013% 1 MΩ to 10 MΩ · CMC 0,05 % 10 MΩ to 100 MΩ · CMC 0,19 % DC resistance |
Internal procedure; At permanent location (S) | L-PW-E08 L-PW-E09 | |
| 8 | Electrostatic discharge (ESD) generator | Calibration — High-frequency electrical quantities test voltage — 1 kV to 2 kV · CMC 4,0 % test voltage — 2 kV to 30 kV · CMC 2,1 % rise time — 0,6 ns to 1 ns · CMC 5,8 % peak discharge current — 3 A to 124 A · CMC 6,2 % discharge current after (30 ÷ 65) ns — 2 A to 78 A · CMC 6,8 % discharge current after (65 ÷ 130) ns — 1 A to 39 A · CMC 6,4 % discharge current after (130 ÷ 800) ns — 0 A to 39 A · CMC 6,4 % High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E20 based on PN-EN 61000-4-2:2011 ISO 10605:2008 + Amd 1:2014 Cor 1:2010 ISO 10605:2023 | |
| 9 | Electrical fast transient/burst (EFT/B) generators | Calibration — High-frequency electrical quantities peak output voltage — 0,2 kV to 5,5 kV · CMC for 50 Ω 5,3 %; for 1000 Ω 6,5 %; at the coupling output 8,0 % pulse rise time — 3 ns to 7 ns · CMC for 50 Ω 4,9 %; for 1000 Ω 6,6 %; at the coupling output 4,9 % pulse duration — 20 ns to 150 ns · CMC for 50 Ω 4,9 %; for 1000 Ω 6,2 %; at the coupling output 4,9 % burst duration — 0,1 ms to 75 ms · CMC 4,8 % pulse repetition frequency — 1 kHz to 1000 kHz · CMC 4,8 % burst period — 200 ms to 400 ms · CMC 4,8 % High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E21 based on PN-EN 61000-4-4:2010 + A1:2010 PN-EN 61000-4-4:2013-05 | |
| 10 | Combination wave (SURGE) generators | Calibration — High-frequency electrical quantities open-circuit voltages — 0,2 kV to 7 kV · CMC 4,0 % peak short-circuit current — 0,01 kA to 3,5 kA · CMC 4,0 % rise time — 0,5 µs to 8 µs · CMC 4,5 % duration — 10 µs to 1000 µs · CMC 3,0 % front time — 0,5 µs to 15 µs · CMC 3,0 % time to half-value — 15 µs to 1000 µs · CMC 4,5 % High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E22 based on PN-EN 61000-4-5:2010 PN-EN 61000-4-5:2014-10 + A1:2018-01 | |
| 11 | Voltage dips, short interruptions and voltage variations generators | Calibration — High-frequency electrical quantities no-load output voltage — 20 V to 600 V · CMC 4,0 % peak inrush current drive capability — 200 A to 1000 A · CMC 6,3 % instantaneous peak overshoot/undershoot — 20 V to 600 V · CMC 4,0 % rise and fall time — 1 µs to 5 µs · CMC 3,0 % phase shift — 0° to 360° · CMC 3,0 % zero-crossing control of the generator — -20° to 20° · CMC 3,0 % High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E23 based on PN-EN 61000-4-11:2007 + A1:2017-09 PN-EN IEC 61000-4-11:2020-11 + AC:2020-12 + AC:2023-01 | |
| 12 | CDN networks | Calibration — High-frequency electrical quantities impedance magnitude (0,01 MHz to 230 MHz) — 150 Ω ±100 Ω · CMC 5% impedance phase (0,01 MHz to 230 MHz) — -30° to 90° · CMC 3° attenuation (0,01 MHz to 230 MHz) — 5 dB to 20 dB · CMC 0,5 dB High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E26 based on PN-EN 61000-4-6:2024-03 PN-EN 55016-1-2:2014-09 + A1:2018-03 | |
| 13 | Artificial networks | Calibration — High-frequency electrical quantities impedance magnitude (9 kHz to 108 MHz) — 3 Ω to 100 Ω · CMC 6% impedance phase (9 kHz to 150 kHz) — -30° to 90° · CMC 3° impedance phase (150 kHz to 108 MHz) — -30° to 90° · CMC 2,5° voltage division factor (9 kHz to 108 MHz) — 0 dB to 10 dB · CMC 0,2 dB voltage division factor (9 kHz to 108 MHz) — 10 dB to 50 dB · CMC 0,5 dB voltage division factor (9 kHz to 108 MHz) — 50 dB to 90 dB · CMC 3,8 dB attenuation factor (9 kHz to 108 MHz) — 30 dB to 50 dB · CMC 1,0 dB attenuation factor (9 kHz to 108 MHz) — 50 dB to 80 dB · CMC 4,0 dB attenuation factor (9 kHz to 108 MHz) — 80 dB to 90 dB · CMC 5,5 dB High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E27 based on PN-EN 55016-1-2:2014-09 + A1:2018-03 | |
| 14 | Attenuators, matching networks, adapters, RF cables | Calibration — High-frequency electrical quantities attenuation (9 kHz to 4 GHz) — 0 dB to 30 dB · CMC 0,3 dB attenuation (9 kHz to 4 GHz) — 30 dB to 60 dB · CMC 1,0 dB High-frequency electrical quantities |
Internal procedure; At permanent location (S) | L-PW-E28 | |
Reference standards (24)
EURAMET cg-15 v 3.0
ISO 10605:2008 + Amd 1:2014 Cor 1:2010
ISO 10605:2023
L-PW-E01
L-PW-E02
L-PW-E03
L-PW-E08
L-PW-E09
L-PW-E20
L-PW-E21
L-PW-E22
L-PW-E23
L-PW-E26
L-PW-E27
L-PW-E28
PN-EN 55016-1-2:2014-09 + A1:2018-03
PN-EN 61000-4-11:2007 + A1:2017-09
PN-EN 61000-4-2:2011
PN-EN 61000-4-4:2010 + A1:2010
PN-EN 61000-4-4:2013-05
PN-EN 61000-4-5:2010
PN-EN 61000-4-5:2014-10 + A1:2018-01
PN-EN 61000-4-6:2024-03
PN-EN IEC 61000-4-11:2020-11 + AC:2020-12 + AC:2023-01