🇵🇱 LabStand Sp. z o.o. PL
Metrology & calibration
LabStand Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 021 (Poznań, Poland). Its accreditation scope covers 30 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electrical thermometers (data loggers), Electrical thermometers (including electronic), Conductometers, Dispensers. Most frequently used calibration principles: Internal procedure; At permanent location (S), At permanent location (S), At permanent location (S) and outside of it (P).
At a glance
- Accreditation number: PCA AP 021
- Calibration items in scope: 30
- Types of calibrated items: 32
- Calibration principles used: 12
- City: Poznań, Poland
- Valid until: 05-11-2028
- Accreditation number
- AP 021
- Accreditation body
- PCA
- Address
- ul. Grunwaldzka 114, 60-308 Poznań
- City
- Poznań, Poland
- Division
- Calibration Laboratory
- Valid until
- 05-11-2028
What this laboratory calibrates (32)
- Electrical thermometers (data loggers) — 4 calibration items
- Electrical thermometers (including electronic) — 4 calibration items
- Conductometers — 2 calibration items
- Dispensers — 2 calibration items
- pH meters — 2 calibration items
- Piston burettes — 2 calibration items
- Barometers — 1 calibration item
- Breath alcohol analyzer — 1 calibration item
- Climatic chambers — 1 calibration item
- Conductivity reference materials — 1 calibration item
- Conductivity sensors — 1 calibration item
- Digital manometers — 1 calibration item
- Furnaces — 1 calibration item
- Glass capillary viscometers — 1 calibration item
- Hygrometers — 1 calibration item
- and 17 more
Calibration principles (12)
- Internal procedure; At permanent location (S) — 7 calibration items
- At permanent location (S) — 6 calibration items
- At permanent location (S) and outside of it (P) — 5 calibration items
- Internal procedure; Electrical method; At permanent location (S) — 2 calibration items
- Internal procedure; Method using reference materials; At permanent location (S) — 2 calibration items
- Value of the redox potential relative to the Ag/AgCl electrode (3 mol KCI); At permanent location (S) — 2 calibration items
- Internal procedure; Direct method Calibration in a liquid thermostat; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method Calibration in the climate chamber; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method Calibration in the temperature calibrator; At permanent location (S) — 1 calibration item
- Internal procedure; Direct method Calibration in the temperature calibrator; At permanent location (S) and outside of it (P) — 1 calibration item
- Internal procedure; Indirect method Direct method; At permanent location (S) — 1 calibration item
- Internal procedure; Mass concentration of ethanol in exhaled air (mg of ethanol in 1 l of air); At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 14.10.2026
- Issue no.
- 27
- Certificate date
- 01.06.2022
- Accreditation cycle
- 14.10.2024 — 05.11.2028
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (30)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | pH meters | Calibration — PH pH — 0 to 14 · CMC 0,0014 PH |
Internal procedure; Electrical method; At permanent location (S) | PT 01/400 | |
| 2 | pH meters | Calibration — PH DC voltage — -2000 mV to 2000 mV · CMC 0,13 mV pH — 1 to 13 · CMC 0,01 PH |
Internal procedure; Method using reference materials; At permanent location (S) | PT 01/400 | |
| 3 | pH electrodes | Calibration — PH slope of the characteristic — 45,0 mV to 65,0 mV · CMC 0,25 mV electrode efficiency — 50,0 % to 110,0 % · CMC 0,3 % zero point (pH for E = 0 mV) — 2,00 to 10,00 · CMC 0,02 PH |
Internal procedure; At permanent location (S) | PT 01/300 | |
| 4 | Conductometers | Calibration — Specific electrical conductivity (conductometry) 0,1 µS/cm to 1,99 µS/cm · CMC 0,001 µS/cm 2 µS/cm to 200 mS/cm · CMC 0,1 % Specific electrical conductivity (conductometry) |
Internal procedure; Electrical method; At permanent location (S) | PT 01/700 | |
| 5 | Conductometers | Calibration — Specific electrical conductivity (conductometry) 0,08 mS/cm to 0,15 mS/cm · CMC 0,4 % 0,15 mS/cm to 115 mS/cm · CMC 0,2 % Specific electrical conductivity (conductometry) |
Internal procedure; Method using reference materials; At permanent location (S) | PT 01/700 | |
| 6 | Conductivity sensors | Calibration — Specific electrical conductivity (conductometry) 0,01 cm⁻¹ to 11 cm⁻¹ · CMC 0,2 % Specific electrical conductivity (conductometry) |
Internal procedure; At permanent location (S) | PT 01/800 | |
| 7 | Breath alcohol analyzer | Calibration — Mass concentration (breath analysis) 0,00 mg/l do 0,40 mg/l · CMC 0,01 mg/l 0,41 mg/l do 1,00 mg/l · CMC 0,02 mg/l 1,01 mg/l to 1,50 mg/l · CMC 0,03 mg/l 1,51 mg/l to 2,00 mg/l · CMC 0,05 mg/l Mass concentration (breath analysis) |
Internal procedure; Mass concentration of ethanol in exhaled air (mg of ethanol in 1 l of air); At permanent location (S) | PT 04/200 based on OIML R126-2:2021 A.1.1, A.2 | |
| 8 | Oxygen meters | Calibration — Volume fraction, mass fraction (liquid analysis) 0 mg/l · CMC 0,02 mg/l 5 mg/l to 13 mg/l1) · CMC 0,11 mg/l 0 % · CMC 0,3 % 100 % · CMC 1,35 % Volume fraction, mass fraction (liquid analysis) |
At permanent location (S) | PN-EN ISO 5814:2013 | |
| 9 | Redox electrodes | Calibration — Redox potential 200 mV to 700 mV · CMC 5 mV Redox potential |
Value of the redox potential relative to the Ag/AgCl electrode (3 mol KCI); At permanent location (S) | ASTM D1498-14 | |
| 10 | pH reference materials | Calibration — Reference materials - measurement 1,00 to 8,00 · CMC 0,01 8,00 to 10,10 · CMC 0,02 10,1 to 12,5 · CMC 0,1 Reference materials - measurement |
Internal procedure; At permanent location (S) | PP 01/200 | |
| 11 | Viscosity reference materials | Calibration — Reference materials - measurement 1 mm²/s to 140 mm²/s · CMC 0,2 % 1 mPa · s to 140 mPa · s · CMC 0,2 % 140 mm²/s to 2000 mm²/s · CMC 0,3 % 140 mPa · s to 2000 mPa · s · CMC 0,3 % Reference materials - measurement |
Internal procedure; At permanent location (S) | PP 02/200 PP 02/201 | |
| 12 | Conductivity reference materials | Calibration — Reference materials - measurement 0,0004 S/m to 12,0 S/m · CMC 0,4 % Reference materials - measurement |
Internal procedure; At permanent location (S) | PP 07/200 | |
| 13 | Redox reference material | Calibration — Reference materials - measurement 200 mV to 700 mV · CMC 5 mV Reference materials - measurement |
Value of the redox potential relative to the Ag/AgCl electrode (3 mol KCI); At permanent location (S) | ASTM D1498-14 | |
| 14 | Pycnometers | Calibration — Density (liquid) volume — 10 cm³ to 51 cm³ · CMC 0,003 cm³ volume — 51 cm³ to 201 cm³ · CMC 0,01 cm³ volume — 202 cm³ to 2000 cm³ · CMC 0,3 cm³ Density (liquid) |
At permanent location (S) | PN-EN ISO 2811-1:2023-03 | |
| 15 | Glass capillary viscometers | Calibration — Viscosity 0,001 mm²/s² to 10,5 mm²/s² · CMC 0,5 % Viscosity |
Internal procedure; At permanent location (S) | PT 02/300 | |
| 16 | Rotational viscometers | Calibration — Viscosity 1 mPa · s to 2000 mPa · s · CMC 2 % 2000 mPa · s to 15000 mPa · s · CMC 3 % Viscosity |
At permanent location (S) and outside of it (P) | PN-EN ISO 2555:2018-07 DIN 53019-2:2001 | |
| 17 | Hygrometers Thermohygrometers |
Calibration — Relative humidity temperature — 15 °C to 20 °C · CMC 0,1 °C relative humidity — 20 %rh to 30 %rh · CMC 0,7 %rh temperature — 20 °C to 40 °C · CMC 0,1 °C relative humidity — 20 %rh to 30 %rh · CMC 0,7 %rh temperature — 10 °C to 42 °C · CMC 0,1 °C relative humidity — 30 %rh to 85 %rh · CMC 0,7 %rh temperature — 7 °C to 10 °C · CMC 0,1 °C relative humidity — 30 %rh to 50 %rh · CMC 0,8 %rh temperature — 5 °C to 10 °C · CMC 0,1 °C relative humidity — 50 %rh to 85 %rh · CMC 0,8 %rh temperature — 42 °C to 50 °C · CMC 0,1 °C relative humidity — 20 %rh to 60 %rh · CMC 0,8 %rh Relative humidity |
Internal procedure; Indirect method Direct method; At permanent location (S) | PT 09/100 | |
| 18 | Non-automatic electronic weighing instruments Non-automatic mechanical weighing instruments |
Calibration — Mass (scales) up 1020 g · CMC 1,8 · 10⁻⁴ % 1020 g to 9000 g · CMC 4,6 · 10⁻⁴ % 9 kg to 31 kg · CMC 1,3 · 10⁻³ % Mass (scales) |
At permanent location (S) and outside of it (P) | Euramet cg-18 v. 4.0 | |
| 19 | Digital manometers Pressure gauges Barometers |
Calibration — Pressure 700 hPa to 1100 hPa · CMC 0,25 hPa Pressure |
At permanent location (S) | Euramet cg-17 v. 4.1 point 5.2.1 point 5.2.2 | |
| 20 | Climatic chambers Thermostatic chambers |
Calibration — Temperature (electric thermometry) -80 °C to 180 °C · CMC 0,3 °C2) 180 °C to 250 °C · CMC 1 °C2) Temperature (electric thermometry) |
At permanent location (S) and outside of it (P) | Euramet cg-20 v. 5.0 | |
| 21 | Liquid baths | Calibration — Temperature (electric thermometry) -40 °C to 250 °C · CMC 0,3 °C Temperature (electric thermometry) |
At permanent location (S) and outside of it (P) | Euramet cg-13 v. 4.0 | |
| 22 | Furnaces | Calibration — Temperature (electric thermometry) 30 °C to 170 °C · CMC 0,5 °C 171 °C to 1100 °C · CMC 1,8 °C Temperature (electric thermometry) |
At permanent location (S) and outside of it (P) | Euramet cg-13 v. 4.0 | |
| 23 | Electrical thermometers (including electronic) Electrical thermometers (data loggers) |
Calibration — Temperature (electric thermometry) -40 °C to 80 °C · CMC 0,1 °C 80 °C to 180 °C · CMC 0,2 °C Temperature (electric thermometry) |
Internal procedure; Direct method Calibration in the climate chamber; At permanent location (S) | PT 05/100 | |
| 24 | Electrical thermometers (including electronic) Electrical thermometers (data loggers) |
Calibration — Temperature (electric thermometry) -80 °C to 180 °C · CMC 0,03 °C Temperature (electric thermometry) |
Internal procedure; Direct method Calibration in a liquid thermostat; At permanent location (S) | PT 05/100 | |
| 25 | Electrical thermometers (including electronic) Electrical thermometers (data loggers) |
Calibration — Temperature (electric thermometry) -100 °C to -80 °C · CMC 0,5 °C -80 °C to 30 °C · CMC 0,2 °C 30 °C to 250 °C · CMC 0,2 °C Temperature (electric thermometry) |
Internal procedure; Direct method Calibration in the temperature calibrator; At permanent location (S) and outside of it (P) | PT 05/100 | |
| 26 | Electrical thermometers (including electronic) Electrical thermometers (data loggers) |
Calibration — Temperature (electric thermometry) 250°C to 660 °C · CMC 0,2 °C 660°C to 700 °C · CMC 0,8 °C 700°C to 1100 °C · CMC 2,1 °C Temperature (electric thermometry) |
Internal procedure; Direct method Calibration in the temperature calibrator; At permanent location (S) | PT 05/100 | |
| 27 | Liquid in glass thermometers | Calibration — Temperature (non-electric thermometry) -40 °C to 180 °C · CMC 0,04 °C Temperature (non-electric thermometry) |
Internal procedure; At permanent location (S) | PT 05/100 | |
| 28 | Pyrometers | Calibration — Temperature (radiation thermometry) -15 °C do 0 °C · CMC 1,6 °C 0 °C to 120 °C · CMC 1,4 °C 120 °C do 300 °C · CMC 2,6 °C 300 °C do 500 °C · CMC 2,8 °C Temperature (radiation thermometry) |
At permanent location (S) | VDI/VDE 3511 Part 4.4 | |
| 29 | Piston pipettes Piston burettes Dispensers |
Calibration — Volume 0,1 µl to 2,5 µl · CMC 0,025 µl 2,5 µl to 20 µl · CMC 0,040 µl 20 µl to 200 µl · CMC 0,30 µl 200 µl to 1000 µl · CMC 1,8 µl 1000 µl to 5000 µl · CMC 9 µl 5000 µl to 10000 µl · CMC 14 µl 10000 µl to 20000 µl · CMC 27 µl Volume |
At permanent location (S) | PN-EN ISO 8655-6:2022-11 | |
| 30 | Piston burettes Dispensers |
Calibration — Volume 20000 µl to 50000 µl · CMC 35 µl Volume |
At permanent location (S) | PN-EN ISO 8655-6:2022-11 | |
Reference standards (24)
ASTM D1498-14
DIN 53019-2:2001
Euramet cg-13 v. 4.0
Euramet cg-17 v. 4.1
Euramet cg-18 v. 4.0
Euramet cg-20 v. 5.0
OIML R126-2:2021 A.1.1, A.2
PN-EN ISO 2555:2018-07
PN-EN ISO 2811-1:2023-03
PN-EN ISO 5814:2013
PN-EN ISO 8655-6:2022-11
PP 01/200
PP 02/200
PP 02/201
PP 07/200
PT 01/300
PT 01/400
PT 01/700
PT 01/800
PT 02/300
PT 04/200
PT 05/100
PT 09/100
VDI/VDE 3511 Part 4.4