🇵🇱 Instytut Łączności - Państwowy Instytut Badawczy PL

PCA AP 016 • Wrocław, Poland • 19 calibration items

Metrology & calibration

Instytut Łączności - Państwowy Instytut Badawczy is a calibration laboratory accredited by PCA under number AP 016 (Wrocław, Poland). Its accreditation scope covers 19 calibration items in the following areas: Metrology & calibration. Calibrated items include: Artificial networks, Calibration pulse generator, CDN networks, CMAD absorbing clamps. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; Direct method; At permanent location (S).

At a glance

  • Accreditation number: PCA AP 016
  • Calibration items in scope: 19
  • Types of calibrated items: 19
  • Calibration principles used: 4
  • City: Wrocław, Poland
  • Valid until: 29-10-2027
Accreditation number
AP 016
Accreditation body
PCA
Address
ul. Swojczycka 38, 51-501 Wrocław
City
Wrocław, Poland
Division
Department of Electromagnetic Compatibility and Radio Networks, EMC Measuring Equipment Laboratory
Valid until
29-10-2027

What this laboratory calibrates (19)

Calibration principles (4)

Certificate information

Issue date
28.10.2026
Issue no.
23
Certificate date
06.05.2026
Accreditation cycle
29.09.2023 — 29.10.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (19)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Discontinuous disturbance analysers Calibration — RF electrical quantities
- accuracy of the disturbance analysis level setting — 20 dB(μV) to 100 dB(μV) · CMC 0,5 dB - duration of a disturbance group — 100 μs to 6 s · CMC 2%
RF electrical quantities
Internal procedure; At permanent location (S) IW3 based on PN-EN 55016-1-1:2019-07
2 MDS absorbing clamps Calibration — RF electrical quantities
- clamp factor — - 25 dB(pW/μV) to 5 dB(pW/μV) · CMC 1,5 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW5 based on PN-EN 55016-1-3:2007 + A1:2016-07 + A2:2020-10
3 CMAD absorbing clamps Calibration — RF electrical quantities
30 MHz to 1 GHz; - reflection coefficient (S11) — 0 to 1 · CMC 0,05 30 MHz to 1 GHz; - transmission (S21) — 0 to 1 · CMC 0,05
RF electrical quantities
Internal procedure; At permanent location (S) IW18 based on PN-EN 55016-1-4:2019-04 + A1:2021-03 + A2:2024-01
4 EFT/B generators Calibration — RF electrical quantities
- burst duration — 0,5 ms to 100 ms · CMC 4% - burst period — 200 ms to 400 ms · CMC 4% - pulse rise time — 3 ns to 10 ms · CMC 5% - pulse duration — 10 ns to 5 s · CMC 5% - pulse peak value — 0,1 kV to 8 kV · CMC 5% - pulse repetition frequency within a burst — 3 kHz to 120 kHz · CMC 2%
RF electrical quantities
Internal procedure; At permanent location (S) IW12 based on PN-EN 61000-4-4:2013 ISO 7637-2:2011 ISO 7637-3:2016
5 ESD generators Calibration — RF electrical quantities
- pulse test voltage — 0,5 kV to 40 kV · CMC 4% - discharge current peak value — -165 A to 165 A · CMC 5% - discharge current — -104 A to 104 A · CMC 5% - discharge pulse rise time — 0,4 ns to 5 ns · CMC 6%
RF electrical quantities
Internal procedure; At permanent location (S) IW11 based on PN-EN 61000-4-2:2011 PN-EN IEC 61000-4-2:2026-03 ISO 10605:2023
6 Combination wave generators Calibration — RF electrical quantities
- pulse front time for open-circuit voltage — 0,5 μs to 15 μs · CMC 6% - pulse duration to half value for open-circuit voltage — 20 μs to 1000 μs · CMC 5% - open-circuit voltage peak value — -6 kV to 6 kV · CMC 4% - pulse front time for short-circuit current — 0,5 μs to 20 μs · CMC 4% - pulse duration to half value for short-circuit current — 5 μs to 500 μs · CMC 5% - short-circuit current peak value — -4 kA to 4 kA · CMC 6%
RF electrical quantities
Internal procedure; At permanent location (S) IW13 based on PN-EN 61000-4-5:2014 + A1:2018-01 PN-EN 61000-4-5:2010
7 Calibration pulse generator Calibration — RF electrical quantities
- frequency — 1 Hz to 10 kHz · CMC 0,1 % - pulse area; band A — 13,5 μVs · CMC 0,2 μVs - pulse area; band B — 0,316 μVs · CMC 0,02 μVs - pulse area; bands C, D — 0,044 μVs · CMC 0,002 μVs - pulse spectral amplitude — 20 dB(μV) to 80 dB(μV) · CMC 1,0 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW2 based on PN-EN 55016-1-1:2019-07
8 Damped oscillatory wave generators Calibration — RF electrical quantities
- open-circuit voltage rise time — 3 ns to 100 ns · CMC 5% - open-circuit voltage oscillation frequency — 0,080 MHz to 35 MHz · CMC 2% - open-circuit waveform repetition frequency — 30 Hz to 6000 Hz · CMC 2% - open-circuit waveform decay — 0,1 to 1 · CMC 4% - open-circuit burst duration — 4 ms to 5 s · CMC 5% - open-circuit voltage peak value — 0,2 kV to 5 kV · CMC 4% - short-circuit current peak value — 1 A to 100 A · CMC 6% - short-circuit waveform decay — 0,1 to 1 · CMC 6% - short-circuit burst duration — 1,5 s to 5 s · CMC 5% - open-circuit burst period — 200 ms to 400 ms · CMC 4% - short-circuit current rise time — 30 ns to 400 ns · CMC 5% - short-circuit oscillation frequency — 2 MHz to 40 MHz · CMC 2%
RF electrical quantities
Internal procedure; At permanent location (S) IW16 based on PN-EN 61000-4-18:2019-08
9 Damped sinusoidal wave (ring wave) generators Calibration — RF electrical quantities
- open-circuit voltage rise time — 0,3 μs to 0,7 μs · CMC 5% - open-circuit voltage oscillation frequency — 80 kHz to 120 kHz · CMC 2% - open-circuit decay — 0,3 to 1,1 · CMC 4% - open-circuit voltage peak value — 0,2 kV to 5 kV · CMC 4% - short-circuit current rise time — 0,1 μs to 1,5 μs · CMC 5% - short-circuit current peak value — 7 A to 400 A · CMC 6%
RF electrical quantities
Internal procedure; At permanent location (S) IW15 based on PN-EN 61000-4-12:2017-12
10 Voltage dip and interruption generators Calibration — RF electrical quantities
- open-circuit output voltage — 20 V to 600 V · CMC 4,0 % - peak inrush current drive capability (Inrush) — 200 A to 1000 A · CMC 6,5 % - instantaneous peak overshoot/undershoot of the actual voltage with the generator loaded with a 100 Ω resistance — -12 V to 600 V · CMC 4,0 % - voltage rise/fall time tr during an abrupt change — 1 µs to 5 µs · CMC 3,0 % - voltage rise/fall time tr during an abrupt change — 0,1 s to 3 s · CMC 3,0 % - phase shifting of dips and interruptions — 0° to 360° · CMC 3,0 % - zero-crossing control — -20° to 20° · CMC 3,0 %
RF electrical quantities
Internal procedure; At permanent location (S) IW17 based on PN-EN 61000-4-11:2020-11
11 Electromagnetic clamps Calibration — RF electrical quantities
9 kHz to 1 GHz; - characteristic impedance (S11) — 10 Ω to 600 Ω · CMC 2Ω 9 kHz to 1 GHz; - coupling factor (S21) — -30 dB to 10 dB · CMC 0,6 dB 9 kHz to 1 GHz; - decoupling factor — -30 dB to 0 dB · CMC 0,6 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW19 based on PN-EN 61000-4-6:2024-03
12 Power meters Calibration — RF electrical quantities
20 Hz to 40 GHz; - accuracy of power indication as a function of Frequency — -30 dB(mW) to 50 dB(mW) · CMC 0,5 dB 20 Hz to 40 GHz; - accuracy of power indication as a function of power level — -30 dB(mW) to 50 dB(mW) · CMC 1 dB 9 kHz to 6 GHz; - standing wave ratio — 1 to 5 · CMC 0,05
RF electrical quantities
Internal procedure; Direct method; At permanent location (S) IW9
13 Measuring receivers/radio disturbance meters Calibration — RF electrical quantities
- tuning frequency accuracy — 10 Hz to 40 GHz · CMC 10⁻³ % - RF voltage indication accuracy — 0 dB(μV) to 120 dB(μV) · CMC 0,5 dB - response to calibration pulses — -25 dB to 40 dB · CMC 1,0 dB - standing wave ratio — 1 to 5 · CMC 0,05 - intermodulation attenuation for band B — 10 dB to 50 dB · CMC 1,0 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW1 based on PN-EN 55016-1-1:2010 + A1:2011 + A2:2014 PN-EN 55016-1-1:2007 (for the RMS detector) PN-EN 55016-1-1:2019-07
14 Artificial networks Calibration — RF electrical quantities
9 kHz to 300 MHz; - impedance magnitude — 2 Ω to 200 Ω · CMC 3% 9 kHz to 300 MHz; - impedance angle — -15° to 90° · CMC 5° 9 kHz to 300 MHz; - voltage division factor — -10 dB to 20 dB · CMC 0,4 dB 9 kHz to 300 MHz; - decoupling factor — 10 dB to 110 dB · CMC 1,5 dB 9 kHz to 300 MHz; - crosstalk factor — 10 dB to 80 dB · CMC 1,5 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW4 based on PN-EN 55016-1-2:2014 + A1:2018-03
15 Voltage probes Calibration — RF electrical quantities
- insertion loss — 10 dB to 80 dB · CMC 0,5 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW7 based on PN-EN 55016-1-2:2014 + A1:2018-03
16 Current probes Calibration — RF electrical quantities
- transfer admittance — -40 dB(S) to 80 dB(S) · CMC 1,0 dB
RF electrical quantities
Internal procedure; At permanent location (S) IW8 based on PN-EN 55016-1-2:2014 + A1:2018-03 PN-EN 61000-4-6:2024-03
17 CDN networks Calibration — RF electrical quantities
- impedance magnitude — 0 Ω to 250 Ω · CMC 5% - phase angle — -40° to 40° · CMC 3°
RF electrical quantities
Internal procedure; At permanent location (S) IW10 based on PN-EN 55016-1-2:2014 + A1:2018-03 PN-EN 55020:2012 + A12:2016-04
18 Passive networks Calibration — RF electrical quantities
- attenuation; 20 Hz to 18 GHz — 0 dB to 80 dB · CMC 0,3 dB - standing wave ratio; 9 kHz to 18 GHz — 1 to 5 · CMC 0,05
RF electrical quantities
Internal procedure; Indirect method; At permanent location (S) IW14
19 Measurement antennas Calibration — Electromagnetic quantities
- magnetic antenna factor (calibration factor) — -50 dB(S/m) to 80 dB(S/m) · CMC 2,0 dB - electric antenna factor (calibration factor) — -20 dB/m to 140 dB/m · CMC 2,0 dB
Electromagnetic quantities
Internal procedure; At permanent location (S) and outside of it (P) IW6 based on PN-EN 55016-1-4:2019-04 + A1:2021-03 + A2:2024-01 ANSI-C63.5 SAE ARP 958 PN-EN 55016-1-6:2015-03 +A1:2017-07 +A2:2022-12

Reference standards (41)

ANSI-C63.5
ISO 10605:2023
ISO 7637-2:2011
ISO 7637-3:2016
IW1
IW10
IW11
IW12
IW13
IW14
IW15
IW16
IW17
IW18
IW19
IW2
IW3
IW4
IW5
IW6
IW7
IW8
IW9
PN-EN 55016-1-1:2007 (dla detektora RMS)
PN-EN 55016-1-1:2010 + A1:2011 + A2:2014
PN-EN 55016-1-1:2019-07
PN-EN 55016-1-2:2014 + A1:2018-03
PN-EN 55016-1-3:2007 + A1:2016-07 + A2:2020-10
PN-EN 55016-1-4:2019-04 + A1:2021-03 + A2:2024-01
PN-EN 55016-1-6:2015-03 +A1:2017-07 +A2:2022-12
PN-EN 55020:2012 + A12:2016-04
PN-EN 61000-4-11:2020-11
PN-EN 61000-4-12:2017-12
PN-EN 61000-4-18:2019-08
PN-EN 61000-4-2:2011
PN-EN 61000-4-4:2013
PN-EN 61000-4-5:2010
PN-EN 61000-4-5:2014 + A1:2018-01
PN-EN 61000-4-6:2024-03
PN-EN IEC 61000-4-2:2026-03
SAE ARP 958