🇵🇱 ZEC DIAGPOM Sp. z o.o. PL
Metrology & calibration
ZEC DIAGPOM Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 007 (Wrocław, Poland). Its accreditation scope covers 19 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Electronic pressure gauges, Spring pressure gauges, Analog current meters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Direct method; At permanent location (S), Internal procedure; Indirect method; At permanent location (S).
At a glance
- Accreditation number: PCA AP 007
- Calibration items in scope: 19
- Types of calibrated items: 20
- Calibration principles used: 8
- City: Wrocław, Poland
- Valid until: 18-11-2030
- Accreditation number
- AP 007
- Accreditation body
- PCA
- Address
- ul. Łowiecka 6/8, 50-220 Wrocław
- City
- Wrocław, Poland
- Division
- Measurement Laboratories Group
- Valid until
- 18-11-2030
What this laboratory calibrates (20)
- Multimeters — 5 calibration items
- Electronic pressure gauges — 4 calibration items
- Spring pressure gauges — 4 calibration items
- Analog current meters — 2 calibration items
- Analog voltage meters — 2 calibration items
- Digital current meters — 2 calibration items
- Digital voltage meters — 2 calibration items
- Pressure transducers — 2 calibration items
- Temperature indicators (meters), including temperature controllers — 2 calibration items
- Base metal thermoelectric sensors — 1 calibration item
- Dial thermometers — 1 calibration item
- Electric thermometers (including electronic) — 1 calibration item
- Furnaces — 1 calibration item
- Precious metal thermoelectric sensors — 1 calibration item
- Resistance thermometer sensors — 1 calibration item
- and 5 more
Calibration principles (8)
- Internal procedure; At permanent location (S) — 6 calibration items
- Internal procedure; Direct method; At permanent location (S) — 5 calibration items
- Internal procedure; Indirect method; At permanent location (S) — 2 calibration items
- Internal procedure; Outside of permanent location (P) — 2 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 1 calibration item
- Internal procedure; Indirect/direct method; At permanent location (S) — 1 calibration item
- Internal procedure; Indirect/direct method; At permanent location (S) and outside of it (P) — 1 calibration item
- Internal procedure; Indirect/direct method; Outside of permanent location (P) — 1 calibration item
Certificate information
- Issue date
- 19.10.2026
- Issue no.
- 22
- Certificate date
- 13.11.2020
- Accreditation cycle
- 19.10.2026 — 18.11.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (19)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Analog voltage meters Digital voltage meters Multimeters |
Calibration — DC voltage 0,2 mV to 100 mV · CMC 0,009 mV 100 mV to 1000 mV · CMC 6 · 10⁻⁵ · U + 0,005 mV 1 V to 1000 V · CMC 0,005 % DC voltage |
Internal procedure; Direct method; At permanent location (S) | PT1 PT3 based on EURAMET cg-15 v.3.0 | |
| 2 | Analog current meters Digital current meters Multimeters |
Calibration — DC current 0,2 mA to 100 mA · CMC 0,03 % 0,1 A to 1 A · CMC 0,04 % 1 A to 3 A · CMC 0,03 % 3 A to 20 A · CMC 0,25 % DC current |
Internal procedure; Direct method; At permanent location (S) | PT1 PT3 based on EURAMET cg-15 v.3.0 | |
| 3 | Analog voltage meters Digital voltage meters Multimeters |
Calibration — AC voltage 50 Hz — 1,0 mV to 100 mV · CMC 5 · 10⁻⁴ · U + 0,01 mV 50 Hz — 0,1 V to 10 V · CMC 0,06 % 50 Hz — 10 V to 1000 V · CMC 0,01 % AC voltage |
Internal procedure; Direct method; At permanent location (S) | PT1 PT3 based on EURAMET cg-15 v.3.0 | |
| 4 | Analog current meters Digital current meters Multimeters |
Calibration — AC current 50 Hz — 0,02 mA to 0,2 mA · CMC 0,4 % 50 Hz — 0,3 mA to 200 mA · CMC 0,15 % 50 Hz — 201 mA to 3 A · CMC 0,35 % 50 Hz — 3 A to 20 A · CMC 0,17 % AC current |
Internal procedure; Direct method; At permanent location (S) | PT1 PT3 based on EURAMET cg-15 v.3.0 | |
| 5 | Multimeters | Calibration — DC resistance 0,1 Ω to 100 Ω · CMC 0,6 % 100 Ω to 1 MΩ · CMC 0,06 % 1 MΩ to 19 MΩ · CMC 0,04 % DC resistance |
Internal procedure; Direct method; At permanent location (S) | PT3 based on EURAMET cg-15 v.3.0 | |
| 6 | Temperature indicators (meters), including temperature controllers Temperature indicators (meters) cooperating with thermoelectric sensors |
Calibration — Electric quantity simulation -180 °C to 0 °C · CMC 0,4 °C 0 °C to 1200 °C · CMC 0,3 °C 1200 °C to 1600 °C · CMC 0,6 °C Electric quantity simulation |
Internal procedure; Indirect/direct method; At permanent location (S) and outside of it (P) | PT5 based on EURAMET cg-11 v. 2.0 | |
| 7 | Temperature indicators (meters), including temperature controllers | Calibration — Electric quantity simulation -180°C to 250°C · CMC 0,1 °C 250°C to 350°C · CMC 0,2 °C 350°C to 750°C · CMC 0,3 °C Electric quantity simulation |
Internal procedure; Indirect/direct method; At permanent location (S) | PT5 based on EURAMET cg-11 v. 2.0 | |
| 8 | Temperature indicators (meters) cooperating with resistance sensors | Calibration — Electric quantity simulation -180°C to 0°C · CMC 0,13 °C 0°C to 250°C · CMC 0,3 °C 250°C to 750°C · CMC 0,7 °C Electric quantity simulation |
Internal procedure; Indirect/direct method; Outside of permanent location (P) | PT5 based on EURAMET cg-11 v. 2.0 | |
| 9 | Temperature transmitters (cooperating with thermoelectric sensors) | Calibration — Electric quantity simulation -25 °C do 0 °C · CMC 0,15 °C 0 °C do 200 °C · CMC 0,23 °C 200 °C to 400 °C · CMC 0,5 °C 400 °C do 600 °C · CMC 0,8 °C 600 °C do 1200 °C · CMC 1,2 °C Electric quantity simulation |
Internal procedure; Indirect method; At permanent location (S) | PT 16 | |
| 10 | Temperature transmitters (cooperating with resistance sensors) | Calibration — Electric quantity simulation -50 °C to 850 °C · CMC 0,14 °C Electric quantity simulation |
Internal procedure; Indirect method; At permanent location (S) | PT 16 | |
| 11 | Electronic pressure gauges Spring pressure gauges Pressure transducers |
Calibration — Pressure relative pressure (gas medium) — -100 kPa to -20 kPa · CMC 0,2 kPa relative pressure (gas medium) — -20 kPa to 20 kPa · CMC 0,02 kPa relative pressure (gas medium) — 20 kPa to 200 kPa · CMC 0,2 kPa relative pressure (gas medium) — 200 kPa to 2000 kPa · CMC 0,7 kPa relative pressure (gas medium) — 2000 kPa to 3500 kPa · CMC 5 kPa Pressure |
Internal procedure; At permanent location (S) | PT6, PT14, PT17 based on EURAMET cg-17 v. 4.1 | |
| 12 | Electronic pressure gauges Spring pressure gauges Pressure transducers |
Calibration — Pressure relative pressure (liquid medium) — 0 MPa to 10 MPa · CMC 7,8 · 10⁻⁴ · p relative pressure (liquid medium) — 10 MPa to 25 MPa · CMC 0,02 MPa relative pressure (liquid medium) — 25 MPa to 40 MPa · CMC 0,03 MPa Pressure |
Internal procedure; At permanent location (S) | PT6, PT14, PT17 based on EURAMET cg-17 v. 4.1 | |
| 13 | Electronic pressure gauges Spring pressure gauges |
Calibration — Pressure relative pressure (gas medium) — -100 kPa to 2000 kPa · CMC 1,3 kPa Pressure |
Internal procedure; Outside of permanent location (P) | PT6 based on EURAMET cg-17 v.4.1 | |
| 14 | Electronic pressure gauges Spring pressure gauges |
Calibration — Pressure relative pressure (liquid medium) — 0 kPa to 2000 kPa · CMC 1,3 kPa relative pressure (liquid medium) — 2 MPa to 40 MPa · CMC 0,03 MPa Pressure |
Internal procedure; Outside of permanent location (P) | PT6 based on EURAMET cg-17 v.4.1 | |
| 15 | Electric thermometers (including electronic) | Calibration — Temperature (electric thermometry) -100 °C to -40 °C · CMC 0,37 °C -40 °C to 155 °C · CMC 0,04 °C 155 °C to 650 °C · CMC 0,14 °C 650 °C to 1100 °C · CMC 0,9 °C Temperature (electric thermometry) |
Internal procedure; At permanent location (S) | PT18 | |
| 16 | Precious metal thermoelectric sensors Base metal thermoelectric sensors |
Calibration — Temperature (electric thermometry) 0 °C to 650 °C · CMC 0,7 °C 650 °C to 1100 °C · CMC 0,8 °C Temperature (electric thermometry) |
Internal procedure; At permanent location (S) | PT12 | |
| 17 | Resistance thermometer sensors | Calibration — Temperature (electric thermometry) -100 °C to -40 °C · CMC 0,37 °C - 40 °C to 155 °C · CMC 0,04 °C 155 °C to 650 °C · CMC 0,13 °C Temperature (electric thermometry) |
Internal procedure; At permanent location (S) | PT11 | |
| 18 | Thermostatic chambers Furnaces |
Calibration — Temperature (electric thermometry) -100 °C to -40 °C · CMC 0,5 °C1) -40 °C to 250 °C · CMC 0,4 °C1) 250 °C to 650 °C · CMC 1,0 °C1) 650 °C to 1100 °C · CMC 1,1 °C1) Temperature (electric thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | PT 19 | |
| 19 | Dial thermometers | Calibration — Temperature (non-electric thermometry) 0 °C to 160 °C · CMC 0,8 °C 160 °C to 250 °C · CMC 2,0 °C Temperature (non-electric thermometry) |
Internal procedure; At permanent location (S) | PT13 | |
Reference standards (16)
EURAMET cg-11 v. 2.0
EURAMET cg-15 v.3.0
EURAMET cg-17 v. 4.1
EURAMET cg-17 v.4.1
PT 16
PT 19
PT1
PT11
PT12
PT13
PT14
PT17
PT18
PT3
PT5
PT6