🇵🇱 „MAWI” Spółka z Ograniczoną Odpowiedzialnością PL
Electronics & EMCEnvironment
„MAWI” Spółka z Ograniczoną Odpowiedzialnością is a laboratory accredited by PCA under number AB 1555 (Rzezawa, Poland). Its accreditation scope covers 38 test methods in the following areas: Electronics & EMC, Environment. Test objects include: Medical electrical equipment, Electrical and electronic equipment, Lighting equipment, Electrical appliances for household and similar use.
At a glance
- Accreditation number: PCA AB 1555
- Methods in scope: 38
- Types of test objects: 8
- City: Rzezawa, Poland
- Accreditation number
- AB 1555
- Accreditation body
- PCA
- Address
- LABORATORIUM BEMC, ul. Przemysłowa 75; 32-765 Rzezawa
- City
- Rzezawa, Poland
What this laboratory tests (8)
- Medical electrical equipment — 12 methods
- Electrical and electronic equipment — 8 methods
- Lighting equipment — 7 methods
- Electrical appliances for household and similar use — 3 methods
- Electrical and electronic equipment, Lighting equipment — 2 methods
- Household appliances, electric tools and similar apparatus — 2 methods
- Industrial, scientific and medical equipment — 2 methods
- Measuring apparatus — 2 methods
Laboratory locations
- LABORATORIUM BEMC — ul. Przemysłowa 75; 32-765 Rzezawa
Accreditation scope — test methods (38)
| No. | Test object | Methodology / Parameters | Technique | Reference document | |
|---|---|---|---|---|---|
| BEMC Laboratory | |||||
| 1 | Industrial, scientific and medical equipment | Radiated disturbance emission Range 9 kHz - 6 GHz |
PN-EN 55011:2016-05 | PN-EN 55011:2016/A1:2017-06 | PN-EN 55011:2016/A11:2020-07 | PN-EN 55011:2016/A2:2021-08 | ||
| 2 | Household appliances, electric tools and similar apparatus | Radiated disturbance emission Range 9 kHz - 6 GHz |
PN-EN 55014-1:2017-06 | PN-EN IEC 55014-1:2021-08 | ||
| 3 | Measuring apparatus | Radiated disturbance emission Range 9 kHz - 6 GHz |
PN-EN 55016-2-3:2017-06 | PN-EN 55016-2-3:2017/ A1:2020-01 | ||
| 4 | Medical electrical equipment | Radiated disturbance emission Range 9 kHz - 6 GHz |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 5 | Lighting equipment | Radiated disturbance emission Range 9 kHz - 6 GHz |
PN-EN 55015:2013-10 | PN-EN 55015:2013/A1:2015-08 | PN-EN IEC 55015:2019-11 | PN-EN IEC 55015:2019/A11-2020-07 | ||
| 6 | Industrial, scientific and medical equipment | Conducted disturbance emission – disturbance voltage Range 9 kHz – 30 MHz |
PN-EN 55011:2016-05 | PN-EN 55011:2016-05/A1:2017-06 | PN-EN 55011:2016-05/A11:2020-07 | PN-EN 55011:2016-05/A2:2021-08 | ||
| 7 | Household appliances, electric tools and similar apparatus | Conducted disturbance emission – disturbance voltage Range 9 kHz – 30 MHz |
PN-EN 55014-1:2017-06 | PN-EN IEC 55014-1:2021-08 | ||
| 8 | Measuring apparatus | Conducted disturbance emission – disturbance voltage Range 9 kHz – 30 MHz |
PN-EN 55016-2-1:2014-09 | PN-EN 55016-2-1:2014/A1:2017-12 | PN-EN 55016-2-1:2014/AC:2020-11 | ||
| 9 | Medical electrical equipment | Conducted disturbance emission – disturbance voltage Range 9 kHz – 30 MHz |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 10 | Lighting equipment | Conducted disturbance emission – disturbance voltage Range 9 kHz – 30 MHz |
PN-EN 55015:2013-10 | PN-EN 55015:2013-10/A1:2015-08 | PN-EN IEC 55015:2019-11 | PN-EN IEC 55015:2019/A11:2020-07 | ||
| 11 | Electrical and electronic equipment | Voltage changes, voltage fluctuations and flicker in public low-voltage supply systems | PN-EN 61000-3-3:2013:10 | PN-EN 61000-3-3:2013/A1:2019-10 | PN-EN 61000-3-3:2013/A2:2022-04 | ||
| 12 | Electrical and electronic equipment | Immunity to discharges electrostatic Range (2 – 15) kV |
PN-EN 61000-4-2:2011 | ||
| 13 | Medical electrical equipment | Immunity to discharges electrostatic Range (2 – 15) kV |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 14 | Lighting equipment | Immunity to discharges electrostatic Range (2 – 15) kV |
PN-EN 61547:2009 | ||
| 15 | Electrical and electronic equipment | Immunity to radiated radio-frequency electromagnetic field Range (80 - 6000) MHz |
PN-EN 61000-4-3:2007 | PN-EN 61000-4-3:2007/A1:2008 | PN-EN 61000-4-3:2007/A2:2011 | PN-EN IEC 61000-4-3:2021-06 | PN-EN 55016-2-4:2005 | ||
| 16 | Medical electrical equipment | Immunity to radiated radio-frequency electromagnetic field Range (80 - 6000) MHz |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 17 | Lighting equipment | Immunity to radiated radio-frequency electromagnetic field Range (80 - 6000) MHz |
PN-EN 61547:2009 | ||
| 18 | Electrical and electronic equipment | Immunity to power-frequency magnetic field Range (0 - 30) A/m |
PN-EN 61000-4-8:2010 | ||
| 19 | Medical electrical equipment | Immunity to electrical fast transients/bursts Range (0,5 – 4 ) kV |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 20 | Electrical and electronic equipment | Immunity to electrical fast transients/bursts Range (0,5 – 4 ) kV |
PN-EN 61000-4-4:2013:05 | ||
| 21 | Lighting equipment | Immunity to electrical fast transients/bursts Range (0,5 – 4 ) kV |
PN-EN 61547:2009 | ||
| 22 | Electrical and electronic equipment | Immunity to surges Range (0,5 – 4) kV |
PN-EN 61000-4-5:2014-10 | PN-EN 61000-4-5:2014/A1:2018-01 | ||
| 23 | Medical electrical equipment | Immunity to surges Range (0,5 – 4) kV |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 24 | Lighting equipment | Immunity to surges Range (0,5 – 4) kV |
PN-EN 61547:2009 | ||
| 25 | Electrical and electronic equipment | Immunity to conducted disturbances induced by radio-frequency fields Range (0,15 - 230) MHz |
PN-EN 61000-4-6:2014-04 | PN-EN 61000-4-6:2024-03 | PN-EN 55016-2-4:2005 | ||
| 26 | Medical electrical equipment | Immunity to conducted disturbances induced by radio-frequency fields Range (0,15 - 230) MHz |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 27 | Electrical and electronic equipment | Immunity to voltage dips, short interruptions and voltage variations Range 0 %UT, 40 %UT, 70 %UT, 80 %UT |
PN-EN 61000-4-11:2007 | PN-EN IEC 61000-4-11:2020-11 PN-EN | IEC 61000-4-11:2020 /AC:2020-12 | ||
| 28 | Medical electrical equipment | Immunity to voltage dips, short interruptions and voltage variations Range 0 %UT, 40 %UT, 70 %UT, 80 %UT |
PN-EN 60601-1-2:2015-11 | PN-EN 60601-1-2:2015/A1:2021-07 | ||
| 29 | Lighting equipment | Immunity to voltage dips, short interruptions and voltage variations Range 0 %UT, 40 %UT, 70 %UT, 80 %UT |
PN-EN 61547:2009 | ||
| 30 | Electrical and electronic equipment, Lighting equipment | Current harmonics injected into the supply mains Range (2 - 40) harmonicznej |
PN-EN 61000-3-2:2014-10 | PN-EN 61000-3-2:2019-04 | PN-EN 61000-3-2:2019/A1:2021-08 | PN-EN 61000-3-2:2019/A2:2024-12 | PN-EN 61000-4-7:2007 | PN-EN 61000-4-7:2007/A1:2011 | ||
| 31 | Electrical and electronic equipment, Lighting equipment | Emission of conducted disturbances – disturbance power Range (30 - 300) MHz |
PN-EN 55014-1:2017-06 | PN-EN IEC 55014-1:2021-08 | PN-EN 55016-2-2:2011 | ||
| 32 | Medical electrical equipment | Leakage current, Touch current, Earth leakage current | PN-EN 60601-1:2011 | PN-EN 60601-1:2011/A1:2014-02 | PN-EN 60601-1:2011/A12:2014-12 p. 8.7 | ||
| 33 | Electrical appliances for household and similar use | Leakage current, Touch current, Earth leakage current | PN-EN 60335-1:2012 p. 13, p. 16 | PN-EN 60335-1:2024-04 p. 13, p. 16 | ||
| 34 | Medical electrical equipment | Electric strength | PN-EN 60601-1:2011 | PN-EN 60601-1:2011/A1:2014-02 | PN-EN 60601-1:2011A12:2014-12 p. 8.8 | ||
| 35 | Electrical appliances for household and similar use | Electric strength | PN-EN 60335-1:2012 p. 13, p. 16 | PN-EN 60335-1:2024-04 p. 13, p. 16 | ||
| 36 | Medical electrical equipment | Earthing resistance | PN-EN 60601-1:2011 | PN-EN 60601-1:2011/A1:2014-02 | PN-EN 60601-1:2011/A12:2014-12 p. 8.6 | ||
| 37 | Electrical appliances for household and similar use | Earthing resistance | PN-EN 60335-1:2012 p. 27 | PN-EN 60335-1:2024-04 p. 27 | ||
| 38 | Medical electrical equipment | Insulation resistance | PN-EN 60601-1:2011 | PN-EN 60601-1:2011/A1:2014-02 | PN-EN 60601-1:2011/A12:2014-12 p. 8.8 | ||
Reference standards (58)
IEC 61000-4-11:2020 /AC:2020-12
PN-EN 55011:2016-05
PN-EN 55011:2016-05/A11:2020-07
PN-EN 55011:2016-05/A1:2017-06
PN-EN 55011:2016-05/A2:2021-08
PN-EN 55011:2016/A11:2020-07
PN-EN 55011:2016/A1:2017-06
PN-EN 55011:2016/A2:2021-08
PN-EN 55014-1:2017-06
PN-EN 55015:2013-10
PN-EN 55015:2013-10/A1:2015-08
PN-EN 55015:2013/A1:2015-08
PN-EN 55016-2-1:2014-09
PN-EN 55016-2-1:2014/A1:2017-12
PN-EN 55016-2-1:2014/AC:2020-11
PN-EN 55016-2-2:2011
PN-EN 55016-2-3:2017-06
PN-EN 55016-2-3:2017/ A1:2020-01
PN-EN 55016-2-4:2005
PN-EN 60335-1:2012 p. 13, p. 16
PN-EN 60335-1:2012 p. 27
PN-EN 60335-1:2024-04 p. 13, p. 16
PN-EN 60335-1:2024-04 p. 27
PN-EN 60601-1-2:2015-11
PN-EN 60601-1-2:2015/A1:2021-07
PN-EN 60601-1:2011
PN-EN 60601-1:2011/A12:2014-12 p. 8.6
PN-EN 60601-1:2011/A12:2014-12 p. 8.7
PN-EN 60601-1:2011/A12:2014-12 p. 8.8
PN-EN 60601-1:2011/A1:2014-02
PN-EN 60601-1:2011A12:2014-12 p. 8.8
PN-EN 61000-3-2:2014-10
PN-EN 61000-3-2:2019-04
PN-EN 61000-3-2:2019/A1:2021-08
PN-EN 61000-3-2:2019/A2:2024-12
PN-EN 61000-3-3:2013/A1:2019-10
PN-EN 61000-3-3:2013/A2:2022-04
PN-EN 61000-3-3:2013:10
PN-EN 61000-4-11:2007
PN-EN 61000-4-2:2011
PN-EN 61000-4-3:2007
PN-EN 61000-4-3:2007/A1:2008
PN-EN 61000-4-3:2007/A2:2011
PN-EN 61000-4-4:2013:05
PN-EN 61000-4-5:2014-10
PN-EN 61000-4-5:2014/A1:2018-01
PN-EN 61000-4-6:2014-04
PN-EN 61000-4-6:2024-03
PN-EN 61000-4-7:2007
PN-EN 61000-4-7:2007/A1:2011
PN-EN 61000-4-8:2010
PN-EN 61547:2009
PN-EN IEC 55014-1:2021-08
PN-EN IEC 55015:2019-11
PN-EN IEC 55015:2019/A11-2020-07
PN-EN IEC 55015:2019/A11:2020-07
PN-EN IEC 61000-4-11:2020-11 PN-EN
PN-EN IEC 61000-4-3:2021-06