🇨🇿 PG electronic s.r.o. PL

ČIA 2417 • Chomutov, Czech Republic • 12 calibration items

Metrology & calibration

PG electronic s.r.o. is a calibration laboratory accredited by ČIA under number 2417 (Chomutov, Czech Republic). Its accreditation scope covers 12 calibration items in the following areas: Metrology & calibration. Calibrated items include: AC current / AC current sources, multifunction testers, DC current / DC current sources, multifunction testers, DC resistance / multifunction testers, DC resistance / resistors. Most frequently used calibration principles: Direct measurement by a standard multimeter, Direct measurement of resistance standards, Direct measurement by a reference milliohmmeter.

At a glance

  • Accreditation number: ČIA 2417
  • Calibration items in scope: 12
  • Types of calibrated items: 8
  • Calibration principles used: 5
  • City: Chomutov, Czech Republic
  • Valid until: 16-10-2030
Accreditation number
2417
Accreditation body
ČIA
Address
Růžová 5363, 430 04 Chomutov
City
Chomutov, Czech Republic
Division
Laboratorium wzorcujące
Valid until
16-10-2030

What this laboratory calibrates (8)

Calibration principles (5)

Certificate information

Certificate date
16.10.2025

Accreditation scope — calibration items (12)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 DC voltage / DC voltage sources, multifunction testers Calibration – electrical quantities
0 mV to 100 mV · CMC 0,003 % + 3 µV 100 mV to 1 V · CMC 0,002 % + 6 µV 1 V to 10 V · CMC 0,002 % + 40 µV 10 V to 100 V · CMC 0,003 % + 0,5 mV 100 V to 1 kV · CMC 0,004 % + 8 mV 1 kV to 6 kV · CMC 0,2 % + 70 mV
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter KA-01
2 AC voltage / AC voltage sources, multifunction testers Calibration – electrical quantities
20 Hz to 1 kHz — 5 mV to 100 mV · CMC 0,05 % + 40 µV 20 Hz to 1 kHz — 100 mV to 1 V · CMC 0,05 % + 0,3 mV 20 Hz to 1 kHz — 1 V to 10 V · CMC 0,05 % + 3 mV 20 Hz to 1 kHz — 10 V to 100 V · CMC 0,05 % + 30 mV 20 Hz to 1 kHz — 100 V to 1 kV · CMC 0,05 % + 0,2 V 20 Hz to 1 kHz — 1 kV to 5 kV · CMC 0,3 % + 4 V
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter KA-02
3 DC current / DC current sources, multifunction testers Calibration – electrical quantities
1 µA to 100 µA · CMC 0,04 % + 0,03 µA 100 µA to 1 mA · CMC 0,04 % + 0,04 µA 1 mA to 10 mA · CMC 0,04 % + 2 µA 10 mA to 100 mA · CMC 0,04 % + 4 µA 100 mA to 400 mA · CMC 0,04 % + 20 µA 400 mA to 1 A · CMC 0,04 % + 0,2 mA 1 A to 3 A · CMC 0,09 % + 0,6 mA 3 A to 10 A · CMC 0,2 % + 0,7 mA
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter KA-03
4 DC current / DC current sources, multifunction testers Calibration – electrical quantities
10 A to 30 A · CMC 0,35 %
CMC for the field of measured quantity: Electrical quantities
Measurement by a reference multimeter with a shunt KA-03
5 AC current / AC current sources, multifunction testers Calibration – electrical quantities
20 Hz to 1 kHz — 10 µA to 100 µA · CMC 0,2 µA 20 Hz to 1 kHz — 100 µA to 1 mA · CMC 0,08 % + 0,4 µA 20 Hz to 1 kHz — 1 mA to 10 mA · CMC 0,2 % + 5 µA 20 Hz to 1 kHz — 10 mA to 100 mA · CMC 0,07 % + 40 µA 20 Hz to 1 kHz — 100 mA to 400 mA · CMC 0,2 % + 0,4 mA 20 Hz to 1 kHz — 400 mA to 1 A · CMC 0,07 % + 0,7 mA 20 Hz to 1 kHz — 1 A to 3 A · CMC 0,09 % + 3 mA 20 Hz to 1 kHz — 3 A to 10 A · CMC 0,2 % + 7 mA
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter KA-04
6 AC current / AC current sources, multifunction testers Calibration – electrical quantities
50 Hz to 60 Hz — 10 A to 30 A · CMC 0,31 % + 18 mA 50 Hz to 60 Hz — 30 A to 100 A · CMC 1,3 % + 24 mA
CMC for the field of measured quantity: Electrical quantities
Measurement by a reference multimeter with current clamps KA-04
7 DC resistance / resistors Calibration – electrical quantities
1 mΩ to 50 mΩ · CMC 0,2 % + 20 µΩ 50 mΩ to 500 mΩ · CMC 0,06 % + 0,2 mΩ 500 mΩ to 3 Ω · CMC 0,06 % + 2 mΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a reference milliohmmeter KA-05
8 DC resistance / resistors Calibration – electrical quantities
3 Ω to 10 Ω · CMC 0,008 % + 3 mΩ 10 Ω to 100 Ω · CMC 0,008 % + 4 mΩ 100 Ω to 1 kΩ · CMC 0,008 % + 8 mΩ 1 kΩ to 10 kΩ · CMC 0,008 % + 0,08 Ω 10 kΩ to 100 kΩ · CMC 0,008 % + 0,8 Ω 100 kΩ to 1 MΩ · CMC 0,008 % + 8 Ω 1 MΩ to 10 MΩ · CMC 0,04 % + 80 Ω 10 MΩ to 100 MΩ · CMC 0,6 % + 7 kΩ 100 MΩ to 1 GΩ · CMC 2 % + 80 kΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a standard multimeter KA-05
9 Protective conductor resistance / multifunction testers Calibration – electrical quantities
50 mΩ · CMC 1,2 mΩ 100 mΩ · CMC 1,2 mΩ 200 mΩ · CMC 1,5 mΩ 300 mΩ · CMC 2 mΩ 400 mΩ · CMC 2 mΩ 1 Ω · CMC 5 mΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement of resistance standards KA-07
10 Insulation resistance / insulation resistance meters, multifunction testers Calibration – electrical quantities
0,5 MΩ · CMC 6 kΩ 1 MΩ · CMC 12 kΩ 2 MΩ · CMC 24 kΩ 3 MΩ · CMC 36 kΩ 5 MΩ · CMC 59 kΩ 9 MΩ · CMC 110 kΩ 10 MΩ · CMC 120 kΩ 20 MΩ · CMC 240 kΩ 30 MΩ · CMC 360 kΩ 45 MΩ · CMC 0,53 MΩ 50 MΩ · CMC 0,63 MΩ 90 MΩ · CMC 1,1 MΩ 220 MΩ · CMC 3,8 MΩ 450 MΩ · CMC 8,4 MΩ 900 MΩ · CMC 11 MΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement of resistance standards KA-06
11 DC resistance / multifunction testers Calibration – electrical quantities
1 Ω · CMC 55 mΩ 5 Ω · CMC 75 mΩ 10 Ω · CMC 150 mΩ 50 Ω · CMC 0,75 Ω 100 Ω · CMC 1,5 Ω 450 Ω · CMC 6,8 Ω 900 Ω · CMC 13,5 Ω
CMC for the field of measured quantity: Electrical quantities
Direct measurement of resistance standards KA-06
12 DC resistance / multifunction testers Calibration – electrical quantities
1 mΩ · CMC 5,8 µΩ 10 mΩ · CMC 58 µΩ 20 mΩ · CMC 9,3 µΩ 50 mΩ · CMC 22 µΩ 200 mΩ · CMC 84 µΩ 500 mΩ · CMC 0,21 mΩ 2 Ω · CMC 0,77 mΩ 5 Ω · CMC 1,9 mΩ 20 Ω · CMC 7,7 mΩ 50 Ω · CMC 19 mΩ 200 Ω · CMC 78 mΩ 500 Ω · CMC 0,19 Ω 2 kΩ · CMC 0,79 Ω 5 kΩ · CMC 1,9 Ω 20 kΩ · CMC 7,9 Ω 50 kΩ · CMC 19 Ω 200 kΩ · CMC 80 Ω 500 kΩ · CMC 0,19 kΩ 2 MΩ · CMC 0,79 kΩ 5 MΩ · CMC 2,5 kΩ
CMC for the field of measured quantity: Electrical quantities
Direct measurement of resistance standards KA-05

Reference standards (7)

KA-01
KA-02
KA-03
KA-04
KA-05
KA-06
KA-07

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2417