🇨🇿 METROTEST s.r.o. PL

ČIA 2414 • Kladno, Czech Republic • 7 calibration items

Metrology & calibration

METROTEST s.r.o. is a calibration laboratory accredited by ČIA under number 2414 (Kladno, Czech Republic). Its accreditation scope covers 7 calibration items in the following areas: Metrology & calibration. Calibrated items include: Hardness – Hardness testers, Calibration of coordinate measuring machines, Calibration of measuring arms, Tensile testing machines. Most frequently used calibration principles: Comparison with a reference force-measuring instrument, ČSN EN ISO 6506-2, ČSN EN ISO 6507-2.

At a glance

  • Accreditation number: ČIA 2414
  • Calibration items in scope: 7
  • Types of calibrated items: 5
  • Calibration principles used: 6
  • City: Kladno, Czech Republic
  • Valid until: 05-06-2029
Accreditation number
2414
Accreditation body
ČIA
Address
T.G.Masaryka 2104, 272 01 Kladno
City
Kladno, Czech Republic
Division
Laboratorium wzorcujące METROTEST
Valid until
05-06-2029

What this laboratory calibrates (5)

Calibration principles (6)

Certificate information

Certificate date
30.09.2025

Accreditation scope — calibration items (7)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Calibration of measuring arms Calibration – length
0 m to 4,5 m · CMC (1,5L + 2) µm
CMC for the field of measured quantity: Length
With a cone standard and a calibration ball KP01 (ASME B89.4.22)
2 Calibration of coordinate measuring machines Calibration – length
0 m to 3 m · CMC (1L + 0,25) µm
CMC for the field of measured quantity: Length
With gauge blocks and a calibration ball KP06 (ČSN EN ISO 10360-2)
3 Tensile testing machines Calibration – force, mechanical tests
tension — 0,04 N to 50 kN · CMC 0,6 %
CMC for the field of measured quantity: Force, mechanical tests
Comparison with a reference force-measuring instrument KP02 (ČSN EN ISO 7501-2; ČSN EN ISO 376; ČSN EN ISO 7500-2; ASTM E4)
4 Testing presses Calibration – force, mechanical tests
compression — 0,04 N to 50 kN · CMC 0,6 %
CMC for the field of measured quantity: Force, mechanical tests
Comparison with a reference force-measuring instrument KP03 (ČSN EN ISO 7501-2; ČSN EN ISO 376; ČSN EN ISO 7500-2; ASTM E4)
5 Hardness – Hardness testers Calibration – force, mechanical tests
Rockwell C — 10 HRC to 80 HRC · CMC 0,5 HRC
CMC for the field of measured quantity: Force, mechanical tests
ČSN EN ISO 6508-2 KP04
6 Hardness – Hardness testers Calibration – force, mechanical tests
Vickers — 20 HV to 3000 HV · CMC 0,5 %
CMC for the field of measured quantity: Force, mechanical tests
ČSN EN ISO 6507-2 KP04
7 Hardness – Hardness testers Calibration – force, mechanical tests
Brinell — 20 HBW to 650 HBW · CMC 0,5 %
CMC for the field of measured quantity: Force, mechanical tests
ČSN EN ISO 6506-2 KP04

Reference standards (11)

ASME B89.4.22
ASTM E4
KP01
KP02
KP03
KP04
KP06
ČSN EN ISO 10360-2
ČSN EN ISO 376
ČSN EN ISO 7500-2
ČSN EN ISO 7501-2

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2414