🇨🇿 Pro Metrology s.r.o. PL
Metrology & calibration
Pro Metrology s.r.o. is a calibration laboratory accredited by ČIA under number 2406 (Rostoklaty, Czech Republic). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electronic thermometers, temperature recorders, Glass thermometers, Infrared non-contact thermometers, Piston volume meters. Most frequently used calibration principles: Direct comparison with a reference digital thermometer, Direct comparison with a reference black body, Direct comparison with a standard stopwatch.
At a glance
- Accreditation number: ČIA 2406
- Calibration items in scope: 5
- Types of calibrated items: 5
- Calibration principles used: 4
- City: Rostoklaty, Czech Republic
- Valid until: 27-05-2027
- Accreditation number
- 2406
- Accreditation body
- ČIA
- Address
- Rostoklaty 30, 287 71 Rostoklaty
- City
- Rostoklaty, Czech Republic
- Division
- PROCALIBRATION – Laboratorium wzorcujące
- Valid until
- 27-05-2027
What this laboratory calibrates (5)
- Electronic thermometers, temperature recorders — 1 calibration item
- Glass thermometers — 1 calibration item
- Infrared non-contact thermometers — 1 calibration item
- Piston volume meters — 1 calibration item
- Time / Digital and mechanical stopwatch — 1 calibration item
Calibration principles (4)
- Direct comparison with a reference digital thermometer — 2 calibration items
- Direct comparison with a reference black body — 1 calibration item
- Direct comparison with a standard stopwatch — 1 calibration item
- Gravimetric method — 1 calibration item
Certificate information
- Certificate date
- 13.05.2026
Accreditation scope — calibration items (5)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Piston volume meters | Calibration – volume Distilled water — 0,1 µl to 10000 µl · CMC 0,12 % + 0,03 µl Distilled water — 10000 µl to 200000 µl · CMC 0,08 % CMC for the field of measured quantity: Volume |
Gravimetric method | PROC_30_000 Objem GM (ČSN EN ISO 8655-6; EURAMET Calibration Guide No. 19, Version 4.1 (12/2025)) | |
| 2 | Electronic thermometers, temperature recorders | Calibration – temperature -30 °C to 0 °C · CMC 0,2 °C 0 °C to 30 °C · CMC 0,15 °C 30 °C to 80 °C · CMC 0,07 °C 80 °C to 130 °C · CMC 0,2 °C 130 °C to 150 °C · CMC 0,3 °C CMC for the field of measured quantity: Temperature |
Direct comparison with a reference digital thermometer | PROC_30_100_Teplota_EL | |
| 3 | Glass thermometers | Calibration – temperature 30 °C to 80 °C · CMC 0,07 °C CMC for the field of measured quantity: Temperature |
Direct comparison with a reference digital thermometer | PROC_30_110_Teplota_SKL | |
| 4 | Infrared non-contact thermometers | Calibration – temperature 20 °C to 37 °C · CMC 1,1 °C 37 °C to 50 °C · CMC 1,3 °C CMC for the field of measured quantity: Temperature |
Direct comparison with a reference black body | PROC_30_300_Teplota_INF | |
| 5 | Time / Digital and mechanical stopwatch | Calibration – time and frequency quantities Scale division 0,1 s — 10 s to 86 400 s · CMC 0,059 s + 7,99·10⁻⁶·ΔT CMC for the field of measured quantity: Time and frequency quantities |
Direct comparison with a standard stopwatch | PROC-30-050 Kalibrační postup ČAS (NIST 960-12, KP 6.1.2/02/14) | |
Reference standards (9)
EURAMET Calibration Guide No. 19, Version 4.1 (12/2025)
KP 6.1.2/02/14
NIST 960-12
PROC-30-050 Kalibrační postup ČAS
PROC_30_000 Objem GM
PROC_30_100_Teplota_EL
PROC_30_110_Teplota_SKL
PROC_30_300_Teplota_INF
ČSN EN ISO 8655-6