🇨🇿 Pro Metrology s.r.o. PL

ČIA 2406 • Rostoklaty, Czech Republic • 5 calibration items

Metrology & calibration

Pro Metrology s.r.o. is a calibration laboratory accredited by ČIA under number 2406 (Rostoklaty, Czech Republic). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electronic thermometers, temperature recorders, Glass thermometers, Infrared non-contact thermometers, Piston volume meters. Most frequently used calibration principles: Direct comparison with a reference digital thermometer, Direct comparison with a reference black body, Direct comparison with a standard stopwatch.

At a glance

  • Accreditation number: ČIA 2406
  • Calibration items in scope: 5
  • Types of calibrated items: 5
  • Calibration principles used: 4
  • City: Rostoklaty, Czech Republic
  • Valid until: 27-05-2027
Accreditation number
2406
Accreditation body
ČIA
Address
Rostoklaty 30, 287 71 Rostoklaty
City
Rostoklaty, Czech Republic
Division
PROCALIBRATION – Laboratorium wzorcujące
Valid until
27-05-2027

What this laboratory calibrates (5)

Calibration principles (4)

Certificate information

Certificate date
13.05.2026

Accreditation scope — calibration items (5)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Piston volume meters Calibration – volume
Distilled water — 0,1 µl to 10000 µl · CMC 0,12 % + 0,03 µl Distilled water — 10000 µl to 200000 µl · CMC 0,08 %
CMC for the field of measured quantity: Volume
Gravimetric method PROC_30_000 Objem GM (ČSN EN ISO 8655-6; EURAMET Calibration Guide No. 19, Version 4.1 (12/2025))
2 Electronic thermometers, temperature recorders Calibration – temperature
-30 °C to 0 °C · CMC 0,2 °C 0 °C to 30 °C · CMC 0,15 °C 30 °C to 80 °C · CMC 0,07 °C 80 °C to 130 °C · CMC 0,2 °C 130 °C to 150 °C · CMC 0,3 °C
CMC for the field of measured quantity: Temperature
Direct comparison with a reference digital thermometer PROC_30_100_Teplota_EL
3 Glass thermometers Calibration – temperature
30 °C to 80 °C · CMC 0,07 °C
CMC for the field of measured quantity: Temperature
Direct comparison with a reference digital thermometer PROC_30_110_Teplota_SKL
4 Infrared non-contact thermometers Calibration – temperature
20 °C to 37 °C · CMC 1,1 °C 37 °C to 50 °C · CMC 1,3 °C
CMC for the field of measured quantity: Temperature
Direct comparison with a reference black body PROC_30_300_Teplota_INF
5 Time / Digital and mechanical stopwatch Calibration – time and frequency quantities
Scale division 0,1 s — 10 s to 86 400 s · CMC 0,059 s + 7,99·10⁻⁶·ΔT
CMC for the field of measured quantity: Time and frequency quantities
Direct comparison with a standard stopwatch PROC-30-050 Kalibrační postup ČAS (NIST 960-12, KP 6.1.2/02/14)

Reference standards (9)

EURAMET Calibration Guide No. 19, Version 4.1 (12/2025)
KP 6.1.2/02/14
NIST 960-12
PROC-30-050 Kalibrační postup ČAS
PROC_30_000 Objem GM
PROC_30_100_Teplota_EL
PROC_30_110_Teplota_SKL
PROC_30_300_Teplota_INF
ČSN EN ISO 8655-6

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2406