ČIA 2405 • Mladá Boleslav, Czech Republic • 17 calibration items

Metrology & calibration

TÜV SÜD Czech s.r.o. is a calibration laboratory accredited by ČIA under number 2405 (Mladá Boleslav, Czech Republic). Its accreditation scope covers 17 calibration items in the following areas: Metrology & calibration. Calibrated items include: Mechanical and electromechanical manometers, Acceleration of linear mechanical vibrations of harmonic wave form…, Angular velocity of the vibration-rotation harmonic waveform / Angu…, Belt motion sensors. Most frequently used calibration principles: comparison with a standard acceleration sensor, comparison with a standard angular velocity sensor, direct measurement.

At a glance

  • Accreditation number: ČIA 2405
  • Calibration items in scope: 17
  • Types of calibrated items: 16
  • Calibration principles used: 14
  • City: Mladá Boleslav, Czech Republic
  • Valid until: 05-02-2029
Accreditation number
2405
Accreditation body
ČIA
Address
UNO TECHNOLOGY PARK Bezděčín, Hala H1a, č.p. 108, 293 01 Mladá Boleslav
City
Mladá Boleslav, Czech Republic
Division
Laboratorium wzorcujące
Valid until
05-02-2029

What this laboratory calibrates (16)

Calibration principles (14)

Certificate information

Certificate date
02.07.2025

Accreditation scope — calibration items (17)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Sensors of distance and position Calibration – length
0 mm to 300 mm · CMC 0,04 mm 300 mm to 1500 mm · CMC 0,30 mm
CMC for the field of measured quantity: Length
comparison with a height gauge I540 – 068 – 10
2 Belt motion sensors Calibration – length
-400 mm to 400 mm · CMC (0,2·L + 0,40) mm
CMC for the field of measured quantity: Length
comparison with a calibration jig I540 – 068 – 10
3 Sensors of position and inclinometers Calibration – plane angle
0 ° to 360 ° · CMC 0,02°
CMC for the field of measured quantity: Plane angle
comparison with a digital inclinometer or rotary index table I540 – 068 – 10
4 Meters and sensors of half-sine wave mechanical shock acceleration Calibration – mechanical motion
100 m·s⁻² to 1471 m·s⁻² · CMC 1,0 % 1471 m·s⁻² to 2000 m·s⁻² · CMC 1,5 % 2000 m·s⁻² to 40000 m·s⁻² · CMC 1,8 %
CMC for the field of measured quantity: Mechanical motion
comparison with a standard acceleration sensor I 540 – 068 – 43
5 Acceleration of linear mechanical vibrations of harmonic wave form / Vibration sensors, vibrometers, vibration calibrators, vibration generators Calibration – mechanical motion
5 Hz to 10 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 2,0 % 10 Hz to 20 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 1,0 % 20 Hz to 80 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 0,75 % 80 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 0,5 % 80 Hz to 1000 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 0,75 % 1000 Hz to 5000 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 1,0 % 5000 Hz to 10000 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 2,0 %
CMC for the field of measured quantity: Mechanical motion
comparison with a standard acceleration sensor I 540 – 068 – 45
6 Sensitivity of vibration sensors and vibrometers Calibration – mechanical motion
5 Hz to 10 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 2,0 % 10 Hz to 20 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 1,0 % 20 Hz to 80 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 0,75 % 80 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 0,5 % 80 Hz to 1000 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 0,75 % 1000 Hz to 5000 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 1,0 % 5000 Hz to 10000 Hz — 0,01 mV/(m·s⁻²) to 10000 mV/(m·s⁻²) · CMC 2,0 % 5 Hz to 10 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 2,0 % 10 Hz to 20 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 1,0 % 20 Hz to 80 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 0,75 % 80 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 0,5 % 80 Hz to 1000 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 0,75 % 1000 Hz to 5000 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 1,0 % 5000 Hz to 10000 Hz — 0,01 pC/(m·s⁻²) to 1000 pC/(m·s⁻²) · CMC 2,0 % 5 Hz to 10 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 2,0 % 10 Hz to 20 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 1,0 % 20 Hz to 80 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 0,75 % 80 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 0,5 % 80 Hz to 1000 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 0,75 % 1000 Hz to 5000 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 1,0 % 5000 Hz to 10000 Hz — 0,01 mV/(m·s⁻¹) to 10000 mV/(m·s⁻¹) · CMC 2,0 % 5 Hz to 10 Hz — 0,01 mV/m to 10000 mV/m · CMC 2,0 % 10 Hz to 20 Hz — 0,01 mV/m to 10000 mV/m · CMC 1,0 % 20 Hz to 80 Hz — 0,01 mV/m to 10000 mV/m · CMC 0,75 % 80 Hz — 0,01 mV/m to 10000 mV/m · CMC 0,5 % 80 Hz to 1000 Hz — 0,01 mV/m to 10000 mV/m · CMC 0,75 % 1000 Hz to 5000 Hz — 0,01 mV/m to 10000 mV/m · CMC 1,0 % 5000 Hz to 10000 Hz — 0,01 mV/m to 10000 mV/m · CMC 2,0 %
CMC for the field of measured quantity: Mechanical motion
comparison with a vibration standard I 540 – 068 – 45
7 Transmission of amplifiers, filters and vibrometers Calibration – mechanical motion
0,2 Hz to 1 Hz — 0,001 mV/pC to 10000 mV/pC · CMC 0,5 % 1 Hz to 5000 Hz — 0,001 mV/pC to 10000 mV/pC · CMC 0,4 % 5000 Hz to 10000 Hz — 0,001 mV/pC to 10000 mV/pC · CMC 0,4 % 10000 Hz to 20000 Hz — 0,001 mV/pC to 10000 mV/pC · CMC 0,6 % 20000 Hz to 50000 Hz — 0,001 mV/pC to 10000 mV/pC · CMC 1,0 % 0,2 Hz to 1 Hz — 0,001 V/V to 1000 V/V · CMC 0,4 % 1 Hz to 20000 Hz — 0,001 V/V to 1000 V/V · CMC 0,3 % up to 50000 Hz — 0,001 V/V to 1000 V/V · CMC 1,0 %
CMC for the field of measured quantity: Mechanical motion
direct measurement I 540 – 068 – 45
8 Vibration measurement / Vibration generators Calibration – mechanical motion
5 Hz to 10 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 2,0 % 10 Hz to 20 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 1,0 % 20 Hz to 80 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 0,75 % 80 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 0,5 % 80 Hz to 1000 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 0,75 % 1000 Hz to 5000 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 1,0 % 5000 Hz to 10000 Hz — 0,1 m·s⁻² to 295 m·s⁻² · CMC 2,0 %
CMC for the field of measured quantity: Mechanical motion
direct measurement I 540 – 068 – 45
9 Speedometers, GPS and radar speedometers, speed sensors Calibration – mechanical motion
distance standard — 5 km·h⁻¹ to 180 km·h⁻¹ · CMC 0,02 % measured distance — 20 km·h⁻¹ to 130 km·h⁻¹ · CMC 0,04 %
CMC for the field of measured quantity: Mechanical motion
comparison with distance and time standard I 540 – 068 – 29
10 Speedometers with a rolling wheel Calibration – mechanical motion
2 m·min⁻¹ to 100 m·min⁻¹ · CMC 0,1 % + 0,01 m·min⁻¹
CMC for the field of measured quantity: Mechanical motion
direct generation of circumferential velocity I 540 – 068 – 05
11 Revolution counters, rpm sensors, stroboscopes Calibration – mechanical motion
500 min⁻¹ to 10 000 min⁻¹ · CMC 0,006 min⁻¹ 10 000 min⁻¹ to 100 000 min⁻¹ · CMC 0,06 min⁻¹
CMC for the field of measured quantity: Mechanical motion
direct measurement of an optical or electrical signal I 540 – 068 – 05
12 Angular velocity of the vibration-rotation harmonic waveform / Angular velocity sensors Calibration – mechanical motion
1 Hz to 200 Hz — 0,01 rad·s⁻¹ to 87,3 rad·s⁻¹ · CMC 1,5 %
CMC for the field of measured quantity: Mechanical motion
comparison with a standard angular velocity sensor I 540 – 068 – 45
13 Sensitivity of angular velocity sensors Calibration – mechanical motion
1 Hz to 200 Hz — 0,001 mV/(rad·s⁻¹) to 10000 mV/(rad·s⁻¹) · CMC 1,5 %
CMC for the field of measured quantity: Mechanical motion
comparison with a standard angular velocity sensor I 540 – 068 – 45
14 Mechanical and electromechanical manometers Calibration – pressure, mechanical stress
gas — -95 kPa to 0 kPa · CMC 0,1 kPa gas / liquid — 0 MPa to 0,7 MPa · CMC 0,03 % + 0,08 kPa gas / liquid — 0,7 MPa to 3,5 MPa · CMC 0,03 % + 0,4 kPa gas / liquid — 3,5 MPa to 7 MPa · CMC 0,03 % + 0,8 kPa gas / liquid — 7 MPa to 20 MPa · CMC 0,03 % + 2,3 kPa
CMC for the field of measured quantity: Pressure, mechanical stress
comparison with a digital pressure gauge I540 – 068 – 3
15 Mechanical and electromechanical manometers Calibration – pressure, mechanical stress
oil — 0 MPa to 6 MPa · CMC 3,5 kPa oil — 6 MPa to 60 MPa · CMC 0,06 %
CMC for the field of measured quantity: Pressure, mechanical stress
comparison with a piston pressure gauge I540 – 068 – 3
16 Frequency measurement, frequency generators, digital speed cameras, synchronization boxes, time bases for measuring instruments and systems, counters and frequency meters Calibration – time, frequency
1 Hz to 1 MHz · CMC 2,3·10⁻⁷ Hz
CMC for the field of measured quantity: Time, frequency
direct measurement with a standard counter I 540 – 068 – 44
17 Time interval / stopwatches, timers and other chronometers Calibration – time, frequency
1 ms to 1000 ms · CMC 1,0·10⁻⁵ s 1 s to 10800 s · CMC 6,0·10⁻⁴ s
CMC for the field of measured quantity: Time, frequency
comparison with a standard counter I 540 – 068 – 02

Reference standards (8)

I 540 – 068 – 02
I 540 – 068 – 05
I 540 – 068 – 29
I 540 – 068 – 43
I 540 – 068 – 44
I 540 – 068 – 45
I540 – 068 – 10
I540 – 068 – 3