🇨🇿 MV Lab, s.r.o. PL

ČIA 2400 • Králův Dvůr, Czech Republic • 5 calibration items

Metrology & calibration

MV Lab, s.r.o. is a calibration laboratory accredited by ČIA under number 2400 (Králův Dvůr, Czech Republic). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Indicating thermometers and temperature measuring chains, Electronic hygrometers and relative humidity measuring chains. Most frequently used calibration principles: Comparison with a standard dew point hygrometer in a humidity gener…, Comparison with a standard digital thermometer around the boiling p…, Comparison with a standard digital thermometer in a calibration oven.

At a glance

  • Accreditation number: ČIA 2400
  • Calibration items in scope: 5
  • Types of calibrated items: 2
  • Calibration principles used: 5
  • City: Králův Dvůr, Czech Republic
  • Valid until: 05-06-2031
Accreditation number
2400
Accreditation body
ČIA
Address
Levínský Vršek 147, Levín, 267 01 Králův Dvůr
City
Králův Dvůr, Czech Republic
Division
Laboratorium metrologiczne
Valid until
05-06-2031

What this laboratory calibrates (2)

Calibration principles (5)

Certificate information

Certificate date
05.06.2026

Accreditation scope — calibration items (5)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Indicating thermometers and temperature measuring chains Calibration – temperature
-196 °C · CMC 0,13 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard digital thermometer around the boiling point of nitrogen PP-M01-1
2 Indicating thermometers and temperature measuring chains Calibration – temperature
-196 °C to -90 °C · CMC 0,3 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard digital thermometer on a customer's device PP-M01-1
3 Indicating thermometers and temperature measuring chains Calibration – temperature
-90 °C to -45 °C · CMC 0,09 °C -45 °C to 0 °C · CMC 0,07 °C 0 °C to 45 °C · CMC 0,07 °C 45 °C to 155 °C · CMC 0,08 °C 155 °C to 250 °C · CMC 0,4 °C 250 °C to 450 °C · CMC 0,6 °C 450 °C to 650 °C · CMC 0,7 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard digital thermometer in a calibration oven PP-M01-1
4 Indicating thermometers and temperature measuring chains Calibration – temperature
0,01 °C · CMC 0,03 °C
CMC for the field of measured quantity: Temperature
Direct measurement at the triple point of water PP-M01-1
5 Electronic hygrometers and relative humidity measuring chains Calibration – air humidity
(10 – 25) °C — 5 % RH to 20 % RH · CMC 0,6 % RH (10 – 25) °C — 20 % RH to 40 % RH · CMC 0,7 % RH (10 – 25) °C — 40 % RH to 50 % RH · CMC 0,8 % RH (10 – 25) °C — 50 % RH to 60 % RH · CMC 0,9 % RH (10 – 25) °C — 60 % RH to 80 % RH · CMC 1,0 % RH (10 – 25) °C — 80 % RH to 90 % RH · CMC 1,1 % RH (10 – 25) °C — 90 % RH to 95 % RH · CMC 1,2 % RH (25 – 40) °C — 5 % RH to 30 % RH · CMC 0,6 % RH (25 – 40) °C — 30 % RH to 40 % RH · CMC 0,7 % RH (25 – 40) °C — 40 % RH to 60 % RH · CMC 0,8 % RH (25 – 40) °C — 60 % RH to 70 % RH · CMC 0,9 % RH (25 – 40) °C — 70 % RH to 90 % RH · CMC 1,0 % RH (25 – 40) °C — 90 % RH to 95 % RH · CMC 1,1 % RH
CMC for the field of measured quantity: Air humidity
Comparison with a standard dew point hygrometer in a humidity generator PP-M02-1

Reference standards (2)

PP-M01-1
PP-M02-1

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2400