🇨🇿 Vysoké učení technické v Brně PL

ČIA 2395 • Brno, Czech Republic • 25 calibration items

Metrology & calibration

Vysoké učení technické v Brně is a calibration laboratory accredited by ČIA under number 2395 (Brno, Czech Republic). Its accreditation scope covers 25 calibration items in the following areas: Metrology & calibration. Calibrated items include: AC Voltage / AC voltage sources and meters, AC voltmeters, multimet…, Acceleration of harmonic vibrations / Vibrometers without a transdu…, Acceleration of linear harmonic mechanical vibrations / Vibrometers…, Acceleration of linear harmonic mechanical vibrations / Vibrometers…. Most frequently used calibration principles: By simulated electrical signal, Comparison with a reference sensor (ČSN ISO 16063-21), Direct measurement or comparison with a reference counter.

At a glance

  • Accreditation number: ČIA 2395
  • Calibration items in scope: 25
  • Types of calibrated items: 25
  • Calibration principles used: 14
  • City: Brno, Czech Republic
  • Valid until: 03-09-2030
Accreditation number
2395
Accreditation body
ČIA
Address
Technická 3082/12, 616 00 Brno
City
Brno, Czech Republic
Division
Laboratorium wzorcujące CVVOZE
Valid until
03-09-2030

What this laboratory calibrates (25)

Calibration principles (14)

Certificate information

Certificate date
03.09.2025

Accreditation scope — calibration items (25)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Acceleration of linear harmonic mechanical vibrations / Vibrometers, seismometers, linear vibration generators, vibration testing systems Calibration – mechanical motion
0,2 Hz to 0,4 Hz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 1,5 % 0,4 to 1 Hz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 1,0 % 1 Hz to 1 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 0,5 % 1 to 5 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 0,7 % 5 to 10 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 1,5 % 10 to 15 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 2,0 % 15 to 20 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 3,0 %
CMC for the field of measured quantity: Mechanical motion
Comparison with a reference sensor (ČSN ISO 16063-21, ČSN ISO 16063-44) KP-01
2 Acceleration of linear harmonic mechanical vibrations / Vibrometers, seismometers, vibration transducer calibration systems, reference linear vibration generators Calibration – mechanical motion
0,2 Hz to 0,4 Hz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 0,7 % 0,4 Hz to 1 Hz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 0,5 % 1 Hz to 1 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 0,3 % 1 kHz to 5 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 0,5 % 5 kHz to 10 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 1,0 % 10 kHz to 15 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 2,0 % 15 kHz to 20 kHz — 0,01 m·s⁻² to 1000 m·s⁻² · CMC 2,5 %
CMC for the field of measured quantity: Mechanical motion
Direct measurement with a laser vibrometer (ČSN ISO 16063-11, ČSN ISO 16063-44) KP-01
3 Phase shift of acceleration of linear harmonic mechanical vibrations / Vibrometers, seismometers, linear vibration generators, vibration testing systems Calibration – mechanical motion
0,2 Hz to 1 Hz — 0 ° to 360 ° · CMC 1,5 ° 1 Hz to 5 kHz — 0 ° to 360 ° · CMC 0,7 ° 5 kHz to 10 kHz — 0 ° to 360 ° · CMC 1,0 ° 10 kHz to 15 kHz — 0 ° to 360 ° · CMC 2,0 ° 15 kHz to 20 kHz — 0 ° to 360 ° · CMC 3,0 °
CMC for the field of measured quantity: Mechanical motion
Comparison with a reference sensor (ČSN ISO 16063-21) KP-01
4 Phase shift of acceleration of linear harmonic mechanical vibrations / Vibrometers, seismometers, vibration transducer calibration systems, reference linear vibration generators Calibration – mechanical motion
0,2 Hz to 0,4 Hz — 0 ° to 360 ° · CMC 0,7 ° 0,4 Hz to 5 kHz — 0 ° to 360 ° · CMC 0,5 ° 5 kHz to 10 kHz — 0 ° to 360 ° · CMC 1,0 ° 10 kHz to 15 kHz — 0 ° to 360 ° · CMC 2,0 ° 15 kHz to 20 kHz — 0 ° to 360 ° · CMC 3,0 °
CMC for the field of measured quantity: Mechanical motion
Direct measurement with a laser vibrometer (ČSN ISO 16063-11) KP-01
5 Sensitivity / Vibration transducers, geophones, vibrometers and vibration measurement chains with an electric output Calibration – mechanical motion
0,2 Hz to 0,4 Hz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 1,5 % 0,4 Hz to 1 Hz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 1,0 % 1 Hz to 1 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 0,5 % 1 kHz to 5 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 0,7 % 5 kHz to 10 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 1,5 % 10 kHz to 15 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 2,0 % 15 kHz to 20 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 3,0 % 0,2 Hz to 0,4 Hz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 1,5 % 0,4 Hz to 1 Hz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 1,0 % 1 Hz to 1 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 0,5 % 1 kHz to 5 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 0,7 % 5 kHz to 10 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 1,5 % 10 kHz to 15 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 2,0 % 15 kHz to 20 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 3,0 % 0,2 Hz to 0,4 Hz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 1,5 % 0,4 Hz to 1 Hz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 1,0 % 1 Hz to 1 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,5 % 1 kHz to 5 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,7 % 5 kHz to 10 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 1,5 % 10 kHz to 15 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 2,0 % 15 kHz to 20 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 3,0 % 0,2 Hz to 0,4 Hz — 0,01 V/m to 10 000 V/m · CMC 1,5 % 0,4 Hz to 1 Hz — 0,01 V/m to 10 000 V/m · CMC 1,0 % 1 Hz to 1 kHz — 0,01 V/m to 10 000 V/m · CMC 0,5 % 1 kHz to 5 kHz — 0,01 V/m to 10 000 V/m · CMC 0,7 % 5 kHz to 10 kHz — 0,01 V/m to 10 000 V/m · CMC 1,5 % 10 kHz to 15 kHz — 0,01 V/m to 10 000 V/m · CMC 2,0 % 15 kHz to 20 kHz — 0,01 V/m to 10 000 V/m · CMC 3,0 %
CMC for the field of measured quantity: Mechanical motion
Comparison with a reference sensor (ČSN ISO 16063-21) KP-01
6 Sensitivity / Vibration transducers, standard vibration transducers, geophones, vibrometers and vibration measurement chains with an electric output Calibration – mechanical motion
0,2 Hz to 0,4 Hz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 0,7 % 0,4 Hz to 1 Hz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 0,5 % 1 Hz to 1 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 0,3 % 1 kHz to 5 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 0,5 % 5 kHz to 10 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 1,0 % 10 kHz to 15 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 2,0 % 15 kHz to 20 kHz — 0,01 mV/m·s⁻² to 10 000 mV/m·s⁻² · CMC 2,5 % 0,2 Hz to 0,4 Hz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 0,7 % 0,4 Hz to 1 Hz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 0,5 % 1 Hz to 1 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 0,3 % 1 kHz to 5 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 0,5 % 5 kHz to 10 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 1,0 % 10 kHz to 15 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 2,0 % 15 kHz to 20 kHz — 0,01 pC/m·s⁻² to 1000 pC/m·s⁻² · CMC 2,5 % 0,2 Hz to 0,4 Hz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,7 % 0,4 Hz to 1 Hz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,5 % 1 Hz to 1 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,3 % 1 kHz to 5 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,5 % 5 kHz to 10 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 1,0 % 10 kHz to 15 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 2,0 % 15 kHz to 20 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 2,5 % 0,2 Hz to 0,4 Hz — 0,01 V/m to 10 000 V/m · CMC 0,7 % 0,4 Hz to 1 Hz — 0,01 V/m to 10 000 V/m · CMC 0,5 % 1 Hz to 1 kHz — 0,01 V/m to 10 000 V/m · CMC 0,3 % 1 kHz to 5 kHz — 0,01 V/m to 10 000 V/m · CMC 0,5 % 5 kHz to 10 kHz — 0,01 V/m to 10 000 V/m · CMC 1,0 % 10 kHz to 15 kHz — 0,01 V/m to 10 000 V/m · CMC 2,0 % 15 kHz to 20 kHz — 0,01 V/m to 10 000 V/m · CMC 2,5 %
CMC for the field of measured quantity: Mechanical motion
Direct measurement with a laser vibrometer (ČSN ISO 16063-11) KP-01
7 Sensitivity / Laser vibrometers with an electric output Calibration – mechanical motion
0,2 Hz to 5 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,25 % 5 kHz to 10 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,3 % 10 kHz to 15 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,5 % 15 kHz to 20 kHz — 0,1 V/m·s⁻¹ to 10 000 V/m·s⁻¹ · CMC 0,7 % 0,2 Hz to 5 kHz — 0,01 V/m to 10 000 V/m · CMC 0,25 % 5 kHz to 10 kHz — 0,01 V/m to 10 000 V/m · CMC 0,3 % 10 kHz to 15 kHz — 0,01 V/m to 10 000 V/m · CMC 0,5 % 15 kHz to 20 kHz — 0,01 V/m to 10 000 V/m · CMC 0,7 %
CMC for the field of measured quantity: Mechanical motion
Direct measurement with a laser vibrometer (ČSN ISO 16063-41) KP-01
8 Sensitivity phase shift / Vibration transducers, geophones, vibrometers and vibration measurement chains with an electric output Calibration – mechanical motion
0,2 Hz to 1 Hz — 0 ° to 360 ° · CMC 1,5 ° 1 Hz to 5 kHz — 0 ° to 360 ° · CMC 0,7 ° 5 kHz to 10 kHz — 0 ° to 360 ° · CMC 1,0 ° 10 kHz to 15 kHz — 0 ° to 360 ° · CMC 2,0 ° 15 kHz to 20 kHz — 0 ° to 360 ° · CMC 3,0 °
CMC for the field of measured quantity: Mechanical motion
Comparison with a reference sensor (ČSN ISO 16063-21) KP-01
9 Sensitivity phase shift / Vibration transducers, standard vibration transducers, geophones, vibrometers and vibration measurement chains with an electric output Calibration – mechanical motion
0,2 Hz to 0,4 Hz — 0 ° to 360 ° · CMC 0,7 ° 0,4 Hz to 5 kHz — 0 ° to 360 ° · CMC 0,5 ° 5 kHz to 10 kHz — 0 ° to 360 ° · CMC 1,0 ° 10 kHz to 15 kHz — 0 ° to 360 ° · CMC 2,0 ° 15 kHz to 20 kHz — 0 ° to 360 ° · CMC 3,0 °
CMC for the field of measured quantity: Mechanical motion
Direct measurement with a laser vibrometer (ČSN ISO 16063-11) KP-01
10 Sensitivity phase shift / Laser vibrometers with an electric output Calibration – mechanical motion
0,2 Hz to 5 kHz — 0 ° to 360 ° · CMC 0,5 ° 5 kHz to 10 kHz — 0 ° to 360 ° · CMC 0,7 ° 10 kHz to 20 kHz — 0 ° to 360 ° · CMC 1,5 °
CMC for the field of measured quantity: Mechanical motion
Direct measurement with a laser vibrometer (ČSN ISO 16063-41) KP-01
11 Frequency of linear harmonic mechanical vibrations / Vibrometers, seismometers, linear vibration generators, vibration testing systems Calibration – mechanical motion
0,2 Hz to 50 kHz · CMC 0,01 %
CMC for the field of measured quantity: Mechanical motion
Measurement or comparison on a standard calibration device or with a reference frequency counter (ČSN ISO 16063-21, ČSN ISO 16063-44) KP-01
12 Frequency of linear harmonic mechanical vibrations / Vibrometers, seismometers, vibration transducer calibration systems, reference linear vibration generators Calibration – mechanical motion
0,2 Hz to 50 kHz · CMC 0,01 %
CMC for the field of measured quantity: Mechanical motion
Measurement or comparison on a standard calibration device or with a standard frequency counter (ČSN ISO 16063-11, ČSN ISO 16063-44) KP-01
13 Transfer / Charge amplifiers, voltage amplifiers, frequency filters, vibration test system controllers Calibration – mechanical motion
0,1 Hz to 1 Hz — 0,001 mV/pC to 1000 mV/pC · CMC 0,4 % 1 Hz to 20 kHz — 0,001 mV/pC to 1000 mV/pC · CMC 0,3 % 20 kHz to 50 kHz — 0,001 mV/pC to 1000 mV/pC · CMC 1,0 % 0,1 Hz to 1 Hz — 0,001 V/V to 1000 V/V · CMC 0,4 % 1 Hz to 20 kHz — 0,001 V/V to 1000 V/V · CMC 0,3 % 20 kHz to 50 kHz — 0,001 V/V to 1000 V/V · CMC 1,0 %
CMC for the field of measured quantity: Mechanical motion
By simulated electrical signal KP-02
14 Transfer phase shift / Charge amplifiers, voltage amplifiers, frequency filters, vibration test system controllers Calibration – mechanical motion
0,1 Hz to 1 Hz — 0 ° to 360 ° · CMC 0,75 ° 1 Hz to 20 kHz — 0 ° to 360 ° · CMC 0,5 °
CMC for the field of measured quantity: Mechanical motion
By simulated electrical signal KP-02
15 Acceleration of harmonic vibrations / Vibrometers without a transducer, measuring chains without a transducer with an output to a display or a scale, vibration test system controllers Calibration – mechanical motion
0,1 Hz to 1 Hz — 0,01 m·s⁻² to 10 000 m·s⁻² · CMC 0,4 % 1 Hz to 20 kHz — 0,01 m·s⁻² to 10 000 m·s⁻² · CMC 0,3 %
CMC for the field of measured quantity: Mechanical motion
By simulated electrical signal KP-02
16 Maximum value of half-sine wave mechanical shock acceleration / shock meters Calibration – mechanical motion
50 m·s⁻² to 200 m·s⁻² · CMC 2,0 % 200 m·s⁻² to 2 km·s⁻² · CMC 1,5 % 2 km·s⁻² to 40 km·s⁻² · CMC 1,8 % 40 km·s⁻² to 100 km·s⁻² · CMC 3,0 %
CMC for the field of measured quantity: Mechanical motion
Direct measurement or comparison with a reference sensor (ČSN ISO 16063-22) KP-01
17 Sensitivity of half-sine wave mechanical shock acceleration sensors / shock sensors, vibration sensors Calibration – mechanical motion
50 m·s⁻² to 200 m·s⁻² — 0,0001 mV/m·s⁻² to 100 mV/m·s⁻² · CMC 2,0 % 200 m·s⁻² to 2 km·s⁻² — 0,0001 mV/m·s⁻² to 100 mV/m·s⁻² · CMC 1,5 % 2 to 40 km·s⁻² — 0,0001 mV/m·s⁻² to 100 mV/m·s⁻² · CMC 1,8 % 40 to 100 km·s⁻² — 0,0001 mV/m·s⁻² to 100 mV/m·s⁻² · CMC 3,0 % 50 m·s⁻² to 200 m·s⁻² — 0,0001 pC/m·s⁻² to 100 pC/m·s⁻² · CMC 2,0 % 200 m·s⁻² to 2 km·s⁻² — 0,0001 pC/m·s⁻² to 100 pC/m·s⁻² · CMC 1,5 % 2 to 40 km·s⁻² — 0,0001 pC/m·s⁻² to 100 pC/m·s⁻² · CMC 1,8 % 40 to 100 km·s⁻² — 0,0001 pC/m·s⁻² to 100 pC/m·s⁻² · CMC 3,0 %
CMC for the field of measured quantity: Mechanical motion
Comparison with a reference sensor (ČSN ISO 16063-22) KP-01
18 Sensitivity of half-sine wave mechanical shock sensors / shock hammers Calibration – mechanical motion
0,1 mV/N to 100 mV/N · CMC 3,0 % 0,1 pC/N to 100 pC/N · CMC 3,0 %
CMC for the field of measured quantity: Mechanical motion
Comparison with a reference sensor using a known reaction mass (ČSN ISO 16063-22) KP-01
19 DC Voltage / DC voltage sources and meters, voltmeters, multimeters, signal analyzers, vibration meters, vibration test system controllers Calibration – electrical quantities
0 mV to 100 mV · CMC 0,010 % + 0,010 mV 100 mV to 1 V · CMC 0,010 % + 0,020 mV 1 V to 10 V · CMC 0,010 % + 0,10 mV 10 V to 100 V · CMC 0,010 % + 1,0 mV
CMC for the field of measured quantity: Electrical quantities
Direct measurement or comparison with a reference multimeter KP-03
20 DC current / DC current sources and meters, ammeters, multimeters Calibration – electrical quantities
0 mA to 1 mA · CMC 0,10 % + 0,10 uA 1 mA to 10 mA · CMC 0,10 % + 5,0 uA 10 mA to 100 mA · CMC 0,10 % + 10 uA 100 mA to 1 A · CMC 0,20 % + 0,20 mA
CMC for the field of measured quantity: Electrical quantities
Direct measurement or comparison with a reference multimeter KP-03
21 DC Resistance / Resistors, resistance boxes, shunts, dividers Calibration – electrical quantities
0 Ω to 100 Ω · CMC 0,020 % + 0,010 Ω 100 Ω to 1 kΩ · CMC 0,020 % + 0,020 Ω 1 kΩ to 10 kΩ · CMC 0,020 % + 0,20 Ω 10 kΩ to 100 kΩ · CMC 0,020 % + 2,0 Ω 100 kΩ to 1 MΩ · CMC 0,025 % + 20 Ω 1 MΩ to 10 MΩ · CMC 0,10 % + 200 Ω
CMC for the field of measured quantity: Electrical quantities
Direct measurement by a reference multimeter KP-03
22 AC Voltage / AC voltage sources and meters, AC voltmeters, multimeters, signal analyzers, vibration meters, vibration sensor calibration systems, signal generators, controllers for vibration test systems Calibration – electrical quantities
10 Hz to 20 kHz — 1 mV to 100 mV · CMC 0,10 % + 0,050 mV 20 kHz to 50 kHz — 1 mV to 100 mV · CMC 0,20 % + 0,10 mV 50 kHz to 100 kHz — 1 mV to 100 mV · CMC 0,80 % + 0,15 mV 10 Hz to 20 kHz — 100 mV to 1 V · CMC 0,10 % + 0,50 mV 20 kHz to 50 kHz — 100 mV to 1 V · CMC 0,20 % + 1,0 mV 50 kHz to 100 kHz — 100 mV to 1 V · CMC 0,80 % + 1,5 mV 10 Hz to 20 kHz — 1 V to 10 V · CMC 0,10 % + 5,0 mV 20 kHz to 50 kHz — 1 V to 10 V · CMC 0,20 % + 10 mV 50 kHz to 100 kHz — 1 V to 10 V · CMC 0,80 % + 15 mV 10 Hz to 20 kHz — 10 V to 100 V · CMC 0,10 % + 50 mV 20 kHz to 50 kHz — 10 V to 100 V · CMC 0,20 % + 100 mV 50 kHz to 100 kHz — 10 V to 100 V · CMC 0,80 % + 150 mV
CMC for the field of measured quantity: Electrical quantities
Direct measurement or comparison with a reference multimeter KP-03
23 Frequency / Signal generators, counters, multimeters, signal analyzers, vibration meters, calibration systems, vibration test system controllers Calibration – time and frequency quantities
15 mV to 250 mV — 0,1 Hz to 100 kHz · CMC 2,0·10⁻⁴/f 250 mV to 50 V — 0,1 Hz to 100 kHz · CMC 2,0·10⁻⁵/f 15 mV to 250 mV — 100 kHz to 1 MHz · CMC 2,0·10⁻⁹ 250 mV to 50 V — 100 kHz to 1 MHz · CMC 2,0·10⁻¹⁰ 15 mV to 50 V — 1 MHz to 10 MHz · CMC 2,0·10⁻¹⁰
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement or comparison with a reference counter KP-03
24 Period / Signal generators, counters, multimeters, signal analyzers, vibration meters, calibration systems, vibration test system controllers Calibration – time and frequency quantities
15 mV to 50 V — 100 ns to 1 µs · CMC 2,0·10⁻¹⁰ 15 mV to 250 mV — 1 µs to 10 µs · CMC 2,0·10⁻⁹ 250 mV to 50 V — 1 µs to 10 µs · CMC 2,0·10⁻¹⁰ 15 mV to 250 mV — 10 µs to 10 s · CMC 2,0·10⁻⁴·T 250 mV to 50 V — 10 µs to 10 s · CMC 2,0·10⁻⁵·T
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement or comparison with a reference counter KP-03
25 Time interval / Signal generators, timers, counters, multimeters, signal analyzers Calibration – time and frequency quantities
250 mV to 50 V — 0,1 ms to 1000 s · CMC 5,0·10⁻⁵
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement or comparison with a reference counter KP-03

Reference standards (3)

KP-01
KP-02
KP-03

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2395