🇨🇿 KALIST AKL s.r.o. PL

ČIA 2394 • Třebětice, Czech Republic • 20 calibration items

Metrology & calibration

KALIST AKL s.r.o. is a calibration laboratory accredited by ČIA under number 2394 (Třebětice, Czech Republic). Its accreditation scope covers 20 calibration items in the following areas: Metrology & calibration. Calibrated items include: Balances with non-automatic function, Indicating thermometers, temperature measuring chains, Calibration of infrared non-contact thermometers, Calibration of thermal equipment with temperature control. Most frequently used calibration principles: Comparison with a reference hygrometer, Comparison with a reference thermometer in a liquid bath, Gravimetric method (ČSN EN ISO 8655-6; EURAMET cg-19).

At a glance

  • Accreditation number: ČIA 2394
  • Calibration items in scope: 20
  • Types of calibrated items: 12
  • Calibration principles used: 16
  • City: Třebětice, Czech Republic
  • Valid until: 04-09-2030
Accreditation number
2394
Accreditation body
ČIA
Address
č.p. 8, 769 01 Třebětice
City
Třebětice, Czech Republic
Division
Laboratorium wzorcujące
Valid until
04-09-2030

What this laboratory calibrates (12)

Calibration principles (16)

Certificate information

Certificate date
04.09.2025

Accreditation scope — calibration items (20)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Piston pipettes Calibration – volume
0,5 µl to 10000 µl · CMC 0,13 % + 0,01 µl
CMC for the field of measured quantity: Volume
Gravimetric method (ČSN EN ISO 8655-6; EURAMET cg-19) KP-05
2 Other piston gauges – burettes, dispensers, dilution devices Calibration – volume
0,1 ml to 100 ml · CMC 0,05 %
CMC for the field of measured quantity: Volume
Gravimetric method (ČSN EN ISO 8655-6; EURAMET cg-19) KP-05
3 Balances with non-automatic function Calibration – mass
0 g · CMC 0,0006 mg
CMC for the field of measured quantity: Mass
KP-01
4 Balances with non-automatic function Calibration – mass
1 mg to 50 mg · CMC 0,005 mg 50 mg to 500 mg · CMC 0,009 mg 500 mg to 1 g · CMC 0,011 mg 1 g to 2 g · CMC 0,014 mg 2 g to 5 g · CMC 0,017 mg 5 g to 10 g · CMC 0,021 mg 10 g to 20 g · CMC 0,027 mg 20 g to 50 g · CMC 0,034 mg 50 g to 66 kg · CMC 5·10⁻⁷
CMC for the field of measured quantity: Mass
Loading using a class E2 reference weight (according to OIML R111-1:2004) KP-01
5 Balances with non-automatic function Calibration – mass
1 mg to 5 g · CMC 0,1 mg 5 g to 50 g · CMC 0,2 mg 50 g to 16 kg · CMC 4·10⁻⁶
CMC for the field of measured quantity: Mass
Loading using a class F1 reference weight (according to OIML R111-1:2004) KP-01
6 Balances with non-automatic function Calibration – mass
1 g to 50 g · CMC 1 mg 50 g to 370 kg · CMC 1·10⁻⁵
CMC for the field of measured quantity: Mass
Loading using a class F2 reference weight (according to OIML R111-1:2004) KP-01
7 Balances with non-automatic function Calibration – mass
200 kg to 1500 kg · CMC 1,2·10⁻⁵
CMC for the field of measured quantity: Mass
Loading using a class F2 reference weight (according to OIML R111-1:2004) with the use of substitute load KP-01
8 Balances with non-automatic function Calibration – mass
10 kg to 11 t · CMC 3,2·10⁻⁵
CMC for the field of measured quantity: Mass
Loading using a class M1 reference weight (according to OIML R111-1:2004) KP-01
9 Balances with non-automatic function Calibration – mass
200 kg to 3 t · CMC 6·10⁻⁴ 3 t to 15 t · CMC 9·10⁻⁴ 15 t to 30 t · CMC 8·10⁻⁴
CMC for the field of measured quantity: Mass
Loading using a class M1 reference weight (according to OIML R111-1:2004) with the use of substitute load KP-01
10 Conventional mass of weights and bodies Calibration – mass
1 mg to 20 mg · CMC 0,007 mg 20 mg to 500 mg · CMC 0,016 mg 1 g · CMC 0,020 mg 2 g · CMC 0,021 mg 5 g · CMC 0,025 mg 10 g · CMC 0,05 mg 20 g · CMC 0,05 mg 50 g · CMC 0,06 mg 100 g · CMC 0,2 mg 200 g · CMC 0,2 mg 500 g · CMC 0,4 mg 1 kg · CMC 2 mg 2 kg · CMC 2 mg 5 kg · CMC 15 mg 10 kg · CMC 20 mg 20 kg · CMC 30 mg 50 kg · CMC 50 mg 60 kg · CMC 80 mg
CMC for the field of measured quantity: Mass
Comparison with a reference weight (according to OIML R111-1:2004) using reference scales and comparators KP-06
11 Speed / rpm gauges Calibration – rotational speed
10 min⁻¹ to 1000 min⁻¹ · CMC 2 min⁻¹ 1000 min⁻¹ to 10000 min⁻¹ · CMC 3 min⁻¹ 10000 min⁻¹ to 50000 min⁻¹ · CMC 0,012 % + 2 min⁻¹
CMC for the field of measured quantity: Rotational speed
Comparison with a reference rpm gauge KP-02
12 Glass thermometers Calibration – temperature
-40 °C to 20 °C · CMC 0,15 °C 20 °C to 150 °C · CMC 0,10 °C 150 °C to 180 °C · CMC 0,12 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a liquid bath KP-03 část C
13 Indicating thermometers, temperature measuring chains Calibration – temperature
-196 °C · CMC 0,70 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in liquid nitrogen KP-03 část A
14 Indicating thermometers, temperature measuring chains Calibration – temperature
-95 °C to -40 °C · CMC 0,20 °C 180 °C to 230 °C · CMC 0,25 °C 230 °C to 415 °C · CMC 0,45 °C 415 °C to 600 °C · CMC 1,5 °C 600 °C to 800 °C · CMC 2,6 °C 800 °C to 1100 °C · CMC 2,8 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a dry block KP-03 část A
15 Indicating thermometers, temperature measuring chains Calibration – temperature
-40 °C to 20 °C · CMC 0,15 °C 20 °C to 150 °C · CMC 0,10 °C 150 °C to 180 °C · CMC 0,12 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a liquid bath KP-03 část A
16 Calibration of thermal equipment with temperature control Calibration – temperature
-196 °C to -70 °C · CMC 0,90 °C -70 °C to -40 °C · CMC 0,60 °C -40 °C to 150 °C · CMC 0,30 °C 150 °C to 230 °C · CMC 0,45 °C 230 °C to 415 °C · CMC 0,50 °C 415 °C to 600 °C · CMC 1,6 °C 600 °C to 800 °C · CMC 2,6 °C 800 °C to 1100 °C · CMC 2,8 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer KP-03 část B
17 Calibration of infrared non-contact thermometers Calibration – temperature
-30 °C to 0 °C · CMC 2,8 °C 0 °C to 20 °C · CMC 1,6 °C 20 °C to 80 °C · CMC 1,1 °C 80 °C to 200 °C · CMC 1,5 °C 200 °C to 350 °C · CMC 2,5 °C 350 °C to 500 °C · CMC 3,7 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer (black body) KP-03-IR
18 Relative humidity / hygrometers and humidity measuring chains Calibration – air humidity
(13 to 33) °C — 10 % RH to 30 % RH · CMC 1,2 % RH (13 to 33) °C — 30 % RH to 60 % RH · CMC 1,5 % RH (13 to 33) °C — 60 % RH to 80 % RH · CMC 1,7 % RH (13 to 33) °C — 80 % RH to 95 % RH · CMC 2,1 % RH
CMC for the field of measured quantity: Air humidity
Comparison with a reference hygrometer KP-04 část A
19 Relative humidity / measuring chains and characterisation of climatic chambers Calibration – air humidity
(13 to 33) °C — 10 % RH to 30 % RH · CMC 1,5 % RH (13 to 33) °C — 30 % RH to 60 % RH · CMC 1,7 % RH (13 to 33) °C — 60 % RH to 80 % RH · CMC 1,8 % RH (13 to 33) °C — 80 % RH to 95 % RH · CMC 2,1 % RH
CMC for the field of measured quantity: Air humidity
Comparison with a reference hygrometer KP-04 část B
20 Time interval / stopwatches, timers and chronometers Calibration – time and frequency quantities
1 s to 86400 s · CMC 0,15 s
CMC for the field of measured quantity: Time and frequency quantities
Comparison with reference stopwatches, manual activation KP-07

Reference standards (11)

KP-01
KP-02
KP-03 část A
KP-03 část B
KP-03 část C
KP-03-IR
KP-04 část A
KP-04 část B
KP-05
KP-06
KP-07

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2394