🇨🇿 KESA, s. r. o. PL
Metrology & calibration
KESA, s. r. o. is a calibration laboratory accredited by ČIA under number 2361 (Skuteč, Czech Republic). Its accreditation scope covers 3 calibration items in the following areas: Metrology & calibration. Calibrated items include: Direct-reading hygrometers – measuring chains, Direct-reading thermometers – measuring chains, incl. field homogen…, Glass thermometers. Most frequently used calibration principles: Comparison with a standard capacitive hygrometer in a climatic chamber, Comparison with a standard digital thermometer in a thermostatic bath, Comparison with a standard digital thermometer in a thermostatic ba….
At a glance
- Accreditation number: ČIA 2361
- Calibration items in scope: 3
- Types of calibrated items: 3
- Calibration principles used: 3
- City: Skuteč, Czech Republic
- Valid until: 12-08-2029
- Accreditation number
- 2361
- Accreditation body
- ČIA
- Address
- Smetanova 846, 539 73 Skuteč
- City
- Skuteč, Czech Republic
- Division
- Laboratorium wzorcujące
- Valid until
- 12-08-2029
What this laboratory calibrates (3)
- Direct-reading hygrometers – measuring chains — 1 calibration item
- Direct-reading thermometers – measuring chains, incl. field homogeneity — 1 calibration item
- Glass thermometers — 1 calibration item
Calibration principles (3)
- Comparison with a standard capacitive hygrometer in a climatic chamber — 1 calibration item
- Comparison with a standard digital thermometer in a thermostatic bath — 1 calibration item
- Comparison with a standard digital thermometer in a thermostatic bath or calibration furnace — 1 calibration item
Certificate information
- Certificate date
- 16.02.2026
Accreditation scope — calibration items (3)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Direct-reading thermometers – measuring chains, incl. field homogeneity | Calibration – temperature -196 °C to -40 °C · CMC 0,20 °C -40 °C to 150 °C · CMC 0,06 °C 150 °C to 300 °C · CMC 0,50 °C 300 °C to 600 °C · CMC 0,60 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in a thermostatic bath or calibration furnace | KP T01, část 9 | |
| 2 | Glass thermometers | Calibration – temperature -80 °C to -40 °C · CMC 0,25 °C -40 °C to 150 °C · CMC 0,15 °C 150 °C to 200 °C · CMC 0,25 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in a thermostatic bath | KP T01, část 10 | |
| 3 | Direct-reading hygrometers – measuring chains | Calibration – physico-chemical quantities – humidity 10 °C to 30 °C — 10 % RH to 30 % RH · CMC 1,2 % RH 10 °C to 30 °C — 30 % RH to 60 % RH · CMC 1,3 % RH 10 °C to 30 °C — 60 % RH to 90 % RH · CMC 1,4 % RH 10 °C to 30 °C — 90 % RH to 95 % RH · CMC 1,8 % RH CMC for the field of measured quantity: Physico-chemical quantities – humidity |
Comparison with a standard capacitive hygrometer in a climatic chamber | KP V01 | |
Reference standards (3)
KP T01, část 10
KP T01, část 9
KP V01