🇨🇿 RHH s.r.o. PL
Metrology & calibration
RHH s.r.o. is a calibration laboratory accredited by ČIA under number 2359 (Pardubice, Czech Republic). Its accreditation scope covers 6 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electronic humidity meters and humidity measuring chains, including…, Electronic thermometers, Electronic thermometers and temperature measuring chains, including…, Gloss meters. Most frequently used calibration principles: Comparison with a standard humidity meter, Comparison with a standard thermometer, Comparison with a standard thermometer in a calibration furnace.
At a glance
- Accreditation number: ČIA 2359
- Calibration items in scope: 6
- Types of calibrated items: 6
- Calibration principles used: 6
- City: Pardubice, Czech Republic
- Valid until: 15-06-2028
- Accreditation number
- 2359
- Accreditation body
- ČIA
- Address
- S. K. Neumanna 1316, 530 02 Pardubice
- City
- Pardubice, Czech Republic
- Division
- RHH s.r.o. – laboratorium wzorcujące
- Valid until
- 15-06-2028
What this laboratory calibrates (6)
- Electronic humidity meters and humidity measuring chains, including climatic chambers — 1 calibration item
- Electronic thermometers — 1 calibration item
- Electronic thermometers and temperature measuring chains, including thermal and climatic chambers — 1 calibration item
- Gloss meters — 1 calibration item
- Gloss standards — 1 calibration item
- Piston pipettes and piston dispensing systems — 1 calibration item
Calibration principles (6)
- Comparison with a standard humidity meter — 1 calibration item
- Comparison with a standard thermometer — 1 calibration item
- Comparison with a standard thermometer in a calibration furnace — 1 calibration item
- Direct measurement of gloss standards — 1 calibration item
- Gravimetric method — 1 calibration item
- Measurement by a standard gloss meter — 1 calibration item
Certificate information
- Certificate date
- 15.06.2023
Accreditation scope — calibration items (6)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Piston pipettes and piston dispensing systems | Calibration – volume distilled water — 0,5 µl to 10 000 µl · CMC 0,13 % + 0,01µl distilled water — 10 000 µl to 150 000 µl · CMC 0,06 % CMC for the field of measured quantity: Volume |
Gravimetric method | RHH-AKP-01 (ČSN EN ISO 8655-6) | |
| 2 | Electronic thermometers | Calibration – temperature -30 °C to 50 °C · CMC 0,12 °C 50 °C to 140 °C · CMC 0,18 °C 140 °C to 250 °C · CMC 0,32 °C 250 °C to 650 °C · CMC 1,2 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard thermometer in a calibration furnace | RHH-AKP-02 | |
| 3 | Electronic thermometers and temperature measuring chains, including thermal and climatic chambers | Calibration – temperature -80 °C to -40 °C · CMC 0,30 °C -40 °C to 50 °C · CMC 0,22 °C 50 °C to 150 °C · CMC 0,30 °C 150 °C to 250 °C · CMC 0,54 °C 250 °C to 500 °C · CMC 1,1 °C 500 °C to 800 °C · CMC 2,1 °C 800 °C to 1000 °C · CMC 2,8 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard thermometer | RHH-AKP-02 | |
| 4 | Gloss meters | Calibration – optical quantities (Gloss) 0 GU to 100 GU · CMC 1,2 GU CMC for the field of measured quantity: Optical quantities (Gloss) |
Direct measurement of gloss standards | RHH-AKP-03 | |
| 5 | Gloss standards | Calibration – optical quantities (Gloss) 0 GU to 100 GU · CMC 1,4 GU CMC for the field of measured quantity: Optical quantities (Gloss) |
Measurement by a standard gloss meter | RHH-AKP-03 | |
| 6 | Electronic humidity meters and humidity measuring chains, including climatic chambers | Calibration – humidity (2 to 20) °C — 10 % RH to 40 % RH · CMC 1,3 % RH (20 to 60) °C — 10 % RH to 40 % RH · CMC 1,1 % RH (60 to 90) °C — 10 % RH to 40 % RH · CMC 1,5 % RH (2 to 20) °C — 40 % RH to 70 % RH · CMC 1,3 % RH (20 to 60) °C — 40 % RH to 70 % RH · CMC 1,2 % RH (60 to 90) °C — 40 % RH to 70 % RH · CMC 1,6 % RH (2 to 20) °C — 70 % RH to 90 % RH · CMC 1,5 % RH (20 to 60) °C — 70 % RH to 90 % RH · CMC 1,3 % RH (60 to 90) °C — 70 % RH to 90 % RH · CMC 1,8 % RH CMC for the field of measured quantity: Humidity |
Comparison with a standard humidity meter | RHH-AKP-04 | |
Reference standards (5)
RHH-AKP-01
RHH-AKP-02
RHH-AKP-03
RHH-AKP-04
ČSN EN ISO 8655-6