🇨🇿 HOLAB, spol. s r.o. PL
Metrology & calibration
HOLAB, spol. s r.o. is a calibration laboratory accredited by ČIA under number 2358 (Blansko, Czech Republic). Its accreditation scope covers 7 calibration items in the following areas: Metrology & calibration. Calibrated items include: AC voltage – peak value / signal analysers, vibrometers, controller…, Acceleration of rectilinear mechanical vibrations of harmonic wavef…, DC voltage / signal analysers, vibrometers, controllers of vibratio…, Frequency meters – signal analysers, vibrometers, counters, control…. Most frequently used calibration principles: Comparison with a reference multimeter, Comparison with a reference sensor, Comparison with a reference digital hygrometer.
At a glance
- Accreditation number: ČIA 2358
- Calibration items in scope: 7
- Types of calibrated items: 7
- Calibration principles used: 4
- City: Blansko, Czech Republic
- Valid until: 22-05-2028
- Accreditation number
- 2358
- Accreditation body
- ČIA
- Address
- Gellhornova 2231/4, 678 01 Blansko
- City
- Blansko, Czech Republic
- Division
- Laboratorium wzorcujące
- Valid until
- 22-05-2028
What this laboratory calibrates (7)
- AC voltage – peak value / signal analysers, vibrometers, controllers of vibration test systems — 1 calibration item
- Acceleration of rectilinear mechanical vibrations of harmonic waveform / vibration standards and vibration test systems, vibrometers with a sensor — 1 calibration item
- DC voltage / signal analysers, vibrometers, controllers of vibration test systems — 1 calibration item
- Frequency meters – signal analysers, vibrometers, counters, controllers of vibration test systems — 1 calibration item
- Humidity meters of thermal and climatic chambers and enclosures, measurement of humidity in equipment with temperature and humidity control — 1 calibration item
- Sensitivity of sensors / vibration sensors — 1 calibration item
- Thermometers of thermal and climatic chambers and enclosures, measurement of temperature in equipment with temperature and humidity control — 1 calibration item
Calibration principles (4)
- Comparison with a reference multimeter — 3 calibration items
- Comparison with a reference sensor — 2 calibration items
- Comparison with a reference digital hygrometer — 1 calibration item
- Comparison with a reference digital thermometer — 1 calibration item
Certificate information
- Certificate date
- 28.12.2023
Accreditation scope — calibration items (7)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Acceleration of rectilinear mechanical vibrations of harmonic waveform / vibration standards and vibration test systems, vibrometers with a sensor | Calibration – mechanical motion 7 Hz to 10 Hz — 0,10 m s⁻² to 200 m s⁻² · CMC 2,0 % 10 Hz to 100 Hz — 0,10 m s⁻² to 200 m s⁻² · CMC 1,4 % 100 Hz — 0,10 m s⁻² to 200 m s⁻² · CMC 1,0 % 100 Hz to 920 Hz — 0,10 m s⁻² to 200 m s⁻² · CMC 1,2 % 920 Hz to 5 kHz — 0,10 m s⁻² to 200 m s⁻² · CMC 1,6 % 5 kHz to 10 kHz — 0,10 m s⁻² to 200 m s⁻² · CMC 2,0 % CMC for the field of measured quantity: Mechanical motion |
Comparison with a reference sensor | KP 2.22 (ČSN ISO 16063-21) | |
| 2 | Sensitivity of sensors / vibration sensors | Calibration – mechanical motion 7 Hz to 10 Hz — 0,10 mV/m s⁻² to 3000 mV/m s⁻² · CMC 2,0 % 10 Hz to 100 Hz — 0,10 mV/m s⁻² to 3000 mV/m s⁻² · CMC 1,4 % 100 Hz — 0,10 mV/m s⁻² to 3000 mV/m s⁻² · CMC 1,0 % 100 Hz to 920 Hz — 0,10 mV/m s⁻² to 3000 mV/m s⁻² · CMC 1,2 % 920 Hz to 5 kHz — 0,10 mV/m s⁻² to 3000 mV/m s⁻² · CMC 1,6 % 5 kHz to 10 kHz — 0,10 mV/m s⁻² to 3000 mV/m s⁻² · CMC 2,2 % 7 Hz to 10 Hz — 0,10 pC/m s⁻² to 100 pC/m s⁻² · CMC 2,0 % 10 Hz to 100 Hz — 0,10 pC/m s⁻² to 100 pC/m s⁻² · CMC 1,4 % 100 Hz — 0,10 pC/m s⁻² to 100 pC/m s⁻² · CMC 1,0 % 100 Hz to 920 Hz — 0,10 pC/m s⁻² to 100 pC/m s⁻² · CMC 1,2 % 920 Hz to 5 kHz — 0,10 pC/m s⁻² to 100 pC/m s⁻² · CMC 1,6 % 5 kHz to 10 kHz — 0,10 pC/m s⁻² to 100 pC/m s⁻² · CMC 2,2 % CMC for the field of measured quantity: Mechanical motion |
Comparison with a reference sensor | KP 2.22 (ČSN ISO 16063-21) | |
| 3 | Thermometers of thermal and climatic chambers and enclosures, measurement of temperature in equipment with temperature and humidity control | Calibration – temperature - 70 °C to 0 °C · CMC 0,5 °C 0 °C to 100 °C · CMC 0,4 °C 100 °C to 200 °C · CMC 0,5 °C 200 °C to 250 °C · CMC 1,0 °C 250 °C to 350 °C · CMC 1,4 °C CMC for the field of measured quantity: Temperature |
Comparison with a reference digital thermometer | KP 1.09 (DKD - R 5 - 7, metoda A, B, C) | |
| 4 | DC voltage / signal analysers, vibrometers, controllers of vibration test systems | Calibration – electrical quantities 0 mV to 100 mV · CMC 0,0025 % + 0,00017 mV 100 mV to 1 V · CMC 0,00090 % + 0,0006 mV 1 V to 10 V · CMC 0,00070 % + 0,006 mV CMC for the field of measured quantity: Electrical quantities |
Comparison with a reference multimeter | KP 3.23 | |
| 5 | AC voltage – peak value / signal analysers, vibrometers, controllers of vibration test systems | Calibration – electrical quantities 10 Hz to 40 Hz — 1 mV to 100 mV · CMC 0,08 % + 0,015 mV 40 Hz to 200 Hz — 1 mV to 100 mV · CMC 0,03 % + 0,009 mV 200 Hz to 2000 Hz — 1 mV to 100 mV · CMC 0,03 % + 0,008 mV 2 kHz to 20 kHz — 1 mV to 100 mV · CMC 0,04 % + 0,010 mV 10 Hz to 40 Hz — 100 mV to 1 V · CMC 0,06 % + 0,10 mV 40 Hz to 200 Hz — 100 mV to 1 V · CMC 0,03 % + 0,15 mV 200 Hz to 2000 Hz — 100 mV to 1 V · CMC 0,02 % + 0,06 mV 2 kHz to 20 kHz — 100 mV to 1 V · CMC 0,04 % + 0,10 mV 10 Hz to 40 Hz — 1 V to 10 V · CMC 0,06 % + 0,10 mV 40 Hz to 200 Hz — 1 V to 10 V · CMC 0,03 % + 0,15 mV 200 Hz to 2000 Hz — 1 V to 10 V · CMC 0,02 % + 0,06 mV 2 kHz to 20 kHz — 1 V to 10 V · CMC 0,04 % + 0,10 mV CMC for the field of measured quantity: Electrical quantities |
Comparison with a reference multimeter | KP 3.23 | |
| 6 | Frequency meters – signal analysers, vibrometers, counters, controllers of vibration test systems | Calibration – frequency 100 mV to 10 V — 1 Hz to 100 Hz · CMC 0,001 % + 0,020 mHz 100 mV to 10 V — 100 Hz to 1000 Hz · CMC 0,001 % + 0,20 mHz 100 mV to 10 V — 1 kHz to 10 kHz · CMC 0,001 % + 2,0 mHz CMC for the field of measured quantity: Frequency |
Comparison with a reference multimeter | KP 3.23 | |
| 7 | Humidity meters of thermal and climatic chambers and enclosures, measurement of humidity in equipment with temperature and humidity control | Calibration – humidity (18 to 32) °C — 10 % RH to 30 % RH · CMC 1,6 % RH (18 to 32) °C — 30 % RH to 60 % RH · CMC 1,9 % RH (18 to 32) °C — 60 % RH to 95 % RH · CMC 2,2 % RH CMC for the field of measured quantity: Humidity |
Comparison with a reference digital hygrometer | KP 1.09 (DKD - R 5 - 7, metoda A, B, C) | |
Reference standards (5)
DKD - R 5 - 7
KP 1.09
KP 2.22
KP 3.23
ČSN ISO 16063-21