🇨🇿 ÚJV Řež, a. s. PL

ČIA 2340 • Husinec, Czech Republic • 11 calibration items

Metrology & calibration

ÚJV Řež, a. s. is a calibration laboratory accredited by ČIA under number 2340 (Husinec, Czech Republic). Its accreditation scope covers 11 calibration items in the following areas: Metrology & calibration. Calibrated items include: Digital thermometers and measuring chains, Thermocouples, Deformation and digital pressure gauges, pressure transducers and p…, Barometric pressure. Most frequently used calibration principles: Comparison with a standard digital pressure gauge, Comparison with a standard resistance thermometer immersed in liqui…, Comparison with a standard resistance thermometer in a calibration….

At a glance

  • Accreditation number: ČIA 2340
  • Calibration items in scope: 11
  • Types of calibrated items: 6
  • Calibration principles used: 6
  • City: Husinec, Czech Republic
  • Valid until: 12-03-2031
Accreditation number
2340
Accreditation body
ČIA
Address
Hlavní 130, Řež, 250 68 Husinec
City
Husinec, Czech Republic
Division
Laboratorium wzorcujące
Valid until
12-03-2031

What this laboratory calibrates (6)

Calibration principles (6)

Certificate information

Certificate date
12.03.2026

Accreditation scope — calibration items (11)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Deformation and digital pressure gauges, pressure transducers and pressure measuring chains Calibration – pressure
medium: gas; gauge pressure — -90 kPa to -1 kPa · CMC 100 Pa medium: gas; gauge pressure — -1 kPa to 1 kPa · CMC 2,5 Pa medium: gas; gauge pressure — 1 kPa to 6,9 kPa · CMC 21 Pa medium: gas; gauge pressure — 6,9 kPa to 200 kPa · CMC 0,04 % + 30 Pa medium: gas; gauge pressure — 0,2 MPa to 2,0 MPa · CMC 1,1 kPa medium: gas; gauge pressure — 2,0 MPa to 4 MPa · CMC 6,9 kPa
CMC for the field of measured quantity: Pressure
Comparison with a standard digital pressure gauge PP 2306 086, PP 2306 088
2 Deformation and digital pressure gauges, pressure transducers and pressure measuring chains Calibration – pressure
medium: liquid; gauge pressure — 0 MPa to 13,8 MPa · CMC 6,9 kPa medium: liquid; gauge pressure — 13,8 MPa to 70 MPa · CMC 0,04 % + 7 kPa
CMC for the field of measured quantity: Pressure
Comparison with a standard digital pressure gauge PP 2306 086, PP 2306 088
3 Barometric pressure Calibration – pressure
70 kPa to 110 kPa · CMC 50 Pa
CMC for the field of measured quantity: Pressure
Comparison with a standard digital pressure gauge PP 2306 088
4 Glass thermometers Calibration – temperature
-30 °C to 0 °C · CMC 0,06 °C 0 °C to 50 °C · CMC 0,08 °C 50 °C to 140 °C · CMC 0,11 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance thermometer in a micro-bath PP 2306 083
5 Thermocouples Calibration – temperature
-196 °C · CMC 0,4 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance thermometer immersed in liquid nitrogen PP 2306 087
6 Thermocouples Calibration – temperature
-30 °C to 250 °C · CMC 0,4 °C 250 °C to 450 °C · CMC 0,6 °C 450 °C to 650 °C · CMC 0,8 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance thermometer in a calibration furnace PP 2306 087
7 Thermocouples Calibration – temperature
650 °C to 850 °C · CMC 1,3 °C 850 °C to 1100 °C · CMC 2,5 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermocouple in a calibration furnace PP 2306 087
8 Digital thermometers and measuring chains Calibration – temperature
-196 °C · CMC 0,15 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance thermometer immersed in liquid nitrogen PP 2306 082, PP 2306 087
9 Digital thermometers and measuring chains Calibration – temperature
-30 °C to 0 °C · CMC 0,06 °C 0 °C to 50 °C · CMC 0,13 °C 50 °C to 150 °C · CMC 0,20 °C 150 °C to 250 °C · CMC 0,27 °C 250 °C to 450 °C · CMC 0,47 °C 450 °C to 550 °C · CMC 0,57 °C 550 °C to 650 °C · CMC 0,67 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance thermometer in a calibration furnace PP 2306 082, PP 2306 087
10 Digital thermometers and measuring chains Calibration – temperature
650 °C to 850 °C · CMC 1,3 °C 850 °C to 1100 °C · CMC 2,5 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermocouple in a calibration furnace PP 2306 082, PP 2306 087
11 Temperature measuring chains including a temperature chamber Calibration – temperature
-190 °C to -100 °C · CMC 1,3 °C -100 °C to 320 °C · CMC 1,2 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermocouple PP 2306 087

Reference standards (5)

PP 2306 082
PP 2306 083
PP 2306 086
PP 2306 087
PP 2306 088

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2340