🇨🇿 PEMIT, s.r.o. PL

ČIA 2337 • Paskov, Czech Republic • 14 calibration items

Metrology & calibration

PEMIT, s.r.o. is a calibration laboratory accredited by ČIA under number 2337 (Paskov, Czech Republic). Its accreditation scope covers 14 calibration items in the following areas: Metrology & calibration. Calibrated items include: Direct-reading thermometers, Temperature sensors with transmitter, Deformation and electromechanical pressure gauges, Resistance temperature sensors. Most frequently used calibration principles: Comparison with a standard resistance sensor in a block furnace, Comparison with a standard resistance sensor in a liquid bath, Comparison with a type S thermocouple in a horizontal furnace.

At a glance

  • Accreditation number: ČIA 2337
  • Calibration items in scope: 14
  • Types of calibrated items: 7
  • Calibration principles used: 6
  • City: Paskov, Czech Republic
  • Valid until: 04-04-2030
Accreditation number
2337
Accreditation body
ČIA
Address
Místecká 845, 739 21 Paskov
City
Paskov, Czech Republic
Division
Metrologia – laboratorium wzorcujące
Valid until
04-04-2030

What this laboratory calibrates (7)

Calibration principles (6)

Certificate information

Certificate date
04.04.2025

Accreditation scope — calibration items (14)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Deformation and electromechanical pressure gauges Calibration – pressure
Gauge pressure; gas — -90 kPa to -67 kPa · CMC 3 Pa Gauge pressure; gas — -67 kPa to -15 kPa · CMC 9,2·10⁻⁵ Gauge pressure; gas — -15 kPa to 15 kPa · CMC 6,9·10⁻⁵ + 0,3 Pa Gauge pressure; gas — 15 kPa to 100 kPa · CMC 5,8·10⁻⁵ + 1,8 Pa Gauge pressure; gas — 100 kPa to 600 kPa · CMC 30 Pa Gauge pressure; gas — 600 kPa to 2 000 kPa · CMC 9,2·10⁻⁵ Absolute pressure; gas — 0 kPa to 33 kPa · CMC 1,7 Pa Absolute pressure; gas — 33 kPa to 110 kPa · CMC 9,2·10⁻⁵ Absolute pressure; gas — 110 kPa to 600 kPa · CMC 30 Pa Absolute pressure; gas — 600 kPa to 2 000 kPa · CMC 9,2·10⁻⁵ Absolute pressure; gas — 2 000 kPa to 10 000 kPa · CMC 5,0·10⁻⁴ + 1,5 kPa
CMC for the field of measured quantity: Pressure
Comparison with a Fluke PPC4 digital calibrator P01, P02, P03
2 Deformation and electromechanical pressure gauges Calibration – pressure
2 000 kPa to 10 000 kPa · CMC 5,0·10⁻⁴ + 1,5 kPa Gauge pressure, absolute pressure; water — 2 MPa to 16 MPa · CMC 3,5·10⁻⁴ + 2,1 kPa Gauge pressure, absolute pressure; water — 16 MPa to 25 MPa · CMC 3,5·10⁻⁴ + 4,5 kPa Gauge pressure, absolute pressure; water — 25 MPa to 60 MPa · CMC 3,5·10⁻⁴ + 9,5 kPa Barometric pressure — 80 kPa to 120 kPa · CMC 58 Pa
CMC for the field of measured quantity: Pressure
Comparison with a Beamex MC5 digital pressure gauge P01, P02, P03
3 Resistance temperature sensors Calibration – temperature
0 °C · CMC 0,06 °C -20 °C to 90 °C · CMC 0,07 °C 90 °C to 200 °C · CMC 0,09 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a liquid bath T01
4 Resistance temperature sensors Calibration – temperature
200 °C to 400 °C · CMC 0,4 °C 400 °C to 600 °C · CMC 0,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a block furnace T01
5 Thermoelectric temperature sensors Calibration – temperature
-20 °C to 90 °C · CMC 0,5 °C 90 °C to 600 °C · CMC 0,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a block furnace T02
6 Thermoelectric temperature sensors Calibration – temperature
600 °C to 1100 °C · CMC 1,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a type S thermocouple in a horizontal furnace T02
7 Direct-reading thermometers Calibration – temperature
-20 °C to 90 °C · CMC 0,07 °C 90 °C to 200 °C · CMC 0,09 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a liquid bath T04
8 Direct-reading thermometers Calibration – temperature
200 °C to 400 °C · CMC 0,4 °C 400 °C to 600 °C · CMC 0,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a block furnace T04
9 Direct-reading thermometers Calibration – temperature
600 °C to 1100 °C · CMC 1,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a type S thermocouple in a horizontal furnace T04
10 Temperature sensors with transmitter Calibration – temperature
-20 °C to 90 °C · CMC 0,07 °C 90 °C to 200 °C · CMC 0,09 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a liquid bath T05
11 Temperature sensors with transmitter Calibration – temperature
200 °C to 400 °C · CMC 0,4 °C 400 °C to 600 °C · CMC 0,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard resistance sensor in a block furnace T05
12 Temperature sensors with transmitter Calibration – temperature
600 °C to 1100 °C · CMC 1,6 °C
CMC for the field of measured quantity: Temperature
Comparison with a type S thermocouple in a horizontal furnace T05
13 Simulation of output signals of thermoelectric temperature sensors Calibration – temperature
TC type B — 20 °C to 500 °C · CMC 2,1 °C TC type B — 500 °C to 1820 °C · CMC 0,9 °C TC type R — -50 °C to 150 °C · CMC 1,1 °C TC type R — 150 °C to 1768 °C · CMC 0,6 °C TC type S — -50 °C to 0 °C · CMC 1,1 °C TC type S — 0 °C to 1768 °C · CMC 0,8 °C TC type J — -200 °C to 0 °C · CMC 0,3 °C TC type J — 0 °C to 1200 °C · CMC 0,4 °C TC type K — -200 °C to 1000 °C · CMC 0,4 °C TC type K — 1000 °C to 1372 °C · CMC 0,5 °C TC type N — -200 °C to -100 °C · CMC 0,5 °C TC type N — -100 °C to 750 °C · CMC 0,4 °C TC type N — 750 °C to 1300 °C · CMC 0,5 °C
CMC for the field of measured quantity: Temperature
Simulation with a Beamex MC5 multifunction calibrator R01
14 Calibration of output signals of resistance sensors Calibration – temperature
Pt 100, Pt 1000 — -200 °C to 0 °C · CMC 0,12 °C Pt 100, Pt 1000 — 0 °C to 850 °C · CMC 2,8·10⁻⁴ + 0,12 °C
CMC for the field of measured quantity: Temperature
Simulation with a Beamex MC5 multifunction calibrator R01

Reference standards (8)

P01
P02
P03
R01
T01
T02
T04
T05