🇨🇿 SAFINA, a.s. PL

ČIA 2329 • Vestec, Czech Republic • 2 calibration items

Metrology & calibration

SAFINA, a.s. is a calibration laboratory accredited by ČIA under number 2329 (Vestec, Czech Republic). Its accreditation scope covers 2 calibration items in the following areas: Metrology & calibration. Calibrated items include: Thermoelectric temperature sensors made of pure and precious metals. Most frequently used calibration principles: Comparison with a reference thermoelectric temperature sensor in a…, Fixed points.

At a glance

  • Accreditation number: ČIA 2329
  • Calibration items in scope: 2
  • Types of calibrated items: 1
  • Calibration principles used: 2
  • City: Vestec, Czech Republic
  • Valid until: 30-01-2028
Accreditation number
2329
Accreditation body
ČIA
Address
Vídeňská 104, 252 50 Vestec
City
Vestec, Czech Republic
Division
Dział produkcji termopar – laboratorium wzorcujące
Valid until
30-01-2028

What this laboratory calibrates (1)

Calibration principles (2)

Certificate information

Certificate date
30.01.2023

Accreditation scope — calibration items (2)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Thermoelectric temperature sensors made of pure and precious metals Calibration – temperature
Zn — 419,527 °C · CMC 0,3 °C Al — 660,323 °C · CMC 0,3 °C Ag — 961,78 °C · CMC 0,5 °C Au — 1064,18 °C · CMC 0,7 °C Cu — 1084,62 °C · CMC 0,5 °C Pd — 1553,5 °C · CMC 1,4 °C
CMC for the field of measured quantity: Temperature
Fixed points PP.31901.070
2 Thermoelectric temperature sensors made of pure and precious metals Calibration – temperature
230 °C to 420 °C · CMC 0,5 °C 420 °C to 660 °C · CMC 0,9 °C 660 °C to 1085 °C · CMC 1,2 °C 1085 °C to 1100 °C · CMC 1,4 °C 1100 °C to 1400 °C · CMC 2,2 °C 1400 °C to 1554 °C · CMC 2,7 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermoelectric temperature sensor in a horizontal furnace PP.31901.071

Reference standards (2)

PP.31901.070
PP.31901.071

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2329