🇨🇿 Všeobecná fakultní nemocnice v Praze PL
Metrology & calibration
Všeobecná fakultní nemocnice v Praze is a calibration laboratory accredited by ČIA under number 2323 (Praha 2, Czech Republic). Its accreditation scope covers 4 calibration items in the following areas: Metrology & calibration. Calibrated items include: Glass and digital thermometers, Non-automatic weighing instruments, Piston pipettes (single- and multi-channel) and piston dispensers, Temperature measuring chains. Most frequently used calibration principles: Comparison with a digital standard thermometer, Comparison with a standard digital thermometer in a liquid bath, Gravimetric method.
At a glance
- Accreditation number: ČIA 2323
- Calibration items in scope: 4
- Types of calibrated items: 4
- Calibration principles used: 4
- City: Praha 2, Czech Republic
- Valid until: 08-09-2027
- Accreditation number
- 2323
- Accreditation body
- ČIA
- Address
- Na Bojišti 1, 128 08 Praha 2
- City
- Praha 2, Czech Republic
- Division
- Ośrodek metrologiczny
- Valid until
- 08-09-2027
What this laboratory calibrates (4)
- Glass and digital thermometers — 1 calibration item
- Non-automatic weighing instruments — 1 calibration item
- Piston pipettes (single- and multi-channel) and piston dispensers — 1 calibration item
- Temperature measuring chains — 1 calibration item
Calibration principles (4)
- Comparison with a digital standard thermometer — 1 calibration item
- Comparison with a standard digital thermometer in a liquid bath — 1 calibration item
- Gravimetric method — 1 calibration item
- Loading with an F1 class standard weight — 1 calibration item
Certificate information
- Certificate date
- 08.09.2022
Accreditation scope — calibration items (4)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Piston pipettes (single- and multi-channel) and piston dispensers | Calibration – volume, flow 5 µl to 10 µl · CMC 0,07 µl 10 µl to 100 µl · CMC 0,16 µl 100 µl to 200 µl · CMC 0,25 µl 200 µl to 1000 µl · CMC 1,2 µl 1000 µl to 5000 µl · CMC 5,8 µl CMC for the field of measured quantity: Volume, flow |
Gravimetric method | KP-06-01 | |
| 2 | Non-automatic weighing instruments | Calibration – mass According to the weights used for calibration, Uᵢ can take the values: U(CMC) = ∑ Uᵢ — 1 mg to 10 kg · CMC U₁ (0,001 g) = 0,019 mg; U₂ (0,002 g) = 0,019 mg; U₃ (0,005 g) = 0,019 mg; U₄ (0,01 g) = 0,024 mg; U₅ (0,02 g) = 0,029 mg; U₆ (0,05 g) = 0,038 mg; U₇ (0,1 g) = 0,048 mg; U₈ (0,02 g) = 0,057 mg; U₉ (0,5 g) = 0,076 mg; U₁₀ (1 g) = 0,095 mg; U₁₁ (2 g) = 0,11 mg; U₁₂ (5 g) = 0,15 mg; U₁₃ (10 g) = 0,19 mg; U₁₄ (20 g) = 0,24 mg; U₁₅ (50 g) = 0,29 mg; U₁₆ (100 g) = 0,48 mg; U₁₇ (200 g) = 0,95 mg; U₁₈ (500 g) = 2,4 mg; U₁₉ (1000 g) = 4,8 mg; U₂₀ (2000 g) = 9,5 mg; U₂₁ (5000 g) = 24 mg; U₂₂ (10000 g) = 48 mg CMC for the field of measured quantity: Mass |
Loading with an F1 class standard weight | KP-08-01 | |
| 3 | Glass and digital thermometers | Calibration – temperature -40 °C to -1 °C · CMC 0,06 °C -1 °C to 1 °C · CMC 0,05 °C 1 °C to 40 °C · CMC 0,07 °C 40 °C to 200 °C · CMC 0,10 °C CMC for the field of measured quantity: Temperature |
Comparison with a standard digital thermometer in a liquid bath | KP-03-01, KP-03-02 | |
| 4 | Temperature measuring chains | Calibration – temperature -40 °C to 50 °C · CMC 0,3 °C 50 °C to 120 °C · CMC 0,4 °C 120 °C to 180 °C · CMC 0,6 °C CMC for the field of measured quantity: Temperature |
Comparison with a digital standard thermometer | KP-03-02 | |
Reference standards (4)
KP-03-01
KP-03-02
KP-06-01
KP-08-01