🇨🇿 TERMOSONDY Kladno, spol. s r.o. PL

ČIA 2305 • Kladno, Czech Republic • 12 calibration items

Metrology & calibration

TERMOSONDY Kladno, spol. s r.o. is a calibration laboratory accredited by ČIA under number 2305 (Kladno, Czech Republic). Its accreditation scope covers 12 calibration items in the following areas: Metrology & calibration. Calibrated items include: Bimetal and direct indicating thermometers, Contact thermometers, Direct current, Direct-current voltage. Most frequently used calibration principles: Comparison with a standard multimeter 34401A, Comparison with a standard thermometer in a liquid bath, -30 °C to 200 °C: comparison with a standard thermometer in a liqui….

At a glance

  • Accreditation number: ČIA 2305
  • Calibration items in scope: 12
  • Types of calibrated items: 12
  • Calibration principles used: 9
  • City: Kladno, Czech Republic
  • Valid until: 11-11-2029
Accreditation number
2305
Accreditation body
ČIA
Address
Dělnická 81, 272 01 Kladno
City
Kladno, Czech Republic
Division
laboratorium wzorcujące
Valid until
11-11-2029

What this laboratory calibrates (12)

Calibration principles (9)

Certificate information

Certificate date
11.11.2024

Accreditation scope — calibration items (12)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Thermoelectric temperature sensors made of precious metals Calibration – temperature
0 °C to 200 °C · CMC 0,8 °C 200 °C to 400 °C · CMC 1,4 °C 400 °C to 600 °C · CMC 1,9 °C 600 °C to 800 °C · CMC 2,3 °C 800 °C to 1000 °C · CMC 2,8 °C 1000 °C to 1200 °C · CMC 3,1 °C 1200 °C to 1554 °C · CMC 4,9 °C
CMC for the field of measured quantity: Temperature
0 °C to 200 °C: comparison with a standard thermometer in a liquid bath; 200 °C to 400 °C and 1200 °C to 1554 °C: comparison with a standard thermometer in a vertical furnace; 400 °C to 1200 °C: comparison with a standard thermometer in a horizontal furnace PP-Th 7/1-5-1
2 Thermoelectric temperature sensors made of common metals Calibration – temperature
-30 °C to 200 °C · CMC 0,5 °C 200 °C to 400 °C · CMC 1,6 °C 400 °C to 600 °C · CMC 1,9 °C 600 °C to 800 °C · CMC 2,3 °C 800 °C to 1000 °C · CMC 2,8 °C 1000 °C to 1200 °C · CMC 3,1 °C
CMC for the field of measured quantity: Temperature
-30 °C to 200 °C: comparison with a standard thermometer in a liquid bath; 200 °C to 400 °C: comparison with a standard thermometer in a vertical furnace; 400 °C to 1200 °C: comparison with a standard thermometer in a horizontal furnace PP-Th 7/1-5-1
3 Resistance temperature sensors Calibration – temperature
0 °C · CMC 0,12 °C -70 °C to -40 °C · CMC 0,35 °C -40 °C to 50 °C · CMC 0,16 °C 50 °C to 150 °C · CMC 0,18 °C 150 °C to 300 °C · CMC 0,60 °C 300 °C to 400 °C · CMC 0,71 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermometer in a liquid bath PP-Th 7/1-5-2
4 Glass thermometers Calibration – temperature
0 °C · CMC 0,18 °C -30 °C to 50 °C · CMC 0,23 °C 50 °C to 150 °C · CMC 0,25 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermometer in a liquid bath PP-Th 7/1-5-4
5 Infrared thermometers Calibration – temperature
50 °C to 200 °C · CMC 4,0 °C 200 °C to 400 °C · CMC 5,0 °C 400 °C to 600 °C · CMC 5,8 °C 600 °C to 800 °C · CMC 7,9 °C 800 °C to 1000 °C · CMC 10 °C 1000 °C to 1100 °C · CMC 11 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermometer on a black body model PP-Th 7/1-5-5
6 Contact thermometers Calibration – temperature
0 °C to 25 °C · CMC 1,5 °C 25 °C to 100 °C · CMC 2,0 °C 100 °C to 200 °C · CMC 4,3 °C 200 °C to 300 °C · CMC 4,9 °C
CMC for the field of measured quantity: Temperature
0 °C to 100 °C: comparison with a standard thermometer on an Al contact plate; 100 °C to 300 °C: comparison with a standard thermometer on a Cu contact plate PP-Th 7/1-5-6
7 Bimetal and direct indicating thermometers Calibration – temperature
0 °C · CMC 0,28 °C -30 °C to 50 °C · CMC 0,30 °C 50 °C to 150 °C · CMC 0,47 °C 150 °C to 300 °C · CMC 1,0 °C 300 °C to 400 °C · CMC 2,2 °C
CMC for the field of measured quantity: Temperature
0 °C and -30 °C to 150 °C: comparison with a standard thermometer in a liquid bath; 150 °C to 400 °C: comparison with a standard thermometer in a vertical furnace PP-Th 7/1-5-17
8 Temperature measuring chains Calibration – temperature
-70 °C to -40 °C · CMC 0,45 °C -40 °C to 150 °C · CMC 0,30 °C 150 °C to 300 °C · CMC 0,70 °C 300 °C to 400 °C · CMC 1,1 °C 400 °C to 600 °C · CMC 2,9 °C 600 °C to 800 °C · CMC 3,3 °C 800 °C to 1000 °C · CMC 3,6 °C 1000 °C to 1200 °C · CMC 4,0 °C 1200 °C to 1554 °C · CMC 5,9 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermometer in the space of a thermal device PP-Th 7/1-5-7a, PP-Th 7/1-5-7b
9 Temperature calibrators (electric method) Calibration – temperature
thermocouples R — 0 °C to 1760 °C · CMC 0,43 °C thermocouples S — 0 °C to 1760 °C · CMC 0,45 °C thermocouples B — 0 °C to 1820 °C · CMC 0,57 °C thermocouples J — -80 °C to 1200 °C · CMC 0,14 °C thermocouples T — 0 °C to 400 °C · CMC 0,09 °C thermocouples E — 0 °C to 1000 °C · CMC 0,12 °C thermocouples K — -80 °C to 1300 °C · CMC 0,19 °C thermocouples N — -80 °C to 1300 °C · CMC 0,18 °C resistance sensors — -80 °C to 850 °C · CMC 0,16 °C
CMC for the field of measured quantity: Temperature
Direct measurement with a standard multimeter 34401A PP-Th 7/1-5-16
10 Direct-current voltage Calibration – electrical quantities
0 mV · CMC 3,5 µV 0 mV to 100 mV · CMC 0,005 % + 3,5 µV 0,1 V to 1 V · CMC 0,004 % + 7 µV 1 V to 10 V · CMC 0,0035 % + 0,05 mV 10 V to 100 V · CMC 0,0045 % + 0,6 mV
CMC for the field of measured quantity: Electrical quantities
Comparison with a standard multimeter 34401A PP-Th 7/1-5-16
11 Direct current Calibration – electrical quantities
0 mA · CMC 2,0 µA 0 mA to 10 mA · CMC 0,05 % + 2 µA 10 mA to 100 mA · CMC 0,05 % + 5 µA 0,1 A to 1 A · CMC 0,10 % + 1 mA 1 A to 3 A · CMC 0,12 % + 6 mA
CMC for the field of measured quantity: Electrical quantities
Comparison with a standard multimeter 34401A PP-Th 7/1-5-16
12 Resistance Calibration – electrical quantities
0 Ω · CMC 4 mΩ 0 Ω to 100 Ω · CMC 0,01 % + 4 mΩ 0,1 kΩ to 1 kΩ · CMC 0,01 % + 10 mΩ 1 kΩ to 10 kΩ · CMC 0,01 % + 100 mΩ 10 kΩ to 100 kΩ · CMC 0,01 % + 1 Ω
CMC for the field of measured quantity: Electrical quantities
Comparison with a standard multimeter 34401A PP-Th 7/1-5-16

Reference standards (9)

PP-Th 7/1-5-1
PP-Th 7/1-5-16
PP-Th 7/1-5-17
PP-Th 7/1-5-2
PP-Th 7/1-5-4
PP-Th 7/1-5-5
PP-Th 7/1-5-6
PP-Th 7/1-5-7a
PP-Th 7/1-5-7b

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2305