🇨🇿 JD Dvořák, s.r.o. PL

ČIA 2298 • Praha 18 – Letňany, Czech Republic • 6 calibration items

Metrology & calibration

JD Dvořák, s.r.o. is a calibration laboratory accredited by ČIA under number 2298 (Praha 18 – Letňany, Czech Republic). Its accreditation scope covers 6 calibration items in the following areas: Metrology & calibration. Calibrated items include: Hygrometers integrated in measuring chains of climatic chambers and…, Thermometers integrated in measuring chains of thermal and climatic…, Direct indication electronic thermometers, thermometers for air tem…, Hygrometers, measuring chains for measuring relative humidity, data…. Most frequently used calibration principles: Comparison with a reference thermometer in a gaseous environment, Comparison with a standard aspiration hygrometer, Comparison with a standard aspiration hygrometer in a climatic chamber.

At a glance

  • Accreditation number: ČIA 2298
  • Calibration items in scope: 6
  • Types of calibrated items: 4
  • Calibration principles used: 4
  • City: Praha 18 – Letňany, Czech Republic
  • Valid until: 26-04-2029
Accreditation number
2298
Accreditation body
ČIA
Address
Toužimská 897/E/3, 199 00 Praha 18 – Letňany
City
Praha 18 – Letňany, Czech Republic
Division
JD Dvořák, s.r.o., laboratorium wzorcujące
Valid until
26-04-2029

What this laboratory calibrates (4)

Calibration principles (4)

Certificate information

Certificate date
26.04.2024

Accreditation scope — calibration items (6)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Thermometers integrated in measuring chains of thermal and climatic chambers and enclosures and special-purpose enclosures, where temperature can be generated Calibration – temperature
- 70 °C to - 45 °C · CMC 0,20 °C - 45 °C to 100 °C · CMC 0,17 °C 100 °C to 200 °C · CMC 0,21 °C 200 °C to 300 °C · CMC 0,9 °C 300 °C to 400 °C · CMC 1,1 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a gaseous environment Interní metoda 1 (DKD-R_5.7 metoda C)
2 Thermometers integrated in measuring chains of thermal and climatic chambers and enclosures and special-purpose enclosures, where temperature can be generated Calibration – temperature
- 70 °C to - 45 °C · CMC 0,35 °C - 45 °C to 100 °C · CMC 0,27 °C 100 °C to 150 °C · CMC 0,37 °C 150 °C to 180 °C · CMC 0,75 °C
CMC for the field of measured quantity: Temperature
Comparison with a reference thermometer in a gaseous environment Interní metoda 3 (DKD-R_5.7 metoda A a B)
3 Direct indication electronic thermometers, thermometers for air temperature measurement, temperature measuring chains, data loggers, outdoor thermometers Calibration – temperature
- 70 °C to 0 °C · CMC 0,45 °C 0 °C to 100 °C · CMC 0,36 °C 100 °C to 150 °C · CMC 0,57 °C 150 °C to 180 °C · CMC 1,0 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard thermometer in a climatic chamber Interní metoda 5
4 Hygrometers integrated in measuring chains of climatic chambers and enclosures and special-purpose enclosures, where humidity and temperature can be generated Calibration – relative humidity
(10 to 20) °C — 10 % RH to 30 % RH · CMC 1,7 % RH (10 to 20) °C — 30 % RH to 95 % RH · CMC 2,3 % RH (20 to 90) °C — 10 % RH to 30 % RH · CMC 1,5 % RH (20 to 90) °C — 30 % RH to 95 % RH · CMC 1,9 % RH (90 to 95) °C — 10 % RH to 30 % RH · CMC 1,4 % RH (90 to 95) °C — 30 % RH to 95 % RH · CMC 1,6 % RH
CMC for the field of measured quantity: Relative humidity
Comparison with a standard aspiration hygrometer Interní metoda 2 (DKD-R_5.7 metoda C)
5 Hygrometers integrated in measuring chains of climatic chambers and enclosures and special-purpose enclosures, where humidity and temperature can be generated Calibration – relative humidity
(10 to 20) °C — 10 % RH to 30 % RH · CMC 1,8 % RH (10 to 20) °C — 30 % RH to 95 % RH · CMC 2,4 % RH (20 to 90) °C — 10 % RH to 30 % RH · CMC 1,6 % RH (20 to 90) °C — 30 % RH to 95 % RH · CMC 2,0 % RH (90 to 95) °C — 10 % RH to 30 % RH · CMC 1,5 % RH (90 to 95) °C — 30 % RH to 95 % RH · CMC 1,7 % RH
CMC for the field of measured quantity: Relative humidity
Comparison with a standard aspiration hygrometer Interní metoda 4 (DKD-R_5.7 metoda A a B)
6 Hygrometers, measuring chains for measuring relative humidity, data loggers for measuring relative humidity Calibration – relative humidity
(10 to 90) °C — 10 % RH to 60 % RH · CMC 2,3 % RH (10 to 90) °C — 60 % RH to 90 % RH · CMC 3,2 % RH (10 to 90) °C — 90 % RH to 95 % RH · CMC 3,5 % RH
CMC for the field of measured quantity: Relative humidity
Comparison with a standard aspiration hygrometer in a climatic chamber Interní metoda 6