🇨🇿 LAB-MET s.r.o. PL

ČIA 2281 • Brno, Czech Republic • 9 calibration items

Metrology & calibration

LAB-MET s.r.o. is a calibration laboratory accredited by ČIA under number 2281 (Brno, Czech Republic). Its accreditation scope covers 9 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electronic and dial thermometers built into equipment, Electronic and direct-reading thermometers, Electronic and hair hygrometers, Electronic hygrometers, electronic hygrometers incl. the loop (chain). Most frequently used calibration principles: Comparison with a standard, Gravimetrically, Comparison with a standard tachometer by the non-contact method.

At a glance

  • Accreditation number: ČIA 2281
  • Calibration items in scope: 9
  • Types of calibrated items: 9
  • Calibration principles used: 3
  • City: Brno, Czech Republic
  • Valid until: 01-11-2027
Accreditation number
2281
Accreditation body
ČIA
Address
Štěpánkova 6, 644 00 Brno
City
Brno, Czech Republic
Division
Laboratorium wzorcujące
Valid until
01-11-2027

What this laboratory calibrates (9)

Calibration principles (3)

Certificate information

Certificate date
01.11.2022

Accreditation scope — calibration items (9)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Piston pipettes Calibration – volume, flow
0,5 µl to 10000 µl · CMC 0,12 % + 0,02 µl
CMC for the field of measured quantity: Volume, flow
Gravimetrically MKO 3/001/98 (ČSN EN ISO 8655-6 Euramet cg-19)
2 Other piston-operated measuring instruments Calibration – volume, flow
100 µl to 50000 µl · CMC 0,05 %
CMC for the field of measured quantity: Volume, flow
Gravimetrically MKO 3/001/98 (ČSN EN ISO 8655-6 Euramet cg-19)
3 Rotational speed meters Calibration – mechanical motion - rotational speed
1,0 min⁻¹ to 600,0 min⁻¹ · CMC 0,2 min⁻¹ 600,1 min⁻¹ to 999,9 min⁻¹ · CMC 1,2 min⁻¹ 1000 min⁻¹ to 30000 min⁻¹ · CMC 2 min⁻¹ 30001 min⁻¹ to 99999 min⁻¹ · CMC 5 min⁻¹
CMC for the field of measured quantity: Mechanical motion - rotational speed
Comparison with a standard tachometer by the non-contact method MKOT 3/027/12
4 Glass thermometers Calibration – temperature
-30 °C to 140 °C · CMC 0,08 °C 140 °C to 250 °C · CMC 0,10 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard MKST 3/002/98
5 Electronic and direct-reading thermometers Calibration – temperature
-30 °C to 0 °C · CMC 0,08 °C 0 °C · CMC 0,05 °C 0 °C to 140 °C · CMC 0,08 °C 140 °C to 250 °C · CMC 0,09 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard MKET 3/003/98
6 Electronic and dial thermometers built into equipment Calibration – temperature
-80 °C to -50 °C · CMC 0,2 °C 0 °C · CMC 0,07 °C 0 °C to 300 °C · CMC 0,1 °C 300 °C to 500 °C · CMC 1,5 °C 500 °C to 800 °C · CMC 1,8 °C 800 °C to 1000 °C · CMC 2,0 °C 1000 °C to 1200 °C · CMC 2,1 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard KEXT 3/030/00
7 Hot-air and steam sterilizers, drying ovens, furnaces, thermostats, freezing and cooling equipment, liquid baths, refrigerated centrifuges, climatic chambers (temperature part only, up to a volume of 2000 dm³), homogenization boilers Calibration – temperature
-80 °C to -50 °C · CMC 0,2 °C -50 °C to -50 °C · CMC 0,1 °C 0 °C · CMC 0,07 °C 0 °C to 300 °C · CMC 0,1 °C 300 °C to 500 °C · CMC 1,5 °C 500 °C to 800 °C · CMC 1,8 °C 800 °C to 1000 °C · CMC 2,0 °C 1000 °C to 1200 °C · CMC 2,1 °C
CMC for the field of measured quantity: Temperature
Comparison with a standard KEXT 3/036/03
8 Electronic and hair hygrometers Calibration – humidity
(18 to 25) °C — 20 % RH to 50 % RH · CMC 1,5 % RH (18 to 25) °C — 50 % RH to 70 % RH · CMC 1,6 % RH (18 to 25) °C — 70 % RH to 90 % RH · CMC 1,7 % RH
CMC for the field of measured quantity: Humidity
Comparison with a standard MKEV 3/039/07
9 Electronic hygrometers, electronic hygrometers incl. the loop (chain) Calibration – humidity
(18 to 25) °C — 20 % RH to 30 % RH · CMC 1,6 % RH (18 to 25) °C — 30 % RH to 70 % RH · CMC 1,7 % RH (18 to 25) °C — 70 % RH to 90 % RH · CMC 1,8 % RH
CMC for the field of measured quantity: Humidity
Comparison with a standard KEXV 3/041/12

Reference standards (10)

Euramet cg-19
KEXT 3/030/00
KEXT 3/036/03
KEXV 3/041/12
MKET 3/003/98
MKEV 3/039/07
MKO 3/001/98
MKOT 3/027/12
MKST 3/002/98
ČSN EN ISO 8655-6