🇨🇿 MEROS, spol. s r.o. PL

ČIA 2249 • Zubří, Czech Republic • 70 calibration items

Metrology & calibration

MEROS, spol. s r.o. is a calibration laboratory accredited by ČIA under number 2249 (Zubří, Czech Republic). Its accreditation scope covers 70 calibration items in the following areas: Metrology & calibration. Calibrated items include: Analogue, digital and proximity revolution counters, induction revo…, DC resistance / Multimeters, resistors, ohmmeters, resistance boxes…, Frequency / Frequency meters, counters, frequency analyzers, functi…, AC current / Power supplies, calibrators, inspection instruments. Most frequently used calibration principles: Direct generation with a calibrator, Direct measurement using a multimeter, Direct measurement by a counter.

At a glance

  • Accreditation number: ČIA 2249
  • Calibration items in scope: 70
  • Types of calibrated items: 50
  • Calibration principles used: 48
  • City: Zubří, Czech Republic
  • Valid until: 08-07-2029
Accreditation number
2249
Accreditation body
ČIA
Address
Starozuberská 1453, 756 54 Zubří
City
Zubří, Czech Republic
Division
Laboratorium wzorcujące MEROS
Valid until
08-07-2029

What this laboratory calibrates (50)

Calibration principles (48)

Certificate information

Certificate date
08.07.2024

Accreditation scope — calibration items (70)

CMC — lowest stated expanded measurement uncertainty (ILAC-P14, EA-4/02 M); if not stated otherwise, coverage probability approx. 95 % and values without a unit are relative to the measured value. The uncertainty of a specific calibration may be higher.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Parallel gauge blocks Calibration – length
0,5 mm to 100 mm · CMC (1 L + 0,1) µm
CMC for the field of measured quantity: Length
Mechanical comparison with a standard using a comparator MKGD KM.2
2 Slide gauges Calibration – length
0 mm to 500 mm · CMC (8 L + 10) µm
CMC for the field of measured quantity: Length
Comparison with parallel gauge blocks and rings MKGD PM.2
3 Micrometers Calibration – length
0 mm to 100 mm · CMC (1 L + 2,5) µm
CMC for the field of measured quantity: Length
Comparison with parallel gauge blocks MKGD MM.2
4 Deviation meters Calibration – length
division 0.01 mm — 0 mm to 30 mm · CMC (1 L + 2,5) µm division 0.001 mm — 0 mm to 30 mm · CMC (1 L + 1) µm
CMC for the field of measured quantity: Length
Direct measurement on a calibration instrument for indicators MKGD UM.2
5 Cylindrical gauges Calibration – length
0,5 mm to 50 mm · CMC (1 L + 2,5) µm
CMC for the field of measured quantity: Length
Direct measurement by a passameter MKGD VK.1
6 Feeler gauges Calibration – length
0,02 mm to 2 mm · CMC (1 L + 2,5) µm
CMC for the field of measured quantity: Length
Direct measurement by a micrometer MKGD LS.1
7 Steel tape measures Calibration – length
0 mm to 3000 mm · CMC (20 L + 100) µm 3000 mm to 5000 mm · CMC (50 L + 200) µm
CMC for the field of measured quantity: Length
Comparison with a steel rule MKGD SM.1
8 Steel rules Calibration – length
0 mm to 1000 mm · CMC (20 L + 100) µm
CMC for the field of measured quantity: Length
Comparison with a steel rule MKGD SM.1
9 Tape measures Calibration – length
0 mm to 5000 mm · CMC 0,4 mm 5000 mm to 20000 mm · CMC 0,6 mm 20000 mm to 30000 mm · CMC 0,8 mm
CMC for the field of measured quantity: Length
Comparison with a tape measure MKGD SM.1
10 Analogue, digital and proximity revolution counters, induction revolution counters, rpm sensors, stroboscopes Calibration – rotational speed
1 min⁻¹ to 100 000 min⁻¹ · CMC 1,0·10⁻⁶
CMC for the field of measured quantity: Rotational speed
Simulation by AC voltage MKF-O1
11 Analogue, digital and proximity revolution counters, induction revolution counters, rpm sensors, stroboscopes Calibration – rotational speed
1 min⁻¹ to 60 min⁻¹ · CMC 0,00020 min⁻¹ 60 min⁻¹ to 100 000 min⁻¹ · CMC 1,5·10⁻⁶
CMC for the field of measured quantity: Rotational speed
Simulation by an optical transducer MKF-O1
12 Analogue, digital and proximity revolution counters, induction revolution counters, rpm sensors, stroboscopes Calibration – rotational speed
1 min⁻¹ to 5 min⁻¹ · CMC 0,25 % 5 min⁻¹ to 20 min⁻¹ · CMC 0,070 % 20 min⁻¹ to 10 000 min⁻¹ · CMC 0,0060 %
CMC for the field of measured quantity: Rotational speed
Direct generation by a speed standard – contact and non-contact method MKF-O1
13 Analogue, digital and proximity revolution counters, induction revolution counters, rpm sensors, stroboscopes Calibration – rotational speed
measurement time at least 4 min — 1 min⁻¹ to 60 min⁻¹ · CMC 0,11 %
CMC for the field of measured quantity: Rotational speed
Comparison with a reference stopwatch MKF-O1
14 Deformation and digital manometers, pressure transducers, differential pressure measuring chains Calibration – pressure, mechanical stress
Gauge pressure, gas — 0 kPa to 1 kPa · CMC 0,1 % + 0,0015 kPa Gauge pressure, gas — 1 kPa to 10 kPa · CMC 0,04 % + 0,004 kPa Gauge pressure, gas — 10 kPa to 200 kPa · CMC 0,05 % + 0,03 kPa Gauge pressure, gas — 200 kPa to 600 kPa · CMC 0,06 % + 0,05 kPa Gauge pressure, gas — 600 kPa to 2 000 kPa · CMC 0,07 % + 0,11 kPa Vacuum, gas — 0 kPa to 1 kPa · CMC 0,1 % + 1,5 Pa Vacuum, gas — 1 kPa to 90 kPa · CMC 0,05 % + 13 Pa Gauge pressure, liquid — 0 MPa to 2 MPa · CMC 0,05 % + 0,3 kPa Gauge pressure, liquid — 2 MPa to 10 MPa · CMC 0,06 % + 1 kPa Gauge pressure, liquid — 10 MPa to 25 MPa · CMC 0,05 % + 5 kPa Gauge pressure, liquid — 25 MPa to 60 MPa · CMC 0,06 % + 9 kPa Absolute pressure, gas — 5 kPa to 2 000 kPa · CMC 0,05 % + 0,3 kPa Absolute pressure, liquid — 2000 kPa to 10000 kPa · CMC 0,07 % + 1,2 kPa Absolute pressure, liquid — 10000 kPa to 25000 kPa · CMC 0,06 % + 4 kPa Barometric pressure — 500 hPa to 1100 hPa · CMC 0,4 hPa
CMC for the field of measured quantity: Pressure, mechanical stress
Comparison with a digital manometer MKMP TL.2
15 Resistance thermometers Calibration – temperature
-80 °C to -30 °C · CMC 0,12 °C -30 °C to 0 °C · CMC 0,08 °C 0 °C · CMC 0,05 °C 0 °C to 150 °C · CMC 0,07 °C 150 °C to 260 °C · CMC 0,08 °C
CMC for the field of measured quantity: Temperature
Comparison with a resistance temperature sensor in a liquid bath MKTT OT.2
16 Resistance thermometers Calibration – temperature
260 °C to 420 °C · CMC 0,4 °C 420 °C to 660 °C · CMC 0,65 °C
CMC for the field of measured quantity: Temperature
Comparison with a resistance temperature sensor in a calibrating oven MKTT OT.2
17 Thermocouple temperature sensors Calibration – temperature
-80 °C to 420 °C · CMC 0,5 °C 420 °C to 660 °C · CMC 0,7 °C
CMC for the field of measured quantity: Temperature
Comparison with a resistance temperature sensor in a calibrating oven MKTT TE.2
18 Thermocouple temperature sensors Calibration – temperature
660 °C to 1 100 °C · CMC 1,5 °C 1 100 °C to 1 300 °C · CMC 2,0 °C
CMC for the field of measured quantity: Temperature
Comparison with a thermoelectric temperature sensor in a horizontal furnace MKTT TE.2
19 Electronic thermometers Calibration – temperature
-80 °C to -20 °C · CMC 0,12 °C -20 °C to 0 °C · CMC 0,08 °C 0 °C · CMC 0,05 °C 0 °C to 150 °C · CMC 0,07 °C 150 °C to 260 °C · CMC 0,08 °C
CMC for the field of measured quantity: Temperature
Comparison with a resistance temperature sensor in a liquid bath MKTT ET.2
20 Electronic thermometers Calibration – temperature
260 °C to 420 °C · CMC 0,4 °C 420 °C to 660 °C · CMC 0,65 °C
CMC for the field of measured quantity: Temperature
Comparison with a resistance temperature sensor in a calibrating oven MKTT ET.2
21 Electronic thermometers Calibration – temperature
660 °C to 1 000 °C · CMC 1,4 °C 1 000 °C to 1 100 °C · CMC 1,5 °C 1 100 °C to 1 300 °C · CMC 2,0 °C
CMC for the field of measured quantity: Temperature
Comparison with a thermoelectric temperature sensor in a horizontal furnace MKTT ET.2
22 Glass thermometers Calibration – temperature
-40 °C to -20 °C · CMC 0,12 °C -20 °C to 200 °C · CMC 0,07 °C 200 °C to 300 °C · CMC 0,15 °C
CMC for the field of measured quantity: Temperature
Comparison with a resistance temperature sensor in a liquid bath MKTT ST.2
23 Non-contact thermometers Calibration – temperature
-20 °C to 300 °C · CMC 1,2 °C 300 °C to 1100 °C · CMC 0,6 %
CMC for the field of measured quantity: Temperature
Comparison with a standard black body MKTT PR.1
24 Temperature measuring chains with thermocouple type B, C, D, E, J, K, N, R, T, S, resistance thermometers, voltage and current output of transducers Calibration – temperature
-200 °C to -100 °C · CMC 0,1 % -100 °C to 100 °C · CMC 0,1 °C 100 °C to 1800 °C · CMC 0,1 %
CMC for the field of measured quantity: Temperature
Comparison with a calibrator or electrical measurement with a multimeter MKTT TX.1
25 Relative humidity / Humidity meters Calibration – air humidity
(18 to 28) °C — 5 % RH to 10 % RH · CMC 2,4 % RH (18 to 28) °C — 10 % RH to 30 % RH · CMC 1,2 % RH (18 to 28) °C — 30 % RH to 70 % RH · CMC 1,3 % RH (18 to 28) °C — 70 % RH to 95 % RH · CMC 1,4 % RH (10 to 18) and (28 to 40) °C — 5 % RH to 50 % RH · CMC 1,8 % RH (10 to 18) and (28 to 40) °C — 50 % RH to 95 % RH · CMC 2,4 % RH (40 to 70) °C — 5 % RH to 50 % RH · CMC 2,8 % RH (40 to 70) °C — 50 % RH to 95 % RH · CMC 3,6 % RH
CMC for the field of measured quantity: Air humidity
Comparison with a humidity meter in a climatic chamber MKRV.1
26 DC voltage / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
0 mV to 1 mV · CMC 0,50 µV 1 mV to 10 mV · CMC 0,032 % 10 mV to 100 mV · CMC 0,0040 % 100 mV to 1 V · CMC 0,0008 % 1 V to 1,9 V · CMC 0,0005 % 1,9 V to 7 V · CMC 0,0007 % 7 V to 19 V · CMC 0,0005 % 19 V to 50 V · CMC 0,0008 % 50 V to 190 V · CMC 0,0007 % 190 V to 1000 V · CMC 0,0009 % 1000 V · CMC 0,0007 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement using a multimeter MKE-USS1, MKE-SRP.1
27 DC voltage / voltmeters, multimeters and inspection equipment Calibration – electrical quantities
1 mV to 10 mV · CMC 0,060 % 10 mV to 100 mV · CMC 0,0065 % 100 mV to 220 mV · CMC 0,0015 % 220 mV to 2,2 V · CMC 0,0012 % 2,2 V to 22 V · CMC 0,0007 % 22 V to 1000 V · CMC 0,0010 %
CMC for the field of measured quantity: Electrical quantities
Direct generation with a calibrator MKE-USS1, MKE-SRP.1
28 AC voltage / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
10 Hz to 10 kHz — 2 mV to 10 mV · CMC 0,40 % 10 kHz to 30 kHz — 2 mV to 10 mV · CMC 0,80 % 30 kHz to 100 kHz — 2 mV to 10 mV · CMC 1,7 % 10 Hz to 100 Hz — 10 mV to 20 mV · CMC 0,080 % 100 Hz to 2 kHz — 10 mV to 20 mV · CMC 0,065 % 2 kHz to 10 kHz — 10 mV to 20 mV · CMC 0,080 % 10 kHz to 30 kHz — 10 mV to 20 mV · CMC 0,14 % 30 kHz to 100 kHz — 10 mV to 20 mV · CMC 0,32 % 10 Hz to 100 Hz — 20 mV to 50 mV · CMC 0,050 % 100 Hz to 2 kHz — 20 mV to 50 mV · CMC 0,045 % 2 kHz to 10 kHz — 20 mV to 50 mV · CMC 0,050 % 10 kHz to 30 kHz — 20 mV to 50 mV · CMC 0,090 % 30 kHz to 100 kHz — 20 mV to 50 mV · CMC 0,20 % 10 Hz to 100 Hz — 50 mV to 100 mV · CMC 0,035 % 100 Hz to 10 kHz — 50 mV to 100 mV · CMC 0,030 % 10k Hz to 30 kHz — 50 mV to 100 mV · CMC 0,060 % 30 kHz to 100 kHz — 50 mV to 100 mV · CMC 0,13 % 10 Hz to 100 Hz — 100 mV to 190 mV · CMC 0,030 % 100 Hz to 10 kHz — 100 mV to 190 mV · CMC 0,025 % 10 kHz to 30 kHz — 100 mV to 190 mV · CMC 0,050 % 30 kHz to 100 kHz — 100 mV to 190 mV · CMC 0,11 % 10 Hz to 40 Hz — 190 mV to 500 mV · CMC 0,026 % 40 Hz to 100 Hz — 190 mV to 500 mV · CMC 0,023 % 100 Hz to 2 kHz — 190 mV to 500 mV · CMC 0,020 % 2 kHz to 10 kHz — 190 mV to 500 mV · CMC 0,022 % 10 kHz to 30 kHz — 190 mV to 500 mV · CMC 0,047 % 30 kHz to 100 kHz — 190 mV to 500 mV · CMC 0,17 % 100 kHz to 300 kHz — 190 mV to 500 mV · CMC 1,4 % 10 Hz to 40 Hz — 500 mV to 1 V · CMC 0,019 % 40 Hz to 100 Hz — 500 mV to 1 V · CMC 0,015 % 100 Hz to 2 kHz — 500 mV to 1 V · CMC 0,013 % 2 kHz to 10 kHz — 500 mV to 1 V · CMC 0,014 % 10 kHz to 30 kHz — 500 mV to 1 V · CMC 0,030 % 30 kHz to 100 kHz — 500 mV to 1 V · CMC 0,10 % 100 kHz to 300 kHz — 500 mV to 1 V · CMC 0,75 % 10 Hz to 40 Hz — 1 V to 1,9 V · CMC 0,016 % 40 Hz to 100 Hz — 1 V to 1,9 V · CMC 0,013 % 100 Hz to 2 kHz — 1 V to 1,9 V · CMC 0,010 % 2 kHz to 10 kHz — 1 V to 1,9 V · CMC 0,012 % 10 kHz to 30 kHz — 1 V to 1,9 V · CMC 0,026 % 30 kHz to 100 kHz — 1 V to 1,9 V · CMC 0,082 % 100 kHz to 300 kHz — 1 V to 1,9 V · CMC 0,55 % 10 Hz to 40 Hz — 1,9 V to 5 V · CMC 0,026 % 40 Hz to 100 Hz — 1,9 V to 5 V · CMC 0,022 % 100 Hz to 2 kHz — 1,9 V to 5 V · CMC 0,019 % 2 kHz to 10 kHz — 1,9 V to 5 V · CMC 0,022 % 10 kHz to 30 kHz — 1,9 V to 5 V · CMC 0,045 % 30 kHz to 100 kHz — 1,9 V to 5 V · CMC 0,17 % 100 kHz to 300 kHz — 1,9 V to 5 V · CMC 1,4 % 10 Hz to 40 Hz — 5 V to 10 V · CMC 0,018 % 40 Hz to 100 Hz — 5 V to 10 V · CMC 0,015 % 100 Hz to 2 kHz — 5 V to 10 V · CMC 0,013 % 2 kHz to 10 kHz — 5 V to 10 V · CMC 0,017 % 10 kHz to 30 kHz — 5 V to 10 V · CMC 0,032 % 30 kHz to 100 kHz — 5 V to 10 V · CMC 0,11 % 100 kHz to 300 kHz — 5 V to 10 V · CMC 0,80 % 10 Hz to 40 Hz — 10 V to 19 V · CMC 0,016 % 40 Hz to 100 Hz — 10 V to 19 V · CMC 0,013 % 100 Hz to 2 kHz — 10 V to 19 V · CMC 0,010 % 2 kHz to 10 kHz — 10 V to 19 V · CMC 0,013 % 10 kHz to 30 kHz — 10 V to 19 V · CMC 0,027 % 30 kHz to 100 kHz — 10 V to 19 V · CMC 0,080 % 100 kHz to 300 kHz — 10 V to 19 V · CMC 0,55 % 10 Hz to 40 Hz — 19 V to 50 V · CMC 0,026 % 40 Hz to 100 Hz — 19 V to 50 V · CMC 0,022 % 100 Hz to 2 kHz — 19 V to 50 V · CMC 0,020 % 2 kHz to 10 kHz — 19 V to 50 V · CMC 0,021 % 10 kHz to 30 kHz — 19 V to 50 V · CMC 0,045 % 30 kHz to 100 kHz — 19 V to 50 V · CMC 0,16 % 10 Hz to 40 Hz — 50 V to 100 V · CMC 0,019 % 40 Hz to 100 Hz — 50 V to 100 V · CMC 0,015 % 100 Hz to 2 kHz — 50 V to 100 V · CMC 0,013 % 2 kHz to 10 kHz — 50 V to 100 V · CMC 0,015 % 10 kHz to 30 kHz — 50 V to 100 V · CMC 0,031 % 30 kHz to 100 kHz — 50 V to 100 V · CMC 0,11 % 10 Hz to 40 Hz — 100 V to 190 V · CMC 0,017 % 40 Hz to 100 Hz — 100 V to 190 V · CMC 0,013 % 100 Hz to 2 kHz — 100 V to 190 V · CMC 0,011 % 2 kHz to 10 kHz — 100 V to 190 V · CMC 0,012 % 10 kHz to 30 kHz — 100 V to 190 V · CMC 0,026 % 30 kHz to 100 kHz — 100 V to 190 V · CMC 0,078 % 40 Hz to 10 kHz — 190 V to 500 V · CMC 0,032 % 10 kHz to 30 kHz — 190 V to 500 V · CMC 0,070 % 40 Hz to 10 kHz — 500 V · CMC 0,023 % 10 kHz to 30 kHz — 500 V · CMC 0,055 % 40 Hz to 10 kHz — 500 V to 1000 V · CMC 0,040 % 10 kHz to 30 kHz — 500 V to 1000 V · CMC 0,16 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement using a multimeter MKE-UST.1, MKE-SRP.1
29 AC voltage / Voltmeters, multimeters, inspection equipment Calibration – electrical quantities
10 Hz to 50 kHz — 1 mV to 2,2 mV · CMC 0,60 % 50 kHz to 100 kHz — 1 mV to 2,2 mV · CMC 0,80 % 10 Hz to 100 kHz — 2,2 mV to 10 mV · CMC 0,32 % 10 Hz to 40 Hz — 10 mV to 22 mV · CMC 0,090 % 40 Hz to 20 kHz — 10 mV to 22 mV · CMC 0,067 % 20 kHz to 50 kHz — 10 mV to 22 mV · CMC 0,082 % 50 kHz to 100 kHz — 10 mV to 22 mV · CMC 0,13 % 10 Hz to 40 Hz — 22 mV to 100 mV · CMC 0,095 % 40 Hz to 20 kHz — 22 mV to 100 mV · CMC 0,050 % 20 kHz to 50 kHz — 22 mV to 100 mV · CMC 0,066 % 50 kHz to 100 kHz — 22 mV to 100 mV · CMC 0,15 % 100 kHz to 300 kHz — 22 mV to 100 mV · CMC 0,23 % 10 Hz to 40 Hz — 100 mV to 220 mV · CMC 0,048 % 40 Hz to 20 kHz — 100 mV to 220 mV · CMC 0,019 % 20 kHz to 50 kHz — 100 mV to 220 mV · CMC 0,033 % 50 kHz to 100 kHz — 100 mV to 220 mV · CMC 0,076 % 100 kHz to 300 kHz — 100 mV to 220 mV · CMC 0,14 % 10 Hz to 40 Hz — 220 mV to 1 V · CMC 0,053 % 40 Hz to 20 kHz — 220 mV to 1 V · CMC 0,011 % 20 kHz to 50 kHz — 220 mV to 1 V · CMC 0,016 % 50 kHz to 100 kHz — 220 mV to 1 V · CMC 0,033 % 100 kHz to 300 kHz — 220 mV to 1 V · CMC 0,093 % 10 Hz to 40 Hz — 1 V to 2,2 V · CMC 0,039 % 40 Hz to 20 kHz — 1 V to 2,2 V · CMC 0,008 % 20 kHz to 50 kHz — 1 V to 2,2 V · CMC 0,012 % 50 kHz to 100 kHz — 1 V to 2,2 V · CMC 0,019 % 100 kHz to 300 kHz — 1 V to 2,2 V · CMC 0,070 % 10 Hz to 40 Hz — 2,2 V to 22 V · CMC 0,060 % 40 Hz to 20 kHz — 2,2 V to 22 V · CMC 0,008 % 20 kHz to 50 kHz — 2,2 V to 22 V · CMC 0,015 % 50 kHz to 100 kHz — 2,2 V to 22 V · CMC 0,023 % 100 kHz to 300 kHz — 2,2 V to 22 V · CMC 0,065 % 10 Hz to 40 Hz — 22 V to 220 V · CMC 0,055 % 40 Hz to 20 kHz — 22 V to 220 V · CMC 0,010 % 20 kHz to 50 kHz — 22 V to 220 V · CMC 0,015 % 50 kHz to 100 kHz — 22 V to 220 V · CMC 0,031 % 50 Hz to 1 kHz — 220 V to 1000 V · CMC 0,011 %
CMC for the field of measured quantity: Electrical quantities
Direct generation with a calibrator MKE-UST1, MKE-SRP.1
30 Peak-to-peak value of square wave AC voltage / Oscilloscopes Calibration – electrical quantities
10 Hz to 100 kHz — 1 mV to 2 mV · CMC 0,30 % + 10 µV 10 Hz to 100 kHz — 2 mV to 100 mV · CMC 0,20 % + 10 µV 10 Hz to 100 kHz — 100 mV to 220 V · CMC 0,15 %
CMC for the field of measured quantity: Electrical quantities
Direct generation with a calibrator MKE-UST1
31 DC current / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
1 nA to 2 nA · CMC 0,40 % 2 nA to 50 nA · CMC 0,30 % 50 nA to 200 nA · CMC 0,20 % 200 nA to 1 µA · CMC 0,25 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with a picoammeter MKE-ISS1, MKE-SRP.1
32 DC current / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
1 µA to 20 µA · CMC 0,011 % 20 µA to 20 mA · CMC 0,0040 % 20 mA to 0,1 A · CMC 0,011 % 0,1 A to 1 A · CMC 0,0040 % 1 A to 2 A · CMC 0,0060 % 2 A to 60 A · CMC 0,011 % 60 A to 200 A · CMC 0,015 % 200 A to 600 A · CMC 0,040 % 600 A to 2000 A · CMC 0,10 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with a multimeter or indirect measurement with a shunt multimeter MKE-ISS1, MKE-SRP.1
33 DC current / Ammeters, multimeters, clamp meters, inspection instruments Calibration – electrical quantities
1 nA to 2 nA · CMC 0,40 % 2 nA to 50 nA · CMC 0,30 % 50 nA to 200 nA · CMC 0,20 % 200 nA to 1 µA · CMC 0,25 %
CMC for the field of measured quantity: Electrical quantities
Direct comparison with a picoammeter MKE-ISS1, MKE-SRP.1
34 DC current / Ammeters, multimeters, clamp meters, inspection instruments Calibration – electrical quantities
1 µA to 20 µA · CMC 0,011 % 20 µA to 20 mA · CMC 0,0040 % 20 mA to 0,1 A · CMC 0,011 % 0,1 A to 1 A · CMC 0,0040 % 1 A to 2 A · CMC 0,0060 % 2 A to 60 A · CMC 0,011 % 60 A to 200 A · CMC 0,015 % 200 A to 600 A · CMC 0,040 % 600 A to 2 000 A · CMC 0,10 %
CMC for the field of measured quantity: Electrical quantities
Direct comparison using a multimeter or indirect comparison using a shunt multimeter MKE-ISS1, MKE-SRP.1
35 DC current / Ammeters, multimeters, clamp meters, inspection instruments Calibration – electrical quantities
22 µA to 100 µA · CMC 0,040 % 100 µA to 2,2 mA · CMC 0,0090 % 2,2 mA to 22 mA · CMC 0,0070 % 22 mA to 220 mA · CMC 0,0090 % 220 mA to 2,2 A · CMC 0,016 %
CMC for the field of measured quantity: Electrical quantities
Direct generation with a calibrator MKE-ISS1, MKE-SRP.1
36 DC current / Clamp meters, inspection instruments Calibration – electrical quantities
200 A to 2000 A · CMC 0,2 %
CMC for the field of measured quantity: Electrical quantities
Indirect measurement with a current coil MKE-ISS1, MKE-SRP.1
37 AC current / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
40 Hz to 1 kHz — 10 µA to 200 µA · CMC 0,050 % + 22 nA
CMC for the field of measured quantity: Electrical quantities
Direct measurement using a multimeter MKE-IST1
38 AC current / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
40 Hz to 1 kHz — 200 µA to 2 mA · CMC 0,035 % + 0,22 µA 40 Hz to 1 kHz — 2 mA to 20 mA · CMC 0,038 % + 2,2 µA 40 Hz to 1 kHz — 20 mA to 200 mA · CMC 0,037 % + 20 µA 40 Hz to 1 kHz — 200 mA to 2 A · CMC 0,082 % + 0,2 mA 40 Hz to 1 kHz — 2 A to 20 A · CMC 0,12 % + 2 mA
CMC for the field of measured quantity: Electrical quantities
Direct measurement using a multimeter MKE-IST1, MKE-SRP.1
39 AC current / Power supplies, calibrators, inspection instruments Calibration – electrical quantities
50 to 60 Hz — 200 mA to 200 A · CMC 0,10 % 50 Hz — 200 A to 2 500 A · CMC 0,20 %
CMC for the field of measured quantity: Electrical quantities
Indirect measurement using a current transformer and multimeter MKE-IST1, MKE-SRP.1
40 AC current / Ammeters, multimeters, inspection instruments Calibration – electrical quantities
40 Hz to 1 kHz — 22 µA to 100 µA · CMC 0,060 % 40 Hz to 1 kHz — 100 µA to 5 mA · CMC 0,035 % 40 Hz to 1 kHz — 5 mA to 22 mA · CMC 0,025 % 40 Hz to 1 kHz — 22 mA to 220 mA · CMC 0,028 % 40 Hz to 1 kHz — 220 mA to 2,2 A · CMC 0,050 % 40 Hz to 1 kHz — 2,2 A to 20 A · CMC 0,06 % + 1 mA
CMC for the field of measured quantity: Electrical quantities
Direct generation with a calibrator MKE-IST1, MKE-SRP.1
41 DC power / Wattmeters, network analyzers, inspection instruments (30 V to 500 V, 0.02 A to 200 A) Calibration – electrical quantities
1 W to 105 kW · CMC 0,060 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with multimeters or with a shunt and multimeter MKE-W1
42 AC power / Power supplies, power calibrators (30 V to 500 V, 0.02 A to 200 A, 50 Hz to 60 Hz, cos φ 0.5 to 1) Calibration – electrical quantities
1 W to 105 kW · CMC 0,060 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with an energy meter MKE-W1
43 AC power / Wattmeters, network analyzers, inspection instruments (30 V to 500 V, 0.02 A to 200 A, 50 Hz to 60 Hz, cos φ 0.5 to 1) Calibration – electrical quantities
1 W to 105 kW · CMC 0,060 %
CMC for the field of measured quantity: Electrical quantities
Direct comparison with an energy meter MKE-W1
44 DC resistance / Multimeters, resistors, ohmmeters, resistance boxes, calibrators, inspection instruments Calibration – electrical quantities
0,1 mΩ · CMC 0,0080 % 1 mΩ · CMC 0,0020 % 0,01 Ω · CMC 0,0025 % 0,1 Ω · CMC 0,0025 % 1 Ω · CMC 0,0020 % 1,9 Ω · CMC 0,010 % 10 Ω · CMC 0,0026 % 19 Ω · CMC 0,0026 % 100 Ω · CMC 0,0014 % 190 Ω · CMC 0,0014 % 1 kΩ · CMC 0,0011 % 1,9 kΩ · CMC 0,0011 % 10 kΩ · CMC 0,0011 % 19 kΩ · CMC 0,0011 % 100 kΩ · CMC 0,0015 % 190 kΩ · CMC 0,0015 % 1 MΩ · CMC 0,0026 % 1,9 MΩ · CMC 0,0026 % 10 MΩ · CMC 0,0050 % 19 MΩ · CMC 0,0050 % 100 MΩ · CMC 0,0050 % 1 GΩ · CMC 0,010 %
CMC for the field of measured quantity: Electrical quantities
Direct generation by resistance standards MKE-RSS1, MKE-SRP.1
45 DC resistance / Multimeters, resistors, ohmmeters, resistance boxes, calibrators, inspection instruments Calibration – electrical quantities
0,1 mΩ to 0,2 mΩ · CMC 0,040 % 0,2 mΩ to 0,5 mΩ · CMC 0,0080 % 0,5 mΩ to 0,9 mΩ · CMC 0,0060 % 0,9 mΩ to 5 Ω · CMC 0,0040 %
CMC for the field of measured quantity: Electrical quantities
Indirect measurement using a shunt and multimeter MKE-RSS1, MKE-SRP.1
46 DC resistance / Multimeters, resistors, ohmmeters, resistance boxes, calibrators, inspection instruments Calibration – electrical quantities
5 Ω to 200 kΩ · CMC 0,0030 % 200 kΩ to 2 MΩ · CMC 0,0040 % 2 MΩ to 20 MΩ · CMC 0,0060 % 20 MΩ to 200 MΩ · CMC 0,040 % 200 MΩ to 300 MΩ · CMC 0,070 % 300 MΩ to 500 MΩ · CMC 0,050 % 500 MΩ to 1 GΩ · CMC 0,035 % 1 GΩ to 2 GΩ · CMC 0,020 % 2 GΩ to 10 GΩ · CMC 0,65 % 10 GΩ to 20 GΩ · CMC 0,20 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement using a multimeter MKE-RSS1, MKE-SRP.1
47 DC resistance / Multimeters, resistors, ohmmeters, resistance boxes, calibrators, inspection instruments Calibration – electrical quantities
up to 10000 V — 20 GΩ to 250 GΩ · CMC 0,50 % up to 10000 V — 250 GΩ to 1 TΩ · CMC 0,70 %
CMC for the field of measured quantity: Electrical quantities
Indirect measurement using a picoammeter and HV voltmeter MKE-RSS, MKE-SRP.1
48 Impedance module / Resistance standards, resistance boxes, resistance meters, clamp meters, inspection meters Calibration – electrical quantities
50 Hz to 60 Hz — 0,1 mΩ to 0,1 Ω · CMC 0,2 % 50 Hz to 1 kHz — 0,1 Ω to 10 Ω · CMC 0,2 %
CMC for the field of measured quantity: Electrical quantities
VA method MKE-Z2, MKE-SRP.1
49 Impedance module, AC resistance / standards, resistance meters, RLC bridges Calibration – electrical quantities
20 Hz to 50 Hz — 100 mΩ · CMC 0,30 % 50 Hz to 10 kHz — 100 mΩ · CMC 0,14 % 20 Hz to 100 Hz — 1 Ω · CMC 0,035 % 100 Hz — 1 Ω · CMC 0,015 % 100 Hz to 1 kHz — 1 Ω · CMC 0,020 % 1 kHz — 1 Ω · CMC 0,015 % 1 kHz to 10 kHz — 1 Ω · CMC 0,025 % 10 kHz — 1 Ω · CMC 0,020 % 10 kHz to 100 kHz — 1 Ω · CMC 0,090 % 20 Hz to 100 Hz — 10 Ω · CMC 0,016 % 100 Hz — 10 Ω · CMC 0,011 % 100 Hz to 1 kHz — 10 Ω · CMC 0,016 % 1 kHz — 10 Ω · CMC 0,011 % 1 kHz to 10 kHz — 10 Ω · CMC 0,016 % 10 kHz — 10 Ω · CMC 0,011 % 10 kHz to 100 kHz — 10 Ω · CMC 0,045 % 100 kHz — 10 Ω · CMC 0,040 % 100 kHz to 1 MHz — 10 Ω · CMC 0,15 % 20 Hz to 100 Hz — 100 Ω · CMC 0,016 % 100 Hz — 100 Ω · CMC 0,011 % 100 Hz to 1 kHz — 100 Ω · CMC 0,016 % 1 kHz — 100 Ω · CMC 0,011 % 1 kHz to 10 kHz — 100 Ω · CMC 0,016 % 10 kHz — 100 Ω · CMC 0,011 % 10 kHz to 100 kHz — 100 Ω · CMC 0,040 % 100 kHz — 100 Ω · CMC 0,035 % 100 kHz to 1 MHz — 100 Ω · CMC 0,15 % 20 Hz to 100 Hz — 1 kΩ · CMC 0,016 % 100 Hz — 1 kΩ · CMC 0,011 % 100 Hz to 1 kHz — 1 kΩ · CMC 0,016 % 1 kHz — 1 kΩ · CMC 0,011 % 1 kHz to 10 kHz — 1 kΩ · CMC 0,016 % 10 kHz — 1 kΩ · CMC 0,011 % 10 kHz to 100 kHz — 1 kΩ · CMC 0,040 % 100 kHz — 1 kΩ · CMC 0,035 % 100 kHz to 1 MHz — 1 kΩ · CMC 0,15 % 20 Hz to 100 Hz — 10 kΩ · CMC 0,016 % 100 Hz — 10 kΩ · CMC 0,011 % 100 Hz to 1 kHz — 10 kΩ · CMC 0,016 % 1 kHz — 10 kΩ · CMC 0,011 % 1 kHz to 10 kHz — 10 kΩ · CMC 0,016 % 10 kHz — 10 kΩ · CMC 0,011 % 10 kHz to 100 kHz — 10 kΩ · CMC 0,040 % 100 kHz — 10 kΩ · CMC 0,035 % 100 kHz to 1 MHz — 10 kΩ · CMC 0,15 % 20 Hz to 100 Hz — 100 kΩ · CMC 0,020 % 100 Hz — 100 kΩ · CMC 0,011 % 100 Hz to 1 kHz — 100 kΩ · CMC 0,016 % 1 kHz — 100 kΩ · CMC 0,011 % 1 kHz to 10 kHz — 100 kΩ · CMC 0,020 % 10 kHz — 100 kΩ · CMC 0,015 % 10 kHz to 100 kHz — 100 kΩ · CMC 0,20 % 20 Hz to 100 Hz — 1 MΩ · CMC 0,035 % 100 Hz — 1 MΩ · CMC 0,018 % 100 Hz to 1 kHz — 1 MΩ · CMC 0,025 % 1 kHz — 1 MΩ · CMC 0,018 % 1 kHz to 10 kHz — 1 MΩ · CMC 0,060 % 10 kHz — 1 MΩ · CMC 0,050 % 10 kHz to 100 kHz — 1 MΩ · CMC 0,50 % 20 Hz to 100 Hz — 10 MΩ · CMC 0,10 % 100 Hz — 10 MΩ · CMC 0,050 % 100 Hz to 1 kHz — 10 MΩ · CMC 0,055 % 1 kHz — 10 MΩ · CMC 0,050 % 1 kHz to 10 kHz — 10 MΩ · CMC 0,18 % 20 Hz to 100 Hz — 100 MΩ · CMC 0,40 % 100 Hz to 1 kHz — 100 MΩ · CMC 0,50 %
CMC for the field of measured quantity: Electrical quantities
Comparison by substitution or direct measurement of impedance standard MKE-Z2
50 Impedance module, AC resistance / Resistance standards, boxes, meters, clamp meters, inspection meters, RLC bridges Calibration – electrical quantities
20 Hz to 50 Hz — 10 mΩ to 100 mΩ · CMC 4 mΩ 50 Hz to 100 Hz — 10 mΩ to 100 mΩ · CMC 3 mΩ 100 Hz to 500 Hz — 10 mΩ to 100 mΩ · CMC 2,2 mΩ 500 Hz to 5 kHz — 10 mΩ to 100 mΩ · CMC 1,5 mΩ 5 kHz to 500 kHz — 10 mΩ to 100 mΩ · CMC 1 mΩ 500 kHz to 1 MHz — 10 mΩ to 100 mΩ · CMC 2 mΩ 20 Hz to 50 Hz — 100 mΩ to 1000 mΩ · CMC 7 mΩ 50 Hz to 100 Hz — 100 mΩ to 1000 mΩ · CMC 6 mΩ 100 Hz to 500 Hz — 100 mΩ to 1000 mΩ · CMC 4 mΩ 500 Hz to 5 kHz — 100 mΩ to 1000 mΩ · CMC 3,5 mΩ 5 kHz to 500 kHz — 100 mΩ to 1000 mΩ · CMC 3 mΩ 500 kHz to 1 MHz — 100 mΩ to 1000 mΩ · CMC 7 mΩ 20 Hz to 50 Hz — 1 Ω to 10 Ω · CMC 0,25 % + 0,007 Ω 50 Hz to 100 Hz — 1 Ω to 10 Ω · CMC 0,20 % + 0,007 Ω 100 Hz to 500 Hz — 1 Ω to 10 Ω · CMC 0,20 % + 0,003 Ω 500 Hz to 5 kHz — 1 Ω to 10 Ω · CMC 0,17 % + 0,003 Ω 5 kHz to 100 kHz — 1 Ω to 10 Ω · CMC 0,17 % + 0,002 Ω 100 kHz to 500 kHz — 1 Ω to 10 Ω · CMC 0,20 % + 0,002 Ω 500 kHz to 1 MHz — 1 Ω to 10 Ω · CMC 0,45 % + 0,002 Ω 20 Hz to 100 Hz — 10 Ω to 100 Ω · CMC 0,15 % + 0,02 Ω 100 Hz to 500 Hz — 10 Ω to 100 Ω · CMC 0,08 % + 0,02 Ω 500 Hz to 100 kHz — 10 Ω to 100 Ω · CMC 0,10 % + 0,02 Ω 100 kHz to 500 kHz — 10 Ω to 100 Ω · CMC 0,13 % + 0,02 Ω 500 kHz to 1 MHz — 10 Ω to 100 Ω · CMC 0,36 % + 0,02 Ω 20 Hz to 100 Hz — 100 Ω to 1 kΩ · CMC 0,16 % 100 Hz to 500 Hz — 100 Ω to 1 kΩ · CMC 0,10 % 500 Hz to 5 kHz — 100 Ω to 1 kΩ · CMC 0,11 % 5 kHz to 100 kHz — 100 Ω to 1 kΩ · CMC 0,12 % 100 kHz to 1 MHz — 100 Ω to 1 kΩ · CMC 0,17 % 1 MHz — 100 Ω to 1 kΩ · CMC 0,35 % 20 Hz to 100 Hz — 1 kΩ to 10 kΩ · CMC 0,16 % 100 Hz to 5 kHz — 1 kΩ to 10 kΩ · CMC 0,10 % 5 kHz to 100 kHz — 1 kΩ to 10 kΩ · CMC 0,12 % 100 kHz to 1 MHz — 1 kΩ to 10 kΩ · CMC 0,27 % 1 MHz — 1 kΩ to 10 kΩ · CMC 0,55 % 20 Hz to 100 Hz — 10 kΩ to 100 kΩ · CMC 0,17 % 100 Hz to 5 kHz — 10 kΩ to 100 kΩ · CMC 0,10 % 5 kHz to 10 kHz — 10 kΩ to 100 kΩ · CMC 0,18 % 10 kHz to 100 kHz — 10 kΩ to 100 kΩ · CMC 0,22 % 100 kHz to 1 MHz — 10 kΩ to 100 kΩ · CMC 0,33 % 1 MHz — 10 kΩ to 100 kΩ · CMC 0,65 % 20 Hz to 50 Hz — 0,1 MΩ to 1 MΩ · CMC (R/5 + 0,20) % 50 Hz to 100 Hz — 0,1 MΩ to 1 MΩ · CMC (R/6 + 0,15) % 100 Hz to 500 Hz — 0,1 MΩ to 1 MΩ · CMC (R/8 + 0,10) % 500 Hz to 1 kHz — 0,1 MΩ to 1 MΩ · CMC (R/10 + 0,10) % 1 kHz to 5 kHz — 0,1 MΩ to 1 MΩ · CMC (R/12 + 0,10) % 5 kHz to 10 kHz — 0,1 MΩ to 1 MΩ · CMC (R/12 + 0,16) % 10 kHz to 100 kHz — 0,1 MΩ to 1 MΩ · CMC (R/15 + 0,26) % 100 kHz to 1 MHz — 0,1 MΩ to 1 MΩ · CMC (R/1,6 + 0,30) % 1 MHz — 0,1 MΩ to 1 MΩ · CMC (R/0,8 + 0,60) % 50 Hz to 100 Hz — 1 MΩ to 10 MΩ · CMC (R/6 + 0,15) % 100 Hz to 500 Hz — 1 MΩ to 10 MΩ · CMC (R/8 + 0,10) % 500 Hz to 1 kHz — 1 MΩ to 10 MΩ · CMC (R/10 + 0,10) % 1 kHz to 5 kHz — 1 MΩ to 10 MΩ · CMC (R/12 + 0,10) % 5 kHz to 10 kHz — 1 MΩ to 10 MΩ · CMC (R/12 + 0,15) % 10 kHz to 100 kHz — 1 MΩ to 10 MΩ · CMC (R/15 + 0,26) % 100 kHz to 1 MHz — 1 MΩ to 10 MΩ · CMC (R/1,6 + 0,30) % 1 MHz — 1 MΩ to 10 MΩ · CMC (R/0,8 + 0,60) % 100 Hz to 500 Hz — 10 MΩ to 100 MΩ · CMC (R/8 + 0,10) % 500 Hz to 1 kHz — 10 MΩ to 100 MΩ · CMC (R/10 + 0,10) % 1 kHz to 100 kHz — 10 MΩ to 100 MΩ · CMC (R/12 + 0,10) %
CMC for the field of measured quantity: Electrical quantities
Direct measurement or comparison by substitution with a reference RLC bridge MKE-Z2
51 Capacitance / Capacitance standards, capacitance meters, RLC bridges Calibration – electrical quantities
50 Hz to 1 kHz — 10 pF · CMC 0,15 % 1 kHz — 10 pF · CMC 0,075 % 1 kHz to 10 kHz — 10 pF · CMC 0,080 % 10 kHz — 10 pF · CMC 0,040 % 10 kHz to 100 kHz — 10 pF · CMC 0,050 % 100kHz to 1 MHz — 10 pF · CMC 0,15 % 50 Hz to 100 Hz — 100 pF · CMC 0,080 % 100 Hz — 100 pF · CMC 0,050 % 100 Hz to 1 kHz — 100 pF · CMC 0,060 % 1 kHz — 100 pF · CMC 0,035 % 1 kHz to 10 kHz — 100 pF · CMC 0,040 % 10 kHz — 100 pF · CMC 0,035 % 10 kHz to 100 kHz — 100 pF · CMC 0,050 % 100 kHz — 100 pF · CMC 0,040 % 100 kHz to 1 MHz — 100 pF · CMC 0,15 % 50 Hz to 100 Hz — 1 nF · CMC 0,040 % 100 Hz — 1 nF · CMC 0,035 % 100 Hz to 1 kHz — 1 nF · CMC 0,040 % 1 kHz — 1 nF · CMC 0,030 % 1 kHz to 10 kHz — 1 nF · CMC 0,040 % 10 kHz — 1 nF · CMC 0,030 % 10 kHz to 100 kHz — 1 nF · CMC 0,050 % 100 kHz — 1 nF · CMC 0,040 % 100 kHz to 1 MHz — 1 nF · CMC 0,15 % 50 Hz to 100 Hz — 10 nF · CMC 0,050 % 100 Hz — 10 nF · CMC 0,035 % 100 Hz to 1 kHz — 10 nF · CMC 0,040 % 1 kHz — 10 nF · CMC 0,035 % 1 kHz to 10 kHz — 10 nF · CMC 0,040 % 10 kHz — 10 nF · CMC 0,035 % 10 kHz to 100 kHz — 10 nF · CMC 0,060 % 100 kHz to 1 MHz — 10 nF · CMC 0,15 % 50 Hz to 100 Hz — 100 nF · CMC 0,040 % 100 Hz — 100 nF · CMC 0,035 % 100 Hz to 1 kHz — 100 nF · CMC 0,040 % 1 kHz — 100 nF · CMC 0,035 % 1 kHz to 10 kHz — 100 nF · CMC 0,040 % 10 kHz — 100 nF · CMC 0,035 % 10 kHz to 100 kHz — 100 nF · CMC 0,060 % 100 kHz to 1 MHz — 100 nF · CMC 0,85 % 20 Hz to 100 Hz — 1 µF · CMC 0,050 % 100 Hz — 1 µF · CMC 0,035 % 100 Hz to 1 kHz — 1 µF · CMC 0,040 % 1 kHz — 1 µF · CMC 0,035 % 1 kHz to 10 kHz — 1 µF · CMC 0,040 % 10 kHz — 1 µF · CMC 0,035 % 10 kHz to 100 kHz — 1 µF · CMC 0,090 % 20 Hz to 1 kHz — 10 µF · CMC 0,15 % 1 kHz to 10 kHz — 10 µF · CMC 0,20 % 20 Hz to 100 Hz — 100 µF · CMC 0,15 % 100 Hz to 1 kHz — 100 µF · CMC 0,20 % 1 kHz to 10 kHz — 100 µF · CMC 0,40 %
CMC for the field of measured quantity: Electrical quantities
Comparison by substitution or direct measurement of capacitance standard MKE-Z2
52 Capacitance / Standards, boxes, capacitance meters, RLC bridges Calibration – electrical quantities
1 kHz to 2 kHz — 1 pF to 10 pF · CMC 0,10 % + 0,135 pF 2 kHz to 5 kHz — 1 pF to 10 pF · CMC 0,10 % + 0,062 pF 5 kHz to 10 kHz — 1 pF to 10 pF · CMC 0,10 % + 0,023 pF 10 kHz to 20 kHz — 1 pF to 10 pF · CMC 0,17 % + 0,011 pF 20 kHz to 50 kHz — 1 pF to 10 pF · CMC 0,22 % + 0,002 pF 50 kHz to 100 kHz — 1 pF to 10 pF · CMC 0,28 % + 0,010 pF 100 kHz to 500 kHz — 1 pF to 10 pF · CMC 0,40 % + 0,005 pF 500 kHz to 1 MHz — 1 pF to 10 pF · CMC 0,55 % + 0,002 pF 500 Hz to 1 kHz — 10 pF to 100 pF · CMC 0,10 % + 0,30 pF 1 kHz to 2 kHz — 10 pF to 100 pF · CMC 0,10 % + 0,15 pF 2 kHz to 5 kHz — 10 pF to 100 pF · CMC 0,10 % + 0,07 pF 5 kHz to 10 kHz — 10 pF to 100 pF · CMC 0,10 % + 0,03 pF 10 kHz to 50 kHz — 10 pF to 100 pF · CMC 0,15 % + 0,015 pF 50 kHz to 500 kHz — 10 pF to 100 pF · CMC 0,15 % + 0,030 pF 500 kHz to 1 MHz — 10 pF to 100 pF · CMC 0,40 % + 0,020 pF 100 Hz to 200 Hz — 100 pF to 1000 pF · CMC 0,10 % + 2 pF 200 Hz to 500 Hz — 100 pF to 1000 pF · CMC 0,10 % + 0,85 pF 500 Hz to 1 kHz — 100 pF to 1000 pF · CMC 0,10 % + 0,30 pF 1 kHz to 2 kHz — 100 pF to 1000 pF · CMC 0,10 % + 0,15 pF 2 kHz to 500 kHz — 100 pF to 1000 pF · CMC 0,15 % 500 kHz to 1 MH — 100 pF to 1000 pF · CMC 0,35 % 50 Hz to 100 Hz — 1 nF to 10 nF · CMC 0,15 % + 0,005 nF 100 Hz to 500 Hz — 1 nF to 10 nF · CMC 0,10 % + 0,002 pF 500 Hz to 5 kHz — 1 nF to 10 nF · CMC 0,10 % 5 kHz to 100 kHz — 1 nF to 10 nF · CMC 0,12 % 100 kHz to 200 kHz — 1 nF to 10 nF · CMC 0,13 % 200 kHz to 500 kHz — 1 nF to 10 nF · CMC 0,15 % 500 kHz to 1 MHz — 1 nF to 10 nF · CMC 0,50 % 50 Hz to 100 Hz — 10 nF to 100 nF · CMC 0,20 % 100 Hz to 5 kHz — 10 nF to 100 nF · CMC 0,10 % 5 kHz to 50 kHz — 10 nF to 100 nF · CMC 0,12 % 50 kHz to 200 kHz — 10 nF to 100 nF · CMC 0,20 % 200 kHz to 500 kHz — 10 nF to 100 nF · CMC 0,25 % 500 kHz to 1 MHz — 10 nF to 100 nF · CMC 0,55 % 50 Hz to 100 Hz — 100 nF to 1000 nF · CMC 0,17 % 100 Hz to 500 Hz — 100 nF to 1000 nF · CMC 0,10 % 500 Hz to 5 kHz — 100 nF to 1000 nF · CMC 0,12 % 5 kHz to 50 kHz — 100 nF to 1000 nF · CMC 0,20 % 50 kHz to 100 kHz — 100 nF to 1000 nF · CMC 0,25 % 100 kHz to 200 kHz — 100 nF to 1000 nF · CMC 0,30 % 200 kHz to 1 MHz — 100 nF to 1000 nF · CMC 0,45 % 20 Hz to 100 Hz — 1 µF to 10 µF · CMC 0,16 % 100 Hz to 500 Hz — 1 µF to 10 µF · CMC 0,10 % 500 Hz to 2 kHz — 1 µF to 10 µF · CMC 0,20 % 2 kHz to 5 kHz — 1 µF to 10 µF · CMC 0,22 % 5 kHz to 10 kHz — 1 µF to 10 µF · CMC 0,25 % 10 kHz to 20 kHz — 1 µF to 10 µF · CMC 0,30 % 20 kHz to 100 kHz — 1 µF to 10 µF · CMC (C/40 + 0,20) % 20 Hz to 50 Hz — 10 µF to 100 µF · CMC 0,17 % 50 Hz to 200 Hz — 10 µF to 100 µF · CMC 0,20 % 200 Hz to 500 Hz — 10 µF to 100 µF · CMC 0,25 % 500 Hz to 1 kHz — 10 µF to 100 µF · CMC 0,30 % 1 kHz to 2 kHz — 10 µF to 100 µF · CMC 0,35 % 2 kHz to 5 kHz — 10 µF to 100 µF · CMC (C/500 + 0,25) % 5 kHz to 10 kHz — 10 µF to 100 µF · CMC (C/250 + 0,25) % 10 kHz to 20 kHz — 10 µF to 100 µF · CMC (C/120 + 0,25) % 20 Hz to 200 Hz — 100 µF to 1000 µF · CMC (C/3500 + 0,15) % 200 Hz to 500 Hz — 100 µF to 1000 µF · CMC (C/2800 + 0,20) % 500 Hz to 1 kHz — 100 µF to 1000 µF · CMC (C/1900 + 0,30) % 1 kHz to 2 kHz — 100 µF to 1000 µF · CMC (C/1000 + 0,30) % 2 kHz to 5 kHz — 100 µF to 1000 µF · CMC (C/500 + 0,30) % 20 Hz to 50 Hz — 1000 µF to 10000 µF · CMC (C/12000 + 0,30) % 50 Hz to 100 Hz — 1000 µF to 10000 µF · CMC (C/9000 + 0,30) % 100 Hz to 200 Hz — 1000 µF to 10000 µF · CMC (C/5000 + 0,20) %
CMC for the field of measured quantity: Electrical quantities
Direct measurement or comparison by substitution with a reference RLC bridge MKE-Z2
53 Inductance / Inductance standards, inductance meters, RLC bridges Calibration – electrical quantities
100 Hz — 10 µH · CMC 0,30 % 1 kHz — 10 µH · CMC 0,040 % 10 kHz — 10 µH · CMC 0,030 % 100 Hz — 100 µH · CMC 0,070 % 1 kHz — 100 µH · CMC 0,020 % 10 kHz — 100 µH · CMC 0,020 % 100 Hz — 1 mH · CMC 0,030 % 1 kHz — 1 mH · CMC 0,020 % 10 kHz — 1 mH · CMC 0,020 % 1 kHz — 2 mH · CMC 0,020 % 1 kHz — 10 mH · CMC 0,020 % 1 kHz — 100 mH · CMC 0,020 % 1 kHz — 1 H · CMC 0,020 % 1 kHz — 10 H · CMC 0,030 %
CMC for the field of measured quantity: Electrical quantities
Comparison by substitution or direct measurement of inductance standard MKE-Z2
54 Inductance / Inductance standards, meters, boxes, RLC bridges Calibration – electrical quantities
1 kHz to 2 kHz — 1 µH to 10 µH · CMC 0,25 % + 0,147 µH 2 kHz to 5 kHz — 1 µH to 10 µH · CMC 0,25 % + 0,063 µH 5 kHz to 10 kHz — 1 µH to 10 µH · CMC 0,25 % + 0,022 µH 10 kHz to 20 kHz — 1 µH to 10 µH · CMC 0,25 % + 0,010 µH 20 kHz to 50 kHz — 1 µH to 10 µH · CMC 0,25 % + 0,005 µH 50 kHz to 100 kHz — 1 µH to 10 µH · CMC 0,25 % + 0,002 µH 100 kHz to 200 kHz — 1 µH to 10 µH · CMC 0,32 % 200 kHz to 500 kHz — 1 µH to 10 µH · CMC 0,28 % 500 Hz to 1 kHz — 10 µH to 100 µH · CMC 0,25 % + 0,35 µH 1 kHz to 2 kHz — 10 µH to 100 µH · CMC 0,25 % + 0,15 µH 2 kHz to 5 kHz — 10 µH to 100 µH · CMC 0,25 % + 0,06 µH 5 kHz to 10 kHz — 10 µH to 100 µH · CMC 0,25 % + 0,02 µH 10 kHz to 20 kHz — 10 µH to 100 µH · CMC 0,18 % + 0,02 µH 20 kHz to 50 kHz — 10 µH to 100 µH · CMC 0,18 % + 0,01 µH 50 kHz to 500 kHz — 10 µH to 100 µH · CMC 0,20 % 100 Hz to 200 Hz — 100 µH to 1000 µH · CMC 0,30 % + 3,0 µH 200 Hz to 500 Hz — 100 µH to 1000 µH · CMC 0,30 % + 1,2 µH 500 Hz to 1 kHz — 100 µH to 1000 µH · CMC 0,30 % + 0,3 µH 1 kHz to 2 kHz — 100 µH to 1000 µH · CMC 0,20 % + 0,25 µH 2 kHz to 5 kHz — 100 µH to 1000 µH · CMC 0,17 % + 0,15 µH 5 kHz to 10 kHz — 100 µH to 1000 µH · CMC 0,22 % 10 kHz to 50 kHz — 100 µH to 1000 µH · CMC 0,20 % 50 kHz to 200 kHz — 100 µH to 1000 µH · CMC 0,13 % 200 kHz to 500 kHz — 100 µH to 1000 µH · CMC 0,15 % 50 Hz to 100 Hz — 1 mH to 10 mH · CMC 0,45 % + 0,007 mH 100 Hz to 200 Hz — 1 mH to 10 mH · CMC 0,30 % + 0,003 mH 200 Hz to 500 Hz — 1 mH to 10 mH · CMC 0,15 % + 0,003 mH 500 Hz to 2 kHz — 1 mH to 10 mH · CMC 0,23 % 2 kHz to 5 kHz — 1 mH to 10 mH · CMC 0,20 % 5 kHz to 10 kHz — 1 mH to 10 mH · CMC 0,15 % 10 kHz to 100 kHz — 1 mH to 10 mH · CMC 0,12 % 100 kHz to 200 kHz — 1 mH to 10 mH · CMC 0,20 % 200 kHz to 500 kHz — 1 mH to 10 mH · CMC 0,25 % 50 Hz to 100 Hz — 10 mH to 100 mH · CMC 0,20 % + 0,03 mH 100 Hz to 200 Hz — 10 mH to 100 mH · CMC 0,22 % 200 Hz to 500 Hz — 10 mH to 100 mH · CMC 0,18 % 500 Hz to 1 kHz — 10 mH to 100 mH · CMC 0,10 % 1 kHz to 20 kHz — 10 mH to 100 mH · CMC 0,12 % 20 kHz to 50 kHz — 10 mH to 100 mH · CMC 0,15 % 50 kHz to 100 kHz — 10 mH to 100 mH · CMC 0,30 % 100 kHz to 500 kHz — 10 mH to 100 mH · CMC 0,50 % 50 Hz to 100 Hz — 0,1 H to 1 H · CMC 0,17 % 100 Hz to 5 kHz — 0,1 H to 1 H · CMC 0,10 % 5 kHz to 10 kHz — 0,1 H to 1 H · CMC 0,11 % 10 kHz to 20 kHz — 0,1 H to 1 H · CMC 0,20 % 20 kHz to 50 kHz — 0,1 H to 1 H · CMC 0,25 % 50 kHz to 100 kHz — 0,1 H to 1 H · CMC (L/2 + 0,25) % 50 Hz to 100 Hz — 1 H to 10 H · CMC 0,16 % 100 Hz to 1 kHz — 1 H to 10 H · CMC 0,10 % 1 kHz to 5 kHz — 1 H to 10 H · CMC 0,12 % 5 kHz to 10 kHz — 1 H to 10 H · CMC 0,22 % 10 kHz to 20 kHz — 1 H to 10 H · CMC (L/200 + 0,20) % 20 kHz to 50 kHz — 1 H to 10 H · CMC (L/40 + 0,20) % 50 Hz to 100 Hz — 10 H to 100 H · CMC 0,16 % 100 Hz to 200 Hz — 10 H to 100 H · CMC 0,12 % 200 Hz to 1 kHz — 10 H to 100 H · CMC 0,15 % 1 kHz to 2 kHz — 10 H to 100 H · CMC 0,20 % 2 kHz to 5 kHz — 10 H to 100 H · CMC (L/400 + 0,10) % 5 kHz to 10 kHz — 10 H to 100 H · CMC (L/240 + 0,20) % 10 kHz to 20 kHz — 10 H to 100 H · CMC (L/110 + 0,20) % 20 kHz to 50 kHz — 10 H to 100 H · CMC (L/45 + 0,20) %
CMC for the field of measured quantity: Electrical quantities
Direct measurement or comparison by substitution with a reference RLC bridge MKE-Z2
55 HF voltage peak-to-peak value / HF voltage meters, oscilloscopes, HF millivoltmeters, frequency analyzers Calibration – electrical quantities
100 kHz to 300 MHz — 5 mV to 5 V · CMC 3,7 % 300 MHz to 550 MHz — 5 mV to 5 V · CMC 4,2 % 550 MHz to 1,1 GHz — 5 mV to 3 V · CMC 5,2 % 1,1 GHz to 2,5 GHz — 5 mV to 3 V · CMC 5,9 % 2,5 GHz to 3,2 GHz — 5 mV to 2 V · CMC 5,9 %
CMC for the field of measured quantity: Electrical quantities
Direct generation with a calibrator in the plane of the connecting BNC connector of 50 Ω MKE-UVF1
56 HF voltage effective value / HF voltage meters, oscilloscopes, HF millivoltmeters, frequency analyzers, function generators, signal generators Calibration – electrical quantities
100 kHz to 100 MHz — 1 mV to 2 mV · CMC 3,6 % 100 kHz to 100 MHz — 2 mV to 10 mV · CMC 2,8 % 100 kHz to 100 MHz — 10 mV to 10 V · CMC 2,5 % 100 MHz to 200 MHz — 1 mV to 2 mV · CMC 4,2 % 100 MHz to 200 MHz — 2 mV to 10 mV · CMC 3,4 % 100 MHz to 200 MHz — 10 mV to 1 V · CMC 3,0 % 100 MHz to 200 MHz — 1 V to 10 V · CMC 4,2 % 200 MHz to 500 MHz — 1 mV to 2 mV · CMC 4,5 % 200 MHz to 500 MHz — 2 mV to 10 mV · CMC 3,7 % 200 MHz to 500 MHz — 10 mV to 1 V · CMC 3,3 % 200 MHz to 500 MHz — 1 V to 10 V · CMC 4,9 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement, comparison with a HF millivoltmeter in the plane of N connector of 50 Ω MKE-UVF1
57 High DC voltage / Power supplies, inspection instruments Calibration – electrical quantities
1 kV to 100 kV · CMC 0,20 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with a HV probe MKE-UVNSS1, MKE-SRP.1
58 High DC voltage / Meters, inspection instruments Calibration – electrical quantities
1 kV to 80 kV · CMC 0,20 %
CMC for the field of measured quantity: Electrical quantities
Direct generation, comparison with a HV probe MKE-UVNSS1, MKE-SRP.1
59 High AC voltage / Meters, inspection instruments Calibration – electrical quantities
50 Hz — 1 kV to 30 kV · CMC 0,16 % 50 Hz — 30 kV to 100 kV · CMC 1,3 %
CMC for the field of measured quantity: Electrical quantities
Direct measurement with a HV probe MKE-UVNST1, MKE-SRP.1
60 High AC voltage / Power supplies, inspection instruments Calibration – electrical quantities
50 Hz — 1 kV to 30 kV · CMC 0,16 % 50 Hz — 30 kV to 50 kV · CMC 1,3 %
CMC for the field of measured quantity: Electrical quantities
Direct generation, comparison with a HV probe MKE-UVNST1, MKE-SRP.1
61 Illuminance / Luxmeters Calibration – optical quantities
1 lx to 30000 lx · CMC 2,5 %
CMC for the field of measured quantity: Optical quantities
Comparison with a luxmeter MK-LUX1
62 Frequency / Frequency meters, counters, frequency analyzers, function generators, signal generators, frequency references, reference oscillators, multimeters, calibrators Calibration – time and frequency quantities
U ≥ 2,2 V TTL/50 Ω — 0,001 Hz to 50 MHz · CMC 3,5·10⁻¹¹ U ≥ 1 V sine / 50 Ω — 50 MHz to 1 GHz · CMC 1,0·10⁻¹⁰ U ≥ 2 Všš / 50 Ω — 1 GHz to 3,2 GHz · CMC 3,0·10⁻¹⁰
CMC for the field of measured quantity: Time and frequency quantities
Direct generation by a GPS receiver or generator MKF-f1
63 Frequency / Frequency meters, counters, frequency analyzers, function generators, signal generators, frequency references, reference oscillators, multimeters, calibrators Calibration – time and frequency quantities
Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ from 100 s to 200 s — 1 MHz · CMC 4,0·10⁻¹¹ Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ from 200 s to 2000 s — 1 MHz · CMC 6,0·10⁻¹² Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ from 100 s to 200 s — 5 MHz · CMC 4,0·10⁻¹¹ Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ from 200 s to 2000 s — 5 MHz · CMC 6,0·10⁻¹² Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ from 100 s to 200 s — 10 MHz · CMC 4,0·10⁻¹¹ Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ from 200 s to 2000 s — 10 MHz · CMC 6,0·10⁻¹²
CMC for the field of measured quantity: Time and frequency quantities
Indirect measurement with a GPS receiver and counter MKF-f1
64 Frequency / Frequency meters, counters, frequency analyzers, function generators, signal generators, frequency references, reference oscillators, multimeters, calibrators Calibration – time and frequency quantities
pulse signals; τ over 10 s — 0,001 Hz to 100 Hz · CMC 8,0·10⁻¹¹ pulse signals; τ over 1 s — 100 Hz to 350 MHz · CMC 8,0·10⁻¹¹ Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ over 10 s — 1 Hz to 30 Hz · CMC 5·10⁻⁵ Hz Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB; τ over 1 s — 30 Hz to 300 Hz · CMC 3·10⁻⁴ Hz Uvst ≥ 1 Vef, signal to noise ratio ≥ 60 dB — 300 Hz to 100 kHz · CMC 5·10⁻⁵ Hz
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement by a counter MKF-f1, MKE-SRP.1
65 Frequency / Frequency meters, counters, frequency analyzers, function generators, signal generators, frequency references, reference oscillators, multimeters, calibrators Calibration – time and frequency quantities
100 kHz to 300 kHz · CMC 4,0·10⁻¹⁰ 300 kHz to 600 kHz · CMC 2,0·10⁻¹⁰ 600 kHz to 6 GHz · CMC 1,0·10⁻¹⁰
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement by a counter MKF-f1
66 Time interval / Time interval meters, counters, oscilloscopes, function generators, pulse generators, digital and mechanical stopwatches, timers, inspection instruments Calibration – time and frequency quantities
Ušš ≥ 1 V; 50 Ω — 1 ns to 4 ns · CMC 0,70 ns Ušš ≥ 1 V; 50 Ω — 4 ns to 10 ns · CMC 4,5 % + 0,5 ns Ušš ≥ 1 V; 50 Ω — 10 ns to 100 ns · CMC 5,5 % + 0,5 ns
CMC for the field of measured quantity: Time and frequency quantities
Direct generation with a calibrator MKF-t1
67 Time interval / Time interval meters, counters, oscilloscopes, function generators, pulse generators, digital and mechanical stopwatches, timers, inspection instruments Calibration – time and frequency quantities
U ≥ 2,2 V TTL; 50 Ω — 100 ns to 1 µs · CMC 2,0 ns U ≥ 2,2 V TTL; 50 Ω — 1 µs to 100 s · CMC 4,0 ns U ≥ 2,2 V TTL; 50 Ω — 100 s to 10⁵ s · CMC 3,5·10⁻¹¹
CMC for the field of measured quantity: Time and frequency quantities
Direct generation by a GPS receiver MKF-t1
68 Pulse signal period / Time interval meters, counters, oscilloscopes, function generators, pulse generators, digital and mechanical stopwatches, timers, inspection instruments Calibration – time and frequency quantities
Uvst ≥ 1 V; τ over 10 s — 2,85 ns to 10 ms · CMC 8,0·10⁻¹¹ Uvst ≥ 1 V; τ over 1 s — 10 ms to 1000 s · CMC 8,0·10⁻¹¹
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement by a counter MKF-t1
69 Pulse signal duration / Time interval meters, counters, oscilloscopes, function generators, pulse generators, digital and mechanical stopwatches, timers, inspection instruments Calibration – time and frequency quantities
Uvst ≥ 1 V — 5 ns to 10 s · CMC 1,0 ns Uvst ≥ 1 V — 10 s to 10⁵ s · CMC 8,0·10⁻¹¹
CMC for the field of measured quantity: Time and frequency quantities
Direct measurement by a counter MKF-t1, MKE-SRP.1
70 Pulse signal duration / Time interval meters, counters, oscilloscopes, function generators, pulse generators, digital and mechanical stopwatches, timers, inspection instruments Calibration – time and frequency quantities
10⁵ s to 4·10⁵ s · CMC 3,6·10⁻⁸
CMC for the field of measured quantity: Time and frequency quantities
Direct comparison with a calibrator MKF-t1

Reference standards (32)

MK-LUX1
MKE-ISS1
MKE-IST1
MKE-RSS
MKE-RSS1
MKE-SRP.1
MKE-USS1
MKE-UST.1
MKE-UST1
MKE-UVF1
MKE-UVNSS1
MKE-UVNST1
MKE-W1
MKE-Z2
MKF-O1
MKF-f1
MKF-t1
MKGD KM.2
MKGD LS.1
MKGD MM.2
MKGD PM.2
MKGD SM.1
MKGD UM.2
MKGD VK.1
MKMP TL.2
MKRV.1
MKTT ET.2
MKTT OT.2
MKTT PR.1
MKTT ST.2
MKTT TE.2
MKTT TX.1

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/CZ/2249