🇵🇱 "SHIM-POL A.M. Borzymowski" E. Borzymowska-Reszka A. Reszka Spółka Jawna PL

PCA AP 225 • Warszawa, Poland • 11 calibration items

Metrology & calibration

"SHIM-POL A.M. Borzymowski" E. Borzymowska-Reszka A. Reszka Spółka Jawna is a calibration laboratory accredited by PCA under number AP 225 (Warszawa, Poland). Its accreditation scope covers 11 calibration items in the following areas: Metrology & calibration. Calibrated items include: Technological devices, Universal Testing Machines, for compressive forces, for tensile forces. Most frequently used calibration principles: Outside of permanent location (P).

At a glance

  • Accreditation number: PCA AP 225
  • Calibration items in scope: 11
  • Types of calibrated items: 11
  • Calibration principles used: 1
  • City: Warszawa, Poland
  • Valid until: 17-02-2029
Accreditation number
AP 225
Accreditation body
PCA
Address
ul. Kochanowskiego 49a, 01-864 Warszawa
City
Warszawa, Poland
Division
SHIM-POL Calibration Laboratory
Valid until
17-02-2029

What this laboratory calibrates (11)

Calibration principles (1)

Certificate information

Issue date
11.06.2026
Issue no.
4
Certificate date
18.02.2025
Accreditation cycle
18.02.2025 — 17.02.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (11)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 %. and is expressed in measurand units. The measurement uncertainty value for CMC expressed as a percentage is the relative measurement uncertainty and refers to the percentage share in the measured value.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Universal Testing Machines
Technological devices
for tensile forces
for compressive forces
Calibration — Force
with weights — 0,01 N to 5 N · CMC 0,1 % with weights — 1 N to 250 N · CMC 0,03 %
Force
Outside of permanent location (P) PN-EN ISO 7500-1:2018-05, PN-EN ISO 12390-4:2020-03 cl. 4.2.2, ASTM E4-24
2 Universal Testing Machines
Technological devices
for compressive forces
Calibration — Force
0,2 N to 1000 kN · CMC 0,12 % 0,2 N to 3000 kN · CMC 0,24 %
Force
Outside of permanent location (P) PN-EN ISO 7500-1:2018-05, PN-EN ISO 12390-4:2020-03 cl. 4.2.2, ASTM E4-24
3 Universal Testing Machines
Technological devices
for tensile forces
Calibration — Force
0,2 N to 1000 kN · CMC 0,12 %
Force
Outside of permanent location (P) PN-EN ISO 7500-1:2018-05, PN-EN ISO 12390-4:2020-03 cl. 4.2.2, ASTM E4-24
4 Extensometers Calibration — Force
0 mm to 100 mm · CMC 0,10 µm 0 mm to 500 mm · CMC 0,23 µm 0 mm to 2500 mm · CMC 2,36 µm
Force
Outside of permanent location (P) PN-EN ISO 9513:2013-06, ASTM E83-23
5 Universal Testing Machines
Technological devices
Crosshead Displacement or Actuator Stroke
Calibration — Force
0 mm to 100 mm · CMC 0,10 µm 0 mm to 500 mm · CMC 0,23 µm 0 mm to 5000 mm · CMC 2,36 µm
Force
Outside of permanent location (P) PN-EN ISO 9513:2013-06 Annex H, ASTM E2309/E2309M-20
6 Universal Testing Machines
Technological devices
Crosshead /Actuator Speed
Calibration — Force
0 mm/min to 10000 mm/min · CMC 0,005 %
Force
Outside of permanent location (P) PN-EN ISO 5893:2015-12, ASTM E2658-15
7 Impact Testing Machines Calibration — Impact
0,5 J to 750 J · CMC 0,18 % K_N K_N – nominal energy of the pendulum in J
Impact
Outside of permanent location (P) PN-EN ISO 148-2:2017-02, PN-EN ISO 13802:2015-07, ASTM E23-24, ASTM D6110-18
8 Brinell testing machines Calibration — Hardness
- hardness — HBW 1/30 · CMC 1,0 % - hardness — HBW 2.5/31.25 · CMC 1,0 % - hardness — HBW 2.5/62.5 · CMC 0,6 % - hardness — HBW 2.5/187.5 · CMC 1,0 % - hardness — HBW 5/250 · CMC 1,0 % - hardness — HBW 5/750 · CMC 1,0 % - hardness — HBW 10/500 · CMC 1,0 % - hardness — HBW 10/1000 · CMC 1,0 % - hardness — HBW 10/3000 · CMC 1,0 % - force — 294,2 N to 29420 N · CMC 0,12 % - length — 0 mm to 10 mm · CMC 0,4 µm - testing time cycle — 1 s to 20 s · CMC 0,04 s
Hardness
Outside of permanent location (P) PN-EN ISO 6506-2:2019-10, ASTM E10-23
9 Vickers testing machines Calibration — Hardness
- hardness — HV 0,025 · CMC 11,2 % - hardness — HV 0,05 · CMC 9,2 % - hardness — HV 0,1 · CMC 6,4 % - hardness — HV 0,2 · CMC 5,2 % - hardness — HV 0,3 · CMC 4,0 % - hardness — HV 0,5 · CMC 4,0 % - hardness — HV 1 · CMC 2,0 % - hardness — HV 2 · CMC 2,0 % - hardness — HV 3 · CMC 1,9 % - hardness — HV 5 · CMC 1,1 % - hardness — HV 10 · CMC 1,3 % - hardness — HV 30 · CMC 1,0 % - hardness — HV 100 · CMC 0,9 % - force — 0,2451 N to 980,7 N · CMC 0,12 % - length — 0 mm to 2 mm · CMC 0,5 µm - testing time cycle — 1 s to 20 s · CMC 0,04 s
Hardness
Outside of permanent location (P) PN-EN ISO 6507-2:2018-05, ASTM E92-23, ASTM E384-22
10 Rockwell testing machines Calibration — Hardness
- hardness — HRA · CMC 0,3 HRA - hardness — HRBW · CMC 1,0 HRBW - hardness — HRC · CMC 0,3 HRC - hardness — HRFW · CMC 0,4 HRFW - hardness — HR15N · CMC 0,4 HR15N - hardness — HR30N · CMC 0,6 HR30N - hardness — HR45N · CMC 0,4 HR45N - hardness — HR15TW · CMC 0,4 HR15TW - hardness — HR30TW · CMC 0,9 HR30TW - hardness — HR45TW · CMC 0,7 HR45TW - force — 29,42 N to 1471 N · CMC 0,12 % - testing time cycle — 1 s to 20 s · CMC 0,04 s
Hardness
Outside of permanent location (P) PN-EN ISO 6508-2:2024-06, ASTM E18-25
11 Rockwell testing machines Calibration — Hardness
- hardness — HR15YW · CMC 0,6 HR15YW - force — 29,42 N to 147,1 N · CMC 0,12 %
Hardness
Outside of permanent location (P) ASTM E18-25

Reference standards (20)

ASTM D6110-18
ASTM E10-23
ASTM E18-25
ASTM E23-24
ASTM E2309/E2309M-20
ASTM E2658-15
ASTM E384-22
ASTM E4-24
ASTM E83-23
ASTM E92-23
PN-EN ISO 12390-4:2020-03 pkt. 4.2.2
PN-EN ISO 13802:2015-07
PN-EN ISO 148-2:2017-02
PN-EN ISO 5893:2015-12
PN-EN ISO 6506-2:2019-10
PN-EN ISO 6507-2:2018-05
PN-EN ISO 6508-2:2024-06
PN-EN ISO 7500-1:2018-05
PN-EN ISO 9513:2013-06
PN-EN ISO 9513:2013-06 zał. H

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20225