🇵🇱 Metroserwis Sp. z o.o. PL
Metrology & calibration
Metroserwis Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 214 (Poznań, Poland). Its accreditation scope covers 2 calibration items in the following areas: Metrology & calibration. Calibrated items include: Coordinate measuring machines with tactile probe, Tactile impulse probes, Tactile scanning probes. Most frequently used calibration principles: Method using a reference sphere; Outside of permanent location (P), Method using a step gauge; Outside of permanent location (P).
At a glance
- Accreditation number: PCA AP 214
- Calibration items in scope: 2
- Types of calibrated items: 3
- Calibration principles used: 2
- City: Poznań, Poland
- Valid until: 17-11-2030
- Accreditation number
- AP 214
- Accreditation body
- PCA
- Address
- ul. Kościańska 29, 60-112 Poznań
- City
- Poznań, Poland
- Division
- Laboratorium Wzorcujące Metroserwis
- Valid until
- 17-11-2030
What this laboratory calibrates (3)
- Coordinate measuring machines with tactile probe — 1 calibration item
- Tactile impulse probes — 1 calibration item
- Tactile scanning probes — 1 calibration item
Calibration principles (2)
- Method using a reference sphere; Outside of permanent location (P) — 1 calibration item
- Method using a step gauge; Outside of permanent location (P) — 1 calibration item
Certificate information
- Issue date
- 21.10.2026
- Issue no.
- 3
- Certificate date
- 18.11.2022
- Accreditation cycle
- 21.10.2026 — 17.11.2030
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (2)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Coordinate measuring machines with tactile probe | Calibration — Length (coordinate measurements) 20 mm to 2333 mm · CMC (0,13 + 0,75 L) µm; L – measured value (m) Length (coordinate measurements) |
Method using a step gauge; Outside of permanent location (P) | PN-EN ISO 10360-2:2010 | |
| 2 | Tactile impulse probes Tactile scanning probes |
Calibration — Length (coordinate measurements) 25 mm · CMC 0,2 µm Length (coordinate measurements) |
Method using a reference sphere; Outside of permanent location (P) | PN-EN ISO 10360-5:2020 | |
Reference standards (2)
PN-EN ISO 10360-2:2010
PN-EN ISO 10360-5:2020