🇵🇱 Technolutions Sp. z o.o. PL

PCA AP 207 • Łowicz, Poland • 13 calibration items

Metrology & calibration

Technolutions Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 207 (Łowicz, Poland). Its accreditation scope covers 13 calibration items in the following areas: Metrology & calibration. Calibrated items include: Static testing machines, - for compressive forces, - for tensile forces, Technological devices. Most frequently used calibration principles: Outside of permanent location (P), Internal procedure; Comparison method using a semi-rigid rule; At p…, Internal procedure; Comparison method using a rigid rule; At perman….

At a glance

  • Accreditation number: PCA AP 207
  • Calibration items in scope: 13
  • Types of calibrated items: 16
  • Calibration principles used: 3
  • City: Łowicz, Poland
  • Valid until: 03-05-2030
Accreditation number
AP 207
Accreditation body
PCA
Address
ul. Wiejska 7, 99-400 Łowicz
City
Łowicz, Poland
Division
Technolutions Calibration Laboratory
Valid until
03-05-2030

What this laboratory calibrates (16)

Calibration principles (3)

Certificate information

Issue date
22.04.2026
Issue no.
6
Certificate date
21.04.2023
Accreditation cycle
22.04.2026 — 03.05.2030
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (13)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Rigid rules Calibration — Length
0 mm to 1000 mm · CMC 0,12 mm
Length
Internal procedure; Comparison method using a rigid rule; At permanent location (S) TLW-M-10
2 Semi-rigid rules Calibration — Length
0 mm to 5000 mm · CMC 0,15 mm
Length
Internal procedure; Comparison method using a semi-rigid rule; At permanent location (S) TLW-M-10
3 Tape measures Calibration — Length
0 mm to 5000 mm · CMC 0,15 mm 0 mm to 10000 mm · CMC 0,20 mm
Length
Internal procedure; Comparison method using a semi-rigid rule; At permanent location (S) TLW-M-10
4 Laser distance meters Calibration — Length
0 m to 10 m · CMC 1,0 mm
Length
Internal procedure; Comparison method using a semi-rigid rule; At permanent location (S) TLW-M-11
5 Static testing machines
- for tensile forces
Technological devices
- for tensile forces
Calibration — Force
0,01 N to 5 N · CMC 0,1 % 1) 1 N to 200 N · CMC 0,03 % 1) 5 N to 250 kN · CMC 0,12 % 2) 5 N to 1000 kN · CMC 0,24 % 3)
Force
Outside of permanent location (P) PN-EN ISO 7500-1:2018-05, ASTM E4-24
6 Static testing machines
- for compressive forces
Technological devices
- for compressive forces
Calibration — Force
0,01 N to 5 N · CMC 0,1 % 1) 1 N to 200 N · CMC 0,03 % 1) 5 N to 250 kN · CMC 0,12 % 2) 5 N to 2000 kN · CMC 0,24 % 3)
Force
Outside of permanent location (P) PN-EN ISO 7500-1:2018-05, ASTM E4-24
7 Extensometers Calibration — Force
0 mm to 10 mm · CMC 0,08 µm 0 mm to 2500 mm · CMC 0,7 µm
Force
Outside of permanent location (P) PN-EN ISO 9513:2013-06, ASTM E83-25
8 Static testing machines
Technological devices for static tests
- crosshead displacement (travel)
Calibration — Force
5000 mm · CMC 0,7 µm
Force
Outside of permanent location (P) PN-EN ISO 9513:2013-06 Annex H, ASTM E2309/E2309M-20
9 Static testing machines
Technological devices for static tests
- crosshead displacement speed
Calibration — Force
0,05 mm/min to 5000 mm/min · CMC 0,03 %
Force
Outside of permanent location (P) PN-EN ISO 5893:2015-12, ASTM E2658-15 (2023)
10 Pendulum impact testers Calibration — Impact strength
- potential energy — 0,5 J to 750 J · CMC 0,25 %
Impact strength
Outside of permanent location (P) PN-EN ISO 148-2:2017-02, ASTM E23-25, PN-EN ISO 13802:2015-07
11 Brinell hardness testers Calibration — Hardness
- hardness — HBW 2.5/31.25 · CMC 1,0 % - hardness — HBW 2.5/62.5 · CMC 1,9 % - hardness — HBW 2.5/187.5 · CMC 0,9 % - hardness — HBW 5/250 · CMC 1,7 % - hardness — HBW 5/750 · CMC 1,5 % - hardness — HBW 10/500 · CMC 1,3 % - hardness — HBW 10/1000 · CMC 1,3 % - hardness — HBW 10/3000 · CMC 0,9 % - force — 306,5 N to 29420 N · CMC 0,12 % - length — 0 mm to 10 mm · CMC 0,27 µm
Hardness
Outside of permanent location (P) PN-EN ISO 6506-2:2019-10
12 Vickers hardness testers Calibration — Hardness
- hardness — HV 0,1 · CMC 5,9 % - hardness — HV 0,2 · CMC 4,5 % - hardness — HV 0,3 · CMC 5,5 % - hardness — HV 0,5 · CMC 4,6 % - hardness — HV 1 · CMC 2,5 % - hardness — HV 3 · CMC 3,4 % - hardness — HV 5 · CMC 2,1 % - hardness — HV 10 · CMC 2,0 % - hardness — HV 30 · CMC 2,3 % - force — 0,9807 N to 294,2 N · CMC 0,12 % - length — 0 mm to 2 mm · CMC 0,27 µm
Hardness
Outside of permanent location (P) PN-EN ISO 6507-2:2018-05
13 Rockwell hardness testers Calibration — Hardness
- hardness — HRA · CMC 0,4 HRA - hardness — HRB · CMC 0,5 HRB - hardness — HRC · CMC 0,4 HRC - force — 98,07 N to 1471 N · CMC 0,12 %
Hardness
Outside of permanent location (P) PN-EN ISO 6508-2:2024-06

Reference standards (16)

ASTM E23-25
ASTM E2309/E2309M-20
ASTM E2658-15 (2023)
ASTM E4-24
ASTM E83-25
PN-EN ISO 13802:2015-07
PN-EN ISO 148-2:2017-02
PN-EN ISO 5893:2015-12
PN-EN ISO 6506-2:2019-10
PN-EN ISO 6507-2:2018-05
PN-EN ISO 6508-2:2024-06
PN-EN ISO 7500-1:2018-05
PN-EN ISO 9513:2013-06
PN-EN ISO 9513:2013-06 zał. H