🇵🇱 Proximo Aero Sp. z o.o. PL
Metrology & calibration
Proximo Aero Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 193 (Rzeszów, Poland). Its accreditation scope covers 9 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Bourdon tube pressure gauges, Digital ammeters, Digital voltmeters. Most frequently used calibration principles: Internal procedure; Direct method; At permanent location (S), At permanent location (S) and outside of it (P), Internal procedure; At permanent location (S) and outside of it (P).
At a glance
- Accreditation number: PCA AP 193
- Calibration items in scope: 9
- Types of calibrated items: 11
- Calibration principles used: 3
- City: Rzeszów, Poland
- Valid until: 15-02-2029
- Accreditation number
- AP 193
- Accreditation body
- PCA
- Address
- ul. Wspólna 2, 35-205 Rzeszów
- City
- Rzeszów, Poland
- Division
- Proximo Aero
- Valid until
- 15-02-2029
What this laboratory calibrates (11)
- Multimeters — 5 calibration items
- Bourdon tube pressure gauges — 2 calibration items
- Digital ammeters — 2 calibration items
- Digital voltmeters — 2 calibration items
- Electronic pressure gauges — 2 calibration items
- Pressure transmitters — 2 calibration items
- - absolute pressure (gas, liquid) — 1 calibration item
- - gauge pressure (gas, liquid) — 1 calibration item
- Digital ohmmeters — 1 calibration item
- Torque screwdrivers — 1 calibration item
- Torque wrenches — 1 calibration item
Calibration principles (3)
- Internal procedure; Direct method; At permanent location (S) — 5 calibration items
- At permanent location (S) and outside of it (P) — 2 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 2 calibration items
Certificate information
- Issue date
- 14.04.2026
- Issue no.
- 8
- Certificate date
- 16.02.2021
- Accreditation cycle
- 16.01.2025 — 15.02.2029
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (9)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Digital voltmeters Multimeters |
Calibration — DC voltage 0 mV to 20 mV · CMC 58 μV 20 mV to 200 mV · CMC 58 μV 200 mV to 2 V · CMC 2,9 · 10⁻⁴ · U 2 V to 20 V · CMC 2,9 · 10⁻⁴ · U 20 V to 100 V · CMC 6,5 · 10⁻⁴ · U - 7,2 mV 100 V to 280 V · CMC 58 mV 280 V to 1050 V · CMC 6,7 · 10⁻⁴ · U - 130 mV DC voltage |
Internal procedure; Direct method; At permanent location (S) | IW-PA-09 based on EURAMET cg-15 v. 3.0 | |
| 2 | Digital ammeters Multimeters |
Calibration — DC current 0 μA to 200 μA · CMC 1,1 · 10⁻⁴ · I + 61 nA 0,2 mA to 2 mA · CMC 3,2 · 10⁻⁴ · I + 35 nA 2 mA to 20 mA · CMC 3 · 10⁻⁴ · I + 0,39 μA 20 mA to 200 mA · CMC 3 · 10⁻⁴ · I + 3,1 μA 0,2 A to 2 A · CMC 3,3 · 10⁻⁴ · I + 35 μA 2 A to 20,5 A · CMC 3,7 · 10⁻⁴ · I + 0,41 mA 20,5 A to 30 A · CMC 4,1 · 10⁻⁴ · I + 3,2 mA DC current |
Internal procedure; Direct method; At permanent location (S) | IW-PA-09 based on EURAMET cg-15 v. 3.0 | |
| 3 | Digital voltmeters Multimeters |
Calibration — AC voltage 15 Hz to 10 kHz — 1 mV to 20 mV · CMC 8,5 · 10⁻⁴ · U + 63 μV 15 Hz to 10 kHz — 20 mV to 200 mV · CMC 2,8 · 10⁻⁴ · U + 69 μV 15 Hz to 10 kHz — 200 mV to 2 V · CMC 3,2 · 10⁻⁴ · U + 72 μV 15 Hz to 10 kHz — 2 V to 20 V · CMC 3,2 · 10⁻⁴ · U + 0,53 mV 15 Hz to 10 kHz — 20 V to 100 V · CMC 6,5 · 10⁻⁴ · U - 2,6 mV 15 Hz to 10 kHz — 100 V to 280 V · CMC 1,1 · 10⁻⁴ · U + 53 mV 15 Hz to 10 kHz — 280 V to 1050 V · CMC 7,1 · 10⁻⁴ · U - 70 mV 10 kHz to 30 kHz — 1 mV to 20 mV · CMC 9,2 · 10⁻⁴ · U + 65 μV 10 kHz to 30 kHz — 20 mV to 200 mV · CMC 4,4 · 10⁻⁴ · U + 82 μV 10 kHz to 30 kHz — 200 mV to 2 V · CMC 3,7 · 10⁻⁴ · U + 86 μV 10 kHz to 30 kHz — 2 V to 20 V · CMC 4,1 · 10⁻⁴ · U + 1,1 mV 10 kHz to 30 kHz — 20 V to 100 V · CMC 7 · 10⁻⁴ · U + 8,7 mV 10 kHz to 30 kHz — 100 V to 280 V · CMC 2,6 · 10⁻⁴ · U + 65 mV 30 kHz to 100 kHz — 1 mV to 20 mV · CMC 1,8 · 10⁻³ · U + 67 μV 30 kHz to 100 kHz — 20 mV to 200 mV · CMC 7,6 · 10⁻⁴ · U + 114 μV 30 kHz to 100 kHz — 200 mV to 2 V · CMC 6,6 · 10⁻⁴ · U + 193 μV 30 kHz to 100 kHz — 2 V to 20 V · CMC 5,6 · 10⁻⁴ · U + 3,9 mV 100 kHz to 300 kHz — 1 mV to 20 mV · CMC 4,9 · 10⁻³ · U + 305 μV 100 kHz to 300 kHz — 20 mV to 200 mV · CMC 5 · 10⁻³ · U + 503 μV 100 kHz to 300 kHz — 200 mV to 2 V · CMC 5 · 10⁻³ · U + 799 μV AC voltage |
Internal procedure; Direct method; At permanent location (S) | IW-PA-09 based on EURAMET cg-15 v. 3.0 | |
| 4 | Digital ammeters Multimeters |
Calibration — AC current 15 Hz to 1 kHz — 0 μA to 200 μA · CMC 1,2 · 10⁻³ · I + 99 nA 15 Hz to 1 kHz — 0,2 mA to 2 mA · CMC 9 · 10⁻⁴ · I + 195 nA 15 Hz to 1 kHz — 2 mA to 20 mA · CMC 4,8 · 10⁻⁴ · I + 1,9 μA 15 Hz to 1 kHz — 20 mA to 200 mA · CMC 4,8 · 10⁻⁴ · I + 19 μA 15 Hz to 1 kHz — 0,2 A to 2 A · CMC 5,6 · 10⁻⁴ · I + 93 μA 15 Hz to 1 kHz — 2 A to 20,5 A · CMC 7,9 · 10⁻⁴ · I + 3,9 mA 15 Hz to 1 kHz — 20,5 A to 30 A · CMC 1,2 · 10⁻³ · I + 4,1 mA 1 kHz to 5 kHz — 0 μA to 200 μA · CMC 3 · 10⁻³ · I + 161 nA 1 kHz to 5 kHz — 0,2 mA to 2 mA · CMC 1,5 · 10⁻³ · I + 497 nA 1 kHz to 5 kHz — 2 mA to 20 mA · CMC 1 · 10⁻³ · I + 4 μA 1 kHz to 5 kHz — 20 mA to 200 mA · CMC 1 · 10⁻³ · I + 50 μA 1 kHz to 5 kHz — 0,2 A to 2 A · CMC 1 · 10⁻³ · I + 495 μA 5 kHz to 10 kHz — 0 μA to 200 μA · CMC 5 · 10⁻³ · I + 208 nA 5 kHz to 10 kHz — 0,2 mA to 2 mA · CMC 4 · 10⁻³ · I + 599 nA 5 kHz to 10 kHz — 2 mA to 20 mA · CMC 2 · 10⁻³ · I + 6 μA 5 kHz to 10 kHz — 20 mA to 200 mA · CMC 2 · 10⁻³ · I + 100 μA AC current |
Internal procedure; Direct method; At permanent location (S) | IW-PA-09 based on EURAMET cg-15 v. 3.0 | |
| 5 | Digital ohmmeters Multimeters |
Calibration — DC resistance 0 Ω to 10 Ω · CMC 1,5 · 10⁻⁴ · R + 66 mΩ 10 Ω to 33 Ω · CMC 1,4 · 10⁻⁴ · R + 66 mΩ 33 Ω to 100 Ω · CMC 8,9 · 10⁻⁵ · R + 66 mΩ 100 Ω to 1000 Ω · CMC 5,8 · 10⁻⁴ · R + 14 mΩ 1 kΩ to 10 kΩ · CMC 5,8 · 10⁻⁴ · R + 12 mΩ 10 kΩ to 100 kΩ · CMC 5,8 · 10⁻⁴ · R + 61 mΩ 100 kΩ to 330 kΩ · CMC 3,9 · 10⁻⁵ · R + 55 Ω 330 kΩ to 1000 kΩ · CMC 7,7 · 10⁻⁴ · R - 0,18 kΩ 1 MΩ to 3,3 MΩ · CMC 7,6 · 10⁻⁵ · R + 0,53 kΩ 3,3 MΩ to 10 MΩ · CMC 7,8 · 10⁻⁴ · R – 1,67 kΩ 10 MΩ to 100 MΩ · CMC 2,1 · 10⁻³ · R + 0,29 kΩ 100 MΩ to 330 MΩ · CMC 3 · 10⁻³ · R + 7,17 kΩ 330 MΩ to 1100 MΩ · CMC 1 · 10⁻² · R – 5,77 kΩ DC resistance |
Internal procedure; Direct method; At permanent location (S) | IW-PA-09 based on EURAMET cg-15 v. 3.0 | |
| 6 | Torque wrenches | Calibration — Torque 0,2 Nm to 2 Nm · CMC 1,5 % 2 Nm to 1000 Nm · CMC 0,5 % Torque |
At permanent location (S) and outside of it (P) | PN-EN ISO 6789-01:2017-03, PN-EN ISO 6789-02:2017-03 | |
| 7 | Torque screwdrivers | Calibration — Torque 0,2 Nm to 2 Nm · CMC 1,5 % 2 Nm to 20 Nm · CMC 0,5 % Torque |
At permanent location (S) and outside of it (P) | PN-EN ISO 6789-01:2017-03, PN-EN ISO 6789-02:2017-03 | |
| 8 | Bourdon tube pressure gauges Electronic pressure gauges Pressure transmitters - gauge pressure (gas, liquid) |
Calibration — Pressure -0,09 MPa to 0,25 MPa · CMC 8,0 · 10⁻⁵ MPa 0,25 MPa to 2,5 MPa · CMC 1,0 · 10⁻³ MPa 2,5 MPa to 7,0 MPa · CMC 4,1 · 10⁻³ MPa 7,0 MPa to 35 MPa · CMC 4,1 · 10⁻² MPa 35 MPa to 100 MPa · CMC 1,2 · 10⁻¹ MPa Pressure |
Internal procedure; At permanent location (S) and outside of it (P) | IW-PA-01 based on EURAMET cg-17 v. 4.1 | |
| 9 | Bourdon tube pressure gauges Electronic pressure gauges Pressure transmitters - absolute pressure (gas, liquid) |
Calibration — Pressure -0,09 MPa to 0,25 MPa · CMC 8 · 10⁻⁵ MPa 0,25 MPa to 2,5 MPa · CMC 1,0 · 10⁻³ MPa 2,5 MPa to 35 MPa · CMC 4,1 · 10⁻² MPa Pressure |
Internal procedure; At permanent location (S) and outside of it (P) | IW-PA-01 based on EURAMET cg-17 v. 4.1 | |
Reference standards (6)
EURAMET cg-15 v. 3.0
EURAMET cg-17 v. 4.1
IW-PA-01
IW-PA-09
PN-EN ISO 6789-01:2017-03
PN-EN ISO 6789-02:2017-03