🇵🇱 Termo-Precyzja Spółka Jawna Barański PL
Metrology & calibration
Termo-Precyzja Spółka Jawna Barański is a calibration laboratory accredited by PCA under number AP 189 (Wrocław, Poland). Its accreditation scope covers 24 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Temperature recorders, Temperature transmitters, Analogue ammeters. Most frequently used calibration principles: Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; At permanent location (S), Internal procedure; Outside of permanent location (P).
At a glance
- Accreditation number: PCA AP 189
- Calibration items in scope: 24
- Types of calibrated items: 32
- Calibration principles used: 3
- City: Wrocław, Poland
- Valid until: 31-07-2027
- Accreditation number
- AP 189
- Accreditation body
- PCA
- Address
- ul. Danuty Siedzikówny 7, 51-214 Wrocław
- City
- Wrocław, Poland
- Division
- Termo-Precyzja Measurement Laboratory
- Valid until
- 31-07-2027
What this laboratory calibrates (32)
- Multimeters — 6 calibration items
- Temperature recorders — 4 calibration items
- Temperature transmitters — 4 calibration items
- Analogue ammeters — 3 calibration items
- Digital ammeters — 3 calibration items
- - type Pt100 — 2 calibration items
- - type RTD — 2 calibration items
- - types: C, A — 2 calibration items
- - types: K, N, J, L, E, T, S, R, B — 2 calibration items
- Analogue voltmeters — 2 calibration items
- Calibrators — 2 calibration items
- Digital voltmeters — 2 calibration items
- Noble and base metal thermocouples — 2 calibration items
- Temperature indicators (meters) for resistance sensors — 2 calibration items
- Temperature indicators (meters) for thermocouples — 2 calibration items
- and 17 more
Calibration principles (3)
- Internal procedure; At permanent location (S) and outside of it (P) — 18 calibration items
- Internal procedure; At permanent location (S) — 4 calibration items
- Internal procedure; Outside of permanent location (P) — 2 calibration items
Certificate information
- Issue date
- 15.06.2026
- Issue no.
- 11
- Certificate date
- 02.03.2021
- Accreditation cycle
- 25.07.2023 — 31.07.2027
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (24)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Calibrators Analogue voltmeters Digital voltmeters Multimeters Transducers |
Calibration — DC voltage 0,1 mV to 10 mV · CMC 2·10⁻⁴ · U + 0,007 mV 10 mV to 100 mV · CMC 2·10⁻⁴ · U + 0,01 mV 0,1 V to 1 V · CMC 1·10⁻⁴ · U + 0,00005 V 1 V to 10 V · CMC 25·10⁻⁵ · U + 0,00035 V 10 V to 100 V · CMC 22·10⁻⁵ · U + 0,005 V 100 V to 1000 V · CMC 22·10⁻⁵ · U + 0,048 V DC voltage |
Internal procedure; At permanent location (S) and outside of it (P) | PW-06 based on Euramet cg-15 v. 3.0 | |
| 2 | Calibrators Analogue ammeters Digital ammeters Multimeters Transducers |
Calibration — DC current 0,1 mA to 1 mA · CMC 3·10⁻⁴ · I + 0,0001 mA 1 mA to 10 mA · CMC 25·10⁻⁵ · I + 0,0005 mA 10 mA to 100 mA · CMC 25·10⁻⁵ · I + 0,0065 mA 0,1 A to 1 A · CMC 3·10⁻⁴ · I + 0,0001 A 1 A to 10 A · CMC 35·10⁻⁵ · I + 0,0014 A DC current |
Internal procedure; At permanent location (S) and outside of it (P) | PW-06 based on Euramet cg-15 v. 3.0 | |
| 3 | Analogue voltmeters Digital voltmeters Multimeters |
Calibration — AC voltage f = 50 Hz — 0,1 mV to 1 mV · CMC 4·10⁻⁴ · U + 0,014 mV f = 50 Hz — 1 mV to 10 mV · CMC 5·10⁻⁴ · U + 0,014 mV f = 50 Hz — 10 mV to 100 mV · CMC 4·10⁻⁴ · U + 0,022 mV f = 50 Hz — 0,1 V to 1 V · CMC 4·10⁻⁴ · U + 0,0004 V f = 50 Hz — 1 V to 10 V · CMC 35·10⁻⁵ · U + 0,0017 V f = 50 Hz — 10 V to 100 V · CMC 35·10⁻⁵ · U + 0,015 V f = 50 Hz — 100 V to 1000 V · CMC 3·10⁻⁴ · U + 0,2 V f = 40 Hz to 500 Hz — 0,1 mV to 100 mV · CMC 21·10⁻⁵ · U + 0,052 mV f = 40 Hz to 500 Hz — 0,1 V to 1 V · CMC 6·10⁻⁴ · U + 0,0007 V f = 40 Hz to 500 Hz — 1 V to 10 V · CMC 6·10⁻⁴ · U + 0,0021 V f = 40 Hz to 500 Hz — 10 V to 100 V · CMC 6·10⁻⁴ · U + 0,016 V f = 40 Hz to 500 Hz — 100 V to 1000 V · CMC 7·10⁻⁴ · U + 0,015 V AC voltage |
Internal procedure; At permanent location (S) and outside of it (P) | PW-06 based on Euramet cg-15 v. 3.0 | |
| 4 | Analogue ammeters Digital ammeters Multimeters |
Calibration — AC current f = 50 Hz — 0,1 mA to 1 mA · CMC 4·10⁻⁴ · I + 0,0021 mA f = 50 Hz — 1 mA to 10 mA · CMC 5·10⁻⁴ · I + 0,0062 mA f = 50 Hz — 10 mA to 100 mA · CMC 3·10⁻⁴ · I + 0,058 mA f = 50 Hz — 0,1 A to 1 A · CMC 4·10⁻⁴ · I + 0,002 A f = 50 Hz — 1 A to 10 A · CMC 4·10⁻⁴ · I + 0,005 A f = 40 Hz to 500 Hz — 0,1 mA to 1 mA · CMC 11·10⁻⁴ · I + 0,0047 mA f = 40 Hz to 500 Hz — 1 mA to 10 mA · CMC 11·10⁻⁴ · I + 0,0072 mA AC current |
Internal procedure; At permanent location (S) and outside of it (P) | PW-06 based on Euramet cg-15 v. 3.0 | |
| 5 | Analogue ammeters Digital ammeters Multimeters |
Calibration — AC current 10 mA to 100 mA · CMC 11·10⁻⁴ · I + 0,034 mA 0,1 A to 1 A · CMC 11·10⁻⁴ · I + 0,0046 A 1 A to 10 A · CMC 11·10⁻⁴ · I + 0,0069 A AC current |
Internal procedure; At permanent location (S) and outside of it (P) | PW-06 based on Euramet cg-15 v. 3.0 | |
| 6 | Analogue ohmmeters Digital ohmmeters Multimeters |
Calibration — DC resistance 0,01 Ω to 1 Ω · CMC 2·10⁻⁶ · R + 0,0058 Ω 1 Ω to 10 Ω · CMC 3·10⁻⁶ · R + 0,0058 Ω 10 Ω to 100 Ω · CMC 15·10⁻⁵ · R + 0,0055 Ω 100 Ω to 1000 Ω · CMC 17·10⁻⁶ · R + 0,0058 Ω 1000 Ω to 11000 Ω · CMC 16·10⁻⁶ · R + 0,0058 Ω 100 kΩ · CMC 13 mΩ 1 MΩ · CMC 220 Ω DC resistance |
Internal procedure; At permanent location (S) and outside of it (P) | PW-06 based on Euramet cg-15 v. 3.0 | |
| 7 | Temperature transmitters Temperature recorders Temperature indicators (meters) for thermocouples - types: K, N, J, L, E, T, S, R, B |
Calibration — Electrical simulation of quantities -270 °C to -100 °C · CMC 0,36 °C -100 °C to 0 °C · CMC 0,24 °C 0 °C to 500 °C · CMC 0,17 °C 500 °C to 1000 °C · CMC 0,19 °C 1000 °C to 1370 °C · CMC 0,24 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) and outside of it (P) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 8 | Temperature transmitters Temperature recorders Temperature indicators (meters) for thermocouples - types: C, A |
Calibration — Electrical simulation of quantities -270 °C to 2500 °C · CMC 0,34 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) and outside of it (P) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 9 | Temperature transmitters Temperature recorders Temperature indicators (meters) for resistance sensors - type Pt100 |
Calibration — Electrical simulation of quantities -200 °C to 500 °C · CMC 0,02 °C 500 °C to 850 °C · CMC 0,03 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) and outside of it (P) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 10 | Temperature transmitters Temperature recorders Temperature indicators (meters) for resistance sensors - type RTD |
Calibration — Electrical simulation of quantities -200 °C to 850 °C · CMC 0,28 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) and outside of it (P) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 11 | Temperature simulators for thermocouples - types: K, N, J, L, E, T, S, R, B |
Calibration — Electrical simulation of quantities -270 °C to -100 °C · CMC 0,36 °C -100 °C to 0 °C · CMC 0,24 °C 0 °C to 500 °C · CMC 0,17 °C 500 °C to 1000 °C · CMC 0,19 °C 1000 °C to 1370 °C · CMC 0,24 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) and outside of it (P) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 12 | Temperature simulators for thermocouples - types: C, A |
Calibration — Electrical simulation of quantities -270 °C to 2500 °C · CMC 0,24 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) and outside of it (P) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 13 | Temperature simulators for resistance sensors - type Pt100 |
Calibration — Electrical simulation of quantities -200 °C to 850 °C · CMC 0,02 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 14 | Temperature simulators for resistance sensors - type RTD |
Calibration — Electrical simulation of quantities -200 °C to 850 °C · CMC 0,13 °C Electrical simulation of quantities |
Internal procedure; At permanent location (S) | PW-07 based on Euramet cg-11 v. 2.0 | |
| 15 | Hygrometers Thermohygrometers Transmitters |
Calibration — Relative humidity 5 °C · CMC 0,24 °C at 5 °C — 15 %rh · CMC 1,2 %rh at 5 °C — 54 %rh · CMC 1,5 %rh at 5 °C — 95 %rh · CMC 1,9 %rh 10 °C · CMC 0,17 °C at 10 °C — 15 %rh · CMC 1,2 %rh at 10 °C — 54 %rh · CMC 1,4 %rh at 10 °C — 95 %rh · CMC 1,7 %rh 23 °C · CMC 0,11 °C at 23 °C — 15 %rh · CMC 1,1 %rh at 23 °C — 54 %rh · CMC 1,2 %rh at 23 °C — 95 %rh · CMC 1,4 %rh 40 °C · CMC 0,23 °C at 40 °C — 15 %rh · CMC 1,1 %rh at 40 °C — 54 %rh · CMC 1,3 %rh at 40 °C — 95 %rh · CMC 1,7 %rh 50 °C · CMC 0,25 °C at 50 °C — 15 %rh · CMC 1,1 %rh at 50 °C — 54 %rh · CMC 1,4 %rh at 50 °C — 95 %rh · CMC 1,9 %rh Relative humidity |
Internal procedure; At permanent location (S) and outside of it (P) | PW-09 | |
| 16 | Electronic pressure gauges Bourdon tube pressure gauges Pressure transmitters |
Calibration — Pressure -95 kPa do 0 kPa · CMC 0,055 kPa 0 kPa to 7 kPa · CMC 0,0021 kPa 7 kPa do 700 kPa · CMC 0,084 kPa 700 kPa do 3400 kPa · CMC 0,42 kPa 3400 kPa do 6900 kPa · CMC 0,72 kPa 6900 kPa do 10000 kPa · CMC 2,4 kPa 10000 kPa do 69000 kPa · CMC 21 kPa Pressure |
Internal procedure; At permanent location (S) and outside of it (P) | PW-08 based on Euramet cg-17 v. 4.1 | |
| 17 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -90 °C to 0 °C · CMC 0,040 °C 0 °C do 180 °C · CMC 0,038 °C 180 °C do 230 °C · CMC 0,059 °C 230 °C do 420 °C · CMC 0,12 °C 420 °C do 650 °C · CMC 0,17 °C 650 °C do 700 °C · CMC 0,38 °C 700 °C do 750 °C · CMC 0,52 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | PW-02 | |
| 18 | Noble and base metal thermocouples | Calibration — Temperature (electrical thermometry) -90 °C to 140 °C · CMC 0,11 °C 140 °C to 650 °C · CMC 0,41 °C 650 °C to 1064 °C · CMC 0,55 °C 1064 °C to 1200 °C · CMC 0,8 °C 1200 °C to 1500 °C · CMC 2,0 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PW-04, PW-05 | |
| 19 | Noble and base metal thermocouples | Calibration — Temperature (electrical thermometry) -90 °C to 140 °C · CMC 0,27 °C 140 °C to 650 °C · CMC 0,47 °C 650 °C to 1064 °C · CMC 0,60 °C 1064 °C to 1200 °C · CMC 0,8 °C Temperature (electrical thermometry) |
Internal procedure; Outside of permanent location (P) | PW-04, PW-05 | |
| 20 | Temperature transmitters (containing resistance sensors) | Calibration — Temperature (electrical thermometry) -90 °C to 0 °C · CMC 0,045 °C 0 °C do 180 °C · CMC 0,039 °C 180 °C do 230 °C · CMC 0,052 °C 230 °C do 420 °C · CMC 0,12 °C 420 °C do 650 °C · CMC 0,17 °C 650 °C do 700 °C · CMC 0,39 °C 700 °C do 750 °C · CMC 0,52 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | PW-02 | |
| 21 | Temperature transmitters (containing thermocouples) | Calibration — Temperature (electrical thermometry) -90 °C to 140 °C · CMC 0,11 °C 140 °C to 650 °C · CMC 0,41 °C 650 °C to 1064 °C · CMC 0,55 °C 1064 °C to 1200 °C · CMC 0,8 °C 1200 °C to 1500 °C · CMC 2,0 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) | PW-02, PW-04, PW-05 | |
| 22 | Temperature transmitters (containing thermocouples) | Calibration — Temperature (electrical thermometry) -90 °C to 140 °C · CMC 0,27 °C 140 °C to 650 °C · CMC 0,47 °C 650 °C to 1064 °C · CMC 0,60 °C 1064 °C to 1200 °C · CMC 0,8 °C Temperature (electrical thermometry) |
Internal procedure; Outside of permanent location (P) | PW-02, PW-04, PW-05 | |
| 23 | Electrical thermometers (including electronic) Electrical thermometers (with temperature recording) |
Calibration — Temperature (electrical thermometry) -90 °C to 0 °C · CMC 0,040 °C 0 °C do 180 °C · CMC 0,037 °C 180 °C do 230 °C · CMC 0,059 °C 230 °C do 420 °C · CMC 0,12 °C 420 °C do 650 °C · CMC 0,17 °C 650 °C do 700 °C · CMC 0,34 °C 700 °C do 1064 °C · CMC 0,59 °C 1064 °C do 1200 °C · CMC 0,82 °C 1200 °C to 1300 °C · CMC 1,8 °C 1300 °C do 1500 °C · CMC 2,0 °C Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | PW-03 | |
| 24 | Pyrometers (including radiation, photoelectric and multiband pyrometers, thermal imaging cameras, line scanners) | Calibration — Temperature (radiation thermometry) -30 °C to -19 °C · CMC 1,2 °C -19 °C to 0 °C · CMC 1,0 °C 0 °C to 35 °C · CMC 0,8 °C 35 °C to 50 °C · CMC 0,6 °C 50 °C to 100 °C · CMC 1,0 °C 100 °C to 150 °C · CMC 1,4 °C 150 °C to 200 °C · CMC 1,8 °C 200 °C to 300 °C · CMC 2,5 °C 300 °C to 500 °C · CMC 4,3 °C 500 °C to 982 °C · CMC 5,0 °C 982 °C to 1500 °C · CMC 8,1 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | PW-01 | |
Reference standards (12)
Euramet cg-11 v. 2.0
Euramet cg-15 v. 3.0
Euramet cg-17 v. 4.1
PW-01
PW-02
PW-03
PW-04
PW-05
PW-06
PW-07
PW-08
PW-09