🇵🇱 Termo-Precyzja Spółka Jawna Barański PL

PCA AP 189 • Wrocław, Poland • 24 calibration items

Metrology & calibration

Termo-Precyzja Spółka Jawna Barański is a calibration laboratory accredited by PCA under number AP 189 (Wrocław, Poland). Its accreditation scope covers 24 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Temperature recorders, Temperature transmitters, Analogue ammeters. Most frequently used calibration principles: Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; At permanent location (S), Internal procedure; Outside of permanent location (P).

At a glance

  • Accreditation number: PCA AP 189
  • Calibration items in scope: 24
  • Types of calibrated items: 32
  • Calibration principles used: 3
  • City: Wrocław, Poland
  • Valid until: 31-07-2027
Accreditation number
AP 189
Accreditation body
PCA
Address
ul. Danuty Siedzikówny 7, 51-214 Wrocław
City
Wrocław, Poland
Division
Termo-Precyzja Measurement Laboratory
Valid until
31-07-2027

What this laboratory calibrates (32)

Calibration principles (3)

Certificate information

Issue date
15.06.2026
Issue no.
11
Certificate date
02.03.2021
Accreditation cycle
25.07.2023 — 31.07.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (24)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Calibrators
Analogue voltmeters
Digital voltmeters
Multimeters
Transducers
Calibration — DC voltage
0,1 mV to 10 mV · CMC 2·10⁻⁴ · U + 0,007 mV 10 mV to 100 mV · CMC 2·10⁻⁴ · U + 0,01 mV 0,1 V to 1 V · CMC 1·10⁻⁴ · U + 0,00005 V 1 V to 10 V · CMC 25·10⁻⁵ · U + 0,00035 V 10 V to 100 V · CMC 22·10⁻⁵ · U + 0,005 V 100 V to 1000 V · CMC 22·10⁻⁵ · U + 0,048 V
DC voltage
Internal procedure; At permanent location (S) and outside of it (P) PW-06 based on Euramet cg-15 v. 3.0
2 Calibrators
Analogue ammeters
Digital ammeters
Multimeters
Transducers
Calibration — DC current
0,1 mA to 1 mA · CMC 3·10⁻⁴ · I + 0,0001 mA 1 mA to 10 mA · CMC 25·10⁻⁵ · I + 0,0005 mA 10 mA to 100 mA · CMC 25·10⁻⁵ · I + 0,0065 mA 0,1 A to 1 A · CMC 3·10⁻⁴ · I + 0,0001 A 1 A to 10 A · CMC 35·10⁻⁵ · I + 0,0014 A
DC current
Internal procedure; At permanent location (S) and outside of it (P) PW-06 based on Euramet cg-15 v. 3.0
3 Analogue voltmeters
Digital voltmeters
Multimeters
Calibration — AC voltage
f = 50 Hz — 0,1 mV to 1 mV · CMC 4·10⁻⁴ · U + 0,014 mV f = 50 Hz — 1 mV to 10 mV · CMC 5·10⁻⁴ · U + 0,014 mV f = 50 Hz — 10 mV to 100 mV · CMC 4·10⁻⁴ · U + 0,022 mV f = 50 Hz — 0,1 V to 1 V · CMC 4·10⁻⁴ · U + 0,0004 V f = 50 Hz — 1 V to 10 V · CMC 35·10⁻⁵ · U + 0,0017 V f = 50 Hz — 10 V to 100 V · CMC 35·10⁻⁵ · U + 0,015 V f = 50 Hz — 100 V to 1000 V · CMC 3·10⁻⁴ · U + 0,2 V f = 40 Hz to 500 Hz — 0,1 mV to 100 mV · CMC 21·10⁻⁵ · U + 0,052 mV f = 40 Hz to 500 Hz — 0,1 V to 1 V · CMC 6·10⁻⁴ · U + 0,0007 V f = 40 Hz to 500 Hz — 1 V to 10 V · CMC 6·10⁻⁴ · U + 0,0021 V f = 40 Hz to 500 Hz — 10 V to 100 V · CMC 6·10⁻⁴ · U + 0,016 V f = 40 Hz to 500 Hz — 100 V to 1000 V · CMC 7·10⁻⁴ · U + 0,015 V
AC voltage
Internal procedure; At permanent location (S) and outside of it (P) PW-06 based on Euramet cg-15 v. 3.0
4 Analogue ammeters
Digital ammeters
Multimeters
Calibration — AC current
f = 50 Hz — 0,1 mA to 1 mA · CMC 4·10⁻⁴ · I + 0,0021 mA f = 50 Hz — 1 mA to 10 mA · CMC 5·10⁻⁴ · I + 0,0062 mA f = 50 Hz — 10 mA to 100 mA · CMC 3·10⁻⁴ · I + 0,058 mA f = 50 Hz — 0,1 A to 1 A · CMC 4·10⁻⁴ · I + 0,002 A f = 50 Hz — 1 A to 10 A · CMC 4·10⁻⁴ · I + 0,005 A f = 40 Hz to 500 Hz — 0,1 mA to 1 mA · CMC 11·10⁻⁴ · I + 0,0047 mA f = 40 Hz to 500 Hz — 1 mA to 10 mA · CMC 11·10⁻⁴ · I + 0,0072 mA
AC current
Internal procedure; At permanent location (S) and outside of it (P) PW-06 based on Euramet cg-15 v. 3.0
5 Analogue ammeters
Digital ammeters
Multimeters
Calibration — AC current
10 mA to 100 mA · CMC 11·10⁻⁴ · I + 0,034 mA 0,1 A to 1 A · CMC 11·10⁻⁴ · I + 0,0046 A 1 A to 10 A · CMC 11·10⁻⁴ · I + 0,0069 A
AC current
Internal procedure; At permanent location (S) and outside of it (P) PW-06 based on Euramet cg-15 v. 3.0
6 Analogue ohmmeters
Digital ohmmeters
Multimeters
Calibration — DC resistance
0,01 Ω to 1 Ω · CMC 2·10⁻⁶ · R + 0,0058 Ω 1 Ω to 10 Ω · CMC 3·10⁻⁶ · R + 0,0058 Ω 10 Ω to 100 Ω · CMC 15·10⁻⁵ · R + 0,0055 Ω 100 Ω to 1000 Ω · CMC 17·10⁻⁶ · R + 0,0058 Ω 1000 Ω to 11000 Ω · CMC 16·10⁻⁶ · R + 0,0058 Ω 100 kΩ · CMC 13 mΩ 1 MΩ · CMC 220 Ω
DC resistance
Internal procedure; At permanent location (S) and outside of it (P) PW-06 based on Euramet cg-15 v. 3.0
7 Temperature transmitters
Temperature recorders
Temperature indicators (meters) for thermocouples
- types: K, N, J, L, E, T, S, R, B
Calibration — Electrical simulation of quantities
-270 °C to -100 °C · CMC 0,36 °C -100 °C to 0 °C · CMC 0,24 °C 0 °C to 500 °C · CMC 0,17 °C 500 °C to 1000 °C · CMC 0,19 °C 1000 °C to 1370 °C · CMC 0,24 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) and outside of it (P) PW-07 based on Euramet cg-11 v. 2.0
8 Temperature transmitters
Temperature recorders
Temperature indicators (meters) for thermocouples
- types: C, A
Calibration — Electrical simulation of quantities
-270 °C to 2500 °C · CMC 0,34 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) and outside of it (P) PW-07 based on Euramet cg-11 v. 2.0
9 Temperature transmitters
Temperature recorders
Temperature indicators (meters) for resistance sensors
- type Pt100
Calibration — Electrical simulation of quantities
-200 °C to 500 °C · CMC 0,02 °C 500 °C to 850 °C · CMC 0,03 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) and outside of it (P) PW-07 based on Euramet cg-11 v. 2.0
10 Temperature transmitters
Temperature recorders
Temperature indicators (meters) for resistance sensors
- type RTD
Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,28 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) and outside of it (P) PW-07 based on Euramet cg-11 v. 2.0
11 Temperature simulators for thermocouples
- types: K, N, J, L, E, T, S, R, B
Calibration — Electrical simulation of quantities
-270 °C to -100 °C · CMC 0,36 °C -100 °C to 0 °C · CMC 0,24 °C 0 °C to 500 °C · CMC 0,17 °C 500 °C to 1000 °C · CMC 0,19 °C 1000 °C to 1370 °C · CMC 0,24 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) and outside of it (P) PW-07 based on Euramet cg-11 v. 2.0
12 Temperature simulators for thermocouples
- types: C, A
Calibration — Electrical simulation of quantities
-270 °C to 2500 °C · CMC 0,24 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) and outside of it (P) PW-07 based on Euramet cg-11 v. 2.0
13 Temperature simulators for resistance sensors
- type Pt100
Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,02 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) PW-07 based on Euramet cg-11 v. 2.0
14 Temperature simulators for resistance sensors
- type RTD
Calibration — Electrical simulation of quantities
-200 °C to 850 °C · CMC 0,13 °C
Electrical simulation of quantities
Internal procedure; At permanent location (S) PW-07 based on Euramet cg-11 v. 2.0
15 Hygrometers
Thermohygrometers
Transmitters
Calibration — Relative humidity
5 °C · CMC 0,24 °C at 5 °C — 15 %rh · CMC 1,2 %rh at 5 °C — 54 %rh · CMC 1,5 %rh at 5 °C — 95 %rh · CMC 1,9 %rh 10 °C · CMC 0,17 °C at 10 °C — 15 %rh · CMC 1,2 %rh at 10 °C — 54 %rh · CMC 1,4 %rh at 10 °C — 95 %rh · CMC 1,7 %rh 23 °C · CMC 0,11 °C at 23 °C — 15 %rh · CMC 1,1 %rh at 23 °C — 54 %rh · CMC 1,2 %rh at 23 °C — 95 %rh · CMC 1,4 %rh 40 °C · CMC 0,23 °C at 40 °C — 15 %rh · CMC 1,1 %rh at 40 °C — 54 %rh · CMC 1,3 %rh at 40 °C — 95 %rh · CMC 1,7 %rh 50 °C · CMC 0,25 °C at 50 °C — 15 %rh · CMC 1,1 %rh at 50 °C — 54 %rh · CMC 1,4 %rh at 50 °C — 95 %rh · CMC 1,9 %rh
Relative humidity
Internal procedure; At permanent location (S) and outside of it (P) PW-09
16 Electronic pressure gauges
Bourdon tube pressure gauges
Pressure transmitters
Calibration — Pressure
-95 kPa do 0 kPa · CMC 0,055 kPa 0 kPa to 7 kPa · CMC 0,0021 kPa 7 kPa do 700 kPa · CMC 0,084 kPa 700 kPa do 3400 kPa · CMC 0,42 kPa 3400 kPa do 6900 kPa · CMC 0,72 kPa 6900 kPa do 10000 kPa · CMC 2,4 kPa 10000 kPa do 69000 kPa · CMC 21 kPa
Pressure
Internal procedure; At permanent location (S) and outside of it (P) PW-08 based on Euramet cg-17 v. 4.1
17 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-90 °C to 0 °C · CMC 0,040 °C 0 °C do 180 °C · CMC 0,038 °C 180 °C do 230 °C · CMC 0,059 °C 230 °C do 420 °C · CMC 0,12 °C 420 °C do 650 °C · CMC 0,17 °C 650 °C do 700 °C · CMC 0,38 °C 700 °C do 750 °C · CMC 0,52 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) and outside of it (P) PW-02
18 Noble and base metal thermocouples Calibration — Temperature (electrical thermometry)
-90 °C to 140 °C · CMC 0,11 °C 140 °C to 650 °C · CMC 0,41 °C 650 °C to 1064 °C · CMC 0,55 °C 1064 °C to 1200 °C · CMC 0,8 °C 1200 °C to 1500 °C · CMC 2,0 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PW-04, PW-05
19 Noble and base metal thermocouples Calibration — Temperature (electrical thermometry)
-90 °C to 140 °C · CMC 0,27 °C 140 °C to 650 °C · CMC 0,47 °C 650 °C to 1064 °C · CMC 0,60 °C 1064 °C to 1200 °C · CMC 0,8 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PW-04, PW-05
20 Temperature transmitters (containing resistance sensors) Calibration — Temperature (electrical thermometry)
-90 °C to 0 °C · CMC 0,045 °C 0 °C do 180 °C · CMC 0,039 °C 180 °C do 230 °C · CMC 0,052 °C 230 °C do 420 °C · CMC 0,12 °C 420 °C do 650 °C · CMC 0,17 °C 650 °C do 700 °C · CMC 0,39 °C 700 °C do 750 °C · CMC 0,52 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) and outside of it (P) PW-02
21 Temperature transmitters (containing thermocouples) Calibration — Temperature (electrical thermometry)
-90 °C to 140 °C · CMC 0,11 °C 140 °C to 650 °C · CMC 0,41 °C 650 °C to 1064 °C · CMC 0,55 °C 1064 °C to 1200 °C · CMC 0,8 °C 1200 °C to 1500 °C · CMC 2,0 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PW-02, PW-04, PW-05
22 Temperature transmitters (containing thermocouples) Calibration — Temperature (electrical thermometry)
-90 °C to 140 °C · CMC 0,27 °C 140 °C to 650 °C · CMC 0,47 °C 650 °C to 1064 °C · CMC 0,60 °C 1064 °C to 1200 °C · CMC 0,8 °C
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) PW-02, PW-04, PW-05
23 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-90 °C to 0 °C · CMC 0,040 °C 0 °C do 180 °C · CMC 0,037 °C 180 °C do 230 °C · CMC 0,059 °C 230 °C do 420 °C · CMC 0,12 °C 420 °C do 650 °C · CMC 0,17 °C 650 °C do 700 °C · CMC 0,34 °C 700 °C do 1064 °C · CMC 0,59 °C 1064 °C do 1200 °C · CMC 0,82 °C 1200 °C to 1300 °C · CMC 1,8 °C 1300 °C do 1500 °C · CMC 2,0 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) and outside of it (P) PW-03
24 Pyrometers (including radiation, photoelectric and multiband pyrometers, thermal imaging cameras, line scanners) Calibration — Temperature (radiation thermometry)
-30 °C to -19 °C · CMC 1,2 °C -19 °C to 0 °C · CMC 1,0 °C 0 °C to 35 °C · CMC 0,8 °C 35 °C to 50 °C · CMC 0,6 °C 50 °C to 100 °C · CMC 1,0 °C 100 °C to 150 °C · CMC 1,4 °C 150 °C to 200 °C · CMC 1,8 °C 200 °C to 300 °C · CMC 2,5 °C 300 °C to 500 °C · CMC 4,3 °C 500 °C to 982 °C · CMC 5,0 °C 982 °C to 1500 °C · CMC 8,1 °C
Temperature (radiation thermometry)
Internal procedure; At permanent location (S) and outside of it (P) PW-01

Reference standards (12)

Euramet cg-11 v. 2.0
Euramet cg-15 v. 3.0
Euramet cg-17 v. 4.1
PW-01
PW-02
PW-03
PW-04
PW-05
PW-06
PW-07
PW-08
PW-09

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20189