🇵🇱 Laboratorium Usługowo-Badawcze "Biochemik" Sp. z o.o. PL
Metrology & calibration
Laboratorium Usługowo-Badawcze "Biochemik" Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 158 (Piła, Poland). Its accreditation scope covers 38 calibration items in the following areas: Metrology & calibration. Calibrated items include: Electrical thermometers (including electronic, Electrical thermometers with temperature recording), Load weights, Dispensers. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P), Internal procedure; Measurement of temperature distribution; At per….
At a glance
- Accreditation number: PCA AP 158
- Calibration items in scope: 38
- Types of calibrated items: 30
- Calibration principles used: 9
- City: Piła, Poland
- Valid until: 06-07-2027
- Accreditation number
- AP 158
- Accreditation body
- PCA
- Address
- ul. Strefowa 15, 64-920 Piła
- City
- Piła, Poland
- Valid until
- 06-07-2027
What this laboratory calibrates (30)
- Electrical thermometers (including electronic — 5 calibration items
- Electrical thermometers with temperature recording) — 5 calibration items
- Load weights — 5 calibration items
- Dispensers — 3 calibration items
- Climatic chambers — 2 calibration items
- Furnaces — 2 calibration items
- Hygrometers — 2 calibration items
- Liquid baths — 2 calibration items
- Piston burettes — 2 calibration items
- Resistance thermometer sensors — 2 calibration items
- Temperature calibrators — 2 calibration items
- Thermohygrometers — 2 calibration items
- Thermostatic chambers — 2 calibration items
- Breath analysers — 1 calibration item
- Burettes — 1 calibration item
- and 15 more
Calibration principles (9)
- Internal procedure; At permanent location (S) — 18 calibration items
- Internal procedure; At permanent location (S) and outside of it (P) — 8 calibration items
- Internal procedure; Measurement of temperature distribution; At permanent location (S) and outside of it (P) — 4 calibration items
- Internal procedure; Calibration in a liquid bath; At permanent location (S) — 2 calibration items
- Internal procedure; Calibration in a temperature calibrator; At permanent location (S) and outside of it (P) — 2 calibration items
- Internal procedure; Calibration in a temperature calibrator; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) — 1 calibration item
- Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) and outside of it (P) — 1 calibration item
- Internal procedure; Mass concentration of ethanol in exhaled air (mg of ethanol per 1 l of air); At permanent location (S) — 1 calibration item
Certificate information
- Issue date
- 12.06.2026
- Issue no.
- 23
- Certificate date
- 26.01.2023
- Accreditation cycle
- 03.07.2023 — 06.07.2027
- Approved by
- Katarzyna Wiśniewska
Accreditation scope — calibration items (38)
| No. | Calibrated quantity / subject of calibration | Nominal range / parameters / CMC | Calibration principle | Calibration procedure | |
|---|---|---|---|---|---|
| Calibration items | |||||
| 1 | Breath analysers | Calibration — Mass concentration (breath analysis) 0,00 mg/l do 0,40 mg/l · CMC 0,01 mg/l 0,41 mg/l to 0,70 mg/l · CMC 0,02 mg/l 0,71 mg/l to 1,00 mg/l · CMC 0,03 mg/l 1,01 mg/l to 1,50 mg/l · CMC 0,04 mg/l 1,51 mg/l to 2,00 mg/l · CMC 0,05 mg/l Mass concentration (breath analysis) |
Internal procedure; Mass concentration of ethanol in exhaled air (mg of ethanol per 1 l of air); At permanent location (S) | Internal procedure PW-05 | |
| 2 | Hygrometers Thermohygrometers |
Calibration — Relative humidity 45 %rh to 85 %rh — 45 %rh · CMC 1,0 %rh 45 %rh to 85 %rh — 85 %rh · CMC 1,5 %rh 45 %rh to 85 %rh — -10 °C to 0 °C · CMC 0,20 °C 20 %rh to 90 %rh — 20 %rh · CMC 0,6 %rh 20 %rh to 90 %rh — 90 %rh · CMC 1,2 %rh 20 %rh to 90 %rh — 0 °C to 4 °C · CMC 0,20 °C 20 %rh to 90 %rh — 4 °C to 10 °C · CMC 0,15 °C 10 %rh to 98 %rh — 10 %rh · CMC 0,5 %rh 10 %rh to 98 %rh — 98 %rh · CMC 1,2 %rh 10 %rh to 98 %rh — 10 °C to 60 °C · CMC 0,15 °C 5 %rh to 95 %rh — 5 %rh · CMC 1,0 %rh 5 %rh to 95 %rh — 95 %rh · CMC 1,5 %rh 5 %rh to 95 %rh — 60 °C to 90 °C · CMC 0,20 °C Relative humidity |
Internal procedure; At permanent location (S) | Internal procedure PW-08 based on NMI / DI publications | |
| 3 | Hygrometers Thermohygrometers |
Calibration — Relative humidity 20 %rh to 89 %rh — 20 %rh · CMC 0,8 %rh 20 %rh to 89 %rh — 89 %rh · CMC 1,0 %rh 20 %rh to 89 %rh — 5 °C to 10 °C · CMC 0,2 °C 89 %rh to 90 %rh — 89 %rh · CMC 1,0 %rh 89 %rh to 90 %rh — 90 %rh · CMC 1,8 %rh 89 %rh to 90 %rh — 5 °C to 10 °C · CMC 0,20 °C 10 %rh to 89 %rh — 10 %rh · CMC 0,8 %rh 10 %rh to 89 %rh — 89 %rh · CMC 1,0 %rh 10 %rh to 89 %rh — 10 °C to 50 °C · CMC 0,2 °C 89 %rh to 92 %rh — 89 %rh · CMC 1,0 %rh 89 %rh to 92 %rh — 92 %rh · CMC 1,7 %rh 89 %rh to 92 %rh — 10 °C to 50 °C · CMC 0,20 °C Relative humidity |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-08 based on NMI / DI publications | |
| 4 | Climatic chambers | Calibration — Relative humidity 25 %rh to 98 %rh — 25 %rh · CMC 2,2 %rh 1) 25 %rh to 98 %rh — 98 %rh · CMC 3,0 %rh 1) 25 %rh to 98 %rh — 10 °C to 20 °C · CMC 0,20 °C 1) 20 %rh to 98 %rh — 20 %rh · CMC 2,2 %rh 1) 20 %rh to 98 %rh — 98 %rh · CMC 3,0 %rh 1) 20 %rh to 98 %rh — 20 °C to 25 °C · CMC 0,20 °C 1) 10 %rh to 98 %rh — 10 %rh · CMC 2,2 %rh 1) 10 %rh to 98 %rh — 98 %rh · CMC 3,0 %rh 1) 10 %rh to 98 %rh — 25 °C to 60 °C · CMC 0,20 °C 1) Relative humidity |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-06 based on EURAMET cg-20 v. 5.0 | |
| 5 | Non-automatic electronic weighing instruments | Calibration — Mass (weighing instruments) up to 2 kg · CMC 8 ∙ 10⁻⁵ % 2 kg to 6 kg · CMC 7 ∙ 10⁻⁴ % 6 kg to 200 kg · CMC 2 ∙ 10⁻³ % Mass (weighing instruments) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-11 based on EURAMET cg-18 v. 4.0 | |
| 6 | Mass standards and weights of accuracy class E2 Load weights |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,002 mg 2 mg · CMC 0,002 mg 5 mg · CMC 0,002 mg 10 mg · CMC 0,002 mg 20 mg · CMC 0,002 mg 50 mg · CMC 0,004 mg 100 mg · CMC 0,004 mg 200 mg · CMC 0,004 mg 500 mg · CMC 0,008 mg 1g · CMC 0,008 mg 2g · CMC 0,008 mg 5g · CMC 0,01 mg 10 g · CMC 0,02 mg 20 g · CMC 0,02 mg 50 g · CMC 0,03 mg 100 g · CMC 0,04 mg 200 g · CMC 0,06 mg 500 g · CMC 0,25 mg 1 kg · CMC 0,5 mg 2 kg · CMC 1,0 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure PW-10 based on OIML R 111-1:2004 Annex C | |
| 7 | Mass standards and weights of accuracy class F1 Load weights |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,006 mg 2 mg · CMC 0,006 mg 5 mg · CMC 0,006 mg 10 mg · CMC 0,008 mg 20 mg · CMC 0,010 mg 50 mg · CMC 0,010 mg 100 mg · CMC 0,013 mg 200 mg · CMC 0,020 mg 500 mg · CMC 0,026 mg 1g · CMC 0,03 mg 2g · CMC 0,04 mg 5g · CMC 0,05 mg 10 g · CMC 0,06 mg 20 g · CMC 0,08 mg 50 g · CMC 0,10 mg 100 g · CMC 0,16 mg 200 g · CMC 0,30 mg 500 g · CMC 0,50 mg 1 kg · CMC 1,0 mg 2 kg · CMC 1,7 mg 5 kg · CMC 8 mg 10 kg · CMC 15 mg 20 kg · CMC 30 mg 25 kg · CMC 40 mg 50 kg · CMC 80 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure PW-10 based on OIML R 111-1:2004 Annex C | |
| 8 | Mass standards and weights of accuracy class F2 Load weights |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,02 mg 2 mg · CMC 0,02 mg 5 mg · CMC 0,02 mg 10 mg · CMC 0,025 mg 20 mg · CMC 0,03 mg 50 mg · CMC 0,04 mg 100 mg · CMC 0,05 mg 200 mg · CMC 0,06 mg 500 mg · CMC 0,08 mg 1g · CMC 0,10 mg 2g · CMC 0,13 mg 5g · CMC 0,15 mg 10 g · CMC 0,20 mg 20 g · CMC 0,25 mg 50 g · CMC 0,30 mg 100 g · CMC 0,50 mg 200 g · CMC 1,0 mg 500 g · CMC 2,5 mg 1 kg · CMC 5 mg 2 kg · CMC 10 mg 5 kg · CMC 25 mg 10 kg · CMC 50 mg 20 kg · CMC 100 mg 25 kg · CMC 120 mg 50 kg · CMC 250 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure PW-10 based on OIML R 111-1:2004 Annex C | |
| 9 | Mass standards and weights of accuracy class M1, M2, M3 Load weights |
Calibration — Mass (weights and mass standards) 1 mg · CMC 0,06 mg 2 mg · CMC 0,06 mg 5 mg · CMC 0,06 mg 10 mg · CMC 0,08 mg 20 mg · CMC 0,10 mg 50 mg · CMC 0,13 mg 100 mg · CMC 0,15 mg 200 mg · CMC 0,20 mg 500 mg · CMC 0,25 mg 1g · CMC 0,30 mg 2g · CMC 0,40 mg 5g · CMC 0,50 mg 10 g · CMC 0,60 mg 20 g · CMC 0,80 mg 50 g · CMC 1,0 mg 100 g · CMC 1,5 mg 200 g · CMC 3,0 mg 500 g · CMC 8,0 mg 1 kg · CMC 15 mg 2 kg · CMC 30 mg 5 kg · CMC 80 mg 10 kg · CMC 150 mg 20 kg · CMC 300 mg 25 kg · CMC 380 mg 50 kg · CMC 800 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure PW-10 based on OIML R 111-1:2004 Annex C | |
| 10 | Load weights | Calibration — Mass (weights and mass standards) 1 mg to 2 mg · CMC 0,004 mg 2 mg to 5 mg · CMC 0,004 mg 5 mg to 10 mg · CMC 0,004 mg 10 mg to 20 mg · CMC 0,005 mg 20 mg to 50 mg · CMC 0,006 mg 50 mg to 100 mg · CMC 0,008 mg 100 mg to 200 mg · CMC 0,01 mg 200 mg to 500 mg · CMC 0,012 mg 0,5 g to 1 g · CMC 0,016 mg 1 g to 2 g · CMC 0,02 mg 2 g to 5 g · CMC 0,024 mg 5 g to 10 g · CMC 0,032 mg 10 g to 20 g · CMC 0,04 mg 20 g to 50 g · CMC 0,05 mg 50 g to 100 g · CMC 0,06 mg 100 g to 200 g · CMC 0,1 mg 200 g to 500 g · CMC 0,2 mg 0,5 kg to 1 kg · CMC 0,5 mg 1 kg to 2 kg · CMC 2,0 mg 2 kg to 5 kg · CMC 8,0 mg 5 kg to 10 kg · CMC 25 mg 10 kg to 20 kg · CMC 40 mg 20 kg to 25 kg · CMC 50 mg 25 kg to 62 kg · CMC 90 mg Mass (weights and mass standards) |
Internal procedure; At permanent location (S) | Internal procedure PP-02 based on OIML R 111-1:2004 Annex C | |
| 11 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -80 °C to 90 °C · CMC 0,01 °C 90 °C to 300 °C · CMC 0,02 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a liquid bath; At permanent location (S) | Internal procedure PW-03 based on PN-EN 60751:2022 | |
| 12 | Resistance thermometer sensors | Calibration — Temperature (electrical thermometry) -100 °C to -30 °C · CMC 0,2 °C -30 °C to 139 °C · CMC 0,1 °C 139 °C to 550 °C · CMC 0,2 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a temperature calibrator; At permanent location (S) and outside of it (P) | Internal procedure PW-03 based on PN-EN 60751:2022 | |
| 13 | Temperature calibrators | Calibration — Temperature (electrical thermometry) -10 °C to 200 °C · CMC 0,15 °C 1) Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-07 based on EURAMET cg-13 v. 4.0 | |
| 14 | Temperature calibrators | Calibration — Temperature (electrical thermometry) -10 °C to 200 °C · CMC 0,15 °C 1) Temperature (electrical thermometry) |
Internal procedure; Measurement of temperature distribution; At permanent location (S) and outside of it (P) | Internal procedure PP-01 based on EURAMET cg-13 v. 4.0 | |
| 15 | Steam sterilisation chambers (e.g. autoclaves) | Calibration — Temperature (electrical thermometry) 60 °C to 140 °C · CMC 0,50 °C 1) Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-14 | |
| 16 | Climatic chambers Thermostatic chambers |
Calibration — Temperature (electrical thermometry) -80 °C to -40 °C · CMC 0,35 °C 1) -40 °C to 180 °C · CMC 0,20 °C 1) 180 °C to 250 °C · CMC 1,0 °C 1) Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-06 based on EURAMET cg-20 v. 5.0 | |
| 17 | Thermostatic chambers | Calibration — Temperature (electrical thermometry) -80 °C to -40 °C · CMC 0,35 °C 1) -40 °C to 180 °C · CMC 0,20 °C 1) 180 °C to 250 °C · CMC 1,0 °C 1) Temperature (electrical thermometry) |
Internal procedure; Measurement of temperature distribution; At permanent location (S) and outside of it (P) | Internal procedure PP-01 based on EURAMET cg-20 v. 5.0 | |
| 18 | Furnaces | Calibration — Temperature (electrical thermometry) 100 °C to 1085 °C · CMC 3,0 °C 1) Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-12 | |
| 19 | Furnaces | Calibration — Temperature (electrical thermometry) 100 °C to 1085 °C · CMC 3,0 °C 1) Temperature (electrical thermometry) |
Internal procedure; Measurement of temperature distribution; At permanent location (S) and outside of it (P) | Internal procedure PP-01 | |
| 20 | Liquid baths | Calibration — Temperature (electrical thermometry) -40 °C to 250 °C · CMC 0,15 °C 1) Temperature (electrical thermometry) |
Internal procedure; At permanent location (S) and outside of it (P) | Internal procedure PW-09 | |
| 21 | Liquid baths | Calibration — Temperature (electrical thermometry) -40 °C to 250 °C · CMC 0,15 °C 1) Temperature (electrical thermometry) |
Internal procedure; Measurement of temperature distribution; At permanent location (S) and outside of it (P) | Internal procedure PP-01 | |
| 22 | Electrical thermometers (including electronic Electrical thermometers with temperature recording) |
Calibration — Temperature (electrical thermometry) -80 °C to 90 °C · CMC 0,01 °C 90 °C to 300 °C · CMC 0,02 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a liquid bath; At permanent location (S) | Internal procedure PW-02 based on NMI / DI publications | |
| 23 | Electrical thermometers (including electronic Electrical thermometers with temperature recording) |
Calibration — Temperature (electrical thermometry) -100 °C to -30 °C · CMC 0,2 °C -30 °C to 139 °C · CMC 0,1 °C 139 °C to 550 °C · CMC 0,2 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a temperature calibrator; At permanent location (S) and outside of it (P) | Internal procedure PW-02 based on NMI / DI publications | |
| 24 | Electrical thermometers (including electronic Electrical thermometers with temperature recording) |
Calibration — Temperature (electrical thermometry) 550 °C to 1085 °C · CMC 0,7 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a temperature calibrator; At permanent location (S) | Internal procedure PW-02 based on NMI / DI publications | |
| 25 | Electrical thermometers (including electronic Electrical thermometers with temperature recording) |
Calibration — Temperature (electrical thermometry) -70 °C to 4 °C · CMC 0,20 °C 4 °C to 60 °C · CMC 0,15 °C 60 °C to 180 °C · CMC 0,20 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) | Internal procedure PW-02 based on NMI / DI publications | |
| 26 | Electrical thermometers (including electronic Electrical thermometers with temperature recording) |
Calibration — Temperature (electrical thermometry) 5 °C to 50 °C · CMC 0,20 °C Temperature (electrical thermometry) |
Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) and outside of it (P) | Internal procedure PW-02 based on NMI / DI publications | |
| 27 | Liquid-in-glass thermometers | Calibration — Temperature (non-electrical thermometry) -80 °C to -40 °C · CMC 0,08 °C -40 °C to 100 °C · CMC 0,04 °C 100 °C to 250 °C · CMC 0,07 °C 250 °C to 300 °C · CMC 0,10 °C Temperature (non-electrical thermometry) |
Internal procedure; At permanent location (S) | Internal procedure PW-01 based on OIML R 133:2002 | |
| 28 | Radiation pyrometers | Calibration — Temperature (radiation thermometry) -15 °C do 100 °C · CMC 1,0 °C 100 °C to 200 °C · CMC 1,4 °C 200 °C do 350 °C · CMC 2,4 °C 350 °C do 500 °C · CMC 3,5 °C Temperature (radiation thermometry) |
Internal procedure; At permanent location (S) | Internal procedure PW-15 | |
| 29 | Piston pipettes Piston burettes Dispensers |
Calibration — Volume 0,1 µl to 1 µl · CMC 0,03 µl 1 µl to 20 µl · CMC 0,05 µl 20 µl to 100 µl · CMC 0,15 µl 100 µl to 250 µl · CMC 0,30 µl 250 µl to 300 µl · CMC 0,40 µl 300 µl to 1000 µl · CMC 1,4 µl 1000 µl to 5000 µl · CMC 6,0 µl 5000 µl to 10000 µl · CMC 14 µl 10000 µl to 20000 µl · CMC 33 µl Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-13 based on ISO 8655-6:2022 | |
| 30 | Piston burettes Dispensers |
Calibration — Volume 20000 µl to 50000 µl · CMC 33 µl 50000 µl to 100000 µl · CMC 280 µl Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-13 based on ISO 8655-6:2022 | |
| 31 | Dispensers | Calibration — Volume 100000 µl to 200000 µl · CMC 580 µl Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-13 based on ISO 8655-6:2022 | |
| 32 | Syringes | Calibration — Volume 0,1 µl to 1 µl · CMC 0,030 µl 1 µl to 10 µl · CMC 0,050 µl 10 µl to 25 µl · CMC 0,15 µl 25 µl to 50 µl · CMC 0,15 µl 50 µl to 100 µl · CMC 0,29 µl 100 µl to 500 µl · CMC 1,4 µl 500 µl to 1000 µl · CMC 3,0 µl 1000 µl to 10000 µl · CMC 15 µl 10000 µl to 50000 µl · CMC 290 µl Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-13 based on ISO 8655-6:2022 | |
| 33 | One-mark glass volumetric flasks | Calibration — Volume 1 ml · CMC 0,0056 ml 2 ml · CMC 0,0056 ml 5 ml · CMC 0,0056 ml 10 ml · CMC 0,0056 ml 20 ml · CMC 0,0080 ml 25 ml · CMC 0,010 ml 50 ml · CMC 0,015 ml 100 ml · CMC 0,025 ml 200 ml · CMC 0,050 ml 250 ml · CMC 0,050 ml 500 ml · CMC 0,070 ml 1000 ml · CMC 0,17 ml 2000 ml · CMC 0,20 ml Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-16 based on ISO 4787:2022-06 | |
| 34 | Graduated measuring cylinders | Calibration — Volume 1 ml to 5 ml · CMC 0,015 ml 5 ml to 10 ml · CMC 0,026 ml 10 ml to 15 ml · CMC 0,045 ml 15 ml to 25 ml · CMC 0,060 ml 25 ml to 50 ml · CMC 0,11 ml 50 ml to 100 ml · CMC 0,13 ml 100 ml to 200 ml · CMC 0,25 ml 200 ml to 250 ml · CMC 0,35 ml 250 ml to 500 ml · CMC 0,70 ml 500 ml to 1000 ml · CMC 1,1 ml 1000 ml to 2000 ml · CMC 2,0 ml Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-16 based on ISO 4787:2022-06 | |
| 35 | One-mark pipettes | Calibration — Volume 0,5 ml · CMC 0,0015 ml 0,5 ml to 1 ml · CMC 0,0030 ml 1 ml to 2 ml · CMC 0,0040 ml 2 ml to 5 ml · CMC 0,0050 ml 5 ml to 10 ml · CMC 0,010 ml 10 ml to 20 ml · CMC 0,010 ml 20 ml to 25 ml · CMC 0,015 ml 25 ml to 50 ml · CMC 0,020 ml 50 ml to 100 ml · CMC 0,030 ml Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-16 based on ISO 4787:2022-06 | |
| 36 | Graduated pipettes | Calibration — Volume 0,5 ml · CMC 0,0015 ml 0,5 ml to 1 ml · CMC 0,0030 ml 1 ml to 2 ml · CMC 0,0040 ml 2 ml to 5 ml · CMC 0,010 ml 5 ml to 10 ml · CMC 0,015 ml 10 ml to 15 ml · CMC 0,040 ml 15 ml to 20 ml · CMC 0,040 ml 20 ml to 25 ml · CMC 0,040 ml Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-16 based on ISO 4787:2022-06 | |
| 37 | Burettes | Calibration — Volume 0,5 ml to 1 ml · CMC 0,0040 ml 1 ml to 2 ml · CMC 0,0040 ml 2 ml to 5 ml · CMC 0,0040 ml 5 ml to 10 ml · CMC 0,010 ml 10 ml to 25 ml · CMC 0,020 ml 25 ml to 50 ml · CMC 0,020 ml 50 ml to 100 ml · CMC 0,050 ml Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-16 based on ISO 4787:2022-06 | |
| 38 | Vessels / instruments of known volume – beakers, conical flasks, one-mark containers, brim-full containers, funnels, receivers, graduated containers, butyrometers | Calibration — Volume 0,5 ml to 10 ml · CMC 0,0056 ml 10 ml to 25 ml · CMC 0,010 ml 25 ml to 100 ml · CMC 0,025 ml 100 ml to 250 ml · CMC 0,050 ml 250 ml to 500 ml · CMC 0,070 ml 500 ml to 2000 ml · CMC 0,22 ml Volume |
Internal procedure; At permanent location (S) | Internal procedure PW-16 based on ISO 4787:2022-06 | |
Reference standards (26)
EURAMET cg-13 v. 4.0
EURAMET cg-18 v. 4.0
EURAMET cg-20 v. 5.0
ISO 4787:2022-06
ISO 8655-6:2022
OIML R 111-1:2004 Załącznik C
OIML R 133:2002
PN-EN 60751:2022
PP-01
PP-02
PW-01
PW-02
PW-03
PW-05
PW-06
PW-07
PW-08
PW-09
PW-10
PW-11
PW-12
PW-13
PW-14
PW-15
PW-16
publikacje NMI / DI