🇵🇱 ENEA Wytwarzanie Sp. z o.o. PL

PCA AP 147 • Kozienice, Poland • 11 calibration items

Metrology & calibration

ENEA Wytwarzanie Sp. z o.o. is a calibration laboratory accredited by PCA under number AP 147 (Kozienice, Poland). Its accreditation scope covers 11 calibration items in the following areas: Metrology & calibration. Calibrated items include: Multimeters, Bourdon tube pressure gauges, Digital ammeters, Digital voltmeters. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Outside of permanent location (P), Internal procedure; At permanent location (S) and outside of it (P).

At a glance

  • Accreditation number: PCA AP 147
  • Calibration items in scope: 11
  • Types of calibrated items: 12
  • Calibration principles used: 3
  • City: Kozienice, Poland
  • Valid until: 15-01-2029
Accreditation number
AP 147
Accreditation body
PCA
Address
Al. Józefa Zielińskiego 1 Świerże Górne , 26-900 Kozienice
City
Kozienice, Poland
Division
Calibration Laboratory of the Measurement Department
Valid until
15-01-2029

What this laboratory calibrates (12)

Calibration principles (3)

Certificate information

Issue date
28.10.2026
Issue no.
14
Certificate date
04.12.2019
Accreditation cycle
18.12.2024 — 15.01.2029
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (11)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. A CMC uncertainty value expressed as a percentage is the relative measurement uncertainty and refers to the percentage of the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Digital voltmeters
Multimeters
Calibration — DC voltage
10 mV to 100 mV · CMC 0,010 mV 100 mV to 1 V · CMC 0,000063 V 1 to 10 V · CMC 0,00055 V 10 V to 100 V · CMC 0,0062 V 100 V to 1000 V · CMC 0,044 V
DC voltage
Internal procedure; At permanent location (S) PW-01.00.00
2 Digital ammeters
Multimeters
Calibration — DC current
1 mA to 10 mA · CMC 0,027 mA 10 mA to 100 mA · CMC 0,036 mA 100 mA to 1 A · CMC 0,00042 mA 1 to 3 A · CMC 0,0064 A
DC current
Internal procedure; At permanent location (S) PW-01.00.00
3 Digital voltmeters
Multimeters
Calibration — AC voltage
f = 50 Hz — 100 mV to 1 V · CMC 0,00062 V f = 50 Hz — 1 V to 10 V · CMC 0,0060 V f = 50 Hz — 10 V to 100 V · CMC 0,062 V f = 50 Hz — 100 V to 750 V · CMC 0,36 V
AC voltage
Internal procedure; At permanent location (S) PW-01.00.00
4 Digital ammeters
Multimeters
Calibration — AC current
f = 50 Hz — 100 mA to 1A · CMC 0,0014 A
AC current
Internal procedure; At permanent location (S) PW-01.00.00
5 Electronic non-automatic weighing instruments Calibration — Mass (weighing instruments)
10 kg to 500 kg · CMC 6 · 10⁻³ % 500 kg to 100 000 kg · CMC 3 · 10⁻² %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) PW-06.00.00 based on EURAMET cg-18 v.4.0
6 Automatic conveyor belt weighers Calibration — Mass (weighing instruments)
up to 100 000 kg · CMC 3 · 10⁻¹ %
Mass (weighing instruments)
Internal procedure; Outside of permanent location (P) PW-07.00.00 based on OIML R50-1:2014 (p. 7), OIML R50-2:2014 (p. 10)
7 Automatic rail weighbridges Calibration — Mass (weighing instruments)
up to 100 000 kg · CMC 6 · 10⁻² %
Mass (weighing instruments)
Internal procedure; Outside of permanent location (P) PW-08.00.00 based on OIML R106-1:2011 (p. 6)
8 Bourdon tube pressure gauges
Pressure transmitters
- gauge pressure (medium gas)
Calibration — Pressure
-100 kPa to 0 kPa · CMC 0,13 kPa 0 kPa to 200 kPa · CMC 0,2 kPa 200 kPa to 250 kPa · CMC 0,4 kPa 250 kPa to 2500 kPa · CMC 3 kPa
Pressure
Internal procedure; At permanent location (S) PW-02.00.00, PW-05.00.00 based on EURAMET cg-17 v.4.1
9 Bourdon tube pressure gauges
Pressure transmitters
- gauge pressure (medium liquid)
Calibration — Pressure
0,06 MPa do 0,6 MPa · CMC 0,0008 MPa 0,6 MPa to 6 MPa · CMC 0,007 MPa 6 MPa do 30 MPa · CMC 0,04 MPa
Pressure
Internal procedure; At permanent location (S) PW-02.00.00, PW-05.00.00 based on EURAMET cg-17 v.4.1
10 Noble and base metal thermocouples (type S, K, J) Calibration — Temperature (electrical thermometry)
200 °C to 600 °C · CMC 1,5 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PW-03.00.00 based on EURAMET cg-8 v.3.1
11 Resistance thermometer sensors Calibration — Temperature (electrical thermometry)
-20 °C to 150 °C · CMC 0,1 °C
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) PW-03.00.00

Reference standards (13)

EURAMET cg-17 v.4.1
EURAMET cg-18 v.4.0
EURAMET cg-8 v.3.1
OIML R106-1:2011 (p. 6)
OIML R50-1:2014 (p. 7)
OIML R50-2:2014 (p. 10)
PW-01.00.00
PW-02.00.00
PW-03.00.00
PW-05.00.00
PW-06.00.00
PW-07.00.00
PW-08.00.00

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20147