🇵🇱 KLIMATEST Trzynadlowski i Wspólnicy Sp. J. PL

PCA AP 144 • Wrocław, Poland • 9 calibration items

Metrology & calibration

KLIMATEST Trzynadlowski i Wspólnicy Sp. J. is a calibration laboratory accredited by PCA under number AP 144 (Wrocław, Poland). Its accreditation scope covers 9 calibration items in the following areas: Metrology & calibration. Calibrated items include: Climatic chambers, Electrical thermometers (including electronic), Hygrometers, Thermohygrometers. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; Outside of permanent location (P), Internal procedure; Calibration in a thermostat; Outside of permane….

At a glance

  • Accreditation number: PCA AP 144
  • Calibration items in scope: 9
  • Types of calibrated items: 6
  • Calibration principles used: 5
  • City: Wrocław, Poland
  • Valid until: 26-06-2028
Accreditation number
AP 144
Accreditation body
PCA
Address
ul. Obrońców Poczty Gdańskiej 5c, 52-204 Wrocław
City
Wrocław, Poland
Division
KLIMATEST Calibration Laboratory
Valid until
26-06-2028

What this laboratory calibrates (6)

Calibration principles (5)

Certificate information

Issue date
15.04.2025
Issue no.
12
Certificate date
22.04.2020
Accreditation cycle
28.05.2024 — 26.06.2028
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (9)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Climatic chambers Calibration — Relative humidity
temperature range 20 °C to 80 °C — 10 %rh to 98 %rh · CMC 2,0 %rh* temperature range 80 °C to 90 °C — 10 %rh to 98 %rh · CMC 2,5 %rh*
Relative humidity
Internal procedure; Outside of permanent location (P) IW/01 based on EURAMET cg-20 v. 5.0
2 Hygrometers
Thermohygrometers
Calibration — Relative humidity
temperature range 35 °C to 65 °C — 20 %rh to 95 %rh · CMC 3,0 %rh
Relative humidity
Internal procedure; Outside of permanent location (P) IW/02
3 Climatic chambers
Thermostatic chambers
Calibration — Temperature (electrical thermometry)
-75 °C to 130 °C · CMC 0,2 °C* 130 °C to 180 °C · CMC 0,3 °C* 180 °C to 250 °C · CMC 0,5 °C*
Temperature (electrical thermometry)
Internal procedure; Outside of permanent location (P) IW/01 based on EURAMET cg-20 v. 5.0
4 Electrical thermometers (including electronic) Calibration — Temperature (electrical thermometry)
30 °C to 95 °C · CMC 0,20 °C
Temperature (electrical thermometry)
Internal procedure; Calibration in a thermostat; Outside of permanent location (P) IW/02
5 Electrical thermometers (including electronic) Calibration — Temperature (electrical thermometry)
35 °C to 65 °C · CMC 1,0 °C
Temperature (electrical thermometry)
Internal procedure; Calibration in a thermostatic chamber; Outside of permanent location (P) IW/02
6 Climatic chambers Calibration — Relative humidity
temperature range 20 °C to 80 °C — 10 %rh to 98 %rh · CMC 2,0 %rh* temperature range 80 °C to 90 °C — 10 %rh to 98 %rh · CMC 2,5 %rh*
Relative humidity
Internal procedure; At permanent location (S) IW/01 based on EURAMET cg-20 v. 5.0
7 Hygrometers
Thermohygrometers
Calibration — Relative humidity
temperature range 10 °C to 70 °C — 20 %rh to 95 %rh · CMC 1,5 %rh
Relative humidity
Internal procedure; At permanent location (S) IS/02
8 Climatic chambers
Thermostatic chambers
Calibration — Temperature (electrical thermometry)
-75 °C to 130 °C · CMC 0,2 °C* 130 °C to 180 °C · CMC 0,3 °C* 180 °C to 250 °C · CMC 0,5 °C*
Temperature (electrical thermometry)
Internal procedure; At permanent location (S) IW/01 based on EURAMET cg-20 v. 5.0
9 Electrical thermometers (including electronic)
Electrical thermometers (with temperature recording)
Calibration — Temperature (electrical thermometry)
-50 °C do 60 °C · CMC 0,2 °C 60 °C do 120 °C · CMC 0,3 °C 120 °C to 180 °C · CMC 0,6 °C
Temperature (electrical thermometry)
Internal procedure; Calibration in a thermostatic chamber; At permanent location (S) IS/01

Reference standards (5)

EURAMET cg-20 v. 5.0
IS/01
IS/02
IW/01
IW/02

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20144