🇵🇱 WAG-LAB Włodzimierz Kozłowski PL

PCA AP 142 • Zbrosławice, Poland • 5 calibration items

Metrology & calibration

WAG-LAB Włodzimierz Kozłowski is a calibration laboratory accredited by PCA under number AP 142 (Zbrosławice, Poland). Its accreditation scope covers 5 calibration items in the following areas: Metrology & calibration. Calibrated items include: Load weights, Mass standards and weights, Mass standards and weights of accuracy classes F1, F2, M1, M2, M3, Mass standards and weights of accuracy classes F2, M1, M2, M3. Most frequently used calibration principles: Internal procedure; At permanent location (S), Internal procedure; At permanent location (S) and outside of it (P).

At a glance

  • Accreditation number: PCA AP 142
  • Calibration items in scope: 5
  • Types of calibrated items: 6
  • Calibration principles used: 2
  • City: Zbrosławice, Poland
  • Valid until: 30-11-2027
Accreditation number
AP 142
Accreditation body
PCA
Address
ul. Morcinka 12c, 42-674 Zbrosławice
City
Zbrosławice, Poland
Division
Laboratorium Wzorcujące WAG-LAB
Valid until
30-11-2027

What this laboratory calibrates (6)

Calibration principles (2)

Certificate information

Issue date
24.09.2024
Issue no.
13
Certificate date
08.09.2023
Accreditation cycle
08.11.2023 — 30.11.2027
Approved by
Katarzyna Wiśniewska

Accreditation scope — calibration items (5)

The uncertainty covered by the CMC is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in measurand units. The measurement uncertainty value for CMC expressed as a percentage is a relative measurement uncertainty and refers to a percentage contribution in the value of the measurand.

No. Calibrated quantity / subject of calibration Nominal range / parameters / CMC Calibration principle Calibration procedure
Calibration items
1 Non-automatic electronic weighing instruments
Non-automatic mechanical weighing instruments
Calibration — Mass (weighing instruments)
up to 300 g · CMC 5,7 · 10⁻⁵ % 300 g to 1 kg · CMC 1,2 · 10⁻⁴ % 1 kg to 10 kg · CMC 3,8 · 10⁻⁴ % 10 kg to 200 kg · CMC 1,0 · 10⁻³ %
Mass (weighing instruments)
Internal procedure; At permanent location (S) and outside of it (P) IO-07, IO-08, IO-09 w oparciu o EURAMET cg-18 v.4.0
2 Mass standards and weights of accuracy classes F1, F2, M1, M2, M3 Calibration — Mass (weights and mass standards)
1 mg · CMC 0,006 mg 2 mg · CMC 0,006 mg 5 mg · CMC 0,006 mg 10 mg · CMC 0,008 mg 20 mg · CMC 0,010 mg 50 mg · CMC 0,012 mg 100 mg · CMC 0,016 mg 200 mg · CMC 0,020 mg 500 mg · CMC 0,025 mg 1 g · CMC 0,03 mg 2 g · CMC 0,04 mg 5 g · CMC 0,05 mg 10 g · CMC 0,06 mg 20 g · CMC 0,08 mg 50 g · CMC 0,10 mg 100 g · CMC 0,16 mg 200 g · CMC 0,3 mg 500 g · CMC 0,8 mg 1 kg · CMC 1,6 mg 2 kg · CMC 3,0 mg 5 kg · CMC 8,0 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) IO-12 w oparci o OIML R 111-1:2004 Załącznik C
3 Mass standards and weights of accuracy classes F2, M1, M2, M3 Calibration — Mass (weights and mass standards)
10 kg · CMC 50 mg 20 kg · CMC 100 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) IO-12 w oparci o OIML R 111-1:2004 Załącznik C
4 Mass standards and weights Calibration — Mass (weights and mass standards)
25 kg · CMC 130 mg
Mass (weights and mass standards)
Internal procedure; At permanent location (S) IO-12 w oparci o OIML R 111-1:2004 Załącznik C
5 Load weights Calibration — Mass (weights and mass standards)
up to 25 kg · CMC 1,5 · 10⁻³ %
Mass (weights and mass standards)
Internal procedure; At permanent location (S) IO-12 w oparci o OIML R 111-1:2004 Załącznik C

Reference standards (6)

EURAMET cg-18 v.4.0
IO-07
IO-08
IO-09
IO-12
OIML R 111-1:2004

The same scope as data (JSON): https://labcoda.pl/api/accreditation-scopes/labs/PL/AP%20142